K Number
K061343
Device Name
INSPIRSTAR GLOVE ELECTRODE
Manufacturer
Date Cleared
2006-05-31

(16 days)

Product Code
Regulation Number
882.1320
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Inspirstar Glove Electrode is intended to for use with legally marketed electrostimulation devices such as TENS to deliver stimulation current for pain relief. Inspirstar Glove Electrode is cutaneous electrode to be used with legally marketed TENS stimulating device. The glove electrodes will deliver the stimulation signals generated by the stimulator to the skin with which they are contact.
Device Description
Inspirstar Glove Electrode is the conductive glove knitted from a continuous silver coated nylon yarn into the form of the glove. The conductivity is derived from the silver material of nylon yarn. Nylon yarn provides elasticity which can ensure skin contact. Inspirstar Glove Electrode has the same conductivity for dry or wet usage. The entire glove electrode is conductive having the resistance of less than 7 ohms per inch consistently.
More Information

No
The device description and intended use focus solely on the conductive properties of the glove electrode for delivering electrical stimulation. There is no mention of AI, ML, or any computational processing of data for diagnosis, treatment planning, or device control.

No
The device is described as an electrode glove intended for use with electrostimulation devices (like TENS) to deliver stimulation current for pain relief. It is an accessory to a therapeutic device, rather than a therapeutic device itself.

No
The device is described as an electrode for delivering stimulation current for pain relief, not for diagnosing a condition. It is intended to be used with electrostimulation devices like TENS.

No

The device description clearly states it is a "conductive glove knitted from a continuous silver coated nylon yarn," which is a physical hardware component.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for delivering stimulation current for pain relief in conjunction with electrostimulation devices like TENS. This is a therapeutic application, not a diagnostic one.
  • Device Description: The description focuses on the physical properties of the glove as a conductive electrode for delivering electrical signals to the skin. It does not mention any components or functions related to analyzing biological samples (blood, urine, tissue, etc.) which is the hallmark of IVD devices.
  • Lack of IVD-related information: The document does not mention any of the typical elements associated with IVD devices, such as:
    • Analysis of biological samples
    • Detection of specific analytes or markers
    • Diagnostic purposes (e.g., identifying a disease or condition)
    • Performance metrics like sensitivity, specificity, etc., in the context of diagnostic accuracy.

The device is a component used in a therapeutic process (electrostimulation for pain relief), not a device used to perform tests on samples taken from the human body to provide information for diagnostic purposes.

N/A

Intended Use / Indications for Use

Inspirstar Glove Electrode is intended to for use with legally marketed electrostimulation devices such as TENS to deliver stimulation current for pain relief.

Product codes

GXY

Device Description

Inspirstar Glove Electrode is the conductive glove knitted from a continuous silver coated nylon yarn into the form of the glove. The conductivity is derived from the silver material of nylon yarn. Nylon yarn provides elasticity which can ensure skin contact. Inspirstar Glove Electrode has the same conductivity for dry or wet usage. The entire glove electrode is conductive having the resistance of less than 7 ohms per inch consistently.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Inspirstar Glove Electrode is tested to verify the device meets the design requirement. Predicate devices are also tested to support substantial equivalence.

Key Metrics

Not Found

Predicate Device(s)

K053214

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).

0

K061343

p. 1/2

Image /page/0/Picture/2 description: The image shows the logo for InspirStar. The logo features a stylized letter "i" with a star at the top, followed by the word "INSPIRSTAR" in bold, uppercase letters. A horizontal line is present under the word "INSPIRSTAR".

Section 5

510(k) Summary

Inspirstar Glove Electrode

MAY 3 1 2006

    1. Submitters Identification
a. Company Name:Inspirstar Inc.
b. Company Address:891 N. Naples Dr.
Chandler, AZ 85226, USA
c. Contact Person:Ning Wu, Vice President
d. Date of Summary Preparation:April 15, 2006

2. Device Identification

  • Common Name: Electrode, Cutaneous a.
  • Inspirstar Glove Electrode Trade Name: b.
  • Class II C. Classification:
  • d. Product Code: GXY
  • Regulation Number: 882.1320 e.

