(266 days)
The above referenced Broadsound transducer assemblies are intended for use, with standard ultrasound systems, in diagnostic ultrasound imaging or fluid flow analysis of the human body and to be operated by or under the direction of a physician.
Their specific indications for use are listed as following:
| Model Name | Specific Indications of Use* | |
|---|---|---|
| 1 | AL3C34A | Abdominal, Fetal |
| 2 | AL7L50A | Small organ, Peripheral vascular |
| 3 | AL3C79B | Abdominal, Fetal, Pediatric |
| 4 | AL7L24B | Small organ, Peripheral vascular |
| 5 | AT3C42B | Abdominal, Fetal, Pediatric |
| 6 | AT3C52B | Abdominal, Fetal, Pediatric |
| 7 | AT5L40B | Abdominal, Pediatric, Peripheral vascular, Musculo-skeletal |
| 8 | AT3P32A | Abdominal, Adult cephalic, Cardiac |
| 9 | AT3P42A | Abdominal, Adult cephalic, Cardiac |
The above referenced devices are replacement ultrasound transducers used with standard ultrasound systems. Each of them consists of piezoelectric crystals covered with an acoustic lens, a scan head that fits around the lens, a cable with strain relief devices on both ends, and a connector to attach the transducer to the ultrasound console.
This document describes the 510(k) summary for Broadsound Corporation's diagnostic ultrasound transducer assemblies. The submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing de novo performance criteria. Therefore, the information provided relates to equivalence criteria and comparative testing against predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here is "substantial equivalence" to predicate devices. The reported device performance is that the Broadsound transducers are "statistically comparable" to their corresponding predicate devices in terms of acoustic output and general performance.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Substantial Equivalence | Similar in features and use parameters to predicate devices. Used on the same diagnostic ultrasound systems. No new technology employed. Compliance with electrical, mechanical, thermal, and biocompatible safety standards. Acoustic output statistically comparable to predicate. | Broadsound transducers are "very similar to [their corresponding predicate device] in terms of features and use parameters." |
| "used on the same diagnostic ultrasound systems." | ||
| "No new technology is employed on these devices." | ||
| Electrical, Mechanical, thermal, and biocompatible safety testing conducted. | "the results are included in the submission." (Implying compliance) | |
| Acoustic output and general performance comparable to predicate devices. | "acoustic output and found statistically comparable to each other." "Performance testing was also conducted and included in the submission." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample size for the test set used for performance testing. It states that "Each of above referenced ultrasound transducer assemblies and its corresponding predicate device were tested," which implies a comparison of individual units of each Broadsound model against individual units of its corresponding predicate model.
The data provenance is not explicitly stated as retrospective or prospective, nor does it specify the country of origin of the data. However, the manufacturer is in Taiwan, so it is likely the testing was conducted there or arranged by the manufacturer.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This submission is for an ultrasound transducer assembly, not a diagnostic algorithm that requires expert interpretation of images for ground truth. Therefore, the concept of "experts" to establish ground truth in the context of image interpretation is not applicable here. The evaluation focuses on the physical and functional performance of the device against established predicate devices and safety standards.
4. Adjudication Method for the Test Set
Not applicable, as this is not a study involving human interpretation with adjudication. The performance testing is based on objective measurements of device characteristics and comparison to predicate devices and safety standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This is a premarket notification for a medical device (ultrasound transducer) demonstrating substantial equivalence to existing devices, not a comparative effectiveness study involving human readers and AI for diagnostic improvement.
6. Standalone (Algorithm Only) Performance Study
No, this is about hardware (ultrasound transducers) and their performance characteristics. It does not involve any algorithms in a standalone performance study as described in AI/ML performance evaluations.
7. Type of Ground Truth Used
The "ground truth" in this context refers to established technical specifications, safety standards, and the performance characteristics of the predicate ultrasound transducers. The Broadsound devices are compared against these benchmarks, rather than against expert consensus, pathology, or outcomes data which are typically used for diagnostic accuracy studies. The ground truth for this submission would involve:
- Predicate Device Specifications: The technical and performance characteristics of the legally marketed predicate ultrasound transducers.
- Safety Standards: Established electrical, mechanical, thermal, and biocompatibility safety standards for medical devices.
- Acoustic Output Measurements: Standardized measurements of acoustic power and intensity.
8. Sample Size for the Training Set
Not applicable. This submission is for hardware (ultrasound transducers) and does not involve a "training set" in the context of machine learning or AI models.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device submission.
