K Number
K061837
Device Name
NXSTAGE CARTRIDGE EXPRESS
Date Cleared
2006-07-31

(32 days)

Product Code
Regulation Number
876.5860
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The NxStage System One is indicated for the treatment of acute and chronic renal failure or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home. All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.
Device Description
The NxStage System One consists of the NxStage Cycler and the NxStage Cartridge Extracorporeal Blood and Fluid Circuit. The NxStage Cycler is an electro-mechanical device that interfaces with the NxStage Cartridge. The NxStage Cartridge is a single-use extracorporeal blood circuit and fluid management device that mounts integrally within the NxStage Cycler. The NxStage Cartridge Express is a single-use extracorporeal blood circuit and fluid management device with a pre-attached high flux (permeability) hollow-fiber filter.
More Information

Not Found

No
The provided text describes an electro-mechanical device for renal failure treatment and does not mention any AI or ML components or capabilities.

Yes
The device is indicated for the treatment of acute and chronic renal failure or fluid overload, which are therapeutic medical interventions.

No.
The device is indicated for the treatment of renal failure or fluid overload, not for diagnosis.

No

The device description explicitly states that the system consists of an electro-mechanical device (NxStage Cycler) and a single-use extracorporeal blood and fluid circuit (NxStage Cartridge), indicating significant hardware components.

Based on the provided information, the NxStage System One is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use is for the treatment of renal failure and fluid overload using techniques like hemofiltration, hemodialysis, and ultrafiltration. These are therapeutic procedures performed on the patient's body, not tests performed on samples taken from the body.
  • Device Description: The device description details an electro-mechanical cycler and a cartridge that forms an extracorporeal blood and fluid circuit. This describes a system that processes blood and fluids outside the body for therapeutic purposes.
  • Lack of IVD Characteristics: There is no mention of analyzing samples (blood, urine, etc.) to provide diagnostic information about a patient's condition. The device's function is to filter and manage fluids, which is a treatment, not a diagnostic test.

IVD devices are specifically designed to examine specimens derived from the human body (like blood, urine, tissue) to provide information for the diagnosis, monitoring, or treatment of a disease or condition. The NxStage System One's purpose is to directly treat a condition by filtering blood and managing fluids.

N/A

Intended Use / Indications for Use

The NxStage System One is indicated for the treatment of acute and chronic renal failure or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Product codes (comma separated list FDA assigned to the subject device)

78 KDI, KDI

Device Description

The NxStage System One consists of the NxStage Cycler and the NxStage Cartridge Extracorporeal Blood and Fluid Circuit.

The NxStage Cycler is an electro-mechanical device that interfaces with the NxStage Cartridge. The NxStage Cartridge is a single-use extracorporeal blood circuit and fluid management device that mounts integrally within the NxStage Cycler.

Cartridge Express is a single-use extracorporeal blood circuit and fluid management device with a pre-attached high flux (permeability) hollow-fiber filter.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

acute or chronic care facility, home.
All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Design validation testing has been performed to ensure that the modified NxStage Cartridge Express meets design specifications.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K050727

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

This submission is a Special 510(k) Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), NxStage has provided certification of compliance to 21 CFR 820.30 Design Control requirements.

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

0

K061837

JUL 3 1 2006

Section 7

510(K) Summary of Safety Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The contents of this 510(k) summary have been provided in conformance with 21 CFR 807.92.

Date:June 28, 2006
Common/Usual Name:Dialyzer with High Permeability Hemodialysis
System
Trade/Proprietary Name:NxStage Cartridge Express
Classification Name:High Permeability Hemodialysis System (21 CFR
876.5860)
Device Classification:Class II
Product Code:78 KD I - Dialyzer, High Permeability with or
without Sealed Dialysate System
Device Panel:Gastroenterology-Urology (GU)/Gastro-Renal
(GRDB)
510(k) Sponsor &
Owner/Operator:NxStage Medical, Inc
439 South Union St, Suite 501
Lawrence, MA 01843
Owner/Operator No. 9045797
Establishment Registration #3003464075
Contact Person:Norma LeMay
Manager, Regulatory Affairs

Device Description:

The NxStage System One consists of the NxStage Cycler and the NxStage Cartridge Extracorporeal Blood and Fluid Circuit.

The NxStage Cycler is an electro-mechanical device that interfaces with the NxStage Cartridge. The NxStage Cartridge is a single-use extracorporeal blood circuit and fluid management device that mounts integrally within the NxStage Cycler. The NxStage

Confidential

1

Cartridge Express is a single-use extracorporeal blood circuit and fluid management device with a pre-attached high flux (permeability) hollow-fiber filter.

Substantial Equivalence:

This submission is a Special 510(k) Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), NxStage has provided certification of compliance to 21 CFR 820.30 Design Control requirements. Design validation testing has been performed to ensure that the modified NxStage Cartridge Express meets design specifications. The modified NxStage Cartridge Express has been compared to the baseline as cleared through K050727 and found to be substantially equivalent.

Conclusion:

Based on the device indications for use, comparison of descriptive and technological characteristics, and design control certification, the modified NxStage Cartridge Express has been shown to meet the minimum requirements that are considered acceptable for its intended use and is substantially equivalent to the baseline device.

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle, emphasizing the department's name and national affiliation.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

JUL 3 1 2006

Ms. Norma LeMay Manager. Regulatory Affairs NxStage Medical, Inc. 439 South Union Street, Suite 501 LAWRENCE MA 01843

Re: K061837

Trade/Device Name: NxStage System One with Cartridge Express Regulation Number: 21 CFR $876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: June 28, 2006 Received: June 29, 2006

Dear Ms. LeMay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a logo with the letters FDA in bold, surrounded by the words "Centennial" and "1906-2006". The logo is encircled by text that reads "In Compliance Protection of the Public Health". Below the logo, the words "Protecting and Promoting Public Health" are written in a decorative font.

3

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

INDICATIONS FOR USE

510(k) Number (if known):

K061839

Device Name:

NxStage System One with Cartridge Express

The NxStage System One is indicated for the treatment of Indications for Use: acute and chronic renal failure or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility. The System is also indicated for hemodialysis with or without ultrafiltration in the home.

All treatments must be administered under physician's prescription, and must be observed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Ingram
Division Sign-Off)

Reproductive, Abdominal, and 510(k) Number

Page 1 of 1