3. Device Description

Inspirstar Glove Electrode is the conductive glove knitted from a continuous silver coated nylon yarn into the form of the glove. The conductivity is derived from the silver material of nylon yarn. Nylon yarn provides elasticity which can ensure skin contact. Inspirstar Glove Electrode has the same conductivity for dry or wet usage. The entire glove electrode is conductive having the resistance of less than 7 ohms per inch consistently.

    1. Intended Use
      Inspirstar Glove Electrode is intended to for use with legally marketed electrostimulation devices such as TENS to deliver stimulation current for pain relief.
    1. Legally Marketed Predicate Devices
      Inspirstar Glove Electrode is substantially equivalent to the following predicate device. See Table 5-1 Predicate Device.

1

Image /page/1/Picture/1 description: The image shows the word "INSPIRSTAR" in a bold, sans-serif font. To the left of the word is a stylized graphic that resembles a lowercase "i" with a star at the top. A horizontal line runs underneath the word "INSPIRSTAR", emphasizing the text.

Section 5

Table 5-1 Predicate Device

Device NameManufacturer510(k) No.Date Cleared
TheraKnit Garment
ElectrodeNeurotron Medical,
IncK05321412/20/2005

6. Substantial Equivalence Summary

Inspirstar Glove Electrode has the same indications for use as the legally marketed predicate device. Inspirstar Glove Electrode has same technological characteristics, the same conductivity performance, the similar connector interface and similar use and care characteristics as the predicate device. Overall, Inspirstar IS02 Microcurrent Stimulator is substantially equivalent to legally marketed predicate devices.

    1. Technological Characteristics
      Inspirstar Glove Electrode is knitted by traditional knitting process from a continuous silver coated nylon yarn into the form of the glove. The conductivity is derived from the silver material of nylon yarn. The resistance measurement on the glove is less than 7 ohms per inch. Inspirstar Glove Electrode can be used as one electrode, positive or negative polarity, on the current path of circuit. Another Inspirstar Glove Electrode or other electrode can be used as opposite polarity on the current path. Nylon yarn provides elasticity which can ensure skin contact.
    1. Testing
      Inspirstar Glove Electrode is tested to verify the device meets the design requirement. Predicate devices are also tested to support substantial equivalence.
    1. Conclusions
      Inspirstar Glove Electrode is substantial equivalent to legally marketed predicative devices.

2

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 3 1 2006

Inspirstar Inc.

c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313

Re: K061343

Trade/Device Name: Inspirstar Glove Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode, Regulatory Class: Class II Product Code: GXY Dated: May 11, 2006 Received: May 15, 2006

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 310(t) premised the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreate) to regard) the enactment date of the Medical Device Amendments, or to commence price to May 20, 1978, the eccordance with the provisions of the Federal Food, Drug, de necs that have boon receise at require approval of a premarket approval application (PMA). alla Cosmetic rior (110.) that as novely subject to the general controls provisions of the Act. The Tourmaly, mercersions of the Act include requirements for annual registration, listing of general voltable provision practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassinod (tional controls. Existing major regulations affecting your device can may or subject to been the code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

  • Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or and I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

3

Page 2 - Mr. Mark Job

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Image /page/4/Picture/0 description: The image shows the word "INSPIRSTAR" in a stylized font. The "I" in "INSPIRSTAR" is stylized with a star at the top and a curved line extending from the bottom. A horizontal line is present under the word "INSPIRSTAR".

Section 4

Indications for Use

510(k) Number (if known):

K061343

Device Name: Inspirstar Glove Electrode

Indications for Use:

Inspirstar Glove Electrode is cutaneous electrode to be used with legally marketed TENS stimulating device. The glove electrodes will deliver the stimulation signals generated by the stimulator to the skin with which they are contact.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

510(k) Number K061343

Page 1 of 1