{0}------------------------------------------------
D9. 1 of 4
Section S
$101) Surmeи y у
PAGE 1 of 13 PAGES
OCT 1 5 2007
510(k) Summary
For BRODASOUND Model AL3C34A, AL7L50A, AL3C79B, AL7L24B, AT3C42B, AT3C52B, AT5L40B, AT3P32A, and AT3P42A Diagnostic Ultrasound Transducer Assemblies
| 510(k) Number: | K070195 |
|---|---|
| Submitter: | Broadsound Corporation5F, No. 31, Shintai RoadJupei City, Hsinchu 30252, TaiwanTel: 886-3-5539868, Fax: 886-3-5539808 |
| Contact Person: | Jiann-Hwa Jeng, Ph.D., President & CEO |
| Date Prepared: | August 30, 2007 |
| Device Name:Trade Name: | BROADSOUND |
| Model Name: | AL3C34A, AL7L50A, AL3C79B, AL7L24BAT3C42B, AT3C52B, AT5L40B, AT3P32A, AT3P42A |
| Common Name: | Diagnostic Ultrasound Transducer Assembly,Ultrasound Imaging Probe,Medical Ultrasound Probe, orDiagnostic Ultrasound Probe |
| Classification Name: | Transducer, Ultrasonic, Diagnostic |
| Classification Regulation: | 21 CRF 892.1570 |
| Product Code: | 90ITX |
| Class: | II |
S10(k) Number: K070195
{1}------------------------------------------------
K070195
pg. 2 of 4
5 HQ(la Summary PAGE Car H FAKES
Device Comparison:
Legally marketed devices for substantial equivalence comparison are listed as following:
| Subject Device | 510(k) Number | Model Name | Manufacturer |
|---|---|---|---|
| AL3C34A | K900805, K880900 | UST-934N-3.5 | Aloka Ltd. |
| AL7L50A | K900805, K880900 | UST-5512U-7.5 | Aloka Ltd. |
| AL3C79B | K983879 | UST-979-3.5 | Aloka Ltd. |
| AL7L24B | K983879 | UST-5524-7.5 | Aloka Ltd. |
| AT3C42B | K002003 | C5-2 | ATL Ultrasound Inc. |
| Marketed product | C4-2 | ATL Ultrasound Inc. | |
| AT3C52B | K002003 | C5-2 | ATL Ultrasound Inc. |
| AT5L40B | K002003 | L7-4 | ATL Ultrasound Inc. |
| AT3P32A | K974269, K961459 | Medison P4-2 | Medison America, Inc. |
| Marketed product | ATL P3-2 | ATL Ultrasound Inc. | |
| AT3P42A | K974269, K961459 | Medison P4-2 | Medison America, Inc. |
| Marketed product | ATL P4-2 | ATL Ultrasound Inc. |
Each of above referenced transducer assemblies is substantially equivalent to its corresponding predicate device, and both of them are very similar to each other in terms of features and use parameters; as well, they are used on the same diagnostic ultrasound systems.
Legally marketed devices that Broadsound replacement transducers are intended to replace with are summarized as following:
| Broadsound NewReplacement Model | Predicate OEMTransducer Model | Indicated System |
|---|---|---|
| AL3C34A | Aloka UST-934N-3.5 | Aloka SSD-500 |
| AL7L50A | Aloka UST-5512U-7.5 | Aloka SSD-500 |
| AL3C79B | Aloka UST-979-3.5 | Aloka SSD-900 |
| AL7L24B | Aloka UST-5524-7.5 | Aloka SSD-900 |
| AT3C42B | ATL C4-2 | ATL HDI 5000 |
| AT3C52B | ATL C5-2 | ATL HDI 5000 |
| AT5L40B | ATL L7-4 | ATL HDI 5000 |
| AT3P32A | ATL P3-2 | ATL HDI 5000 |
| AT3P42A | ATL P4-2 | ATL HDI 5000 |
Description of Devices:
The above referenced devices are replacement ultrasound transducers used with standard ultrasound Each of them consists of piezoelectric crystals covered with an acoustic lens, a scan head systems. that fits around the lens, a cable with strain relief devices on both ends, and a connector to attach the transducer to the ultrasound console.
