K Number
K070320

Validate with FDA (Live)

Date Cleared
2007-06-07

(125 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.

The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.

Device Description

The Edwards Aquarius System needs to be used in conjunction with a tubing set and a filter to provide Hemofiltration treatment to the patient.

The Aquarius system is an Automated Fluid Balance Monitor, designed to be used with various extracorporeal treatments in the field of renal replacement therapies or plasma therapies. All therapies must be prescribed by a physician.

The Aquarius system is divided into three circuits: the extracorporeal (blood) circuit, the substitution/dialysate circuit and the filtrate circuit.

Toxic substances are removed by filters and clean blood is returned to the patient.

The Aquarius system allows the patient to be positioned left or right of the instrument.

The Aquarius system uses two scales to accurately measure and precisely balance filtration and substitution volumes.

Heparin may be supplied to the extracorporeal circuit via an anticoagulant pump (Heparin pump). The prescribing physician may select continuous or intermittent options.

The Aquarius protective system is designed as a 2-channel system to protect the patient from foreseeable danger.

At the back of the scale system a removable hand-crank is mounted. This can be used to manually turn the blood pump in case a pump stops.

The Aquarius system is portable. It has a wheeled base connected with a handle to move or carry the Aquarius.

AI/ML Overview

Acceptance Criteria and Device Performance for Edwards Aquarius System

The Edwards Aquarius System is an Automated Fluid Balance Monitor designed for extracorporeal treatments in renal replacement therapies or plasma therapies. The device's performance was evaluated against a set of specifications to demonstrate substantial equivalence to predicate devices (Baxter Accura System, Gambro Prisma System, B. Braun Diapact).

1. Acceptance Criteria and Reported Device Performance

The provided document details various performance specifications for the Edwards Aquarius System. The study supporting these claims is described as "Bench Testing."

ParameterAcceptance Criteria (Low Blood Flow)Reported Device Performance (Low Blood Flow)Acceptance Criteria (Regular Blood Flow)Reported Device Performance (Regular Blood Flow)
Fluids Management
Pump accuracy±5%±5% (Implied, as reported performance matches criteria)±5%±5% (Implied, as reported performance matches criteria)
Blood Flow10 to 200 ml/min by 2 ml/min10 to 200 ml/min by 2 ml/min30 to 450 ml/min by 10 ml/min30 to 450 ml/min by 10 ml/min
Pre-dilution0-100 to 6000 ml/h0-100 to 6000 ml/h0-100 to 10000 ml/h0-100 to 10000 ml/h
Post-dilution0-100 to 4000 ml/h0-100 to 4000 ml/h0-100 to 10000 ml/h0-100 to 10000 ml/h
Dialysate (CVVHD only)0-100 to 10000 ml/h0-100 to 10000 ml/h0-100 to 10000 ml/h0-100 to 10000 ml/h
Filtrate0-100 to 11000 ml/h0-100 to 11000 ml/h0-100 to 12000 ml/h0-100 to 12000 ml/h
Net fluid loss0-10 to 1000 ml/h by 1 ml0-10 to 1000 ml/h by 1 ml0-10 to 2000 ml/h0-10 to 2000 ml/h
Heparin pump accuracy± 0.2 ml/h± 0.2 ml/h± 0.2 ml/h± 0.2 ml/h
Heparin pump settings0-15 ml/h by 0.5 ml/h0-15 ml/h by 0.5 ml/h0-15 ml/h by 0.5 ml/h0-15 ml/h by 0.5 ml/h
Heparin syringe size30 or 50 ml30 or 50 ml30 or 50 ml30 or 50 ml
Bolus function0.5 to 2.5 ml by 0.5 ml bolus increment0.5 to 2.5 ml by 0.5 ml bolus increment0.5 to 2.5 ml by 0.5 ml bolus increment0.5 to 2.5 ml by 0.5 ml bolus increment
Scales
Fluid balance alarm20g20g50g50g
Substitution scale capacity (max. load)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)
Filtrate scale capacity (max. load)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)0-30kg (max. load: 23kg)
Fluid Warmer
Accuracy± 0.3 °C± 0.3 °C± 0.3 °C± 0.3 °C
Working rangeOff or 35°C to 39°C by 0.5°COff or 35°C to 39°C by 0.5°COff or 35°C to 39°C by 0.5°COff or 35°C to 39°C by 0.5°C
Pressure Monitoring
Sensor accuracy± 5 mmHg± 5 mmHg± 5 mmHg± 5 mmHg
Access Pressure-250 to +200 mmHg-250 to +200 mmHg-250 to +200 mmHg-250 to +200 mmHg
Return Pressure-50 to +350 mmHg-50 to +350 mmHg-50 to +350 mmHg-50 to +350 mmHg
Pre-filter Pressure-150 to +500 mmHg-150 to +500 mmHg-150 to +500 mmHg-150 to +500 mmHg
Filtrate Pressure-250 to +400 mmHg-250 to +400 mmHg-250 to +400 mmHg-250 to +400 mmHg
Degassing unit Pressure-300 to +30 mmHg-300 to +30 mmHg-300 to +30 mmHg-300 to +30 mmHg
Gas removal Max.10 ml/min10 ml/min10 ml/min10 ml/min
Monitor / Detection Parameters
Blood leak detector2 or 4 ml blood/1l filtrate at HCT 32%2 or 4 ml blood/1l filtrate at HCT 32%2 or 4 ml blood/1l filtrate at HCT 32%2 or 4 ml blood/1l filtrate at HCT 32%

