K Number
K061031
Date Cleared
2006-06-02

(49 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The D 100 L001 Ph.I.S.I.O. Newborn-Hollow Fiber Membrane Oxygenator with Phosphorilcholine Coating is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 0.7 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation, to always assure the proper oxygenation capability of the device. The D 100 L001 Ph.I.S.I.O. should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

Device Description

The D 100 L001 Ph.I.S.I.O Newborn Hollow Fiber Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir with phosphory/choline coating (hereafter referred to as the D 100 Ph.I.S.I.O) is a high efficiency microporous newborn hollow fiber membrane oxygenator integrated with an heat exchanger and connected to an hardshell cardiotomy/venous reservoir. The device is a modified version of the currently marketed D 901 Ph.I.S.I.O. Newborn Hollow Fiber Membrane Oxygenator with phosphoryIcholine coating unmodified device (K991737, K010478) (hereafter referred to as the D 901 Ph.I.S.I.Q.), The modification is limited to an overall reduction in the size of the device, design change to the integrated heat exchanger and hardshell cardiotomy/venous reservoir and consequent updating of product specifications in the IFUs. The reduction in size enables the device to be better suited for surgical procedures on smaller pediatric patients (newborn) where a minimal priming volume is required. Other than this change the D 100 Ph.I.S.I.O. and the D 901 Ph.I.S.I.O. are identical in intended use, materials and manufacturing processes.

AI/ML Overview

The provided text describes a 510(k) summary for the Sorin Group Italia S.r.l. D 100 L001 Ph.I.S.I.O. Newborn Hollow Fiber Oxygenator. This device is a modified version of an existing oxygenator (D 901 Ph.I.S.I.O.) with changes primarily focused on reduced size and design alterations to the heat exchanger and reservoir, intended for use in smaller pediatric patients (newborns).

Here's an analysis of the acceptance criteria and study information based on the provided document:

Acceptance Criteria and Device Performance

The document does not explicitly list a table of quantifiable acceptance criteria with corresponding performance metrics for the D 100 Ph.I.S.I.O. The studies conducted were primarily comparative, aiming to demonstrate substantial equivalence to the predicate device (D 901 Ph.I.S.I.O.) rather than meeting discrete, pre-defined numerical thresholds for a novel device.

However, the text implies that the acceptance criteria are met if the device's performance is either "substantially equivalent" to the predicate, or if observed differences are explainable by the design changes and still suitable for the intended use. The "results of these tests met established specifications" is a recurring phrase, indicating that internal specifications were indeed met, although these are not detailed in the provided summary.

General implied performance areas and their qualitative results:

Performance AreaReported Device Performance (D 100 Ph.I.S.I.O.)
Gas Transfer CharacteristicsTested; results met established specifications; slightly different from D 901 Ph.I.S.I.O. due to size and design changes.
Pressure DropTested; results met established specifications; slightly different from D 901 Ph.I.S.I.O. due to size and design changes.
Plasma Leakage DataTested; results met established specifications.
Heat Exchanger Performance EvaluationTested; results met established specifications; different design compared to D 901 Ph.I.S.I.O.
In Vitro Hemolysis/Cell DepletionTested; results met established specifications.
Operating Blood VolumesTested; results met established specifications.
Structural IntegrityTested; results met established specifications.
Mechanical IntegrityTested; results met established specifications.
Venous, Cardiotomy Reservoir Characterization (breakthrough times, volumes, hold up, graduated scale accuracy, residual blood volume, defoaming capacity, filtration efficiency, microembolic activity, housing integrity)Tested; results met established specifications; different design compared to D 901 Ph.I.S.I.O.
Flaking and Leaching Studies CharacterizationTested; results met established specifications.
Biocompatibility (Hemolysis, Cytotoxicity, Irritation, Acute Systemic Toxicity, Mutagenicity)Tested on D 901 Ph.I.S.I.O.; results met established specifications. Considered applicable to D 100 Ph.I.S.I.O. due to no new materials.
Sterility, Pyrogenicity, ETO residualsTested; results met established specifications. Effectiveness of production techniques to assure sterility and non-pyrogenic nature demonstrated.
Package IntegrityTested; results met established specifications.
Overall Comparison to D 901 Ph.I.S.I.O."Slightly different" but performs in a manner "substantially equivalent" with respect to biocompatibility and intended use. Performance differences are expected due to smaller size, more packed design, and different heat exchanger/reservoir design, which offer advantages for newborn patients (reduced priming volume, less hemodilution, etc.).

