AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

Device Description

The Expeditionary Deployable Oxygen Concentration System (EDOCS) is a family of devices that draw in normal air and produce oxygen. The EDOCS separates nitrogen from the air, producing an output of concentrated oxygen at 93 percent by the molecular sieve process. The final product contains not less than 90 percent and not more than 96 percent oxygen, by volume. The remainder of the product consists mostly of argon and nitrogen. EDOCS employs an integral oxygen analyzer to provide control of oxygen purity. The EDOCS device is an electromechanical device consisting primarily of a molecular sieve type oxygen concentrator, a compressor module(s), and one or more oxygen collection cylinders. The oxygen concentrator operates by adsorbing water and nitrogen from filtered air. Depending on the unit size, the resulting gas has increased oxygen at flow rates from 30 to 500 liters per minute at a pressure of 50 - 100 psi nominal and is capable of high pressure cylinder filling.

The Mobile Oxygen Storage Tank (MOST) is a oxygen storage and distribution system consisting of lightweight, high-pressure, composite wound, brass lined cylinders and a distribution manifold secured within a rugged waterproof case. The MOST contains 10,000 liters of oxygen when filled to 2,250 psig (13,400 liters at 3,000 psig). The MOST has a maximum flow capability of 225 LPM, which is equivalent to 75 LPM per Oxygen Output fitting (3 fittings per MOST). These three oxygen output flows would typically be subdivided further by the customer's equipment.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Expeditionary Deployable Oxygen Concentration System (EDOCS) and Mobile Oxygen Storage Tank (MOST), based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

DevicePerformance AreaAcceptance Criteria (Source)Reported Device Performance
EDOCSOxygen PurityUSP grade 93% +/- 3% oxygen (USP 24-NF 19, US Army specifications)"The final product contains not less than 90 percent and not more than 96 percent oxygen, by volume." (Meets 93% ± 3%) / "In all instances the device met all required performance criteria and functioned as intended."
Flow RateUS Army specifications (USAMMA), voluntary standards"In all instances the device met all required performance criteria and functioned as intended."
Electrical SafetyIEC 60601-1-2 (Electromagnetic Compatibility tests)"has passed the required Electromagnetic Compatibility tests."
MechanicalUS Army specifications (USAMMA), voluntary standards"In all instances the device met all required performance criteria and functioned as intended."
ControlsUS Army specifications (USAMMA), voluntary standards"In all instances the device met all required performance criteria and functioned as intended."
Device PerformanceUS Army specifications (USAMMA), voluntary standards"In all instances the device met all required performance criteria and functioned as intended."
MOSTOxygen PurityUSP grade 93% (+7%/-3%) oxygenIntended to dispense USP grade 93% (+7%/-3%) oxygen. "In all instances the device met all required performance criteria and functioned as intended."
Flow RateUSAF specifications (AFMESA)Intended to have a maximum flow capability of 225 LPM. "In all instances the device met all required performance criteria and functioned as intended."
MechanicalDOT-CFFC, NASA, MIL-STD-810F"In all instances the device met all required performance criteria and functioned as intended."
ControlsUSAF specifications (AFMESA), DOT-CFFC, NASA, MIL-STD-810F"In all instances the device met all required performance criteria and functioned as intended."
Device PerformanceUSAF specifications (AFMESA), DOT-CFFC, NASA, MIL-STD-810F"In all instances the device met all required performance criteria and functioned as intended."
SafetyISO 14971 (Risk Analysis)"all identified risks have been managed to the degree possible via design, engineering controls or labeling." / "In all instances the device met all required performance criteria and functioned as intended."
Storage Capacity10,000 liters of oxygen when filled to 2,250 psig (13,400 liters at 3,000 psig)Claims adherence to this capacity. "In all instances the device met all required performance criteria and functioned as intended."

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not specify the exact sample size for the test sets. It only mentions "Verification and validation testing activities were conducted."
  • Data Provenance: The data provenance is not specified. It does not indicate the country of origin or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • The document does not mention the use of experts or any process for establishing a "ground truth" in the context of human interpretation or diagnostic accuracy. The testing appears to be focused on device performance against engineering and regulatory standards rather than subjective human assessment.

4. Adjudication Method for the Test Set:

  • Not applicable as the testing described is not related to subjective assessments requiring adjudication (e.g., image interpretation). The evaluation is based on objective measurements against predefined technical specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study typically involves human readers evaluating cases with and without AI assistance to assess changes in diagnostic performance. The devices in question (oxygen concentrator and storage tank) are physical medical devices, not diagnostic AI algorithms.

6. Standalone (Algorithm Only) Performance Study:

  • Yes, in essence. The performance testing described for both EDOCS and MOST relates to the standalone performance of the devices themselves, without direct human-in-the-loop intervention for the performance evaluation. The devices generate or store oxygen, and their output and characteristics are measured directly against predefined criteria.

