K Number
K071642
Manufacturer
Date Cleared
2008-05-02

(319 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only.

WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

Device Description

The Mediaid Vital Signs monitor Pulse Oximeter Model M960P provides real time monitoring and display of ox vgen saturation of arteriolar hemoglobin (SpO2), Non Invasive Blood Pressure, pulse rate and body temperature. Where as the Mediaid Pulse Oximeter Model M900 provides real time monitoring and display of oxygen saturation of arteriolar hemoglobin (SpO2) and pulse rate only.

Noninvasive arterial oxygen saturation measurement is obtained by directing red and infra red light through a pulsating vascular bed. The pulsating arterioles in the path of the light beam cause a change in the amount of light detected by a photodiode. The pulse oximeter determines the oxygen saturation of arterial blood by measuring the ratio of transmitted red to infrared light within the pulse waveform. The non-pulsatile signal is removed electronically for the purpose of calculation. Therefore, skin, bone, and other non-pulsating substances do not interfere with the measurement of arterial oxygen saturation.

Non invasive blood pressure is intended to non invasively measure systolic, diastolic and mean arterial pressures (MAP).

Temperature- Thermometer takes patient temperature in oral, axillary or rectal mode.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Mediaid M960/M900 series devices:

1. Table of Acceptance Criteria and Reported Device Performance

Device ComponentAcceptance CriteriaReported Device Performance
Pulse Oximeter"equivalent to predicate device accuracy claims" (for SpO2 accuracy)Mediaid Pulse Oximeter Model 900 series readings were compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be "equivalent to predicate device accuracy claims." Bench testing was performed to verify pulse rate accuracy.
NIBP Monitor"claimed accuracy is met as per the requirement of IEC 60601-2-30 and AAMI SP10"Clinical study was conducted by Suntech Medical Inc on their OEM module (which Mediaid uses in Model 960 series) and the results show that the claimed accuracy is met as per the requirement of IEC 60601-2-30 and AAMI SP10.
Clinical Thermometer"meets the claimed accuracy as per ASTM E 1112:00 and EN 12470-4"Clinical Tests for Clinical thermometer is conducted on an external hospital and the results show that the thermometer module of Model 960 series meets the claimed accuracy as per ASTM E 1112:00 and EN 12470-4.
GeneralCompliance with IEC60601-1-2 and other applicable standardsApplicable environmental and non-clinical testing was performed per IEC60601-1-2 as well as other applicable standards and procedures. The Mediaid Pulse Oximeter Model 960 series passed all tests as per the requirements of various applicable international standards (including Biocompatibility, electrical safety, and EMC testing). This demonstrates conformance with established industry standards.

2. Sample Size Used for the Test Set and Data Provenance

  • Pulse Oximeter (SpO2): The document states "clinical hypoxia studies conducted in an independent research lab." It does not specify the sample size for the patients involved in these studies. The data provenance is described as "independent research lab," implying a controlled clinical environment, likely prospective for the specific hypoxia challenge. No country of origin is explicitly mentioned.
  • NIBP Monitor: No specific sample size is provided. The study was conducted by "Suntech Medical Inc on their OEM module." This suggests the test data was generated by the OEM, not Mediaid directly. The provenance would be wherever Suntech Medical Inc. conducted their studies.
  • Clinical Thermometer: No specific sample size is provided. The clinical tests were "conducted on an external hospital." The provenance would be the location of that hospital.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not explicitly provided in the document for any of the components.

  • Pulse Oximeter (SpO2): Ground truth was established by "arterial blood samples analyzed on a laboratory co-oximeter." This implies the co-oximeter is the "expert" or standard reference, not human experts.
  • NIBP Monitor: Ground truth would be derived from the reference method used in compliance with IEC 60601-2-30 and AAMI SP10 (e.g., intra-arterial measurement or a highly accurate sphygmomanometer with expert oversight), but the document does not elaborate on the specific experts or their qualifications.
  • Clinical Thermometer: Ground truth would be derived from a highly accurate reference thermometer in the hospital setting, but the document does not elaborate on specific experts or their qualifications for establishing this ground truth.