Study Details

  1. Sample size used for the test set and the data provenance:

    • The document describes in vitro testing. There is no mention of a "test set" in the context of clinical trials or patient data. Instead, the "test set" refers to the devices themselves that were subjected to various in vitro performance assessments.
    • The sample size for these in vitro tests (e.g., number of devices tested for gas transfer, pressure drop, etc.) is not specified.
    • Data provenance: All testing described is in vitro (laboratory-based) and "carried out in accordance with the requirements of 'Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions – Final Guidance for Industry and FDA Staff' ... and ... ISO 7199 (1996)." This implies the studies were conducted by the manufacturer, Sorin Group Italia S.r.l., likely in Italy.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable to this type of submission. The studies described are in vitro performance tests of a medical device, not evaluations of human-interpretable data (like medical images) where expert ground truth is typically established. The "ground truth" for these tests would be the physical/chemical properties and functional performance of the device as measured by standardized laboratory methods.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This question is not applicable. Adjudication methods are relevant for resolving discrepancies among multiple human readers/reviewers of data (e.g., medical images). The studies here are objective laboratory measurements.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/CAD devices where human interpretation is involved. The device here is a physical medical device (oxygenator) used in cardiopulmonary bypass.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This is a physical medical device, not an algorithm or AI. The tests performed are "standalone" in the sense that they assess the device's technical performance directly, independent of a human operator's skill in interpreting outputs, beyond standard laboratory procedures.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the in vitro tests is derived from objective laboratory measurements performed according to established international and FDA guidance documents (e.g., ISO 7199, FDA Guidance for Oxygenators). The "truth" is whether the device's measured performance parameters fall within pre-defined specifications or are comparable to the predicate device's performance. For biocompatibility, the ground truth relates to the absence of toxic effects as per ISO 10993.
  7. The sample size for the training set:

    • This question is not applicable. This device is a physical medical device. It does not involve a "training set" in the context of machine learning.
  8. How the ground truth for the training set was established:

    • This question is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

510(k) SUMMARY

SUBMITTER:Sorin Group Italia S.r.l.86, Via Statale 12 Nord41037 Mirandola (MO) Italy
CONTACT PERSON:Luigi VecchiPhone: 011 39 0535 29811Fax: 011 39 0535 25229
DATE PREPARED:April 13, 2006
JUN - 2 2006
DEVICE TRADE NAME:D 100 L001 Ph.I.S.I.O: D 100 L001Ph.I.S.I.O Newborn Hollow FiberOxygenator with Integrated HardshellCardiotomy/Venous Reservoir withphosphorylcholine coating (hereafterreferred to as D 100 Ph.I.S.I.O.)
COMMON NAME:Hollow Fiber Membrane Oxygenator withHardshell Cardiotomy/Venous Reservoir
CLASSIFICATION NAME:Cardiopulmonary Bypass Oxygenator /Cardiopulmonary Bypass Heat Exchanger /Cardiopulmonary Bypass BloodReservoir / Cardiopulmonary BypassDefoamer
UNMODIFIED DEVICED 901 Lilliput 1 Ph.I.S.I.O. NewbornHollow Fiber Membrane Oxygenator withphosphorylcholine coating (open andclosed system) (hereafter referred to as D901 Ph.I.S.I.O.) (K991737, K010478).
PREDICATE DEVICE:D 901 Lilliput 1 Ph.I.S.I.O. NewbornHollow Fiber Membrane Oxygenator withphosphorylcholine coating (open andclosed system) (K991737, K010478).D 920 Lilliput 1 Twin Reservoir Ph.I.S.I.O.,Hardshell Venous Cardiotomy Reservoirwith phosphorylcholine coating (K001602)