7. Type of Ground Truth Used:

  • The "ground truth" for the performance testing of these devices is based on objective technical specifications and accepted industry/regulatory standards. These include:
    • USP grade 93% +/- 3% oxygen (United States Pharmacopeia) for purity.
    • US Army specifications (USAMMA) and USAF specifications (AFMESA) for various performance aspects.
    • Voluntary standards.
    • IEC 60601-1-2 for Electrical Safety (Electromagnetic Compatibility).
    • ISO 14971 for Risk Analysis.
    • DOT-CFFC, NASA, and MIL-STD-810F for mechanical and safety aspects.

8. Sample Size for the Training Set:

  • Not applicable. These are physical medical devices, not AI algorithms that require a "training set" in the machine learning sense. The devices' operational parameters are set during manufacturing and design, not learned from data.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle with the letters "PCI" embedded within it. To the right of the logo is the text "Pacific Consolidated Industries" in a bold, sans-serif font.

K061414

510(k) Summary

JAN 2 2 2007

Expeditionary Deployable Oxygen Concentration System (EDOCS) & Mobile Oxygen Storage Tank (MOST)

Type of FDA SubmissionTraditional 510(k)
Submitter Information
Submitter's Name:Pacific Consolidated Industries
Submitter's Address:12201 Magnolia AvenueRiverside, California 92503-4820
Owner/Operator ID:9049531
Contact Person:Gary W. Clark
Submitter's Phone:(951) 479-0872 (Phone)(951) 479-0861 (Fax)
Date of Preparation:January 12, 2007
Device Classification
Classification Name:Portable Oxygen Generator
Device Classification:868.5440
Product Code:CAW
Panel:Anesthesiology and Respiratory Therapy Devices

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle with the letters "P", "C", and "I" incorporated into the design. To the right of the logo is the text "Pacific Consolidated Industries" in a bold, sans-serif font.

Expeditionary Deployable Oxygen Concentration System (EDOCS)

Predicate Legally Marketed Device Equivalence

Substantial equivalence is claimed to the legally marketed predicate device previously submitted under K020330 - Deployable Oxygen Concentration System (DOCS).

Safe Medical Devices Act (SMDA) Statement

The oxygen supplied by the oxygen concentrator is supplemental and is not considered to be life supporting or life sustaining. The failure of the device would not have serious health consequences to the user.

Intended Use of Device

The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to . generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

Description of Device

The Expeditionary Deployable Oxygen Concentration System (EDOCS) is a family of devices that draw in normal air and produce oxygen. The EDOCS separates nitrogen from the air, producing an output of concentrated oxygen at 93 percent by the molecular sieve process. The final product contains not less than 90 percent and not more than 96 percent oxygen, by volume. The remainder of the product consists mostly of argon and nitrogen.

EDOCS employs an integral oxygen analyzer to provide control of oxygen purity.

The EDOCS device is an electromechanical device consisting primarily of a molecular sieve type oxygen concentrator, a compressor module(s), and one or more oxygen collection cylinders. The oxygen concentrator operates by adsorbing water and nitrogen from filtered air. Depending on the unit size, the resulting gas has increased oxygen at flow rates from 30 to 500 liters per minute at a pressure of 50 - 100 psi nominal and is capable of high pressure cylinder filling.

Device Labeling

The EDOCS is clearly labeled with the following:

  • 93% ±3% Oxygen .
  • Contraindications, for example, oil-rich or flammable atmosphere .
  • Rx Only

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized triangle on the left, with the words "Pacific Consolidated Industries" to the right of the triangle. The triangle is made up of three horizontal lines, with the top line being the shortest and the bottom line being the longest.

Comparison of Technological Characteristics

The technological characteristics of these devices and their intended use to supply supplemental oxygen are the same as the predicate devices and raise no new questions of safety and effectiveness.

Special Controls/Conformance to Recognized Standards

The Expeditionary Deployable Oxygen Concentration System (EDOCS) conforms to the recognized standard USP 24-NF 19 (through Supplement Four, December 31, 2001) and employs an integral oxygen analyzer to provide control of oxygen purity in accordance with the standard.

The EDOCS has been tested under IEC 60601-1-2 and has passed the required Electromagnetic Compatibility tests.

A Risk Analysis, as described in ISO 14971, has been conducted on the EDOCS and all identified risks have been managed to the degree possible via design, engineering controls or labeling.

Summary of Performance Testing

Verification and validation testing activities were conducted to establish the performance and reliability characteristics of the EDOCS, to demonstrate performance as intended and substantial equivalency to predicate device. Testing involved the following areas:

  • Purity I
  • 피 Flow Rate
  • Electrical Safety ■
  • 트 Mechanical
  • 해 Controls
  • 피 Device Performance

Acceptance criteria were based on US Army specifications (USAMMA), and those established in voluntary standards.

