K Number
K063243
Manufacturer
Date Cleared
2007-12-14

(414 days)

Product Code
Regulation Number
862.1215
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VIDAS® Troponin I Ultra is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in human serum or plasma (lithium heparin) using the ELFA (Enzymc-Linked Fluorescent Assay) technique. VIDAS Troponin I Ultra is intended to be used as an aid in the diagnosis of myocardial infarction.

Device Description

The VIDAS Troponin I Ultra (TNIU) Assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated VIDAS® instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are in the sealed TNIU Reagent Strips. The sample is transferred into the wells containing anti-cardiac troponin i antibodies labeled with alkaline phosphatase (conjugate). The sample/conjugate mixture is cycled in and out of the SPR for a specified length of time. Troponin I present in the specimen will bind to the anticardiac troponin i immunoglobulin coating the interior of the SPR. Unbound sample components are washed away. A fluorescent substrate, 4-methylumbellifery| phosphate, is introduced into the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The optical scanner in the instrument measures the intensity of fluorescence. When the VIDAS TNIU assay is completed, the results are analyzed automatically by the computer, a test value is generated, and a report is printed for each sample.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the VIDAS® Troponin I Ultra (TNIU) Assay, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a separate section with specific numerical targets. Instead, it presents performance data for both the VIDAS® TNIU and the predicate device (Dimension RxL® CTNI) for comparison, implying that similar or improved performance relative to the predicate device is the goal for most parameters, and certain established clinical cut-offs are considered.

Given this, I've constructed a table focusing on the performance characteristics that were measured and compared, as these implicitly serve as the basis for demonstrating substantial equivalence.

Acceptance Criteria (Implied / Comparison Target)Reported Device Performance (VIDAS® TNIU)Predicate Device Performance (Dimension RxL® CTNI)
Expected Values (Healthy Patients)99% of 747 patients with no cardiac symptoms had values of <0.01 µg/l97.5% of 101 apparently healthy patients had values of 0.00 -0.05 µg/L
Cut-off0.11 µg/l0.6-1.5 ng/mL (Note: unit difference)
Specificity (Cardiac Troponin T) (at 1000 µg/L)0.2%0.34%
Specificity (Cardiac Troponin C) (at 1000 µg/L)<0.0010.00%
Specificity (Skeletal Troponin I) (at 1000 µg/L)<0.0010.04%
Analytical Detection Limit< 0.01 µg/l0.04 µg/L
Dilution - Recovery Test80-120%98.6-106.4%
Interference (Bilirubin)No significant interference (up to 510 µM)No significant interference (up to 20 mg/dL)
Interference (Hemoglobin)No significant interference (up to 332 µM)No significant interference (up to 1000 mg/dL)
Interference (Lipemia/Triglycerides)No significant interference (up to 30 mg/ml)No significant interference (up to 3000 mg/dL)
Hook Effect (up to stated concentration)No hook effect found up to 1000 µg/LNo hook effect found up to 1800 ng/mL
Clinical Method Comparison (Correlation Coefficient)0.97(N/A - This is a comparison between devices)
Clinical Method Comparison (Slope)0.42 (Confidence Interval: 0.38-0.44)(N/A - This is a comparison between devices)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Method Comparison: 534 samples
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). It is described as a "sample comparison study," implying patient samples were collected and tested.
  • Sample Size for Expected Values (Healthy Patients):
    • VIDAS® TNIU: 747 patients with no cardiac symptoms
    • Predicate Device: 101 apparently healthy patients

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not Applicable (N/A) / Not explicitly stated. This device is an immunoassay for a biomarker (Troponin I). The "ground truth" for analytical performance studies would typically be established by reference measurements (e.g., gold standard analytical methods, spiked samples) or by clinical diagnosis for the "expected values" data, not by expert interpretation of images or symptoms in the same way as, for example, an imaging AI device.
  • For the "expected values in healthy patients," the ground truth is simply the clinical status of the patient (e.g., "no cardiac symptoms" or "apparently healthy"). The document doesn't mention expert review of these patient status determinations beyond general clinical criteria.

