K Number
K063427
Date Cleared
2007-07-25

(254 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LUMINA Intense Pulsed Light (& Laser) System is intended for use in applications requiring the selective ablation, vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology.

The LUMINA Intense Pulsed Light (& Laser) System, when using the IPL Handpieces, is indicated for:

  • The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation and ephelides (freckles).
  • The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
  • The removal of unwanted hair and to effect stable long-term or permanent hair removal.
  • Indicated for use on all skin types (Fitzpatrick I-IV)

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of:

  • Benign vascular lesions such as, but not limited to treatment of:
    • Port wine stains
    • Hemangiomas
    • Superficial and deep telangiectasias (venulectasias)
    • Reticular veins (0.1 - 4.0mm diameter) of the leg
    • Rosacea
    • Venus Lake
    • Leg veins
    • Spider veins
    • Poikiloderma of Civatte
    • Angiomas
  • The removal of blue or black tattoos (significant reduction in the intensity of black and/or blue/black tattoos)
  • The non-ablative treatment of facial wrinkles, such as, but not limited to:
    • Periocular wrinkles
    • Perioral wrinkles
  • The removal of unwanted hair, for stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles.
  • Removal or lightening of unwanted hair (with and without adjuvant preparation).
  • Pseudofolliculitis barbae (PFB)
  • Indicated for use on all skin types (Fitzpatrick I-IV)

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 532nm (Long Pulsed and Q-Switched), is indicated for:

  • Tattoo removal:
    • Light blue
    • Yellow
    • Red
  • Benign pigmented lesions such as, but not limited to:
    • Café au lait macules
    • Lentigines (senile and solar)
    • Freckles (ephelides)
    • Chloasma
    • Nevi
    • Nevus spillus
    • Nevus of Ota
    • Becker's nevi
Device Description

The LUMINA Intense Pulsed Light (& Laser) System is a platform for a range of Intense Pulse Light (IPL) Handpieces and an internal Nd:YAG Laser Accessory for a number of applications in general and plastic surgery and dermatology. The System includes:

  • A floor-standing main cabinet unit (including the central control electronics) that controls timing and dosing parameters. This unit also contains a water-cooling system used to remove heat from the IPL Handpieces and the internal Nd:YAG Laser Accessory (if fitted).
  • A control and display panel including a touchscreen.
  • A range of optional IPL handpieces that include the flashlamp light source, electrical and cooling water connections.
AI/ML Overview

The provided document does not contain a study that proves the device meets specific acceptance criteria in the way a clinical trial or performance study would for an AI/ML device. Instead, this is a 510(k) summary for the LUMINA Intense Pulsed Light (& Laser) System, which demonstrates substantial equivalence to a predicate device.

Therefore, many of the requested fields cannot be directly extracted from this document, as they pertain to a different type of performance evaluation (e.g., for AI/ML algorithms).

Here's a breakdown of what can be gleaned:

1. A table of acceptance criteria and the reported device performance

This document does not present "acceptance criteria" in the typical sense of numerical thresholds for performance metrics. Instead, the "performance" described is the substantial equivalence to a predicate device based on its intended use, indications for use, technological characteristics, and safety testing.

Acceptance Criteria (Implicit)Reported Device Performance (LUMINA System)
Safe and Effective for stated Indications for UseSubstantially equivalent to predicate device (Lovely System, K033946, K042000) for all stated indications.
Compliance with relevant Electromedical StandardsComplies with European Medical Devices Directive 93/42/EEC (Annex II) and standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
Similar Technological Characteristics to predicateSimilar technological characteristics, including laser output values, to the predicate device.
No new safety and/or effectiveness issuesThe system "raises no new safety and/or effectiveness issues" compared to the predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This is not a study involving a test dataset for an AI/ML device. The "test" here refers to safety and electrical performance testing, not clinical data evaluation for performance metrics like sensitivity or specificity.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No ground truth establishment for a test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, and no MRMC study is mentioned. This device is a laser/light system for medical treatment.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. No ground truth, in the context of AI/ML performance evaluation, is described. The "truth" in this context is established medical understanding and clinical outcomes from previously cleared, similar devices.