4. Adjudication Method for the Test Set

This information is not provided for any of the tests. The measurements are compared against reference standards rather than subject to adjudication by multiple human experts.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed and is not applicable to this type of device (vital signs monitor). This document describes the performance of a medical device itself, not an AI or imaging diagnostic tool that would typically involve human reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are essentially standalone performance evaluations. The devices (pulse oximeter, NIBP, thermometer) are designed to provide direct measurements, and their accuracy is assessed against reference standards without a human interpreting an image or output that then needs to be compared to a human baseline. The "human-in-the-loop" here is the clinician using the device for patient assessment, but the device's accuracy is evaluated independently of that use.

7. The Type of Ground Truth Used

  • Pulse Oximeter (SpO2): "Arterial blood samples analyzed on a laboratory co-oximeter." This is a high-precision, gold-standard laboratory reference.
  • NIBP Monitor: Based on compliance with IEC 60601-2-30 and AAMI SP10, the ground truth would be established through defined reference methods within those standards (e.g., invasive arterial pressure measurement or a validated oscillometric reference).
  • Clinical Thermometer: Based on compliance with ASTM E 1112:00 and EN 12470-4, the ground truth would be established using a calibrated reference thermometer in a clinical setting.

8. The Sample Size for the Training Set

This information is not provided. Since these are vital signs monitors and not AI/machine learning algorithms that are "trained," the concept of a "training set" as typically understood in AI development does not directly apply here. The devices are based on established physiological principles and calibrated during manufacturing.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, the concept of a training set in the AI sense is not applicable. The devices are calibrated based on physical principles and validated against reference standards.

{0}------------------------------------------------

page 1 of 4

MAY - 2 2008

Section 5 - 510(k) Summary of Safety and Effectiveness

K07/642

  1. Submitter Information:

Mediaid, Inc. 17517 Fabrica Way #H Cerritos, CA 90703 Registration # 2087439

  1. Contact:

Jayesh Patel, CEO Telephone: 714-367-2848 Fax: 714-367-2852

  1. Date Prepared: April 1st, 2008

4. Description:

Mediaid M960 Series, Vital Signs Monitor and Mediaid M900 Series, Pulse Oximeter Trade Name: Common Name: Vital signs measurement devices Name: Non-invasive pulse Oximeter, SpO2 (21CFR870.2700), also contains NIBP measurement system (21CFR870.1130) and clinical electronic thermometer (21CFR880.2910) Classification: Class II

Mediaid M900 Series, Pulse Oximeter Trade Name: Common Name: Pulse Oximeter Name: Non-invasive pulse Oximeter, SpO2 (21CFR870.2700) Classification: Class II

The Mediaid Vital Signs monitor Pulse Oximeter Model M960P provides real time monitoring and display of ox vgen saturation of arteriolar hemoglobin (SpO2), Non Invasive Blood Pressure, pulse rate and body temperature. Where as the Mediaid Pulse Oximeter Model M900 provides real time monitoring and display of oxygen saturation of arteriolar hemoglobin (SpO2) and pulse rate only.

Noninvasive arterial oxygen saturation measurement is obtained by directing red and infra red light through a pulsating vascular bed. The pulsating arterioles in the path of the light beam cause a change in the amount of light detected by a photodiode. The pulse oximeter determines the oxygen saturation of arterial blood by measuring the ratio of transmitted red to infrared light within the pulse waveform. The non-pulsatile signal is removed electronically for the purpose of calculation. Therefore, skin, bone, and other non-pulsating substances do not interfere with the measurement of arterial oxygen saturation.

{1}------------------------------------------------

Non invasive blood pressure is intended to non invasively measure systolic, diastolic and mean arterial pressures (MAP).

