K Number
K060559

Validate with FDA (Live)

Date Cleared
2006-08-17

(168 days)

Product Code
Regulation Number
870.5310
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Defibrillation: Use of the R Series products in the manual and semiautomatic modes for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heartbeats. In manual mode, the unit can also be used for synchronized cardioversion to atrial fibrillation (AF) or ventricular tachycardia (VT) by using the R wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate. The advisory function should be used to confirm ventricular fibrillation and wide complex ventricular tachycardia in patients meeting the three conditions indicating lack of circulation (previously listed).

External Pacemaker (Pacer Version Only): This product may be used for temporary external, demand or non-demand pacing, as an alternative to endocardial stimulation. External cardiac pacing is indicated for conscious or unconscious patients in asystole, profound bradycardia or any other conditions determined by a clinician to require external pacing.

ECG Monitoring: The device is indicated for monitoring a patient's electrocardiogram (ECG), via the 3 or 5 lead patient cable, MFE Pads, or through the paddles, for the detection of salient cardiac rhythms and dysrhythmias and calculating heart rate.

CPR Monitoring: The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients. The CPR monitoring function is not intended for use on patients under 8 years of age.

SpO2 Option (if equipped): The SpO2 Option with Masimo Set Technology is indicated for the continuous noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate for adult, pediatric and neonatal patients, during both no motion and patient motion conditions, and for patients who are well or poorly perfusing in the hospital or pre-hospital environments.

Device Description

The ZOLL R Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport. The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing an adult or pediatric patient's ECG signal via the Advisory feature on the device. The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned. Therapy is provided by using defibrillation electrode products specifically designed for use with the R Series. When used in conjunction with ZOLL Ready padz One-step CPR or Ready padz One-step Complete Electrodes, the R SERIES provides CPR compression performance feedback using displayed symbols, text messages and voice prompts.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text for the ZOLL R Series Defibrillator:

The provided document is a 510(k) summary for the ZOLL R Series Defibrillator, which focuses on demonstrating substantial equivalence to previously cleared devices rather than presenting a novel device with extensive new performance studies. Therefore, much of the requested information regarding detailed performance studies for new acceptance criteria is not explicitly stated in this type of submission. The document relies heavily on the fact that existing, cleared algorithms and technologies are being re-used.


Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" in the way a new device's performance study might. Instead, it states that the device reuses existing algorithms and technologies from predicate devices (ZOLL E Series and ZOLL AED Pro Defibrillator).

The primary acceptance criteria are implicitly met by demonstrating that the ECG Analysis Algorithms for shock/no-shock advisory and CPR Aid feature are "identical" or "similar" to those in the predicate devices which have already been cleared by the FDA.

Table 1: Implicit Acceptance Criteria and Reported Device Performance

Feature/FunctionImplicit Acceptance Criteria (based on predicate equivalence)Reported Device Performance (as stated in document)
ECG Analysis Algorithm (Advisory feature)Must determine if acquired heart rhythm is shockable or non-shockable with accuracy comparable to predicate ZOLL AED Pro Defibrillator.Utilizes "identical ECG Analysis Algorithms" to the predicate ZOLL AED Pro Defibrillator (K041892).
CPR Performance FeedbackMust provide performance feedback for compression depth and rate comparable to predicate ZOLL AED Pro.Provides performance feedback for compression depth and rate "similar to the ZOLL AED Pro."
Multi-parameter Monitoring (ECG & Pulse Oximetry)Technologies for ECG Monitoring and Pulse Oximetry must be similar to predicate ZOLL E Series.Uses technology "very similar to that of the ZOLL E Series" (K042007).
Overall Functional Requirements & Performance SpecificationsDevice must meet all functional requirements and performance specifications."Extensive performance and software testing ensures that the ZOLL R Series Defibrillator meets all of its functional requirements and performance specifications." (No specific numerical data provided here).
SafetyDevice must comply with applicable sections of recognized industry and safety standards."Safety testing assures the device complies with applicable sections of recognized industry and safety standards."

Study Details

The provided document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence rather than presenting a full, detailed clinical study report for a novel algorithm. For this reason, many of the requested details about specific performance studies (like sample sizes, expert qualifications, etc.) are not explicitly stated for the ZOLL R Series itself, but rather inferred from the predicate devices.