{2}------------------------------------------------
K070195
pg. 3 of 4
Section 510(k) Summary 彩金彩金 4 01 13 PAGES
| Specification | AL3C34A | AL7L50A | AL3C79B | AL7L24B |
|---|---|---|---|---|
| Frequency range | 5-2 MHz | 10-5 MHz | 5-2 MHz | 10-5 MHz |
| Array type | Convex | Linear | Convex | Linear |
| Field of view | 60 degree | 38 mm | 70 degree | 42 mm |
| Element number | 72 | 72 | 120 | 120 |
| Pitch | 1.0 mm | 0.6 mm | 0.6 mm | 0.375 mm |
The specifications of each transducer assembly are listed as following:
| Specification | AT3C42B | AT3C52B | AT5L40B | AT3P32A | AT3P42A |
|---|---|---|---|---|---|
| Frequency range | 4-2 MHz | 5-2 MHz | 7-4 MHz | 3-2 MHz | 4-2 MHz |
| Array type | Convex | Convex | Linear | Phased | Phased |
| Field of view | 75 degree | 75 degree | 40 mm | 90 degree | 90 degree |
| Element number | 128 | 128 | 128 | 64 | 64 |
| Pitch | 0.4 mm | 0.4 mm | 0.6 mm | 0.32 mm | 0.32 mm |
Intended Use:
The above referenced Broadsound transducer assemblies are intended for use, with standard ultrasound systems, in diagnostic ultrasound imaging or fluid flow analysis of the human body and to be operated by or under the direction of a physician.
Their specific indications for use are listed as following:
| Model Name | Specific Indications of Use* | |
|---|---|---|
| 1 | AL3C34A | Abdominal, Fetal |
| 2 | AL7L50A | Small organ, Peripheral vascular |
| 3 | AL3C79B | Abdominal, Fetal, Pediatric |
| 4 | AL7L24B | Small organ, Peripheral vascular |
| 5 | AT3C42B | Abdominal, Fetal, Pediatric |
| 6 | AT3C52B | Abdominal, Fetal, Pediatric |
| 7 | AT5L40B | Abdominal, Pediatric, Peripheral vascular, Musculo-skeletal |
| 8 | AT3P32A | Abdominal, Adult cephalic, Cardiac |
| 9 | AT3P42A | Abdominal, Adult cephalic, Cardiac |
{3}------------------------------------------------
K070195
pg-Hof 4
3 (A) (2) 3 (2) (2007) (2007) PAGE of 13 CAGE
Characteristics of Technology:
Each of above referenced Broadsound transducer assemblies is substantially equivalent to its corresponding predicate device, and both of them are very similar to each other in terms of features and use parameters; as well, they are used on the same diagnostic ultrasound systems. No new technology is employed on these devices.
Safety Testing:
Electrical, Mechanical, thermal, and biocompatible safety testing were conducted on these devices; the results are included in the submission.
Performance Testing:
Each of above referenced ultrasound transducer assemblies and its corresponding predicate device were tested for acoustic output and found statistically comparable to each other. Performance testing was also conducted and included in the submission. -
CE Mark:
All above referenced devices have been granted CE0197 mark of European Union by TÜV Rheinland Product Safety GmbH, Germany since April 2006, and the pertinent information is included in the submission.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 5 2007
Jiann-Hwa Jeng, Ph.D. President & CEO Broadsound Corporation 5F, No.31, Shintai Road JUPEI CITY HSINCHU 30252 TAIWAN
Re: K070195
Trade/Device Name: BROADSOUND Regulation Number: 21 CFR §892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: August 31, 2007 Received: September 4, 2007
Dear Dr. Jeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the BROADSOUND, as described in your premarket notification:
| Transducer Model Number | |
|---|---|
| and and the first and the first of the first of the American American American | I |
| AL3C34A | AT3C52B |
|---|---|
| AL7L50A | AT5L40B |
| AL3C79B | AT3P32A |
| AL7L24B | AT3P42A |
| AT3C42B |
{5}------------------------------------------------
Page 2 - Dr. Jeng
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{6}------------------------------------------------
Page 3 – Dr. Jeng
If you have any questions regarding the content of this letter, please contact Ewa Czerska, MD at (240) 276-3666.
Sincerely yours,
Jouyi Tri Whang
for Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
Section 4 Indications For Use Statement PAGE: 1 of 1 PAGES
Indications For Use Statement
| 510(k) Number: | K070195 |
|---|---|
| Common Name: | Diagnostic Ultrasound Transducer Assembly,Ultrasound Imaging Probe,Medical Ultrasound Probe, orDiagnostic Ultrasound Probe |
| Device Name | |
| Trade Name: | BROADSOUND |
| Model Name: | AL3C34A, AL7L50A, AL3C79B, AL7L24B,AT3C42B, AT3C52B, AT5L40B, AT3P32A, AT3P42A |
The above referenced devices are replacement ultrasound transducers intended to be used with standard ultrasound systems in diagnostic ultrasound imaging or fluid flow analysis of the human body and to be operated by or under the direction of a physician.
