(127 days)
The SAVe is intended for use on adults that need ventilatory support. Specifically, the SAVe is intended to provide short term ventilatory support for individuals during CPR or when positive-pressure ventilation is required to manage acute respiratory failure. The SAVe device is appropriate for individuals that weigh at least 45 kilograms. It is intended to be used in field hospitals, transport and prehospital environments.
The SAVe is a time cycled ventilator that delivers a set tidal volume. If, at any point, an upper threshold pressure is reached, the device immediately triggers the exhale mode. The SAVe consists of an internal pump (blower), rechargeable sealed lead acid battery, circuit board, manifold, face plate, EMI shielded upper and lower enclosure, air intake port with debris filter, patient port, pressure port, supplemental air / oxygen port, electric port for recharging and / or running, charger / power supply, patient circuit, pressure tubing, oxygen tubing and an activation switch. The device has audio and visual alarms that trigger for low pressure, disconnect, blockage, low battery, high temperature or internal failure. The device not only alarms for stacked breaths but also cuts off the pump until adequate pressure has been released before triggering the next breath. The SAVe comes in three distinct models. Each has a set tidal volume and respiratory rate. Model 550x10 delivers 550 mL of air 10 times per minute. Model 600x10 delivers 600 mL of air 10 times per minute and Model 600x12 delivers 600 mL of air 12 times per minute.
The provided text describes a 510(k) submission for a medical device called the "Simplified Automated Ventilator - SAVe." The submission focuses on demonstrating substantial equivalence to a predicate device, the Impact Uni-Vent 754.
Acceptance Criteria and Device Performance:
The document does not explicitly state numerical "acceptance criteria" as would be found in a performance goal for a diagnostic device. Instead, the performance evaluation is based on demonstrating substantial equivalence to the predicate device, the Impact Uni-Vent 754.
The "reported device performance" is a qualitative statement of equivalence to the predicate device.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Equivalence to Predicate Device (Impact Uni-Vent 754) in: | The SAVe (at its fixed settings) is as safe and effective as the Uni-Vent 754. |
| Fixed tidal volume and respiratory rate capability. | The SAVe's fixed tidal volume and respiratory rate are a subset of the Uni-Vent 754's capabilities. |
| Intended Use and Indications for Use. | The SAVe's intended use and indications for use are a subset of the predicate device. |
| Technological Characteristics and Principles of Operation. | Both devices use similar technological characteristics and principles of operation. Minor technological differences raise no new safety/effectiveness issues. |
| Safety and Effectiveness based on performance data. | Bench generated performance data demonstrate that the SAVe is as safe and effective as the Uni-Vent 754 at the prescribed settings. |
| Compliance with FDA's Draft Reviewer Guidance for Ventilators (July 1995). | The SAVe was thoroughly tested as per all applicable parts of the FDA's Draft Reviewer Guidance for Ventilators dated July 1995. |
| Electromagnetic compatibility and air quality standards. | External tests by third parties confirmed the SAVe met various standards for electromagnetic compatibility and air quality. The SAVe functioned as intended, and all observed results were as expected. |
Study Details:
The provided text describes a substantial equivalence study for a medical device (ventilator), not a diagnostic algorithm. Therefore, many of the requested elements (like sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth types for test/training sets, and training set sample size) are not applicable or not detailed in this type of submission.
Here's an analysis based on the information provided:
- Sample size used for the test set and the data provenance: Not explicitly stated. The evaluation was primarily based on bench tests and potentially external third-party tests. The provenance of this test data is implied to be from the manufacturer (Automedx) and external testing agencies.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For a ventilator, ground truth isn't established by experts in the same way as for a diagnostic imaging algorithm. Performance is assessed against engineering specifications, simulated physiological conditions, and comparison to a predicate device.
- Adjudication method for the test set: Not applicable. Performance is measured against physical parameters and predicate device performance.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a ventilator, not a diagnostic AI algorithm for human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device is a "Simplified Automated Ventilator," implying it operates autonomously in its core function of delivering ventilation. Performance data was generated from bench tests, which would assess its standalone mechanical and electronic function.
- The type of ground truth used:
- Engineering specifications/performance metrics: The device's ability to deliver preset tidal volumes and respiratory rates, trigger exhale mode at an upper threshold pressure, and its alarm functions.
- Predicate device performance: Comparative data showing the SAVe performs "as safe and effective as the Uni-Vent 754" at the prescribed settings.
- Compliance with standards: Functioning as intended according to standards for electromagnetic compatibility and air quality.
