K Number
K060962
Date Cleared
2006-04-19

(12 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Endoscopic irrigation for use with washing catheters integral endoscope water jet channels and endoscope working channels.

Device Description

The Byrne Medical, Inc., EndoGator Endoscopy Irrigation Pump is used for Endoscopic irrigation for use with washing catheters integral endoscope water jet channels and endoscope working channels.

The EGP-100 pump will provide an adjustable flow-rate from 0-550ml/min, which is based on a variable motor RPM. The motor is controlled by the use of a potentiometer and control dial. A foot pedal activated air switch controls the running or stopping of the purnp motor and a peristaltic pump head. These parts will not cause contamination of sterile water during deliverance.

AI/ML Overview

The provided document states that the Byrne Medical, Inc. EndoGator Endoscopy Irrigation Pump is substantially equivalent to its predicate device, the Meditron Endolav Model EL-100 (K882048), based on a comparison of specifications and the successful completion of EMI and Safety Testing.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" for the device's performance in the traditional sense of a clinical study. Instead, it compares the specifications of the EndoGator EGP-100 to its predicate device, the Meditron Endovav EL-100, as part of demonstrating substantial equivalence. The "acceptance criteria" here are implied to be achieving comparable or superior performance to the predicate device for relevant parameters, and meeting safety/EMC standards.

ParameterAcceptance Criteria (Predicate Device Performance)Reported Device Performance (EndoGator EGP-100)Meets Criteria?
Indications for UseEndoscopic irrigation for use with washing catheters, integral endoscope water jet channels, and endoscope working channels.Endoscopic irrigation for use with washing catheters, integral endoscope water jet channels, and endoscope working channels.Yes
Regulation #21 CFR 876.150021 CFR 876.1500Yes
Target PopulationMale/Female pediatric to adultMale/Female pediatric to adultYes
Prescription DeviceYesYesYes
Dimensions (H x W xD)4.5" x 5" x 8"3.5" x 6" x 8"Not directly "same," but considered acceptable for substantial equivalence (smaller size)
ClassificationClass I, Type BF, Ordinary Equipment for Continuous operationClass I, Type BF, Ordinary Equipment for Continuous operationYes
Min flow rate15 ml/min0 ml/minNot directly "same," but an improvement (lower minimum) and considered acceptable.
Max flow rate650 ml/min550 ml/minNot directly "same," but within a comparable range and considered acceptable.
Min Pressure2 PSI0 PSINot directly "same," but an improvement (lower minimum) and considered acceptable.
Max Pressure40 PSI48 PSINot directly "same," but an improvement (higher maximum) and considered acceptable.
Pump typePeristalticPeristalticYes
EMI and Safety TestingMeet specified acceptance criteriaSuccessfully conducted and expected to meet specified acceptance criteria (as stated in Conclusions)Yes, planned and accepted as part of the overall submission.

2. Sample size used for the test set and the data provenance

The document does not describe a "test set" in the context of clinical data or patient samples. The evaluation is based on a comparison of device specifications and the results of engineering tests (EMI and Safety). Therefore, there is no sample size for a test set, nor data provenance (country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. As there is no clinical "test set" with ground truth established by experts, this information is not provided. The substantial equivalence relies on comparing technical specifications and engineering test results.

4. Adjudication method for the test set

Not applicable. There is no clinical "test set" requiring an adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an irrigation pump, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a mechanical pump, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance, in the context of this 510(k) submission, is established through:

  • Comparison to a legally marketed predicate device's established performance and specifications.
  • Engineering standards and testing: Specifically, EMI (Electromagnetic Interference) and Safety Testing (referred to as "Test Review" and affirmed in the "Conclusions"). These tests verify the device's electrical and mechanical integrity against recognized standards.

8. The sample size for the training set

Not applicable. This device is a mechanical pump, not a learning algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable for the same reason as above.

{0}------------------------------------------------

K060962

Attachment H Reply to deficiency

AFR 1 9 2006

Date: 3/28/06

510(k) Summary 21 CFR 807.92

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR part 807.92.

