(23 days)
For in vitro diagnostic use only. VITROS Chemistry Products CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma on the DT60/DT60 II analyzer. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
For in vitro diagnostic use only. VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK, Cl, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems.
The VITROS CRSC assay is performed using the VITROS Chemistry Products CRSC DT Slides and the VITROS Chemistry Products DT Calibrator Kit on VITROS DT60/DT60II Chemistry Systems.
The VITROS CRSC Slide is a multilayered, analytical element coated on a polyester support.
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glvcine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product.
Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting rate of change in reflection density is proportional to the concentration of creatinine in the sample.
Once a calibration has been performed for each slide lot, creatinine concentration in unknown samples can be determined using the softwareresident rate math model and the change in reflectance calculated for each unknown test slide.
VITROS DT Calibrator Kit contains four levels of lyophilized standards with corresponding diluents. The standards are prepared from processed bovine serum to which organic analytes, electrolytes, stabilizers, and preservatives have been added. The diluents are prepared from processed water to which inorganic salts have been added. Once reconstituted, the appropriate standards are used to calibrate VITROS DT60/DT60 II Chemistry Systems for the quantitative measurement of creatinine in serum and plasma.
Calibration of the VITROS CRSC assay requires the use of bottles 1, 2 and 4 of the VITROS DT Calibrator Kit.
The provided document is a 510(k) summary for the VITROS Chemistry Products CRSC DT Slides and VITROS Chemistry Products DT Calibrator Kit. This document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with specific acceptance criteria and performance data in the format of a clinical trial for a standalone diagnostic device.
Therefore, much of the requested information regarding "acceptance criteria," "study that proves the device meets the acceptance criteria," "sample size for the test set," "data provenance," "number of experts," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for the training set," and "how ground truth for the training set was established" cannot be found in this document.
The document details the technical characteristics and intended use of the modified device and compares them to the predicate device to establish substantial equivalence. It does not present a de novo clinical study with specific performance metrics and acceptance thresholds.
However, based on the information available, I can extract the following:
1. A table of acceptance criteria and the reported device performance:
The document doesn't explicitly state "acceptance criteria" in a numerical or categorical format with direct performance results for the modified device. Instead, it argues for "substantial equivalence" to a predicate device. The performance is assessed by demonstrating consistency in intended use, calibration traceability, and similar principles of operation.
| Device Characteristic | Predicate Device (Original VITROS CRSC assay - K875191) | Modified Device (VITROS CRSC assay) |
|---|---|---|
| Intended Use | For in vitro diagnostic use only. VITROS CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma. For in vitro diagnostic use only. VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK, Cl, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems. | No change from predicate. For in vitro diagnostic use only. VITROS CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma. For in vitro diagnostic use only. VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK, Cl, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems. |
| Calibration traceability | Traceable to Certified NIST (National Institute of Standards and Technology) Reference Material SRM® (Standard Reference Material) 914a. | The values assigned to the VITROS Chemistry Products DT Calibrator Kit for Creatinine are traceable to a Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method¹ and NIST SRM®914, creatinine standard reference material. |
| Reference Interval (Serum) | Males: 0.8 – 1.5 mg/dL; Females: 0.7 – 1.2 mg/dL | Males: 0.7 – 1.3 mg/dL; Females: 0.6 – 1.0 mg/dL |
| Reportable range (Serum) | 0.1 – 16.5 mg/dL | 0.2-11.2 mg/dL |
| Sample type | Serum and Plasma | No change from predicate. Serum and Plasma. |
| Basic principle | Colorimetric rate | No change from predicate. Colorimetric rate. |
| Instrumentation | VITROS DTSC/DTSCII Module | No change from predicate. VITROS DTSC/DTSCII Module. |
The conclusion states: "Equivalence was demonstrated using manufactured reagents along with quality control fluids, proficiency samples, and human serum samples with measured creatinine values spanning the assay range." This implies that the performance of the modified device, when evaluated against these samples, was found to be consistent with or equivalent to the predicate device. However, specific numerical acceptance criteria (e.g., % agreement, bias limits) are not provided.
2. Sample sized used for the test set and the data provenance:
- Sample size: The document mentions "human serum samples with measured creatinine values spanning the assay range," but it does not specify the exact number of samples used for the test set.
