(136 days)
Not Found
No
The device description and performance studies focus on the mechanical and fluid dynamics of a safety fistula needle, with no mention of AI or ML technologies.
No
The device is described as a "blood access device for blood purification and for other treatments requiring an extracorporeal circuit," which facilitates a medical procedure but does not directly deliver therapy or treat a condition itself.
No
The device is a blood access device used for blood purification and treatments requiring extracorporeal circuits, and it has a safety feature to prevent needlesticks. It does not mention any function related to diagnosing a condition or disease.
No
The device description clearly outlines physical components like needles, tubing, and adaptors, and the performance studies focus on the mechanical properties of these physical components. There is no mention of software as a component or function of the device.
Based on the provided text, the NIPRO SafeTouch TULIP™ Safety Fistula Needle is not an IVD (In Vitro Diagnostic) device.
Here's why:
- Intended Use: The intended use is described as a "blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood." This describes a device used on the patient to access their blood for treatment, not a device used to test samples outside the body to diagnose a condition.
- Device Description: The description focuses on the physical components of a needle and tubing system for accessing blood vessels. It mentions it's a "non-implanted blood access device" and operates on the "principles of a blood access device."
- Lack of IVD Characteristics: The text does not mention any of the typical characteristics of an IVD, such as:
- Analyzing biological samples (blood, urine, tissue, etc.)
- Detecting or measuring specific substances in samples
- Providing information for diagnosis, monitoring, or screening of diseases
- Reagents, calibrators, or controls
The device is clearly intended for direct patient use to facilitate medical treatment, which falls outside the scope of In Vitro Diagnostics.
N/A
Intended Use / Indications for Use
The NIPRO SafeTouch TULIP™ Safety AVF Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood. Secondly, it is designed with an active sharp safety feature requiring physical action by the clinician to aid in the prevention of accidental needlesticks. The compatibility of available configurations is the responsibility of the physician in charge.
The NIPRO SafeTouch TULIP™ Safety Fistula Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood. Secondly, it is designed with an anti-stick needle protector requiring physical action by the clinician to aid in the prevention of accidental needlesticks. The compatibility of available configurations is the responsibility of the physician in charge.
Product codes (comma separated list FDA assigned to the subject device)
FIE, MPB, KOC, LLB
Device Description
The NIPRO SafeTouch TULIP™ Safety Fistula Needle is a sterile, single use, safety AVF needle. It consists of an arterial and venous adaptor, flexible tube and needle with an active sharps safety feature (non-implanted blood access device) as described in 21 CFR 876.5540.
The NIPRO SafeTouch TULIP™ Safety Fistula Needle includes 2 basic types of designs; fixed wing type (stationary) and turnable wing type (rotating). These two designs are offered in 64 configurations with options that include needle gauge, needle length, type of needle (with or without backeye), and tubing length.
The integrated sharps injury prevention feature requires physical action by the clinician to activate and is designed to cover the cannula after treatment. Correct uses of this anti-stick feature will eliminate accidental needlesticks.
These devices operate on the principles of a blood access device. They are sterile, single use only, non-toxic and non-pyrogenic.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
clinician
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests submitted:
The results of biocompatibility data support the equivalence of the predicate device and include sterility, bacterial endotoxin, systemic injection. intracutaneous reactivity, hemolysis, and implantation testing. Performance testing was also conducted to verify:
- The tensile, flexural, and elongation strength of the materials.
- Force to attach and detach connections
- Rate of fluid flow simulating extremes of pressure
- Force to activate and deactivate the safety feature
- Strength of joints, bonds, connections, hinges, valves, locking mechanisms.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
0
Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a combination of alphanumeric characters and symbols. The text includes "K071145" on the top line, followed by "pg 1 of 2" on the second line. The handwriting is somewhat stylized, with some characters connected and others distinct.
Image /page/0/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized symbol resembling two interlocking letter "N" shapes on the left, followed by the word "NIPRO" in bold, sans-serif font on the right. The logo is black and white.
