K Number
K061225

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-09-14

(135 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker Disposable Laparoscopic Scissors are intended for cutting, dissecting and performing monopolar electrosurgery during any laparoscopic procedure.

Device Description

The Stryker Disposable Laparoscopic Scissors are composed of an ABS plastic handle, 304 and 420 grade stainless steel, silicon, and medical grade PPSU (Radel). All materials are tested for biocompatibility as any of these materials may come into contact with both the patient and surgeon. The device is distributed sterile and intended for single use only.

AI/ML Overview

The provided text is a 510(k) summary for the Stryker Disposable Laparoscopic Scissors. It states the device's intended use and claims substantial equivalence to a predicate device (Stryker Multi Cut Scissors, K935237).

However, the document does not contain information regarding an AI medical device, performance criteria, or any study involving AI or human readers. The information requested in the prompt, such as acceptance criteria table, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance, is not present in the provided text, as this is a regulatory submission for a conventional surgical instrument, not an AI-powered diagnostic or assistive tool.

Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the document pertains to a physical surgical tool and lacks any relevant data for an AI-based system.

{0}------------------------------------------------

5900 Optical Court San Jose, CA 95138 t: 408 754 2491 f: 408 754 2521 www.stryker.com

Image /page/0/Picture/3 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black against a white background. There is a small circle above and to the right of the letter "r", which is likely a trademark symbol.

Endoscor

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Device Name

Proprietary Name:Stryker Disposable Laparoscopic Scissors
Common and Usual Name:Laparoscopic Scissors
Classification Name:Electrosurgical, Cutting & Coagulation & Accessories

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of the SMDA 1990.

The Stryker Disposable Laparoscopic Scissors are substantially equivalent in terms of safety and effectiveness to currently marketed devices, including the Stryker Multi Cut Scissors (K935237).

The Stryker Disposable Laparoscopic Scissors are composed of an ABS plastic handle, 304 and 420 grade stainless steel, silicon, and medical grade PPSU (Radel). All materials are tested for biocompatibility as any of these materials may come into contact with both the patient and surgeon. The device is distributed sterile and intended for single use only.

The Stryker Disposable Laparoscopic Scissors are intended for cutting, dissecting and performing monopolar electrosurgery during any laparoscopic procedure.

The Stryker Disposable Laparoscopic Scissors are tested to conform to the following voluntary safety and performance standards: ISO 10993-1:2003 Biological Evaluation of Medical Devices, ASTM D4169:2004 Standard Practice for Performance Testing of Shipping Containers and Systems, AAM HF-18 Electrosurgical Devices, IEC 60601-1:1988 Medical Equipment: General Requirements for Safety, ISO7741:1986 Instruments for Surgery - Scissors and Shears: General Requirements and Test Methods and EN 552:1994 Sterilization of Medical Devices - Validation and Routine Control of Irradiation.

There are no significant technological or performance differences between the Stryker Disposable Laparoscopic Scissor and the identified predicate devices [Stryker Multi Cut Scissors (K935237)] nor are there any new questions raised regarding safety or effectiveness, the Stryker Disposable Laparoscopic Scissor is substantially equivalent to the identified predicate devices and surgery systems.

Desiree Mae Crisolo Regulatory Affairs Representative Date:

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo features a stylized eagle with three lines forming its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 4 2006

Stryker Corporation % Stryker Endoscopy Ms. Desiree Mae Crisolo Regulatory Affairs Representative 5900 Optical Court San Jose, California 95138

Re: K061225

Trade/Device Name: Stryker Disposable Laparoscopic Scissors Regulation Number: 21 CFR 878.4400 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GEI Dated: August 21, 2006 Received: August 21, 2006

Dear Ms. Crisolo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{2}------------------------------------------------

Page 2 - Ms. Desiree Mae Crisolo

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510 (K) Number: K061225

Device Name: Stryker Disposable Laparoscopic Scissors

Indications for Use:

The Stryker Disposable Laparoscopic Scissors are intended for cutting, dissecting and performing monopolar electrosurgery during any laparoscopic procedure.

Prescription Use_ x (Per 21 CFR 801.109) OR

Over-The-Counter-Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Paha

(Division Sign-Olf) Division of General, Restorative, and Neurological Devices

510(k) Number_ K061225

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.