- Data provenance: Not explicitly stated, but it would typically be from laboratory testing using the device. No country of origin or retrospective/prospective nature is specified, but regulatory submissions like this generally involve prospective data collection for equivalence studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device is an in vitro diagnostic assay, and "ground truth" for creatinine measurement is typically established through reference methods or clinical laboratory assays, not expert consensus in the way image analysis or clinical diagnosis might be. The document refers to "measured creatinine values" but doesn't detail how these were established beyond mentioning traceability to NIST SRM®914a and GC/IDMS.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable for this type of IVD device. Adjudication methods are typically used in studies involving human interpretation or subjective assessments.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic tool that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device (VITROS CRSC assay with VITROS Chemistry Products CRSC DT Slides) is an automated system for measuring creatinine. Its performance is inherently "standalone" in the sense that the measurement is performed by the instrument and reagents, not by human interpretation of raw data. The study mentioned to demonstrate equivalence would thus evaluate its standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for creatinine measurement in the equivalence study would be established by reference methods or highly accurate laboratory measurements, as indicated by the mention of traceability to NIST SRM®914a and a Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method. These are highly precise and accurate analytical techniques considered gold standards for creatinine measurement.
8. The sample size for the training set:
Not applicable. This is a chemical assay, not a machine learning or AI model that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. As above, there is no "training set."
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Ortho-Clinical Diagnostics. The logo consists of a symbol on the left and the text "Ortho-Clinical Diagnostics" on the right. The text is underlined. The symbol on the left appears to be a stylized graphic.
100 Indigo Creek Drive Rochester, New York 14626-5101
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _ Ko 7/2/6
-
- Submitter Ortho-Clinical Diagnostics, Inc. name, ad-100 Indigo Creek Drive MC00881 dress, con-Rochester, New York 14626-5101 tact Phone: (585) 453-3143 FAX: (585) 453-3368 Email: mbyrne3@ocdus.jnj.com
MAY 2 5 2007
- Submitter Ortho-Clinical Diagnostics, Inc. name, ad-100 Indigo Creek Drive MC00881 dress, con-Rochester, New York 14626-5101 tact Phone: (585) 453-3143 FAX: (585) 453-3368 Email: mbyrne3@ocdus.jnj.com
Michael M. Byrne Contact Person:
- May 1, 2007 2. Preparation
- date
- Trade or Proprietary Name: 3. Device name VITROS Chemistry Products CRSC DT Slides VITROS Chemistry Products DT Calibrator Kit
Common Name: VITROS CRSC assay Classification Name: Creatinine test system (21 CFR 862.1225).
VITROS DT Calibrator Kit Common Name: Classification Name: Calibrator (21 CFR 862.1150)
-
- Predicate The VITROS CRSC assay (modified device) is substantially equivalent to the device VITROS CRSC assay (original device). The FDA cleared the VITROS Chemistry Products CRSC DT Slides on and VITROS Chemistry Products DT Calibrator Kit on February 24, 1988 (K875191) under the product names:
KODAK EKTACHEM DT SLIDES (CRSC) and
- Predicate The VITROS CRSC assay (modified device) is substantially equivalent to the device VITROS CRSC assay (original device). The FDA cleared the VITROS Chemistry Products CRSC DT Slides on and VITROS Chemistry Products DT Calibrator Kit on February 24, 1988 (K875191) under the product names:
KODAK EKTACHEM DT Calibrator Kit
Note: With the purchase of KODAK Clinical Products Division by Johnson and Johnson, the product branding was revised to VITROS Chemistry Products CRSC DT Slides and VITROS Chemistry Products DT Calibrator Kit.
The most recent FDA clearance for the VITROS Chemistry Products DT Calibrator Kit was March 21, 1994 (K934071).
Continued on next page
SPECIAL 510(K) SUBMISSION
VITROS CRSC DT Slides and VITROS DT Calibrator Kit
{1}------------------------------------------------
510(k) Summary, Continued
scription
The VITROS CRSC assay is performed using the VITROS Chemistry 5. Device de-Products CRSC DT Slides and the VITROS Chemistry Products DT Calibrator Kit on VITROS DT60/DT60II Chemistry Systems.
The VITROS CRSC Slide is a multilayered, analytical element coated on a polyester support.
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glvcine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product.
Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting rate of change in reflection density is proportional to the concentration of creatinine in the sample.
Once a calibration has been performed for each slide lot, creatinine concentration in unknown samples can be determined using the softwareresident rate math model and the change in reflectance calculated for each unknown test slide.
VITROS DT Calibrator Kit contains four levels of lyophilized standards with corresponding diluents. The standards are prepared from processed bovine serum to which organic analytes, electrolytes, stabilizers, and preservatives have been added. The diluents are prepared from processed water to which inorganic salts have been added. Once reconstituted, the appropriate standards are used to calibrate VITROS DT60/DT60 II Chemistry Systems for the quantitative measurement of creatinine in serum and plasma.
Calibration of the VITROS CRSC assay requires the use of bottles 1, 2 and 4 of the VITROS DT Calibrator Kit.
- Device in-For in vitro diagnostic use only. tended use VITROS Chemistry Products CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma.