4. Adjudication Method for the Test Set

  • Not Applicable (N/A) / Not mentioned. Adjudication methods like 2+1 or 3+1 are typically used for studies where multiple human readers interpret data (e.g., medical images) and their interpretations need to be reconciled to establish a consensus ground truth. For an immunoassay, the measurement itself is the primary focus, and while clinical diagnosis (the outcome) might involve expert consensus, the assay performance itself is not adjudicated in this manner.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is not relevant for this type of in vitro diagnostic device (immunoassay). MRMC studies compare the diagnostic performance of different readers (humans, AI, or human+AI) on a set of cases. This submission is for an automated immunoassay for measuring a biomarker.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, by nature of the device. The VIDAS® TNIU Assay is a standalone automated quantitative test. The results are generated automatically by the instrument, and then a report is printed. There is no explicit "human-in-the-loop" component in the generation of the quantitative result itself, though a clinician would interpret the result in the context of a patient's overall clinical picture. The performance data presented (e.g., correlation, detection limits) directly reflects the standalone performance of the assay.

7. The Type of Ground Truth Used

  • Analytical Measurement Comparisons and Clinical Status:
    • For Analytical Detection Limit, Dilution-Recovery, Interference, Hook Effect, Specificity: Ground truth is established by the known concentrations of analytes, spiked samples, or absence/presence of interfering substances under laboratory conditions.
    • For Expected Values (Healthy Patients): Clinical documentation of the patient's health status (e.g., "no cardiac symptoms" or "apparently healthy") serves as the ground truth for that specific study component.
    • For the Clinical Testing (Method Comparison): The "ground truth" is one device's measurement (Dimension RxL® CTNI Assay, the predicate) against which the new device (VIDAS® TNIU Assay) is compared to demonstrate correlation and substantial equivalence.

8. The Sample Size for the Training Set

  • Not explicitly stated in relation to algorithm training. This device is an immunoassay, not an AI/machine learning algorithm that requires a "training set" in the conventional sense of computational models. The assay is based on chemical and biological reactions, not statistical pattern recognition trained on a large dataset.
  • However, the development of such assays involves extensive research and development with numerous samples to optimize reagents, protocols, and establish performance characteristics, which could be conceptually considered akin to "training" in a broader sense, but not with a distinct "training set" as understood in AI.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable (N/A). As explained above, this is an immunoassay, not an AI/machine learning device that uses a "training set" with ground truth in the typical computational sense. The "ground truth" for optimizing the assay would involve using reference materials, known concentrations, and clinical samples whose true values (or clinical diagnoses) were independently established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for BIOMÉRIEUX. The logo consists of a stylized globe with vertical lines on the left half and a solid black right half. The company name, "BIOMÉRIEUX", is printed in a stylized font below the globe.

K063243

DEC 1 4 2007

510(k) SUMMARY

VIDAS® Troponin | Ultra (TNIU) Assay

A. Submitter Information

Submitter's Name:Address:bioMérieux, Inc.595 Anglum RoadHazelwood, MO 63042
Contact Person:Nikita S. Mapp
Phone Number:314-731-7474
Fax Number:314-731-8689
Date of Preparation:August 11, 2006

B. Device Name

Trade Name:VIDAS® Troponin I Ultra (TNIU) Assay
Common Name:Troponin I Enzyme Immunoassay
Classification Name:21 CFR 862.1215, Product Code MMIImmunoassay method, Troponin Subunit

C. Predicate Device Name

Trade Name:Dimension RxL® Cardiac Troponin I (CTNI) Assay
-------------------------------------------------------------

D. Device Description

The VIDAS Troponin I Ultra (TNIU) Assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated VIDAS® instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are in the sealed TNIU Reagent Strips.

The sample is transferred into the wells containing anti-cardiac troponin i antibodies labeled with alkaline phosphatase (conjugate). The sample/conjugate mixture is cycled in and out of the SPR for a specified length of time. Troponin I present in the specimen will bind to the anticardiac troponin i immunoglobulin coating the interior of the SPR. Unbound sample components are washed away.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a stylized globe split vertically, with one half shaded with vertical lines and the other half solid black. Below the globe, the company name "bioMérieux" is written in a stylized font with spaces between the letters.

A fluorescent substrate, 4-methylumbellifery| phosphate, is introduced into the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The optical scanner in the instrument measures the intensity of fluorescence. When the VIDAS TNIU assay is completed, the results are analyzed automatically by the computer, a test value is generated, and a report is printed for each sample.

