K Number
K070209

Validate with FDA (Live)

Date Cleared
2007-08-08

(198 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FB-100U Series Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the upper arm. The device is indicated for use by people over 15 years old in home use.

Device Description

FB-100U Series DIGITAL Blood Pressure Monitor is designed to measure the systolic and diastolic blood pressure, and pulse rate(heart of an individual). All values can be read out in one LCD DISPLAY. Measurement position is on adult arm only.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaDevice Performance (FB-100U Series)
Conformance to ANSI/AAMI SP-10Meets requirements
Conformance to IEC 60601-1Meets requirements
Conformance to IEC 60601-1-2Meets requirements
Intended Use: Measure systolic and diastolic blood pressure, and pulse rate, in adults over 15 years old for home useSame as predicate device

2. Sample Size Used for the Test Set and Data Provenance:

The provided 510(k) summary does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It mentions "bench testing" to demonstrate conformance to standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the given document. The study relies on conformance to established standards (ANSI/AAMI SP-10, IEC 60601-1, IEC 60601-1-2) rather than expert-established ground truth for specific measurements.

4. Adjudication Method for the Test Set:

This information is not applicable or not provided. Since the primary method of demonstrating performance is conformance to standards through "bench testing," a formal adjudication method like 2+1 or 3+1 typically used for subjective clinical assessments is not mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which doesn't apply to a digital blood pressure monitor that directly provides readings.

6. If a Standalone Study Was Done:

Yes, a standalone performance evaluation was done through "bench testing" to demonstrate the device's conformance to the specified standards (ANSI/AAMI SP-10, IEC 60601-1, and IEC 60601-1-2). This refers to the algorithm/device operating independently to perform its intended function.

7. The Type of Ground Truth Used:

The ground truth used for this device's evaluation is primarily the requirements and specifications outlined in the referenced performance standards: ANSI/AAMI SP-10 (Manual, electronic, or automated sphygmomanometers), IEC 60601-1 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance), and IEC 60601-1-2 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests).

8. The Sample Size for the Training Set:

This information is not applicable or not provided. For a blood pressure monitor, the device's algorithms are typically designed based on established physiological principles and engineering standards, not through training on a large dataset in the way a machine learning algorithm would be.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable or not provided. As mentioned above, the device's functionality is based on established engineering and physiological principles rather than a "training set" with ground truth in the context of machine learning.

{0}------------------------------------------------

510(K) SUMMARY

AUG - 8 2007

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

1. Submitter's Name:FEGO Precision Industrial Co., Ltd.
Address:947 Lin-Sen Rd., Wu-Fong, Taichung County., 413 Taiwan, R.O.C.
Phone:+886-4-23391802
Fax:+886-4-23399374
Contact:Mr. Bruce Chang / General Manager
2. Device Name :
Trade Name:FB-100U Series DIGITAL Blood Pressure Monitor
(Model No.: FB-1XYU, XY=00 ~ 99 represent type of case housing )
Common Name:Non-Invasive Blood Pressure Monitor
Classification nameSystem , Measurement , Blood-Pressure , Non-Invasive
3. DEVICE CLASSFB-100U Series DIGITAL Blood Pressure Monitor have been classified as
Regulatory Class: II
Panel: 74
Product Code: DXN
Regulation Number: 2ICFR 870.1130
4. Predicate Device:The predicate device is the Rossmax Automatic Blood PressureMonitor(K021225) marketed by ROSSMAX INTERNATIONAL LTD.
5. Device Description:FB-100U Series DIGITAL Blood Pressure Monitor is designed tomeasure the systolic and diastolic blood pressure, and pulse rate(heartof an individual). All values can be read out in one LCD DISPLAY.Measurement position is on adult arm only.
FB-100U Series DIGITAL Blood Pressure Monitor meets therequirements specified in ANSI/AAMI SP-10 - Manual, electronic, orautomated sphygmomanometers. Full responsibility for theconformance of this product to standard is assumed by FEGO PrecisionIndustrial Co., Ltd.

{1}------------------------------------------------

K070209

  • FB-100U Series DIGITAL Blood Pressure Monitor is intended to 6. Intended Use: measure the systolic and diastolic blood pressure, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the upper arm.
    The device is indicated for use by people over 15 years old in home use.

  • In terms of operating specification, Safety & EMC requirements, the 7. Performance device conforms to applicable standards included ANSI/AAMI SP-10, Summary: IEC 60601-1 and IEC 60601-1-2 requirements.

8. Conclusions:

FB-100U Series DIGITAL Blood Pressure Monitor has the same intended use and similar technological characteristics as the Rossmax Automatic Blood Pressure Monitor(K021225) marketed by ROSSMAX INTERNATIONAL LTD. Moreover, bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, FB-100U Series DIGITAL Blood Pressure Monitor is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three overlapping, curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 8 2007

Fego Precision Industrial Co. Ltd. c/o Ms. Jennifer Reich Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001

Re: K070209

Trade/Device Name: FB-100U Series Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: July 17, 2007 Received: July 19, 2007

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Jennifer Reich

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

B. Gemmer for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: FB-100U Series Digital Blood Pressure Monitor FEGO Precision Industrial Co., Ltd.

Indications For Use:

FB-100U Series Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the upper arm.

The device is indicated for use by people over 15 years old in home use.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use V (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blumuna

Page 1 of 1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).