For in vitro diagnostic use only.
VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK, Cl, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems.
SPECIAL 510(K) SUBMISSION
Continued on next page
13
Ortho-Clinical Diagnostics, Inc.
VITROS CRSC DT Slides and VITROS DT Calibrator Kit
{2}------------------------------------------------
510(k) Summary, Continued
- The VITROS CRSC assay (modified device) is substantially equivalent to the 7. Comparison to predicate predicate, VITROS CRSC assay (original), which was cleared by the FDA device (K875191) for in vitro diagnostic use.
| DeviceCharacteristic | Modified DeviceVITROS CRSC assay(modified) | Predicate DeviceVITROS CRSC assay (original) |
|---|---|---|
| Intended Use | No change | For in vitro diagnostic use only.VITROS CRSC DT Slides quantita-tively measure creatinine concentrationin serum and plasma. |
| For in vitro diagnostic use only.VITROS Chemistry Products DT Calibrator Kit is specially formulated for |
Tahle 1 I ist of A y Characteristics: Comparison to Predicate Devic
| (modified) | ||
|---|---|---|
| Intended Use | No change | For in vitro diagnostic use only.VITROS CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma.For in vitro diagnostic use only.VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBIL, BUN/UREA, Ca, CHOL, CK, Cl, CO2, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems. |
| Calibration traceability | The values assigned to the VITROS Chemistry Products DT Calibrator Kit for Creatinine are traceable to a Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method¹ and NIST SRM®914, creatinine standard reference material. | Traceable to Certified NIST (National Institute of Standards and Technology) Reference Material SRM® (Standard Reference Material) 914a. |
| Reference Interval (Serum) | Males: 0.7 – 1.3 mg/dLFemales: 0.6 – 1.0 mg/dL | 0.8 – 1.5 mg/dL0.7 – 1.2 mg/dL |
| Reportable rangeSerum | 0.2-11.2 mg/dL | 0.1 – 16.5 mg/dL |
| Sample type | No change | Serum and Plasma |
| Basic principle | No change | Colorimetric rate |
| Instrumentation | No change | VITROS DTSC/DTSCII Module |
Continued on next page
{3}------------------------------------------------
510(k) Summary, Continued
The information presented in the pre-market notification demonstrates that the 8. Conclusions performance of the VITROS CRSC assay (modified device) for use with human serum and plasma is substantially equivalent to the cleared predicate device.
Equivalence was demonstrated using manufactured reagents along with quality control fluids, proficiency samples, and human serum samples with measured creatinine values spanning the assay range.
The information presented in the premarket notification provides a reasonable assurance that the VITROS CRSC assay (modified device) for use with human serum and plasma is safe and effective for the stated intended use.
SPECIAL 510(K) SUBMISSION
VITROS CRSC DT Slides and VITROS DT Calibrator Kit
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY 2 5 2007
Ortho-Clinical Diagnostics, Inc. c/o Michael M. Byrne, RAC Regulatory Affairs Associate 100 Indigo Creek Drive Rochester, NY 14626-5101
Re: K071216
Trade/Device Name: VITROS Chemistry Products CRSC DT Slides, VITROS Chemistry Products DT Calibrator Kit Regulation Number: 21 CFR §862.1215 Regulation Name: Creatinine Test System Regulatory Class: Class II Product Code: JFY, JIX Dated: May 1, 2007 Received: May 2, 2007
Dear Mr. Byrne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K071216 |
|---|---|
| Device Name: | VITROS Chemistry Products CRSC DT Slides |
| Indications for Use: | For in vitro diagnostic use only. VITROS Chemistry Products CRSC DT Slides quantitatively measure creatinine concentration in serum and plasma on the DT60/DT60 II analyzer.Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. |
| Prescription Use | X |
| (Part 21 CFR 801 Subpart D) | AND/OR |
| Over-The-Counter Use | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol C Benson
Sign-Off
Page 1 of 2
୍ରାଚe of In Vitro Diagnostic Device los duation and Safety
{7}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
VITROS Chemistry Products DT Calibrator Kit
Indications for Use:
For in vitro diagnostic use only. VITROS Chemistry Products DT Calibrator Kit is specially formulated for use as calibrators for ALB, ALKP, ALT, AMYL, AST, TBIL, NBL, BUN/UREA, Ca, CHOL, CK, Cl, C02, CREA, CRSC, Fe, GGT, GLU, HDLC, K+, LAC, LDH, LIPA, Mg, Na+, NH3, PHOS, TP, TRIG, urCR, and URIC on VITROS DT Chemistry Systems.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
1071216
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol C. Benson
ion Sign-O
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 2 of 2
§ 862.1225 Creatinine test system.
(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.