Note: The reported device performance aligns perfectly with the acceptance criteria as presented, which is typical for a summary demonstrating compliance. The "Bench Testing" section explicitly states "This concludes that the two machines [Edwards Aquarius and Baxter Accura] have substantially equivalent performance," indicating the Edwards Aquarius met these criteria.

2. Sample size used for the test set and the data provenance

The document specifies "Comparative bench testing" was conducted using the Edwards Aquarius System and the Baxter Accura System, along with an Edwards Aqualine Sterile Tubing Set K063293.

  • Sample size: The specific number of tests or runs performed during the bench testing is not explicitly stated in the provided text.
  • Data provenance: The testing appears to be internal lab testing ("bench testing"). The parent company, Edwards Lifesciences, is based in Germany, and the device has been "marketed and developed in Europe for six years," suggesting the testing was likely conducted in Europe. The document does not specify if the data is retrospective or prospective, but bench testing typically involves prospective data collection under controlled conditions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This device is a medical machine with performance specifications related to fluid management, pressure monitoring, temperature control, and detection of operational issues (e.g., blood leak). The ground truth for such technical performance is established through engineering and calibration standards, not typically through expert consensus from medical professionals. The document does not mention any expert review process for establishing the ground truth of the technical specifications.

4. Adjudication method for the test set

Not applicable. As described above, the ground truth for performance specifications in bench testing is based on engineering standards and measurements, not subjective expert judgment that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Edwards Aquarius System is a hardware medical device (Hemofiltration System) and does not involve AI or human "readers" in its primary function, nor does it generate diagnostic images requiring interpretation. The study mentioned is a "comparative bench testing" for performance against a predicate device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a Hemofiltration System, a medical machine, not an algorithm. Bench testing evaluates the machine's inherent functionality and performance characteristics.

7. The type of ground truth used

The ground truth used for the performance parameters would be absolute physical measurements and engineering specifications. For example, a pump's accuracy is measured against known flow rates, scales are calibrated against known weights, and temperature sensors are verified against standard temperature references. The document explicitly mentions "Bench Testing" and "Electrical Safety and Electromagnetic Compatibility: Testing has been conducted following the IEC 60601 series of standards," further supporting the use of objective technical standards as ground truth.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI-driven device, so there is no concept of a "training set" in the context of its performance validation. The device is validated based on its physical design and adherence to manufacturing and performance specifications.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Center for Devices and Radiological Health

KC 7032C
Page 1 of 8
1/31/2007

Edwards Lifesciences Edisonstrasse 6 85716 Unterschleissheim Germany Telephone: +49 89 95475 203 JUN - 7 2007

Image /page/0/Picture/4 description: The image shows a logo with the word "Edwards" written in a serif font at the bottom. Above the word, there is a graphic element that appears to be a stylized letter "E" or a design composed of squares and rectangles. The logo has a simple and clean design.