- The sample size for the training set: Not applicable. Ventilators are typically designed and calibrated based on engineering principles and preclinical testing, not "trained" on a dataset in the way a machine learning algorithm is.
- How the ground truth for the training set was established: Not applicable for this type of device. The design specifications and performance parameters derive from medical requirements for ventilation and the characteristics of the predicate device.
{0}------------------------------------------------
Contact: James Evans President & CEO 12321 Middlebrook Rd Ste 150 Germantown, Md 20874 (301) 916-9508
Submission Date: May 1, 2007
Manufacturer: Automedx, Incorporated 12321 Middlebrook Rd. Ste 150 Germantown, MD 20874 P: (301) 916-9508 F: (301) 916-7155
Classification Name Ventilator, Emergency, Powered, (Resuscitator)
Regulation Number 868.5925
Predicate Device Impact Uni-Vent 754 (K931473)
Intended Use / Indications for Use
The SAVe is intended for use on adults that need ventilatory support. Specifically, the SAVe is intended to provide short term ventilatory support for individuals during CPR or when positive-pressure ventilation is required to manage acute respiratory failure. The SAVe device is appropriate for individuals that weigh at least 45 kilograms. It is intended to be used in field hospitals, transport and pre-hospital environments.
Technological Characteristics
The SAVe is a time cycled ventilator that delivers a set tidal volume. If, at any point, an upper threshold pressure is reached, the device immediately triggers the exhale mode.
The SAVe consists of an internal pump (blower), rechargeable sealed lead acid battery, circuit board, manifold, face plate, EMI shielded upper and lower enclosure, air intake port with debris filter, patient
26
JAN 1 8 2 ...
Received
FDA CDKA THE
{1}------------------------------------------------
port, pressure port, supplemental air / oxygen port, electric port for recharging and / or running, charger / power supply, patient circuit, pressure tubing, oxygen tubing and an activation switch.
The device has audio and visual alarms that trigger for low pressure, disconnect, blockage, low battery, high temperature or internal failure. The device not only alarms for stacked breaths but also cuts off the pump until adequate pressure has been released before triggering the next breath.
The SAVe comes in three distinct models. Each has a set tidal volume and respiratory rate. Model 550x10 delivers 550 mL of air 10 times per minute. Model 600x10 delivers 600 mL of air 10 times per minute and Model 600x12 delivers 600 mL of air 12 times per minute.
Performance Data
The SAVe was thoroughly tested as per all applicable parts of the FDA's Draft Reviewer Guidance for Ventilators dated July 1995. Its performance vis-à-vis the Impact Uni-Vent 754 showed substantial equivalence to the predicate device. The results of that test can be found in Appendix I. In addition to the bench tests performed, external tests were conducted by third parties to see if the SAVe met various standards for electromagnetic compatibility and air quality. In all instances, the SAVe functioned as intended and all observed results were as expected.
Substantial Equivalence
The SAVe has a fixed tidal volume and respiratory rate that are a subset of the capabilities of the Uni-Vent 754. At these settings, the SAVe is as safe and effective as the Uni-Vent 754. The SAVe's intended use and indications for use are also a subset of the predicate device. Both devices use similar technological characteristics and principles of operation. The minor technological differences between the SAVe and its predicate device raise no new issues of safety or effectiveness. Bench generated performance data demonstrate that the SAVe is as safe and effective as the Uni-Vent 754 at the prescribed settings. Thus, the SAVe is substantially equivalent to the Uni-Vent 754.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 6 2007
Mr. James D. Evans President and Chief Executive Officer Automedx, Incorporated 12321 Middlebrook Road, Suite 150 Germantown, Maryland 20874
Re: K071221
Trade/Device Name: Simplified Automated Ventilator - SAVe (Models 550x10, 600x10, 600x12) Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: August 29, 2007 Received: August 29, 2007
Dear Mr. Evans:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -Mr. Evans
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Snute Y. Michim Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Simplified Automated Ventilator - SAVe (Models 550x10, 600x10, 600x12)
The SAVe is intended for use on adults that need ventilatory support. Specifically, the SAVe is intended to provide short term ventilatory support for individuals during CPR or when positive-pressure ventilation is required to manage acute respiratory failure. The SAVe device is appropriate for individuals that weigh at least 45 kilograms. It is intended to be used in field hospitals, transport and prehospital environments.
Prescription Use _X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CBRH, Office of Device Evaluation (ODE)
Suette G. Mrohin D.mD.
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K071221
§ 868.5925 Powered emergency ventilator.
(a)
Identification. A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).