AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator wayeforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Device Description

The INOmax DS provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS uses a "dual-channel" design to ensure the safe delivery of INOmax.

The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.

The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder regulator is attached via tubing to the INOmax DS using one of the two NO/N2 quick connect inlets on the back of the machine.

The INOmax enters the back of the INOmax DS, passes through a filter. then a safety shutoff valve, which is open under normal operation.

An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.

A flow sensor inside the INOmax DS also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.

The INOmax DS gas monitoring system provides monitored values for inspired NO, NO2, and O2. The sample gas is withdrawn form the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.

The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of the gas present.

If the delivery system does go into shut down, the INOmax DS has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

AI/ML Overview

The INOmax DS is a Nitric Oxide Administration Apparatus designed to deliver INOmax (nitric oxide for inhalation) therapy gas to neonates in clinical settings such as the Neonatal Intensive Care Unit (NICU) and during transport. The device ensures a constant concentration of nitric oxide (NO) as set by the user, throughout the inspired breath, utilizing a dual-channel design for delivery and monitoring. It also includes integrated monitoring of O2, NO2, and NO, and a comprehensive alarm system. A key safety feature is a 6-hour battery backup and a fixed-flow backup NO delivery capability.

Here's an analysis of its acceptance criteria and the study proving it:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/SpecificationAcceptance Criteria (Predicate: Ohmeda INOvent Delivery System)INOmax DS Reported Performance
Indications for UseSimilar to INOmax DSExcept battery backup is 6 hours and provides an integrated backup system for delivering a fixed flow of NO into a constant flow of breathing circuit gas to produce a constant concentration of NO. (Also includes continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. Backup NO delivery is 250 mL/min of NO which, with 10 L/min of oxygen, provides 20 ppm; can also use INOblender for backup.) Target population: neonates; Clinical setting: NICU and transport.
Physical DimensionsHeight 215 mm, Width 350 mm, Depth 430 mm, Weight 21 KgHeight 220 mm, Width 350 mm, Depth 160 mm, Weight 5.5 Kg
Ventilator CompatibilitySimilar to INOmax DSExcept 2-60 L/min for inspiratory flow for neonatal and 4-120 L/min for adult.
NO Delivery ControlSimilar to INOmax DS (implicitly, the INOvent delivers a constant concentration as set by the user via an injector module)NO set resolution is 0.1 / 1 / 5 ppm depending on range, accuracy is +/- 20% indicated or 2 ppm whichever is greater, and NO inlet pressure is 1.7 to 2.3 Bar (25 to 33 psig). (Delivers a constant concentration of NO as set by the user, throughout the inspired breath, tracking ventilator waveforms).
NO CylinderSame as INOmax DSSame (INOmax ™)
Gas MonitoringSimilar to INOmax DS (implicitly, includes O2, NO2, and NO monitoring)NO range resolution is 0 to 10ppm +/- (20% of reading +0.5 ppm) 10 to 100ppm +/- (10% of reading +0.5 ppm), calibration span is daily zero with span at pre-use test if needed, and NO accuracy is +/- 20% of reading +0.5 ppm. (Continuously monitors inspired O2, NO2, and NO).
Injector ModuleSame as INOmax DSSame (Enables tracking of ventilator waveforms and synchronized/proportional dose of NO).
NO Delivery Shut OffSame as INOmax DSSame (dual-channel design with monitoring system able to shutdown INOmax delivery if fault detects NO concentration > 100 ppm).
Calibration Gas CylindersNO Cal gas 40 - 80 ppm +/- 4%, NO₂ Cal gas 10 -15 ppm +/- 10%INOcal calibration gas, INO Therapeutics LLC; NO Cal Gas 45 ppm +/- 4%, NO₂ Cal Gas 10 ppm +/- 10%
Electrical SpecificationsSimilar toINOMax DSExcept no nurse call and battery back-up is 6 hours.
Environmental SpecsSimilar to INOmax DSExcept ambient operation and storage pressure are 57 to 110 kPa (430 to 825 mmHg).
NO Back-up Delivery20 ppm @ 15 L/min0.25 L/min when delivered into 10 L/min provides 20 ppm, plus INOblender may also be used.
Safety and EffectivenessImplied to be safe and effective for its intended uses by the predicate device (Ohmeda INOvent Delivery System), which was legally marketed prior to May 28, 1976, or reclassified.The INOmax DS met its design input specifications, design output specifications, and risk analysis requirements, as demonstrated by performance bench testing, integrity testing, environmental and mechanical testing, validation against ventilators, and product standards certification testing. Deemed safe and effective, and "as safe and effective as the predicate INOvent Delivery System, while performing as well or better than this predicate device."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set" in the context of clinical or patient data. The testing described includes:

  • Performance bench testing
  • Integrity testing
  • Environmental and mechanical testing
  • Validation against ventilators testing
  • Product standards certification testing

These are technical engineering and laboratory tests, not tests with a human patient cohort. Therefore, concepts like sample size, country of origin, retrospective/prospective data provenance (for patient data) are not applicable here. The data provenance would be from internal lab testing by INC Therapeutics.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a piece of medical equipment (Nitric Oxide Administration Apparatus). The "ground truth" for its performance is established through objective engineering specifications, performance standards (e.g., IEC 60601-1), and comparisons to a predicate device's measured performance. Validation is against these technical specifications, not against expert medical diagnoses or interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable for the type of technical performance testing described. Adjudication methods are typically used in clinical studies for ambiguous cases, often involving multiple experts to reach a consensus on diagnostic interpretations. The performance of the INOmax DS is measured against defined technical parameters and standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The INOmax DS is a medical device for delivering therapy gas and monitoring physiological parameters, not an AI-powered diagnostic tool requiring human reader interpretation of images or other data. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are irrelevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable as the device is not an algorithm or AI system. It is a standalone medical device that performs its intended function independently, delivering nitric oxide and monitoring gases. While a human operator sets parameters and monitors the device, the core function (gas delivery, monitoring, alarms) is performed by the device itself (algorithm only without human-in-the-loop performance is a concept for AI driven diagnostics).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the INOmax DS's performance is based on:

  • Design input specifications: Pre-defined technical requirements for the device.
  • Design output specifications: The technical characteristics and performance generated by the design process.
  • Risk analysis requirements: Ensuring safety and mitigating identified risks.
  • Applicable product standards: Compliance with industry standards like FDA Guidance, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, CGA V-1, ASTM F-1462-93, ASTM F-1054-87.
  • Predicate device performance: Comparison to the established performance of the Ohmeda INOvent Delivery System.

These are primarily objective, quantitative engineering and regulatory criteria, rather than clinical 'ground truth' derived from patient outcomes or pathology.

8. The sample size for the training set

This is not applicable. The INOmax DS is a hardware-based medical device with integrated software/firmware for control and monitoring, not a machine learning or AI model. Therefore, there is no "training set" in the context of data used to train an algorithm.

9. How the ground truth for the training set was established

This is not applicable for the reasons outlined in point 8. The device's functionality is designed, engineered, and then verified against its specifications and relevant standards through testing.

{0}------------------------------------------------

K 061901

Image /page/0/Picture/1 description: The image shows the logo for INC Therapeutics. The letters "INC" are in bold, black font, with the "C" partially encircling the word "Therapeutics." The word "Therapeutics" is in a smaller, black font and is underlined.

7601-B Murphy Drive
Middleton, WI 53562 USA Tel. 608-831-4401 / Fax. 608-831-4409 DEC 1 4 2006

Date:September 18, 2006
Contact Person:Frederick Montgomery, Vice President, Medical Devices
Subject:510(k) Summary of Safety and Effectiveness Information for theINO Therapeutics INOmax DS. This summary of safety and effectivenessinformation is being submitted in accordance with the requirements ofSMDA 1990 and 1992.
Proprietary Name:INOmax DS
Common Name(s):Nitric Oxide Administration Apparatus (primary)Nitric Oxide Administration Apparatus, Back-up SystemNitric Oxide AnalyzerNitrogen Dioxide Analyzer
Classification(s):Class II - 21CFR868.5165, MROClass II - 21CFR868.5165, MRN (primary)Class II - 21CFR868.2380, MRPClass II - 21CFR868.2385, MRQ
Panel:Anesthesiology
Predicate Device(s):Ohmeda INOvent Delivery System (now GE Healthcare) (primary)510(k) Number K974562Datex-Ohmeda INOvent Delivery System (now GE Healthcare)510(k) Number K000186INO Therapeutics INOblender510(k) Number K052663Pulmonox Vianox Delivery System (now SensorMedics)510(k) Number K023014Pulmonox Aeronox Delivery System (now SensorMedics)

510(k) Number K000653

{1}------------------------------------------------

Patient Population:The targeted patient population is neonates. This is derived from theINOmax drug package labeling.
Clinical Setting:The primary targeted clinical setting is the Neonatal Intensive Care Unit(NICU) and secondary targeted clinical setting is the transport of neonatesThis is derived from the INOmax drug package labeling.
Technology:The technology utilized in the INOmax DS is very similar to the Datex-Ohmeda INOvent Delivery System (now GE Healthcare), with many of

the subsystems and components being unchanged.

