K Number
K061675
Manufacturer
Date Cleared
2006-08-02

(49 days)

Product Code
Regulation Number
870.1025
Panel
CV
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vision V x C series 4300 central station monitoring system is intended to be used to provide, using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia/ST monitoring for a variable number of Escort Bedside Monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.

The ST algorithm has been tested for accuracy of the ST segment measurement data. The significance of the ST segment changes must be determined by a physician.

Device Description

The Vision VxC Central Station is designed specifically to provide centralized display for up to 16 patients, storage and recording (or printing) of patient vital sign and waveform data that are being monitored at the bedside by Invivo monitors and telemetry devices.

The Vision VxC Central Station can provide alarm detection and reporting for all vital sign parameters available to the central station for patient alarm surveillance. This alarm surveillance includes alarms reported by the bedside monitors and repeated to the central station as well as primary alarm surveillance for the patient worn WMTS telemetry transmitter device where there is no alarm notification capability on the transmitter worn by the patient.

Arrhythmia monitoring and ST segment detection capability is available as an option. The arrhythmia feature is equipped with password protection to prevent unauthorized users from turning off arrhythmia when a patient is being monitored. When a patient is monitored by arrhythmia the system will provide continuous monitoring of lifethreatening alarms.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Vision VxC Central Station, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Requirement)Reported Device Performance (Results)
Central Station
Display TypeFlatpanel 19" LCDPass
Central ProcessorIntel PentiumPass
User InterfaceTouchscreen and/or mousePass
Operating SystemMicrosoft Windows XP EmbeddedPass
Number of Patients Monitored1 to 16 (up to 16 designated for telemetry)Pass
Parameters MonitoredECG, Resp, IBP (SYS, DIA, MEAN), NIBP (SYS, DIAS, MEAN), SpO2, ETCO2, TempPass
Max Parameters Monitored20Pass
TrendingTabular for 16 patients, all parameters; up to 72 hours at 1, 2, 3, 4, 5, 15, 30, 60, 120, and 180 min. intervalsPass
Alarm History Storage1000 events within 72 hours per patient; 20 sec per eventPass
DocumentationThermal array recorder and/or laser printerPass
AC Main90-130/180-260 VAC, 47-63 Hz selectable, 6 amps@115VPass
Power Supply235 WPass
Operating Temperature10 to 40° CPass
Storage Temperature-40 to 75° CPass
Relative Humidity5 to 95 %Pass
StandardsUL 60950Pass
Telemetry Receiver Platform
Alarm History Storage1000 events within 72 hours per patient; 20 sec per eventPass
DocumentationThermal array recorder and/or laser printerPass
AC Main115/230 VAC, 60/50 Hz selectable, 4 amps@115VPass
Power Supply235 WPass
Input Voltage100-120 VAC / 200 - 240 VAC, selectable, 50/60 HzPass
Input Current6 Amp max @ 115V (20A Max inrush cold start), 3 Amp max @ 230 V (10A Max inrush cold start)Pass
Telemetry BandFCC WMTS (608-614 MHz)Pass
Operating Temperature10 to 40° CPass
Relative Humidity10 to 90 %Pass
StandardsUL 60950, FCC Part 15 (Spread Spectrum)Pass
Arrhythmia Analysis Option
Number of Arrhythmia Channels1 to 16 (dual vector)Pass
Types of Detected EventsAsystole, VFIB, VTACH, Couplet, High and Low Heart Rate, High Abnormal Count, Bigeminy, Trigeminy, V.RUN, V.Rhythm, Multi-Focal, R-ON-T, PausePass
Type of AlgorithmHeuristic algorithm using template matching and feature extractionPass
QRS Detection SensitivityAHA ≈99.88%, MIT ≈99.93%Pass
QRS Detection Positive PredictivityAHA ≈99.89%, MIT ≈99.85%Pass
PVC Detection SensitivityAHA ≈94.07%, MIT ≈95.44%Pass
PVC Detection Positive PredictivityAHA ≈97.72%, MIT ≈96.60%Pass
PVC Detection False Positive RateAHA ≈0.22%, MIT ≈0.23%Pass
Alarm Displaysarrhythmia alarm must displayPass
Alarm Tonesarrhythmia alarm must soundPass
Alarm Recordingsarrhythmia alarm must generate an alarm recording if configuredPass
Alarm Event Historyarrhythmia alarm must generate event history recordPass
Maximum Patient Load16 patientPass
StandardAAMI/ANSI EC 57: 1998Compliant
ST Analysis Option
Number of ST Channels1 to 16, but no more than arrhythmia channelsPass
Alarmshigh and low for both vectors; can be recorded and/or stored as eventsPass
StandardAAMI/ANSI EC 57: 1998Compliant

2. Sample Size Used for the Test Set and Data Provenance

The document states, "This device was validated using patient simulators under simulated use conditions." This indicates that the test set did not consist of real patient data. The provenance for this simulated data is not specified (e.g., country of origin). The testing would be prospective in the sense that the device was evaluated against controlled simulated conditions, but not retrospective using historical clinical patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Given that "patient simulators" were used, there were likely no human experts required to establish "ground truth" in the traditional sense of clinical diagnosis. The "ground truth" would have been the pre-programmed and known states of the patient simulator, which were then compared against the device's measurements and detections. The qualifications of the individuals who programmed and operated the simulators are not specified.

4. Adjudication Method for the Test Set

No adjudication method is mentioned for the test set, as the evaluation was against known simulator outputs rather than human interpretation that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The performance data focuses on the device's technical specifications and algorithm accuracy (for arrhythmia and ST analysis) against predefined standards and simulated conditions, not on comparing human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance evaluation was done, particularly for the arrhythmia and ST analysis algorithms. The document explicitly lists performance metrics for QRS and PVC detection (Sensitivity, Positive Predictivity, False Positive Rate) based on the AHA and MIT databases. This demonstrates the algorithm's performance independent of human interaction.

7. The Type of Ground Truth Used

For the arrhythmia analysis, the ground truth was based on established benchmarks from the AHA (American Heart Association) and MIT (Massachusetts Institute of Technology) databases. These databases contain annotated ECG recordings where events (like QRS complexes and PVCs) have been meticulously identified and marked, often through expert review and consensus. For other general device functions (display, processing, etc.), the ground truth would be conformance to the stated technical specifications and environmental standards.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for the training set used for the arrhythmia and ST analysis algorithms. It mentions the "current arrhythmia detection and ST segment analysis algorithm library was replaced with the Mortara Instrument Incorporated product." It further states that "This same library is used in the Welch Allyn Acuity Central Station, which was cleared to market via 510(k) K022453." While the training set data is not provided, the implication is that the Mortara algorithm was trained on a significant dataset to achieve the reported performance on standard benchmarks.

9. How the Ground Truth for the Training Set Was Established

The document does not directly describe how the ground truth for the training set of the Mortara algorithms was established. However, given that these algorithms are evaluated against AHA and MIT databases for their "ground truth" performance metrics (as seen in the "Arrhythmia Analysis Option" table), it can be inferred that the training process would have also relied on highly curated and expert-annotated ECG datasets similar to, or including, segments of these recognized benchmark databases. The "heuristic algorithm using template matching and feature extraction" also points to an approach that would benefit from large, labeled datasets for developing and refining performance.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.