Temperature- Thermometer takes patient temperature in oral, axillary or rectal mode.

5. Substantial Equivalence:

The Mediaid Pulse Oximeter Model 900 is substantially equivalent to the following devices:

  • Model 305 of Mediaid Palco marketed under K943842 .
  • Model 340 of Mediaid Palco marketed under K920066 .
  • . Model 120 of Mediaid Palco marketed under K994372

The Mediaid Vital Signs Monitor Model 960 series is substantially equivalent to the following devices:

For Pulse Oximeter :

  • Model 305 of Mediaid Palco marketed under K943842
  • Model 340 of Mediaid Palco marketed under K920066
  • Model 120 of Mediaid Palco marketed under K994372

For Ni-BP Monitor:

  • Welch Allyn Spot Ultra Vital Signs marked under K040490 .
  • Tango+Automatic Blood pressure and Oxygen Saturation Measurement System of Suntech Medical . Inc Marketed under K053209.

For Clinical Thermometer:

  • Welch Allyn Spot Ultra Vital Signs marked under K040490 .
    12

Intended Use:

Indications for Use

The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only.

{2}------------------------------------------------

WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

510(k) Number (if known): K071642

Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediad 900P Pulse Oximeter

Comparison to Predicate Device:

Mediaid Pulse Oximeter Model 900 series and pulse Oximeter module of Model 960 series uses the same theory and principle of operation as the predicate device. Design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.

Mediaid uses Suntech Medical Inc's Advantage OEM BP Module in Model 960 series. This module is compared with Tango+Automatic Blood Pressure and oxygen saturation measurement system of Suntech Medical Inc. NiBP module of Welch Allyn Spot Ultra Vital Signs monitor. It is observed that design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.

Mediaid use YSI compatible temperature probe for temperature measurement and its performance is compared with the thermometer module of Welch Allyn Spot Ultra Vital Signs monitor. It is observed that design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.

Technological Characteristics

A comparison of the technological characteristics of Model 960 Vital Signs Monitor and the predicate devices has been performed. The results of this comparison demonstrate that the Model 960 series is substantially equivalent to the marketed predicate devices in technological characteristics.

Environmental and non-clinical testing

Applicable environmental and non-clinical testing was performed per IEC60601-1-2 as well as other applicable standards and procedures. The Mediaid Pulse Oximeter Model 960 series passed all tests as per the requirements of various applicable international standards.

Performance Data & Conclusions:

{3}------------------------------------------------

Performance testing was conducted during clinical hypoxia studies conducted in an independent research lab. Mediaid Pulse Oximeter Model 900 scries readings were compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be equivalent to predicate device accuracy claims. Bench testing was performed to verify pulse rate accuracy.

For Ni-BP, clinical study was conducted by Suntech Medical Inc on their OEM module and the results show that the claimed accuracy is met as per the requirement of IEC 60601-2-30 and AAMI SP10.

Clinical Tests for Clinical thermometer is conducted on an external hospital and the results shows that the thermometer module of Model 960 series meets the claimed accuracy as per ASTM E 1112:00 and EN 12470-4

Biocompatibility, electrical safety, and EMC testing were also performed to demonstrate conformance with established industry standards.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three wing-like shapes, oriented towards the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 2 2008

Mr. Jayesh Patel Chief Executive Officer Mediaid, Incorporated 17517 Fabrica Way, Suite H Cerritos, California 90703

Re: K071642

Trade/Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediaid 900P Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, DXN Dated: April 24, 2008 Received: April 28, 2008

Dear Mr. Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Mr. Patel

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Section 4 - Indications for Use Statement

Detailed instructions are given in the user's manual.

Indications for Use

The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only.

WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

510(k) Number (if known): K071642

Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediad 900P Pulse Oximeter

Prescription Use XX · (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mi Zhang

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K071692 510(k) Number: ___

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).