8. The sample size for the training set

Not applicable. This is not an AI/ML device with a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

K063427

Appendix A11 The 510(k) Summary

. .

Applicant & Submitter :Lynton Lasers LimitedJUL 25 2007
Address :Lynton House, Manor Lane,Holmes Chapel, Cheshire, CW4 8AF, UK
Telephone :+44 (0)1477 536977
Fax :+44 (0)1477 536978
Contact Person :Dr. Andrew J BerryAJBerry@lynton.co.uk
Preparation Date :9th November 2006
Device Submitted :LUMINA Intense Pulsed Light (& Laser)System
Common Name :Light Therapy Device
Classification Name :Laser surgical instrument for use in generaland plastic surgery and in dermatology.
Product Code :GEX
Predicate Device :Lovely System (Lovely II, or "Harmony"),manufactured by MSq(M2) Ltd.7 Haeshel St. P.O.B 3021Caesarea Industrial Park, 38900 Israel(K033946, K042000)
Device Description :The LUMINA Intense Pulsed Light (& Laser)System is a platform for a range of IntensePulse Light (IPL) Handpieces and an internalNd:YAG Laser Accessory for a number ofapplications in general and plastic surgery anddermatology, including :Hair RemovalVascular Lesion treatmentsPigmented Lesion treatmentsTattoo removalFacial wrinkle reductionThe System includes :

{1}------------------------------------------------

  • . A floor-standing main cabinet unit (including the central control electronics) that controls timing and dosing parameters. This unit also contains a water-cooling system used to remove heat from the IPL Handpieces and the internal Nd:YAG Laser Accessory (if fitted).
  • A control and display panel including a ● touchscreen.
  • A range of optional IPL handpieces that . include the flashlamp light source, electrical and cooling water connections.
  • The LUMINA Intense Pulsed Light (& Laser) Intended Use : System is intended for use in applications requiring the selective ablation, vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology.
  • The LUMINA Intense Pulsed Light (& Laser) Indications for Use : stem, when using the IPL Handpieces, is indicated for :
    • The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation and ephelides (freckles).
    • The treatment of benign cutaneous . vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
    • The removal of unwanted hair and to effect . stable long-term or permanent hair removal.
    • Indicated for use on all skin types (Fitzpatrick I-IV)

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of :

{2}------------------------------------------------

  • . Benign vascular lesions such as, but not limited to treatment of :
    • Port wine stains

    • Hemangiomas

Superficial and deep telangiectasias (venulectasias)

Reticular veins (0.1 - 4.0mm diameter) of the leg

  • Rosacea

  • Venus Lake

  • Leg veins

  • Spider veins

  • Poikiloderma of Civatte

  • Angiomas

  • . The removal of blue or black tattoos (significant reduction in the intensity of black and/or blue/black tattoos)
  • The non-ablative treatment of facial ● wrinkles, such as, but not limited to: > Periocular wrinkles
    • Perioral wrinkles

  • The removal of unwanted hair, for stable ● long-term or permanent hair reduction through selective targeting of melanin in
  • hair follicles. Removal or lightening of unwanted hair . (with and without adjuvant preparation).
  • Pseudofolliculitis barbae (PFB) .
  • Indicated for use on all skin types . (Fitzpatrick I-IV)

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 532nm (Long Pulsed and Q-Switched), is indicated for :

  • Tattoo removal :
    • Light blue

    • Yellow

    • Red

  • Benign pigmented lesions such as, but not . limited to :
    • Café au lait macules

    • Lentigines (senile and solar)

    • Freckles (ephelides)