1. Company making the submission:

Byrne Medical, Inc. Company Name: 2021 Airport Road, Conroe, TX 77301 Address: Telephone: (936) 539-0381 (936) 639-0393 Fax Christopher Meador Contact: e-mail: cmeador@byrnemedical.com

2. Device:

Proprietary Name:EndoGator Endoscopy Irrigation Pump
Common Name:Endoscopy Irrigation Pump
Classification Name:Endoscope and/or Accessories

3. Predicate Devices:

Meditron Endolav Model EL-100, K882048, Xylog Corp., 83 Hobart St, Hackensack, NJ 07601

র্বা Classifications Names & Citations:

  • 21 CFR 876.1500, Class II, KOG, Gastroenterology and Urology.

5. Description:

The Byrne Medical, Inc., EndoGator Endoscopy Irrigation Pump is used for Endoscopic irrigation for use with washing catheters integral endoscope water jet channels and endoscope working channels.

The EGP-100 pump will provide an adjustable flow-rate from 0-550ml/min, which is based on a variable motor RPM. The motor is controlled by the use of a potentiometer and control dial. A foot pedal activated air switch controls the running or stopping of the purnp motor and a peristaltic pump head. These parts will not cause contamination of sterile water during deliverance.

{1}------------------------------------------------

6. Indications for Use:

Endoscopic irrigation for use with washing catheters integral endoscope water jet channels and endoscope working channels.

7. Contra-indications:

None noted at the time of this submission.

8. Comparison:

The general method of operation and construction between the predicate device and the Byrne Medical, Inc., EndoGator pump is simulator. Indications for use are the same for both devices.

The Byrne Medical, Inc., EndoGator pump is smaller and the weight is less.

Specifications comparison:

ParameterEndoGator EGP-100This SubmissionMeditron EndovavEL-100K882048Parameteris thesame
Indications for UseEndoscopic irrigation foruse with washingcatheters, integralendoscope water jetchannels, andendoscope workingchannels.Endoscopic irrigationfor use with washingcatheters, integralendoscope water jetchannels, andendoscope workingchannels.Yes
Regulation #21 CFR 876.150021 CFR 876.1500Yes
Target PopulationMale/Female pediatric to adultMale/Female pediatric to adultYes
Prescription DeviceYesYesYes
Dimensions (H x W xD)3.5" x 6" x 8"4.5" x 5" x 8"No
ClassificationClass I, Type BF,Ordinary Equipment forContinuous operationClass I, Type BF,Ordinary Equipment forContinuous operationYes
Min and Max flow rateMin - 0 ml/minMax - 550 ml/minMin - 15 ml/minMax - 650 ml/minNo
PressureMin - 0 PSIMax - 48 PSIMin - 2 PSIMax - 40 PSINo
Pump typePeristalticPeristalticYes

். Test Review:

The EMI and Safety Testing shall be conducted and meet spacified acceptance criteria before the device is marketed.

{2}------------------------------------------------

Conclusions:

The conclusion drawn trom this comparison, together with required passage of electromagnetic compatibility and safety testing, is that the Byrne Medical, Inc. EndoGator pump is equivalent in safety and efficacy to its predicated device.

T. Ricards

Christopher Meador Regulatory Affairs and Quality Control Manager

Date: 3/28/06

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or other bird-like figure. The symbol is composed of three curved lines that converge at the bottom.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Byrne Medical, Inc. c/o Mr. Daniel W. Lehtonen Staff Engineer - Medical Devices and Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Road, Unit B7 TWINSBURG OH 44087

K060962 Rc:

Trade/Device Name: EndoGator Model # EGP-100 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: KOG Dated: April 6, 2006 Received: April 7, 2006

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

APR 19 2006

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/9 description: The image shows the FDA Centennial logo, which includes the years 1906-2006. Below the logo, the text "Protecting and Promoting Public Health" is written in a decorative font. The logo is circular and contains the letters FDA in a stylized font. The word "Centennial" is written below the letters FDA.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Nancy C. Brogdon

Nancv C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment C Reply to deficiencies

Indications for Use Statement

510(k) Number K060962

EndoGator Model # EGP-100 Device Name:

Indications for Use: Endoscopic irrigation for use with washing catheters integral endoscope water jet channels and endoscope working channels.

Prescription Use ત્ત્વન્ય (Part 21 CFR 801 Subpart D) ANDIOR

Over The-Counter Use (21 CFR 807 Subpart C)

(PLEASE: DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Larson
Division Sign Off

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K060962

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.