E. Intended Use

VIDAS Troponin I Ultra (TNIU) Assay is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in serum or plasma (lithium heparinate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique.

Similarities
ItemVIDAS® TNIUDimension RxL® CTNI
Assay PrincipleOne-step automated immunoassay based onsandwich principleSame
Intended UseQuantitative determination of human cardiactroponin I in human serum or plasma (lithiumheparinate)Same
Indications for UseAn aid in the diagnosis of mycocardial infarctionSame
Sample TypeHuman serum or plasma (lithium heparin)Same
AntibodyCapture: mouse monoclonal antibodyConjugate: mouse monoclonal antibodylabeled with alkaline phosphataseSame

F. Technological Characteristics Summary

Differences
ItemVIDAS® TNIUDimension RxL® CTNI
Solid PhaseSolid Phase Receptacle (SPR)Chrome
Final Detection oftroponin-I antigenFluorescence (ELFA) of 4-methyl-umbelliferyl measured at 450 nmColorimetric rate measurement at 510nm
Measurement range0.01 to 30 µg/L0.04 to 40 µg/L
Analytical Detection Limit0.01 µg/L0.04 µg/L
Hook effectNo hook effect found up toconcentrations of 1000 µg/LNo hook effect found up toconcentrations of 1800 µg/L
Sample Volume200 µl50 µl
Assay Time~20 minutes~17 minutes

228

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of a circle that is half black and half white with lines. Below the circle is the name of the company, "bioMerieux".

G. Performance Data

Nonclinical Testing

Dimension® RxL CTNIVIDAS® TNIU
Expected Values97.5% of 101 apparentlyhealthy patients had valuesof 0.00 -0.05 µg/L99% of 747 patients with nocardiac symptoms hadvalues of <0.01 µg/l
Cut-off0.6-1.5 ng/mL0.11 µg/l
SpecificityCardiac Troponin T1000 µg/L: 0.34%1000 µg/L: 0.2%
Cardiac Troponin C1000 µg/L: 0.00%1000 µg/L: <0.001
Skeletal Troponin I1000 µg/L: 0.04%1000 µg/L: <0.001
Analytical Detection Limit0.04 µg/L< 0.01 µg/l
Dilution - Recovery Test98.6-106.4%80-120%
InterferenceNo significant interferenceNo significant interference
Bilirubin20 mg/dL510 µM
Hemoglobin1000 mg/dL332 µM
Lipemia (triglycerides)3000 mg/dL30 mg/ml
Hook Effect1800 ng/mL1000 µg/l

Clinical Testing

Five-hundred and thirty-four samples were tested with both the VIDAS® TNIU Assay (Y) and the Dimension RxL ® CTNI Assay (X). Data from the sample comparison study was evaluated with a Passing-Bablok method and correlation coefficient and produced the following results:

Y = 0.42 X

Confidence interval for the slope: 0.38-0.44 Correlation coefficient = 0.97

H. Conclusion

The VIDAS Troponin I Ultra (TNIU) Assay is substantially equivalent to the Dimension RxL® Cardiac Troponin I (CTNI) Assay.

The 510(k) summary includes only information that is also covered in the body of the 510(k). The summary does not contain any puffery or unsubstantiated labeling claims. The summary does not contain any raw data, i.e., contains only summary data. The summary does not contain any trade secret or confidential commercial information. The summary does not contain any patient identification information.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 4 2007

bioMérieux, Inc. c/o Ms. Nikita S. Mapp Senior Regulatory Affairs Specialist 595 Anglum Road Hazelwood, MO 63042

Re: K063243 Trade/Device Name: Vidas Troponin I Ultra (TNIU) Assay, Model 30 448 21 CFR 862.1215 Regulation Number: Regulation Name: Creatine Phosphokinase/Creatine Kinase or Isoenzymes Test System. Regulatory Class: Class II Product Code: MMI Dated: September 14, 2007 Received: September 17, 2007

Dear Ms. Mapp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Coopes, M.S., D.V.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number (if known): K063243

Device Name: VIDAS Troponin I Ultra (TNIU)

Indication For Use: VIDAS® Troponin I Ultra is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in human serum or plasma (lithium heparin) using the ELFA (Enzymc-Linked Fluorescent Assay) technique. VIDAS Troponin I Ultra is intended to be used as an aid in the diagnosis of myocardial infarction.

Prescription Use (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K063243

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.