510(k) Summary as required by 21 CFR 807.92 (c)

Date Summary Prepared:

January 31, 2007

Owner's Name:

Fax: +49 89 95475 301

Edwards Lifesciences Services GmbH
Address:Edisonstrasse 6UnterschleissheimD-85717Germany
Tel:+49 89 95475 203
Fax:+49 89 95475 301
Contact Person:Robert MadjnoDirector of Regulatory and Quality Affairs

510(k) Contact Persons:

Neil R. Armstrong Regulatory Affairs Advisor MeddiQuest Limited Business and Technology Centre Bessemer Drive Stevenage SG1 2DX United Kingdom Tel: +44 1763 222505 Fax: +44 1763 222049

or

{1}------------------------------------------------

Robert Madino Director of Regulatory and Quality Affairs Edwards Lifesciences Services GmbH Edisonstrasse 6 85716 Unterschleissheim Germany +49 89 95475 203 Tel: +49 89 95475 301 Fax:

Type of 510(k):

Traditional

Classification Name:

High Permeability Hemodialysis Systems

78 KDI

Common/Usual Name:

Hemofiltration System

Proprietary Name:

Edwards Aquarius System

Establishment Registration Number:

The device will be manufactured for: Edwards Lifesciences Services GmbH Edisonstrasse 6 Unterschleissheim D-85717 Germany Establishment Registration Number applied for

by

MeSys GmbH Beneckallee 30 Hannover D-30419 Germany Establishment Registration Number 9680719

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows handwritten text on a white background. The text includes "KC70320C", "Page 3 of 8", and the date "1/31/2007". The handwriting appears to be cursive and slightly slanted. The text is arranged in three lines, with the date at the bottom.

Substantial Equivalence:

The Edwards Aguarius system is based on product manufactured by MeSys, which has also been used as the basis for the Baxter Accura System - K021615.

The Edwards Aquarius system has been marketed and developed in Europe for six years. The software has been significantly improved from that used in the Baxter Accura System and the hardware has been similarly improved reflecting the European field experience.

The Edwards Aquarius system is substantially equivalent in design, use and materials to the:

Baxter Accura System Gambro Prisma System B. Braun Diapact

K021615 K993064, K981681, K946279 K973322

The Edwards Aquarius system is made of the same materials as the Baxter Accura System (K021615).

The Edwards Aquarius system is manufactured by the same processes as the Baxter Accura System (K021615).

Comparison of Technological Features:

FeatureActual Device:Edwards AquariusPredicate Device A:Baxter AccuraSystem,K021615
HousingFront panelDifferent - repositioning
Rear panelSame, except color
Interior (including cardslots)Same, except ADU andnew BLD
CartTrolley on rollersequipped with brakesSame, except differenttube surface treatment
Computer systemsSame
DisplaySolid State ColormonitorSame
Monitor casingSame
MotherboardSame, except Accurabattery
Master extensionExtension for ADUNot available
FeatureActual Device:Edwards AquariusPredicate Device A:Baxter AccuraSystem,K021615
board
Serial connector(RS232)Optical connectorSame
Power supplySame
WiringCable formSame, except ADUconnection
Blood pump withrotation sensorSame
Heparin pumpHardware and SoftwareDifferent
Arterial pressuresensorSame
Prefilter pressuresensorSame
Filtrate pressuresensorMembrane unit holderPressure transducerEvaluation electronicsSoftwareSameSameSame with later AccuraSW ver. 2.05Same
Bubble trap andvenous tube clampSame
Blood leak detectorHousingEvaluation electronicsSoftwareSameDifferent, change toSMD technologyDifferent, changedvalues for alarm limits
Predilution pumpSame
Filtrate pumpSame
Postdilution pumpSame
Substitution fluidweighing scalesWeighing scales hooksDifferent, moreergonomic withAquarius
Weighing deviceEvaluation electronics(18 bit converter)SoftwareSameDifferent, change toSMD technologySame
Filtrate weighingscalesWeighing scales hooksDifferent, moreergonomic withAquarius
FeatureActual Device:Edwards AquariusPredicate Device A:Baxter AccuraSystem,K021615
Weighing deviceSame
Evaluation electronics(18 bit converter)Different, change toSMD technology
SoftwareSame
ADUPressure monitoringand pumpNot available atAccura
HeaterMechanical structureSame
Plate heaterSame
Temperature MonitorDifferent, integrated toheparin pump housing
Driver and evaluationelectronicsDifferent, change toSMD technology
SoftwareSame
AC PowerOperating VoltageSame
Input Line CurrentSame

