K Number
K071526
Device Name
UNI-VENT 731 SERIES VENTILATORS, MODEL 731, 731EMV
Date Cleared
2008-12-10

(555 days)

Product Code
Regulation Number
868.5895
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.
Device Description
The Uni-Vent @ Model 731EMV is a portable, microprocessor controlled, electrically or pneumatically powered intensive care ventilator designed to use either oxygen (02) from a 55 psig source or ambient air using an internal compressor power to deliver a positive pressure breaths. The unit can be electrically powered from an external alternating current source, external direct current (DC) source or the internal DC battery. An intuitive point-turn-and click interface allows the operator to set and monitor ventilation in all operating environments. A series of alarms alert the user operator to all conditions that affect the ventilator's operation and/or performance and provide context sensitive help relevant to the alarm condition. Ambient air is filtered using a particulate filter or when the operating environment requires either a bacterial/viral or chemical/biologic (NATO No: 4240-01-361-1319) filter. The unit is contained in an impact resistant polycarbonate case which protects of the controls from damage and inadvertent manipulation. The Uni-Vent @ Model 731EMV internal pulse oximeter connects to the patient using noninvasive sensors to monitor oxygen saturation and pulse rate. Pulse oximeter specific alarms and instructions are presented to the operator through the user interface. Isolated DC power is provided to the pulse oximeter.
More Information

Not Found

No
The document describes a microprocessor-controlled ventilator with a user interface and alarms, but there is no mention of AI or ML capabilities in the intended use, device description, or the sections specifically checked for AI/ML terms.

Yes
The device is a ventilator, indicated for use in the management of patients with acute or chronic respiratory failure or during resuscitation, which are therapeutic interventions.

No

The device is a ventilator, which provides respiratory support, and includes a pulse oximeter for monitoring oxygen saturation and pulse rate. While it monitors physiological parameters, its primary function is therapeutic (providing ventilation) rather than diagnostic (determining medical conditions).

No

The device description clearly outlines a physical ventilator with hardware components such as a microprocessor, internal compressor, power sources (AC, external DC, internal battery), filters, a polycarbonate case, and a pulse oximeter with sensors. While it is microprocessor-controlled, it is not solely software.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description clearly states that the Model 731EMV is a ventilator. Its function is to provide positive-pressure ventilation to patients with respiratory failure. It also includes a pulse oximeter to monitor oxygen saturation and pulse rate, which are physiological measurements taken directly from the patient, not from a sample.
  • Intended Use: The intended use describes managing patients with respiratory failure by providing ventilation. This is a direct medical intervention, not a diagnostic test performed on a sample.

The device's purpose is to support a patient's breathing, which is a life-sustaining treatment, not a diagnostic procedure.

N/A

Intended Use / Indications for Use

The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.

Product codes

CBK, DQA

Device Description

The Uni-Vent @ Model 731EMV is a portable, microprocessor controlled, electrically or pneumatically powered intensive care ventilator designed to use either oxygen (02) from a 55 psig source or ambient air using an internal compressor power to deliver a positive pressure breaths. The unit can be electrically powered from an external alternating current source, external direct current (DC) source or the internal DC battery. An intuitive point-turn-and click interface allows the operator to set and monitor ventilation in all operating environments. A series of alarms alert the user operator to all conditions that affect the ventilator's operation and/or performance and provide context sensitive help relevant to the alarm condition. Ambient air is filtered using a particulate filter or when the operating environment requires either a bacterial/viral or chemical/biologic (NATO No: 4240-01-361-1319) filter. The unit is contained in an impact resistant polycarbonate case which protects of the controls from damage and inadvertent manipulation.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adolescent and adult patients weighing ≥30 kg

Intended User / Care Setting

skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.
hospitals, outside the hospital, during transport, and in austere environments

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K870861/B, K931473, K032386, K051476, K033296, K061627, K032451

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).

0

IMPACT Instrumentation, Inc.

27 Fairfield Place, West Caldwell, NJ 07006 P.O. Box 508, West Caldwell, NJ 07007-0508

Image /page/0/Picture/2 description: The image shows the word "IMPACT" in large, bold letters. The word is positioned in the center of the image, with a world map outline above and below it. There is also a heart rate line to the left of the word, and some handwritten text in the upper right corner.

DEC10 2008

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Impact, Uni-Vent® Model 731EMV 510(k) Number K071526

Manufacturer:

Impact Instrumentation, Inc. P.O. Box 508/27 Fairfield Place West Caldwell, New Jersey 07006 Phone: 973-882-1212 Fax: 973-882-4993

Contact Person:

December 10, 2008 Date Summary Prepared:

Uni-Vent 10 Model 731EMV Trade Name:

Continuous Ventilator (per 21 CFR 868.5895) Classification Name:

Mr. Leslie H. Sherman

Classification: Class II

Product Code: CBK, DQA

Device Description:

The Uni-Vent @ Model 731EMV is a portable, microprocessor controlled, electrically or pneumatically powered intensive care ventilator designed to use either oxygen (02) from a 55 psig source or ambient air using an internal compressor power to deliver a positive pressure breaths. The unit can be electrically powered from an external alternating current source, external direct current (DC) source or the internal DC battery. An intuitive point-turn-and click interface allows the operator to set and monitor ventilation in all operating environments. A series of alarms alert the user operator to all conditions that affect the ventilator's operation and/or performance and provide context sensitive help relevant to the alarm condition. Ambient air is filtered using a particulate filter or when the operating environment requires either a bacterial/viral or chemical/biologic (NATO No: 4240-01-361-1319) filter. The unit is contained in an impact resistant polycarbonate case which protects of the controls from damage and inadvertent manipulation.

Page 1 of 2

electro-medica

1

K071526
p. 2 of 2

Description of Noninvasive Pulse Oximeter

The Uni-Vent @ Model 731EMV internal pulse oximeter connects to the patient using noninvasive sensors to monitor oxygen saturation and pulse rate. Pulse oximeter specific alarms and instructions are presented to the operator through the user interface. Isolated DC power is provided to the pulse oximeter.

Intended Use:

The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positivepressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.

Substantial Equivalence:

The Impact, Model 731EMV, is substantially equivalent to the predicate devices listed below:

Predicate Devices:

    1. Impact Uni-Vent- Eagle, Model 754. 510(k) #K870861/B and K931473.
    1. Impact Uni-Vent Model 730. 510(k) #K032386
    1. Impact Uni-Vent Model 73X. 510(k) #K051476.
    1. Masimo SET RAD 5 MS-11 PCB 510(k) #K033296, Masimo Corp, Irvine, CA
    1. Versamed SmartVent™ 201, 510(k) #K061627, Versamed Medical Systems, Inc., Pearl River, NY
    1. Vela Bird Products, Palm Springs, CA, 510(k) # K032451

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 0 2008

Mr. Leslie H. Sherman President PACT Instrumentation, Incorporated 27 Fairfield Place West Caldwell, New Jersey 07006

Re: K071526

Trade/Device Name: Uni-Vent ® Model 731 EMV Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, DQA Dated: December 3, 2008 Received: December 3, 2008

Dear Mr. Sherman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

3

Page 2 - Mr. Sherman

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sybil y. Mckain PhD.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

INDICATIONS FOR USE

510(k) Number (if known): K071526 Device Name: Uni-Vent @ Model 731EMV

Indications for Use:

The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.

X Prescription Use (Per 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Suage y. Michie Ous

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:_大07 15 こん

Page 1 of 1