{3}------------------------------------------------

  • Chloasma

  • Nevi

  • Nevus spillus

  • Nevus of Ota

  • Becker's nevi

  • The LUMINA Intense Pulsed Light (& Laser) Performance Data : System complies with the European Medical Devices Directive 93/42/EEC (Annex II) and the standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22 and IEC 60825-1
  • Substantial Equivalence : The LUMINA Intense Pulsed Light (& Laser) System is substantially equivalent in terms of technological characteristics, performance, intended use, indications for use and operator interface to the previously cleared Lovely System (Lovely II, or "Harmony" System), cleared under K033946, K042000. There are no unique applications, indications, materials or specifications presented. Evidence of equivalence has been demonstrated through :
    • · The LUMINA Intense Pulsed Light (& Laser) System's intended use and indications for use were previously cleared by the FDA for the predicate device.
    • The technical characteristics of The LUMINA Intense Pulsed Light (& Laser) System are similar to those of the predicate device.
    • Laser output values of the LUMINA Intense ● Pulsed Light (& Laser) System are similar to those of the predicate device (and other FDA cleared devices on the market)
    • The predicate device, and other previously ● cleared devices with similar output specifications, have a proven safety and effectiveness in the treatment of the claimed indications.
    • Safety and performance testing. .

Therefore, the LUMINA Intense Pulsed Light (& Laser) System is substantially equivalent to the predicate device cited above and raises no new safety and/or effectiveness issues.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract, overlapping shapes that resemble human profiles or waves, creating a sense of depth and movement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 5 2007

Lynton Lasers Ltd. % Dr. Andrew J. Berry Technical Director Lynton House, Manor Lane Holmes Chapel, Cheshire CW4 8AF United Kingdom

Re: K063427

Trade/Device Name: LUMINA Intense Pulsed Light (& Laser) System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgerv and in dermatology Regulatory Class: II Product Code: GEX Dated: June 21, 2007 Received: June 25, 2007

Dear Dr. Berry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Dr. Andrew J. Berry

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 6338-241 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/suport/index20th11.

Sincerely yours,

Mark A. Millikusen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known) : (063427

LUMINA Intense Pulsed Light (& Laser) System Device Name :

Indications for Use :

The LUMINA Intense Pulsed Light (& Laser) System is intended for use in applications requiring the selective ablation. vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology, as follows :

The LUMINA Intense Pulsed Light (& Laser) System, when using the IPL Handpieces, is indicated for :

  • . The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation and ephelides (freckles).
  • The treatment of benign cutaneous vascular lesions including port wine . stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
  • The removal of unwanted hair and to effect stable long-term or . permanent hair removal.
  • Indicated for use on all skin types (Fitzpatrick I-IV) .

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Mark N. Milliken

Division of General, Restorative, and Neurological Devices

510(k) Number K064327

{7}------------------------------------------------

510(k) Number (if known) :K(063427

Device Name :

Indications for Use : - Continued from previous page -

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of :

  • Benign vascular lesions such as, but not limited to treatment of :
    • Port wine stains

    • Hemangiomas

    • Superficial and deep telangiectasias (venulectasias)

    • Reticular veins (0.1 4.0mm diameter) of the leg

    • Rosacea

    • Venus Lake

    • Leg veins

    • Spider veins

    • Poikiloderma of Civatte

    • Angiomas

  • The removal of blue or black tattoos (significant reduction in the . intensity of black and/or blue/black tattoos)

Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

{8}------------------------------------------------

510(k) Number (if known) : (063427

Device Name :

Indications for Use : - Continued from previous page -

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of - continued - :

  • . The non-ablative treatment of facial wrinkles, such as, but not limited to:
    • Periocular wrinkles

    • Perioral wrinkles

  • The removal of unwanted hair, for stable long-term or permanent hair . reduction through selective targeting of melanin in hair follicles.
  • Removal or lightening of unwanted hair (with and without adjuvant preparation).
  • Pseudofolliculitis barbae (PFB) ◆
  • Indicated for use on all skin types (Fitzpatrick I-IV) ●

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

{9}------------------------------------------------

510(k) Number (if known) :)(063427

Device Name :

Indications for Use : - Continued from previous page -

The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 532nm (Long Pulsed and Q-Switched), is indicated for :

  • Tattoo removal : ●
    • Light blue = = =

    • Yellow

    • Red

  • Benign pigmented lesions such as, but not limited to : �
    • Café au lait macules

    • Lentigines (senile and solar)

    • Freckles (ephelides)

    • Chloasma

    • Nevi

    • Nevus spillus

    • Nevus of Ota

    • Becker's nevi

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.