(54 days)
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent ertapenem at concentrations of 0.25-32 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Erapenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli
Klebsiella pneumoniae
Active In Vitro:
Citrobacter freundii
Citrobacter koseri
Enterobacter aerogenes
Enterobacter cloacae
Klebsiella oxytoca
Morganella morganii
Proteus mirabilis
Proteus vulgaris
Providencia rettgeri
Providencia stuartii
Serratia marcescens
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, or resistant).
Here's a breakdown of the acceptance criteria and study information based on the provided text, adapted as if it were for an AI device for clarity in a modern context.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Target Performance | Reported Device Performance (BD Phoenix™ System) |
|---|---|---|
| Site Reproducibility (Intra-site) | >90% | >90% |
| Site Reproducibility (Inter-site) | >95% | >95% |
| Essential Agreement (EA) | Not explicitly stated in percentage | Achieved (compared to CLSI reference method) |
| Category Agreement (CA) | Not explicitly stated in percentage | Achieved (compared to CLSI reference method) |
| Accuracy (Overall for Ertapenem) | Not explicitly stated in percentage | Demonstrated (compared to CLSI reference method) |
| Reproducibility (Overall for Ertapenem) | ≥95% within ±1 dilution | 95% or greater within ±1 dilution |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document indicates that "Clinical, stock and challenge isolates were tested." However, specific numerical sample sizes for each type of isolate are not provided in the summary.
- Data Provenance: The data was collected from "multiple geographically diverse sites across the United States." The study used a combination of:
- Clinical isolates: Presumably prospective as they represent real-world samples.
- Stock isolates: Likely retrospective or pre-existing lab strains.
- Challenge isolates: Likely retrospective or pre-existing, used for specific performance verification against known results.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set. Instead, the ground truth was established by:
- CLSI reference broth microdilution method: This is a standardized, established laboratory method, not reliant on individual expert interpretation for each case.
- Expected results for Challenge set isolates: These "expected results" would be pre-determined values for known strains, which could have been established by consensus or by previous rigorous testing using reference methods.
4. Adjudication Method for the Test Set
No explicit adjudication method (e.g., 2+1, 3+1) is described for the test set. The comparison was made directly against the CLSI reference broth microdilution method or expected results for challenge isolates, which serve as the definitive standard.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This study evaluates the performance of an automated microbiology system (BD Phoenix™) against a reference standard (CLSI broth microdilution method) for antimicrobial susceptibility testing. It does not involve human readers or assess the improvement of human readers with AI assistance.
6. Standalone (Algorithm Only) Performance Study
- Yes, a standalone performance study was done. The entire study is a standalone evaluation of the BD Phoenix™ Automated Microbiology System (which incorporates an algorithm to interpret results) by comparing its outputs (MIC values and category interpretations) against the CLSI reference method. There is no human-in-the-loop component described in the performance evaluation.
7. Type of Ground Truth Used
The primary ground truth used was the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). For challenge isolates, "expected results" were used.
8. Sample Size for the Training Set
The document does not provide information on the sample size used for the training set for the BD Phoenix™ system's underlying algorithms. This is typical for such submissions, as the focus is on the performance of the final, already-developed device.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. Similar to the training set sample size, this information is not typically included in the summary for a 510(k) submission for a device like this, which is an automated system rather than a de novo AI model with publicly disclosed training data. It's assumed the system was developed and validated internally using appropriate microbiological standards.
{0}------------------------------------------------
510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410)-316-4161Fax: (410)-316-4499 | NOV 21 2006 |
|---|---|---|
| CONTACT NAME: | Vicki L. KennedyRegulatory Affairs Specialist | |
| DATE PREPARED: | September 29, 2006 | |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System –Ertapenem 0.25-32 µg/mL | |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645 | |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002, and K060324, May 25, 2006), Ofloxacin(K020323, April 14, 2002), and Levofloxacin (K020322, March27, 2002). | |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin. |
DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, or resistant).
DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.