{3}------------------------------------------------

KC 70320
Page 4 of 8
1/31/2007

{4}------------------------------------------------

KC7032C
Page 5 of 8
1/31/2007

Description of Product:

The Edwards Aquarius System needs to be used in conjunction with a tubing set and a filter to provide Hemofiltration treatment to the patient.

The tubing set has already been the subject of a PreMarket Notification, K063293 Edwards Aqualine Sterile Tubing Sets, and the filter will be the subject of a separate PreMarket Notification, currently in preparation.

The Aquarius system is an Automated Fluid Balance Monitor, designed to be used with various extracorporeal treatments in the field of renal replacement therapies or plasma therapies. All therapies must be prescribed by a physician.

The Aquarius system is divided into three circuits: the extracorporeal (blood) circuit, the substitution/dialysate circuit and the filtrate circuit.

Toxic substances are removed by filters and clean blood is returned to the patient.

The Aquarius system allows the patient to be positioned left or right of the instrument.

{5}------------------------------------------------

KC70320
Page 6 of 8
1/31/2007

The Aquarius system uses two scales to accurately measure and precisely balance filtration and substitution volumes.

Heparin may be supplied to the extracorporeal circuit via an anticoagulant pump (Heparin pump). The prescribing physician may select continuous or intermittent options.

The Aquarius protective system is designed as a 2-channel system to protect the patient from foreseeable danger.

At the back of the scale system a removable hand-crank is mounted. This can be used to manually turn the blood pump in case a pump stops.

The Aquarius system is portable. It has a wheeled base connected with a handle to move or carry the Aquarius,

Fluids ManagementLow Blood FlowRegular Blood Flow
Pump accuracy±5%±5%
Blood10 to 200 ml/min30 to 450 ml/min
by 2 ml/minby 10 ml/min
Pre-dilution0-100 to 6000 ml/h0-100 to 10000 ml/h
Post-dilution0-100 to 4000 ml/h0-100 to 10000 ml/h
Dialysate0-100 to 10000 ml/h(CVVHD only)0-100 to 10000 ml/h
Filtrate0-100 to 11000 ml/h0-100 to 12000 ml/h
Net fluid loss0-10 to 1000 ml/hby 1 ml0-10 to 2000 ml/h
Heparin pump accuracy± 0.2 ml/h± 0.2 ml/h
Heparin pump settings0-15 ml/h by 0.5 ml/h0-15 ml/h by 0.5 ml/h
Heparin syringe size30 or 50 ml30 or 50 ml
Bolus function0.5 to 2.5 ml by0.5 to 2.5 ml by
0.5 ml bolus increment0.5 ml bolus increment
ScalesLow Blood FlowRegular Blood Flow
Fluid balance alarm20g50g
Substitution scale0-30kg0-30kg
capacity(max. load: 23kg)(max. load: 23kg)
Filtrate scale capacity0-30kg0-30kg
(max. load: 23kg)(max. load: 23kg)
Fluid warmerLow Blood FlowRegular Blood Flow

