K Number
K070572
Date Cleared
2007-06-26

(118 days)

Product Code
Regulation Number
876.5540
Panel
GU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ash Advance™ Hemodialysis Catheter and Procedure Kit is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. This catheter is indicated for > 30 days long-term placement.

Device Description

The Ash Advance™ Hemodialysis Catheter and Procedure Kit is a dual lumen, 15FR catheter available in multiple lengths. The catheter lumens are D-shaped and made from radiopaque Carbothane. The distal end design is a fixed length split-tip, without rom radiopaque Caro venous lumen extends past the arterial lumen, and includes a guidewire slit for insertion by the optional guidewire placement technique. The proximal device contains a fixed polyester cuff, an integrated hub, suture wing, and extension set with color coded occlusion clamps and luer connectors (red and blue for the arterial and venous lumens respectively). The lumen priming volumes are printed on the clamps. The procedure kit includes the necessary accessories to correctly insert the catheter.

AI/ML Overview

This document describes the Ash Advance™ Hemodialysis Catheter and Procedure Kit, a medical device for long-term vascular access in hemodialysis and apheresis. The 510(k) summary (K070572) focuses on non-clinical testing for substantial equivalence to predicate devices, rather than a clinical study evaluating specific performance metrics against pre-defined acceptance criteria. Therefore, much of the requested information regarding acceptance criteria and study design for performance evaluation of AI/algorithm-only performance is not available in these documents.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Tensile StrengthSatisfied
Joint StrengthSatisfied
LeakageSatisfied
Flow RateSatisfied
Kit Component CompatibilitySatisfied
Reliable performance as a conventional hemodialysis catheter for obtaining blood access for hemodialysis and apheresis"Results of testing show that the Ash Advance™ Hemodialysis Catheter and Procedure Kit can reliably perform as a conventional hemodialysis catheter for obtaining blood access for hemodialysis and apheresis."

Explanation:
The document does not explicitly state numerical acceptance criteria for these tests. Instead, it broadly indicates that "The safety and performance... have been substantiated through extensive non-clinical testing, including [listed tests]." The conclusion is that the device "can reliably perform" and raises "no new questions of safety or effectiveness," implying that the results met internal or industry standards for these non-clinical tests.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The testing described is "non-clinical," which typically refers to bench testing and in-vitro studies, not human clinical trials. Thus, there is no "data provenance" in terms of country of origin of data or retrospective/prospective study design as would apply to clinical performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable/provided. The evaluation was based on non-clinical engineering and material tests, not on expert assessment of clinical cases.

4. Adjudication Method for the Test Set

This information is not applicable/provided. As the testing was non-clinical, there was no need for expert adjudication of results in the way it would be applied to clinical outcomes or image interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. An MRMC study is a clinical study involving human readers evaluating cases, often with and without AI assistance. This document describes non-clinical testing of a physical medical device (catheter) and does not involve AI or human image interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No. This device is a physical medical catheter, not an algorithm or AI system. Therefore, standalone algorithm performance is not applicable.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests would be defined by engineering specifications, industry standards, and regulatory requirements for medical devices of this type. For example, for tensile strength, the ground truth would be a defined minimum tensile strength value that the catheter must withstand. For leakage, it would be a maximum allowable leakage rate.

8. The Sample Size for the Training Set

This information is not applicable/provided. The document describes non-clinical testing of a physical device, not an AI or machine learning model that would require a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided. As above, there is no training set for a physical medical catheter.

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.