  1. Sample size used for the test set and the data provenance:

    • Not explicitly stated for the ZOLL R Series device. The document states that the ECG analysis algorithms are "identical" to the predicate ZOLL AED Pro. The CPR Aid feature is "similar." The performance of these predicate devices would have been evaluated with test sets, but those details are not provided in this summary for the R Series.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not explicitly stated for the ZOLL R Series device. This information would typically be available in the detailed reports for the original predicate devices' algorithm validation, which are not included here.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not explicitly stated for the ZOLL R Series device. Again, this detail would pertain to the validation of the algorithms in the predicate devices.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable/Not stated for this submission. The device includes an "Advisory function" where the algorithm determines shockable/non-shockable rhythms and prompts the user. This is a form of AI assistance, but the submission doesn't describe a comparative effectiveness study in an MRMC format comparing human performance with and without this specific device's assistance. The focus is on the algorithm's performance equivalence to the predicate.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Implicitly yes, for the algorithm itself. The statement "The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable" indicates a standalone assessment of the algorithm's performance, which is then presented to a human. However, the specific details of that standalone study (like sensitivity/specificity metrics) are not provided, only the assertion that the algorithm is "identical" to a previously cleared one.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not explicitly stated for the ZOLL R Series device. For ECG rhythm analysis, ground truth is typically established by expert cardiologists reviewing the ECG waveforms. This would have been done for the predicate device.
  7. The sample size for the training set:

    • Not explicitly stated for the ZOLL R Series device. As the algorithms are "identical" or "similar" to predicate devices, the training was done for those prior iterations, and the specifics are not in this document.
  8. How the ground truth for the training set was established:

    • Not explicitly stated for the ZOLL R Series device. Similar to test set ground truth, for ECG algorithms, this would involve expert cardiologist review and labeling of ECG data.

{0}------------------------------------------------

ZOLL Medical Corpora

Worldwide Headquarters 269 Mill Boad Chelmsford, Massachusetts 01824-4105 978 421-9655 978 421-0025 Main Fax

510(k) Summary:

Submitter's Name and Address:

ZOLL Medical Corporation Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655

Contact Person:

Sean Reynolds (978) 421-9655, Ext. 9386

Date Summary Prepared:

February 24, 2006

Device:

ZOLL R Series

Classification:

Defibrillator, Low-energy -- DC: Class II (21 CFR 870.5300)

Automated External Defibrillator: Class III (21 CFR 870.5310)

Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200)

Cardiac Monitors (including Cardiotachometers and Rate Alarms): Class II (21 CFR 870.2300

External Transcutaneous Cardiac Pacemakers (Non-invasive): Class II (21CFR 870.5550)

Oximeters: Class II (21 CFR 870.2700)

Description:

The ZOLL R Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport.

The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing an adult or pediatric patient's ECG signal via the Advisory feature on the device. The Advisory Alqorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in

AUG 1 7 2006

{1}------------------------------------------------

the event of a shock advised determination. The user will be prompted to rethe overte of a choolhe event a no shock advised determination is returned.

Therapy is provided by using defibrillation electrode products specifically Therapy 1s "provided" by "acing" usens" with a series. When used in conjunction with ZOLL Ready padz One-step™ CPR or Ready padz One-step™ Complete 2OEL Tready paul - One SERIES provides CPR compression performance Manufation Elootrodoo the 17 & Lisplayed symbols, text messages and voice prompts.

Substantial Equivalence:

The features and functions of the ZOLL R SERIES are substantially equivalent to The Touth of the ZOLL E SERIES Defibrillator: 510(k) No. K042007, cleared 1/07/2005, and the ZOLL AED Pro Defibrillator, 510(k) No. K041892, cleared 2/04/2005.

Intended Use:

Defibrillation

Use of the R Series products in the manual and semiautomatic modes for OSC of the 11 'Golfoo productims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • Unconsciousness .
  • Absence of breathing .
  • Absence of pulse. .

This product should be used only by qualified medical personnel for converting rms product entricular and rapid ventricular tachycardia to sinus rhythm or other venthoular Inbhilation all of producing hemodynamically significant heartbeats.

In manual mode, the unit can also be used for synchronized cardioversion to in manufal fibrillation (AF) or ventricular tachycardia (VT) by using the R wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate.

The advisory function should be used to confirm ventricular fibrillation and wide The advisory function cardia in patients meeting the three conditions indicating lack of circulation (previously listed).