Significant Physical and Performance Characteristics:

{6}------------------------------------------------

Accuracy$\pm$ 0.3 °C$\pm$ 0.3 °C
Working rangeOff or 35°Cto 39°C by 0.5°COff or 35°Cto 39°C by 0.5°C
Pressure Monitoring
Sensor accuracy$\pm$ 5 mmHg
Access-250 to +200 mmHg
Return-50 to +350 mmHg
Pre-filter-150 to +500 mmHg
Filtrate-250 to +400 mmHg
Degassing unit
Pressure-300 to +30 mmHg
Gas removal Max.10 ml/min
Weight and Dimensions
H x W x D170 cm x 50 cm x 60 cm
Floor spaceApprox. 55 cm (W) x 65 cm (D)
WeightApprox. 75 kg
Power Requirements
European voltage230 V (alternating voltage) $\pm$ 10%,50 Hz
Power consumption500 VA
North American voltage115 V (alternating voltage) $\pm$ 10%,60 Hz
North American consumption500 VA
Monitor / Detection Parameters
Air detectorUltrasonic measurement
Blood leak detector2 or 4 ml blood/1l filtrate atHCT 32%
Display monitor10.4" TFT color

Intended use:

The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.

The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows handwritten text on a white background. The text includes "K070320", "Page 8 of 8", and "1/31/2007". The handwriting is cursive and slightly slanted. The text appears to be part of a document or record.

Testing Summary:

Electrical Safety and Electromagnetic Compatibility:

Testing has been conducted following the IEC 60601 series of standards, which are FDA recognized voluntary consensus standards.

Bench Testing:

Comparative bench testing of the Edwards Aquarius System and Baxter Accura System, using an Edwards Aqualine Sterile Tubing Set K063293, is included. This concludes that the two machines have substantially equivalent performance.

Animal Testing:

No animal testing is included.

There are no patient or patient's blood contacting materials.

Clinical Testing:

No clinical testing is included.

Other Information:

A summary of European surveillance and vigilance activities is included.

Currently approximately 1,500 machines have been placed on the European market and it is estimated these now (2006 figures) perform approximately 64,000 treatments per year (based on sales of tubing sets).

- H . TILOncoSATEANNO AND AN E & .------------900000A CHINE A MCareerC
P. L. 1 . Ph.and the delivery of Children1M-Added and see a from the commend of the commend of theCHANGE OF ANCALLER ARASALES CONSULER AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND ANDA PRODUCT OF COLL CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAStatus Controllers of Children1
www.go.government.com/shoulersA PROPERTY COLLECTION OF THE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A A A C C Status Concess ofMARK AND ANNAL BARREN BARREN BALLERYParcellers of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution ofControllation Company ofPhone Charmore------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Robert Madjno Director of Regulatory and Quality Affairs Edwards Lifesciences Services GmbH

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image is a seal for the Department of Health & Human Services (USA). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Edwards Lifesciences Services GmbH c/o Neil R. Armstrong Regulatory Affairs Advisor MeddiQuest Limited Business and Technology Centre Bessemer Drive Stevenage SG1 2DX UNITED KINGDOM

JUN - 7 2007

Re: K070320

Trade/Device Name: Edwards Aquarius System Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: May 14, 2007 Received: May 21, 2007

Dear Mr. Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/8/Picture/10 description: The image shows a circular logo with the text "1906-2006 FDA Centennial" in the center. The letters "FDA" are prominently displayed in a bold, stylized font. Three stars are arranged below the word "Centennial". The text around the circle reads "U.S. Department of Health and Human Services".

Protecting and Promoting Public Health

{9}------------------------------------------------

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21. CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours, .

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use

4070320 510(k) Number (if known):

Edwards Aquarius System Device Name:

Indications for Use:

The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.

The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Posted November 13, 2003)

Page 1 of 1
Nancy Brogdon

Division of Repr ductive, J and Radiological Devi

Radiological Devices
510(k) Number K070320

Indications for Use Statement 4

Page 4.2

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”