{1}------------------------------------------------

DEVICE DESCRIPTION:

The D 100 L001 Ph.I.S.I.O Newborn Hollow Fiber Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir with phosphory/choline coating (hereafter referred to as the D 100 Ph.I.S.I.O) is a high efficiency microporous newborn hollow fiber membrane oxygenator integrated with an heat exchanger and connected to an hardshell cardiotomy/venous reservoir. The device is a modified version of the currently marketed D 901 Ph.I.S.I.O. Newborn Hollow Fiber Membrane Oxygenator with phosphoryIcholine coating unmodified device (K991737, K010478) (hereafter referred to as the D 901 Ph.I.S.I.Q.), The modification is limited to an overall reduction in the size of the device, design change to the integrated heat exchanger and hardshell cardiotomy/venous reservoir and consequent updating of product specifications in the IFUs. The reduction in size enables the device to be better suited for surgical procedures on smaller pediatric patients (newborn) where a minimal priming volume is required. Other than this change the D 100 Ph.I.S.I.O. and the D 901 Ph.I.S.I.O. are identical in intended use, materials and manufacturing processes.

INDICATION FOR USE:

The D 100 L001 Ph.I.S.I.O. Newborn Hollow Fiber Membrane Oxygenator with Phosphorilcholine coating is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 0.7 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation, to always assure the proper oxygenation capability of the device. The D 100 L001 Ph.I.S.I.O. should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

TECHNOLOGICAL CHARACTERISTICS:

The D 100 Ph.I.S.I.O. is similar to the D 901 Ph.I.S.I.O. unmodified device with respect to intended use, materials, biocompatibility and performance of the PmMl2 coating, operating principles, control mechanisms, fundamental scientific technology and manufacturing process. The D 100 Ph.I.S.I.O. oxygenating module shares the same basic design, operating principles and control mechanisms of the D 901 Ph.I.S.I.O. module. The only modifications consist of a heat exchanger design revision with respect to D 901 Ph.I.S.I.O.. The hardshell cardiotomy/venous reservoir present in both D 100 Ph.I.S.I.O. and D 901 Ph.I.S.I.O. share the same technological characteristics, operating principles, materials except for the design. The hardshell reservoir type has been changed form a dual chamber with two distinct sections (cardiotomy and reservoir respectively) similar to the D 920 Twin Ph.I.S.I.O. to a single chamber with cardiotomy and venous section in the same chamber. These design changes

{2}------------------------------------------------

update the D 100 Ph.I.S.I.O. device with technology implemented in existing oxygenators manufactured by Sorin Group Italia S.r.l. An overall reduction of its dimensions with consequent less hollow fiber membrane material and reduced priming volume as well as some external features revision, complete the revision of the device. The D 100 Ph.I.S.I.O. is substantially equivalent to D 901 Ph.I.S.I.O. in intended use, patient population and performance specifications.

The coating of the D 100 Ph.I.S.I.O. is identical to the phosphorylcholine coating used on D 901 Ph.I.S.I.O. The oxygenator is ethylene oxide sterilized and has a nonpyrogenic fluid path. It is for single use only.

NON CLINICAL TEST RESULTS:

A complete battery of tests were carried out in accordance with the requirements of ISO 10993-1:2003 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing was performed on the D 901 Ph.I.S.I.O. The aged device was tested for Hemolysis, Cytotoxicity, Irritation, Acute Systemic Toxicity and Mutagenicity. Sterility, Pyrogenicity, ETO residuals. Package integrity testing was also conducted. The results of this testing met established specifications. As no new materials are used in the D 100 L001 Ph.I.S.I.O. newborn oxygenator with respect to the D 901 Ph.I.S.I.O. data collected are considered applicable to D 100 Ph.I.S.I.O. oxygenator.