Their specific indications for use are listed as following:
| Model Name | Specific Indications of Use* | |
|---|---|---|
| 1 | AL3C34A | Abdominal, Fetal |
| 2 | AL7L50A | Small organ, Peripheral vascular |
| 3 | AL3C79B | Abdominal, Fetal, Pediatric |
| 4 | AL7L24B | Small organ, Peripheral vascular |
| 5 | AT3C42B | Abdominal, Fetal, Pediatric |
| 6 | AT3C52B | Abdominal, Fetal, Pediatric |
| 7 | AT5L40B | Abdominal, Pediatric, Peripheral vascular, Musculo-skeletal |
| 8 | AT3P32A | Abdominal, Adult cephalic, Cardiac |
| 9 | AT3P42A | Abdominal, Adult cephalic, Cardiac |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part21 CFR801 Subpart D)
(Division Sign-Off)
Division of Reproductive. Abdominal, and Radiological Devices 510(k) Number
Over-The-Counter Use _ (Part21 CFR801 Subpart C)
{8}------------------------------------------------
510(K) Number (if known): K070195 Device Name (transducer): BROADSOUND AL3C34A Ultrasound System: Aloka SSD 500 series
Indications for Use:
Diagnostic ultrasound imaging of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | ||||||||||
| Fetal | P | P | Note 1 | |||||||
| Abdominal | P | P | Note 1 | |||||||
| Intraoperative (specify) | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal | ||||||||||
| Conventional | ||||||||||
| Musculo-skeletal | ||||||||||
| Superficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K900805 and K88090; E = added under Appendix E Note 1: Combined application includes B/M mode.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{9}------------------------------------------------
Section 5 510(k) Summary PAGE 6 of 13 PAGES
.
Diagnostic Ultrasound Indications for Use Form
| 510(K) Number (if known): | K070195 |
|---|---|
| Device Name (transducer): | BROADSOUND AL7L50A |
| Ultrasound System: | Aloka SSD 500 series |
Indications for Use:
Aloka SSD 500 series
Diagnostic ultrasound imaging of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative(specify) | ||||||||||
| Pediatric | ||||||||||
| Small Organ(specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | P | P | Note 1 | |||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal | ||||||||||
| Conventional | ||||||||||
| Musculo-skeletal | ||||||||||
| Superficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K900805 and K88090; E = added under Appendix E Note 1: Combined application includes B/M mode.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Wha
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{10}------------------------------------------------
| 510(K) Number (if known): | |
|---|---|
| Device Name (transducer): | |
| Ultrasound System: |
K070195 BROADSOUND AL3C79B Aloka SSD 900 series
Indications for Use:
Diagnostic ultrasound imaging of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | P | P | Note 1 | |||||||
| Abdominal | P | P | Note 1 | |||||||
| Intraoperative(specify) | ||||||||||
| Pediatric | P | P | Note 1 | |||||||
| Small Organ(specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletalSuperficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K983879; E = added under Appendix E
Note 1: Combined application includes B/M mode.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, a Radiological Devices 510(k) Number
{11}------------------------------------------------
Section 5 510(k) Summary PAGE 8 of 13 PAGES
Diagnostic Ultrasound Indications for Use Form
510(K) Number (if known): K070195 Device Name (transducer): BROADSOUND AL7L24B Ultrasound System: Aloka SSD 900 series
Indications for Use: Diagnostic ultrasound imaging of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | ||||||||||
| Intraoperative(specify) | ||||||||||
| Pediatric | ||||||||||
| Small Organ(specify) | P | P | Note 1 | |||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | P | P | Note 1 | |||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal | ||||||||||
| Conventional | ||||||||||
| Musculo-skeletal | ||||||||||
| Superficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K983879; E = added under Appendix E
Note 1: Combined application includes B/M mode.
Small organ applications: Breast, Testes, Thyroid
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
louis Whan
(Divided Ben. Off)
(Division Sign-Off) Division of Reproductive. Abdominal, a Radiological Devices 510(k) Number
{12}------------------------------------------------
510(K) Number (if known): Device Name (transducer): Ultrasound System:
K070195 BROADSOUND AT3C42B ATL HDI 5000 series
Indications for Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | ||||||||||
| Fetal | P | P | P | P | P | Note 1 | ||||
| Abdominal | P | P | P | P | P | Note 2 | ||||
| Intraoperative (specify) | ||||||||||
| Pediatric | P | P | P | P | P | Note 2 | ||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal | ||||||||||
| Conventional | ||||||||||
| Musculo-skeletal | ||||||||||
| Superficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K002003; E = added under Appendix E Note 1: Combined application includes B/M mode.