In all instances the device met all required performance criteria and functioned as intended, meeting the acceptance criteria.

Conclusions

In summary, Pacific Consolidated Industries has demonstrated that the EDOCS meets its specifications, is safe and effective for its intended use, and is substantially equivalent to the currently marketed devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized letter "A" on the left, followed by the company name "Pacific Consolidated Industries" in bold, sans-serif font. The letter "A" is formed by three horizontal lines stacked on top of each other, creating a triangular shape.

Mobile Oxygen Storage Tank (MOST)

Predicate Legally Marketed Device Equivalence

Substantial equivalence is claimed to the legally marketed predicate device previously submitted under K040738 - Mobile Oxygen Storage Tank (MOST).

Safe Medical Devices Act (SMDA) Statement

The oxygen supplied by the MOST is supplemental and is not considered to be life supporting or life sustaining. The failure of the device would not have serious health consequences to the user.

Description of Device

The Mobile Oxygen Storage Tank (MOST) is a oxygen storage and distribution system consisting of lightweight, high-pressure, composite wound, brass lined cylinders and a distribution manifold secured within a rugged waterproof case. The MOST contains 10,000 liters of oxygen when filled to 2,250 psig (13,400 liters at 3,000 psig). The MOST has a maximum flow capability of 225 LPM, which is equivalent to 75 LPM per Oxygen Output fitting (3 fittings per MOST). These three oxygen output flows would typically be subdivided further by the customer's equipment.

The small footprint, relatively light weight and easy carrying of the MOST system make it optimal for emergency use and military deployments. Eight handles, two on each long side of the MOST and two on the top at each end are provided for ease of transportation. The MOST contains the Apparatus Kit, which includes the equipment required to distribute the oxygen to up to three patients concurrently. The exterior dimensions of the MOST are: a length of 37.25", a width of 27.18", and height of 15.44." The MOST weighs approximately 200 pounds when charged to 2,250 psig.

Intended Use of Device

The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50 psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

Device Labeling

The MOST is clearly labeled with the following:

  • 93% +7%1-3% Oxygen .
  • Contraindications, for example, oil-rich or flammable atmosphere
  • Rx Only

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Pacific Consolidated Industries. The logo consists of a stylized letter "A" on the left, followed by the company name in a sans-serif font. The letter "A" is formed by three horizontal lines connecting the two sides of the triangle.

Comparison of Technological Characteristics

There are no significant changes in the MOST unit in this submission from that of the original submission.

Special Controls/Conformance to Recognized Standards

The Mobile Oxygen Storage Tank (MOST) conforms to the DOT-CFFC standard.

A Risk Analysis, as described in ISO 14971, has been conducted on the MOST and all identified risks have been managed to the degree possible via design, engineering controls or labeling.

Summary of Performance Testing

Verification and validation testing activities were conducted to establish the performance, reliability and safety characteristics of the MOST and to demonstrate performance as intended. Testing involved the following areas:

  • Purity
  • Flow Rate
  • Mechanical
  • 트 Controls
  • Device Performance
  • Safety

Acceptance criteria were based on USAF specifications (AFMESA), DOT-CFFC, NASA and MIL-STD-810F.

In all instances the device met all required performance criteria and functioned as intended, meeting the acceptance criteria.

Conclusions

In summary, Pacific Consolidated Industries has demonstrated that the MOST meets its specifications, is safe and effective in providing a portable supply of USP oxygen to the intended markets, and is substantially equivalent to the currently marketed device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, clutching a staff with a snake wrapped around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 2 2007

Mr. Gary Clark Vice President-Regulatory Affairs/ Ouality Assurance Pacific Consolidated Industries, LLC 12201 Magnolia Avenue Riverside, California 92503-4820

Re: K061414

Trade/Device Name: Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) Regulation Number: 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: January 12, 2007 Received: January 16, 2007

Dear Mr. Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Clark

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Statement of Indications for Use

Name: Expeditionary Deployable Oxygen Concentration System (EDOCS-120B)

The Expeditionary Deployable Oxygen Concentration System (EDOCS-120B) is intended to generate and deliver USP grade 93% +/- 3% oxygen for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

Name: Mobile Oxygen Storage Tank (MOST)

The Mobile Oxygen Storage Tank (MOST) is intended to store and dispense USP grade 93% (+7%/-3%) oxygen at 50psig nominal pressure for supplemental oxygen use. This device is to be used only by trained personnel in disaster relief situations where bottled oxygen is not readily available.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801.109)

Mell Zelen

Anesthesiology, General Hospital
Pain Control, Cranial Devices
Number: K061414

§ 868.5440 Portable oxygen generator.

(a)
Identification. A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).(b)
Classification. Class II (performance standards).