IN VITRO TEST RESULTS:

In vitro testing was carried out in accordance with the requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions – Final Guidance for Industry and FDA Staff" issued on November 13, 2000 - "Guidance for Blood Extracorporeal Blood Circuit Defoamer 510(k) Submission" Final Guidance for industry and FDA issued on November 29, 2000 - and FDA issued on November 29, 2000 and when applicable, following the ISO 7199 (1996) standard for "Cardiovascular Implants and Artificial Organs – Extra Corporeal Blood-Gas Exchangers (Oxygenator)" for providing the data necessary to demonstrate both the substantial equivalence with the unmodified and predicate devices and also complying with safety and effectiveness requirements. The device aged up to 1 year was tested for gas transfer characteristics, pressure drop, plasma leakage data, heat exchanger performance evaluation, in vitro hemolysis/cell depletion, operating blood volumes, structural integrity, mechanical integrity, venous, cardiotomy reservoir characterization (including breakthrough times and volumes, hold up, reservoir graduated scale accuracy, residual blood volume, defoaming capacity, filtration efficiency, microembolic activity and reservoir housing intearrity) and flaking and leaching studies characterization. The results of these texts my, established specifications. For comparative purposes all tests, when applicable, were carried out on sterilized aged devices comparing the D 100 Ph.I.S.I.O. and D 901 Ph.I.S.I.O. The result of the study showed that D 100 Ph.I.S.I.O. was slightly different to D 901 Ph.I.S.I.O. D 100 Ph.I.S.I.O. is smaller in overall size high a

{3}------------------------------------------------

more packed design, and contains a different heat exchanger/reservoir design than D 901 Ph.I.S.I.O., therefore these results were expected.

CONCLUSIONS:

The results of in vitro studies demonstrate that the design modification affect the D 100 Ph.1.S.I.O. performance with respect to the relevant functional parameters as compared to the D 901 Ph.I.S.I.O. unmodified device. However, the results are expected because the D 100 Ph.I.S.I.O. is smaller in overall size, has a more packed design, and contains a different heat exchanger/reservoir design as compared to the unmodified device. Although these design differences affect the above performance parameters, they are intended to make the D 100 Ph.I.S.I.O. more suitable for use on newborn patients than the unmodified and predicate devices, The smaller size offers theoretical advantages in terms of reduced priming volume and consequently less hemodilution. A lower priming volume is desirable as it results in advantageous patient hemodynamic, reduced exposure of the blood cells and plasma proteins to large surface areas and reduced need of transfusion which has potential risk of donor transmitted disease.

Furthermore, the D 100 Ph.I.S.I.O. performs in a manner substantially equivalent to the D 901 Ph.I.S.I.O. with respect to biocompatibility according to its intended use. Additional testing has also demonstrated the effectiveness of production techniques to assure that the oxygenator is sterile and non-pyrogenic.

{4}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 2 2006

Sorin Group Italla S.R.L. c/o Mr. Parry Sall Principal Consultant Parexel Consulting 200 West Street Waltham, MA 02451-1163

Re: K061031

D 100 L001 Ph.I.S.I.O. Newborn Hollow Fiber Membrane Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir with phosphorylcholine coating Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: May 23, 2006 Received: May 24, 2006

Dear Mr. Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Parry Sall

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Brimmer for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Sorin Group, a medical technology company. The logo features a stylized graphic to the left of the company name, "SORIN GROUP," which is written in a bold, sans-serif font. Below the company name is the tagline "AT THE HEART OF MEDICAL TECHNOLOGY" in a smaller font size.

Sorin Group Italia S.r.l.

510(k) Number (if known):

Device Name: D 100 L001 Ph.I.S.I.O. Newborn Hollow Fiber Membrane Oxygenator with Phosphoriicholine Coating

Indications for Use:

The D 100 L001 Ph.I.S.I.O. Newborn-Hollow Fiber Membrane Oxygenator with Phosphorilcholine Coating is intended for use in infants who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 0.7 liters/minute. It provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation, to always assure the proper oxygenation capability of the device. The D 100 L001 Ph.I.S.I.O. should not be used longer than 6 hours. Contact with blood for longer periods is not advised.

Bhummo

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”