3150 N.W. 107 Avenue Miami, Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
SEP - 7 2007
510(k) Summary of Safety and Effectiveness for NIPRO SafeTouch TULIP™ Safety Fistula Needle
807.92(a)(1)
Contact Person:
Jessica Oswald Regulatory Affairs Specialist
Date of summary preparation: April 10, 2007
807.92(a)(2)
:
Trade Name: NIPRO SafeTouch TULIP™ Safety Fistula Needle Common Name: Safety AVF Needle Classification Name: Blood access device and accessories Product Code: FIE
807.92(a)(3)
Legally marketed substantial equivalent device: Medisystems Masterguard™ AVF Needle (K932074) Nipro AVF Needle (K955182)
807.92(a)(4)
Description of device:
The NIPRO SafeTouch TULIP™ Safety Fistula Needle is a sterile, single use, safety AVF needle. It consists of an arterial and venous adaptor, flexible tube and needle with an active sharps safety feature (non-implanted blood access device) as described in 21 CFR 876.5540.
The NIPRO SafeTouch TULIP™ Safety Fistula Needle includes 2 basic types of designs; fixed wing type (stationary) and turnable wing type (rotating). These two designs are offered in 64 configurations with options that include needle gauge, needle length, type of needle (with or without backeye), and tubing length.
The integrated sharps injury prevention feature requires physical action by the clinician to activate and is designed to cover the cannula after treatment. Correct uses of this anti-stick feature will eliminate accidental needlesticks.
These devices operate on the principles of a blood access device. They are sterile, single use only, non-toxic and non-pyrogenic.
1
K071145
pg 2 of 2
Image /page/1/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized, intertwined double helix symbol to the left of the word "NIPRO" in bold, sans-serif font. The double helix symbol is black with white space in between the curves, and the word "NIPRO" is also black.
PRO MEDICAL CORPORATION 3150 N.W. 107 Avenue Miami. Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
807.92(a)(5)
Indications for Use:
The NIPRO Safetouch TULIP™ Safety AVF Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood. Secondly, it is designed with an active sharp safety feature requiring physical action by the clinician to aid in the prevention of accidental needlesticks. The compatibility of available configurations is the responsibility of the physician in charge.
807.92(a)(6)
Comparison of technological characteristics:
The NIPRO Safetouch TULIP™ Safety AVF Needle is substantially equivalent to the Medisystems Masterguard™ AVF Needle (K932074) with regard to intended use, operational characteristics, labeling, and overall performance characteristics. It is identical to the NIPRO AVF needle (K955182), with the addition of the TULIP needle protector, in regards to materials of construction.
807.92(b)(1)
Non-clinical tests submitted:
The results of biocompatibility data support the equivalence of the predicate device and include sterility, bacterial endotoxin, systemic injection. intracutaneous reactivity, hemolysis, and implantation testing. Performance testing was also conducted to verify:
-
- The tensile, flexural, and elongation strength of the materials.
-
- Force to attach and detach connections
-
- Rate of fluid flow simulating extremes of pressure
-
- Force to activate and deactivate the safety feature
-
- Strength of joints, bonds, connections, hinges, valves, locking mechanisms.
807.92(b)(3)
Conclusions drawn from non-clinical and clinical tests:
The results of the performance testing and the comparison of technological characteristics demonstrate that the NIPRO Safetouch TULIP™ Safety AVF Needle performs equivalent to the predicate devices and is safe and effective when used as intended.
2
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings spread, symbolizing strength and freedom. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
SEP - 7 2007
Ms. Jessica Oswald Regulatory Affairs Specialist NIPRO Medical Corporation 3150 N.W. 107 Avenue MIAMI FL 33172
Re: K071145
Trade/Device Name: NIPRO SafeTouch TULIP™ Safety Fistula Needle Regulation Number: 21 CFR §876.5540 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: MPB, KOC and LLB Dated: August 15, 2007 Received: August 17, 2007
Dear Ms. Oswald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top, and the word "Centennial" is at the bottom. The letters "FDA" are in the center of the logo. There are three stars below the letters "FDA". The logo is surrounded by a dotted border.
Protecting and Promoting Public Health
3
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-011 |
---|---|---|
21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-011 |
21 CFR 894.xxx | (Radiology) | 240-276-012 |
Other | 240-276-010 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Ko71/45
Indications for Use
510(k) Number: K071145
Device Name: NIPRO SafeTouch TULIP™ Safety Fistula Needle
Indications for Use:
The NIPRO SafeTouch TULIP™ Safety Fistula Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood. Secondly, it is designed with an anti-stick needle protector requiring physical action by the clinician to aid in the prevention of accidental needlesticks. The compatibility of available configurations is the responsibility of the physician in charge.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K071145
NIPRO SafeTouch TULIP™ Safety Fistula Needle
-4.1-