{2}------------------------------------------------
Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD PhoenixTM Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
Performance of BD Phoenix System for Gram-Negative Organisms by Drug Table 1:
| A montant of became to any and the country a come a come of the comments of the comments of the comments of the comments of the comments of the comments of the comments of thAntimicrobia | And Company of Children Company ofoncentration | A100come of the commend of the count of the continues and consideration | colorado chokodon Romation it a Follow ligging ! The accompany | .n | 0)Householded of the Art of Children would be |
|---|---|---|---|---|---|
| Concession AND LAND STATUS CONSULTERSErtapenem | стание продательные в своемника в собсавления в собставляется в собставляется подательные(t Carlos Career Career (Carrier Comments) (Career (Carrier (Carrier) (Carrier (Carrier) (Carrier (Carrier (Carrier) (Carrier (Carrier) (Carrier (Carrier (Carrier) (Career (Ca11014/1l | 100. | 1 | 200 | 00 |
Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002, and K060324, May 25, 2006), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
SUMMARY INFORMATION FOR ERTAPENEM Available Range 0.25-32 µg/mL
Performance
Accuracy
| APor course of the Cases of the States andAntimierohiaof the first and the country of the distance of the country of the country of | COSCHOOLperson deliveres designations in considered in aoncentration | nand a come a marrier with | A Charles contrast of the work | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------77 | A . But the program and and the support of the status of the status ofﺑC |
|---|---|---|---|---|---|
| A Mch Busin' MP3 PP PFrianenem | t the contraction of the contributed to the contribution of the contribution of the first of the first of the first of the first of the first of the first of the first of thet100 | 000 | . 500 | 00 |
Reproducibility
Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution.
Breakpoints - CLSI
| THE PERSONAL OF THE RESERVED TO THE COLLECT ANDand the same and and and and any and any and the family of the first and "service"" " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "-------------------1000 100 1000------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. A S | .Call Property of Children And CharlesCarles Call of ChildrenAdvanced by Concellent of Concession------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and a man a pay the land of the can and a many aCONTRACTOR COLLECTION CONCELLENTAND A LAND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND ALL PROPERTY OF ALL PROPERTY OF ALL PROPERTY OF ALL PROPERTY OF ALL PROPERT------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of Children College of Children Company of Children & Free |
|---|---|---|---|
| And College of Carres of Carres of Careerof the case the no the a sens of the United onSERVE SERVE IN FREED I DESIGN I BEREAK WIT E A LE T17 14 11 14 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 1A country the call the called to the a comments on a comments of the comments------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of the Comments of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children and | -commender a contribution and the couple of the comments of the compares of the may be any andon the first and an all an the can and an and any cases and second station and come and compares and compares and compares and compares and compares and compares and compares | ---------------------------------------------------------------------------------MARK & AND : A SEE MENSEL : A corporation : Check A CHI | contract couple of the course and an any any any any any any any any any any any any any any any any any any any any any any any any any any anaona any anao anay any anaona a |
Recommended Quality Control Organisms
| Quality Control Strain | MIC Range | Source |
|---|---|---|
| Escherichia coli ATCC 25922 | 0.004 - 0.015 $µg/mL$ | CLSI |
| Pseudomonas aeruginosa ATCC 27853 | 2-8 $µg/mL$ | CLSI |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Vicki L. Kennedy Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
NOV 2 1 2006
K062945 Re: Trade/Device Name: BD Phoenix" M Automated Microbiology System Ertapenem (0.25-32 µg/mL) - Gram Negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: September 27, 2006 Received: September 28, 2006
Dear Ms. Kennedy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Sally, attorn
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: KO6 2945
Device Name: BD PhoenixTM Automated Microbiology System for use with the antimicrobial agent ertapenem (0.25-32 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent ertapenem at concentrations of 0.25-32 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Erapenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli Klebsiella pneumoniae
Active In Vitro:
Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Enterobacter cloacae
Prescription Use V (Per 21 CFR 801.109) Klebsiella oxytoca Morganella morganii Proteus mirabilis Proteus vulgaris
Providencia rettgeri Providencia stuartii Serratia marcescens
Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Luddie L. Poole
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
SIO(E) K062945
BD Diagnostic Systems Becton, Dickinson and Company
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”