SpecificationINOvent Delivery SystemINOmax DS
Indications forUseSimilar to INOmax DSExcept battery backup is 6 hours and provides anintegrated backup system for delivering a fixed flow ofNO into a constant flow of breathing circuit gas toproduce a constant concentration of NO.
PhysicalDimensionsHeight 215 mmWidth 350 mmDepth 430 mmWeight 21 KgHeight 220 mmWidth 350 mmDepth 160 mmWeight 5.5 Kg
VentilatorCompatibilitySimilar to INOmax DSExcept 2-60 L/min for inspiratory flow for neonatal and4-120 L/min for adult.
NO DeliverySimilar to INOmax DSExcept NO set resolution is 0.1 / 1 / 5 ppm depend onrange, accuracy is +/- 20% indicated or 2 ppmwhichever greater, and NO inlet pressure is 1.7 to 2.3Bar (25 to 33 psig).
NO CylinderINOmax ™Same as INOmax DSSame
Gas MonitoringSimilar to INOmax DSExcept NO range resolution is 0 to 10ppm +/- (20% ofreading +0.5 ppm) 10 to 100ppm +/- (10% of reading+0.5 ppm), calibration span is daily zero with span atpre-use test if needed, and NO accuracy is+/- 20% of reading +0.5 ppm.
Injector ModuleSame as INOmax DSSame
NO deliveryShut OffSame as INOmax DSSame
Calibration GasCylindersNO Cal gas 40 - 80 ppm +/- 4%NO₂ Cal gas 10 -15 ppm +/- 10%INOcal calibration gas, INO Therapeutics LLCNO Cal Gas 45 ppm +/- 4%NO₂ Cal Gas 10 ppm +/- 10%
ElectricalSpecificationsSimilar to INOmax DSExcept no nurse call and battery back-up is 6 hours.
EnvironmentalSpecificationsSimilar to INOmax DSExcept ambient operation and storage pressure are 57to 110 kPa (430 to 825 mmHg) and
NO Back-upDelivery20 ppm @ 15 L/min0.25 L/min when delivered into 10 L/min provides 20ppm, plus INOblender may also be used

Material(s)

The materials selected were based upon the Datex-Ohmeda INOvent Delivery System (now GE Healthcare) and the INOblender.

{2}------------------------------------------------

The INOmax DS provides a constant dose of INOmax (nitric oxide) Device Description: therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS uses a "dual-channel" design to ensure the safe delivery of INOmax.

The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.

The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder regulator is attached via tubing to the INOmax DS using one of the two NO/N2 quick connect inlets on the back of the machine.

The INOmax enters the back of the INOmax DS, passes through a filter. then a safety shutoff valve, which is open under normal operation.

An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.

A flow sensor inside the INOmax DS also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.

The INOmax DS gas monitoring system provides monitored values for inspired NO, NO2, and O2. The sample gas is withdrawn form the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.

{3}------------------------------------------------

The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of the gas present.

If the delivery system does go into shut down, the INOmax DS has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

The INOmax DS delivery system delivers INOmax® (nitric oxide for Indications for Use: inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

{4}------------------------------------------------

Testing:The testing indicated the INOmax DS met its design input specifications,design output specifications and risk analysis requirements. Testingcompleted included:Performance bench testingIntegrity testingEnvironmental and mechanical testingValidation against ventilators testingProduct standards certification testing
Standard(s):The INOmax DS was designed to comply with the applicable portions ofthe following product standards
1. FDA Guidance Document for Premarket Notification Submissionsfor Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer andNitrogen Dioxide Analyzer (Special Controls).
2. IEC 60601-1, plus amendments: Medical Electrical Equipment
3. IEC 60601-1-2: EMC/EMI
4. IEC 60601-1-4: Programmable Systems
5. IEC 60601-1-8: Alarms
6. CGA V-1: Medical Cylinder Connections
7. ASTM F-1462-93: Oxygen Analyzers
8. ASTM F-1054-87: Conical Fittings
Conclusion:As a result of the testing of the INOmax DS, INO therapeutics deems thedevice safe and effective of its indicated uses. INO Therapuetics furtherdeems the INOmax DS, based on testing, to be as safe and effective as thepredicate INOvent Delivery System, while performing as well or betterthan this predicate device.

{5}------------------------------------------------

Image /page/5/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Frederick Montgomery Vice President, Medical Devices INO Therapeutics 7601-B Murphy Drive Middleton, Wisconsin 53562

DEC 1 4 2006

Re: K061901

Trade/Device Name: INO Therapeutics INOmax DS Regulation Number: 21 CFR 868,5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN, MRO, MRP and MRQ Dated: December 4, 2006 Received: December 5, 2006

Dear Mr. Montgomery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Montgomery

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Ching-Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

INO Therapeutics INOmax DS

Indications For Use:

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator wayeforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Prescription Use (Part 21 CFR 801 Subpart D) and/or

Over-the-Counter-Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michael Nichols 12/14/06

Page 1 of 1

195

§ 868.5165 Nitric oxide administration apparatus.

(a)
Identification. The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”