Note 2: Combined applications include B/M, PWD/Color Doppler, and PWD/ Power Doppler modes.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Tony M. White
ctive. Abdomin Badiolo 510(k) Num
{13}------------------------------------------------
510(K) Number (if known): K070195 Device Name (transducer): Ultrasound System:
Indications for Use:
BROADSOUND AT3C52B ATL HDI 5000 series
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | P | P | P | P | P | Note 1 | ||||
| Abdominal | P | P | P | P | P | Note 2 | ||||
| Intraoperative(specify) | ||||||||||
| Pediatric | P | P | P | P | P | Note 2 | ||||
| Small Organ(specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletalConventional | ||||||||||
| Musculo-skeletalSuperficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K002003; E = added under Appendix E Note 1: Combined application includes B/M mode.
Note 2: Combined applications include B/M, PWD/Color Doppler, and PWD/ Power Doppler modes.
Concurrence of CDRH, Office of Device Evaluation (ODE)
/Prescription Use
(Per 21 CFR 801.109)
(Divisio Division of Reproductive, Abdominal, Radiological Devices 510(k) Number
{14}------------------------------------------------
510(K) Number (if known): K070195 Device Name (transducer): BROADSOUND AT5L40B Ultrasound System: ATL HDI 5000 series
Indications for Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Modes of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | ColorDoppler | P | Note 1 | ||||
| Intraoperative(specify) | ||||||||||
| Pediatric | P | P | P | ColorDoppler | P | Note 1 | ||||
| Small Organ(specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | ||||||||||
| Cardiac | ||||||||||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | P | P | P | ColorDoppler | Note 1 | |||||
| Laparoscopic | ||||||||||
| Musculo-skeletal | P | P | P | ColorDoppler | Note 1 | |||||
| Conventional | ||||||||||
| Musculo-skeletalSuperficial | P | P | P | ColorDoppler | ||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K002003; E = added under Appendix E Note 1: Combined applications include B/M, PWD/Color Doppler, and PWD/ Power Doppler modes.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Vorge 7. Wh
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{15}------------------------------------------------
510(K) Number (if known): K070195 Device Name (transducer): Ultrasound System: ATL HDI 5000 series
Indications for Use:
BROADSOUND AT3P32A
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||
| Fetal | P | P | P | P | P | Note 1 | ||||
| Abdominal | P | P | P | P | P | P | Note 2 | |||
| Intraoperative (specify) | ||||||||||
| Pediatric | ||||||||||
| Small Organ (specify) | ||||||||||
| Neonatal Cephalic | ||||||||||
| Adult Cephalic | P | P | P | P | P | P | Note 2 | |||
| Cardiac | P | P | P | P | P | P | Note 2 | |||
| Transesophageal | ||||||||||
| Transrectal | ||||||||||
| Transvaginal | ||||||||||
| Transurethral | ||||||||||
| Intravascular | ||||||||||
| Peripheral Vascular | ||||||||||
| Laparoscopic | ||||||||||
| Musculo-skeletal Conventional | ||||||||||
| Musculo-skeletal Superficial | ||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K974269 & K961459; E = added under Appendix E Note 1: Combined application includes B/M mode.
Note 2: Combined applications include PWD/Color Doppler and PWD/ Power Doppler modes..
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{16}------------------------------------------------
510(K) Number (if known): K070195 Device Name (transducer): BROADSOUND AT3P42A Ultrasound System: ATL HDI 5000 series
Indications for Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Modes of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | A | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined(specify) | Other(specify) | |
| Ophthalmic | |||||||||||
| Fetal | P | P | P | P | P | Note 1 | |||||
| Abdominal | P | P | P | P | P | P | Note 2 | ||||
| Intraoperative | |||||||||||
| (specify) | |||||||||||
| Pediatric | |||||||||||
| Small Organ | |||||||||||
| (specify) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | P | P | P | P | Note 2 | ||||
| Cardiac | P | P | P | P | P | P | Note 2 | ||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intravascular | |||||||||||
| Peripheral Vascular | |||||||||||
| Laparoscopic | |||||||||||
| Musculo-skeletal | |||||||||||
| Conventional | |||||||||||
| Musculo-skeletalSuperficial | |||||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA under K974269 & K961459; E = added under Appendix E Note 1: Combined application includes B/M mode.
Note 2: Combined applications include PWD/Color Doppler and PWD/ Power Doppler modes..
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Sign-Off) on of Reproductive, Abdominal, and cal Devices
§ 892.1570 Diagnostic ultrasonic transducer.
(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.