External Pacemaker (Pacer Version Only)

This product may be used for temporary external, demand or non-demand This product may be about to endocardial stimulation. External cardiac pacing is pating, as an alternatious or unconscious patients in asystole, profound indical for any other conditions determined by a clinician to require external pacing.

ECG Monitoring

The device is indicated for monitoring a patient's electrocardiogram (ECG), via the 3 or 5 lead patient cable, MFE Pads, or through the paddles, for the the 5 of 5 lead pation. Salian cardiac rhythms and dysrhythmias and calculating heart rate.

{2}------------------------------------------------

CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

The CPR monitoring function is not intended for use on patients under 8 years of age.

SpO2 Option (if equipped)

The SpO2 Option with Masimo Set Technology is indicated for the continuous noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate for adult, pediatric and neonatal patients, during both no motion and patient motion conditions, and for patients who are well or poorly perfusing in the hospital or pre-hospital environments.

Comparison of Technological Characteristics

The ZOLL R Series design characteristics are fundamentally the same scientific technologies as those of the indicated predicate devices; the technology is very similar to that of the ZOLL E Series. The ZOLL R Series provides monitoring of patient vital signs, including: ECG Monitoring and Pulse Oximetry, using technology very similar to that of the predicate ZOLL E Series.

The ZOLL R Series advises users to deliver a shock, perform CPR or conduct patient assessment through audible and visual prompts nearly identical to those used by the indicated predicate devices. The ZOLL R Series acquires and analyzes ECG signals and provides shock advisory determinations for adult and pediatric patients, utilizing the identical ECG Analysis Algorithms that are incorporated into the predicate ZOLL AED Pro Defibrillator. ZOLL's proprietary CPR Aid feature provides users with performance feedback for compression depth and rate similar to the ZOLL AED Pro.

Performance Testing:

Extensive performance and software testing ensures that the ZOLL R Series Defibrillator meets all of its functional requirements and performance specifications. Safety testing assures the device complies with applicable sections of recognized industry and safety standards.

Conclusion

Performance and safety testing of the ZOLL R Series Defibrillator demonstrates that its features, functions and incorporated interpretive algorithm are substantially equivalent to that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness.

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 7 2006

ZOLL Medical Corporation c/o Mr. Sean Reynolds Regulatory Affairs Engineer Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824-4105

Re: K060559

TradeDevice Name: Zoll R Series Defibrillator Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillators Regulatory Class: Class III (three) Product Code: MKJ, LDD, DRO, DQA, LIX Dated: August 10, 2006 Received: August 14, 2006

Dear Mr. Reynolds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -- Mr. Sean Reynolds

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised frat FDA s issualtes or a succession of the requirements of the Act
that FDA has made a determination that your device complies with of any must that FDA has made a determination that your as resear agencies. You must and or any Federal statues and regulations administered of other to: registration and listing (21 l
eomply with all the Act's requirements, including, but not limited to: repiram comply with an the Act S requiremont, me; good manufacturing practice requirements as sett CFR Part 807), labeling (21 CFR 1 at 607), good market 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Rates (2) 11 CFR 1000 1 forth in the quanty systems (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections of I stills) in your Section 100(k)
This letter will allow you to begin marketing your device of your davice to 1egal This letter will allow you to begin makemig of substantial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivales and thus, premarket notification. The FDA inding of substantal equive and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the many of the many on 1000 - All any also regulation entitled If you desire specific advice ion your de vice of an also, please note the regulation entitled. contact the Office of Compilance at (217) 276 - 12 - 17 - 17 807.97). You may obtain
"Misbranding by reference to premarket notification" (21 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - "Misbranding by relefence to premarket tionneation" (1) 35 oct from the Division of Small other general information on your responsionals at its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance as to too working to word index.html.

Sincerely yours,

Q.S.U

Con Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 4 - Indications for Use

510(k) Number (if known):

Device Name: ZOLL R Series

Defibrillator Function

Intended Use - Manual Operation

Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • Unconsciousness .
  • Absence of breathing .
  • Absence of pulse. .

This product should be used only by qualified medical personnel for converting ventricular fibrillation This product should be used only by qualified percently the cardiac rhythms capable of producing hemodynamically significant heart beats.

In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or in manual noot, the unt our also by ician must decide when synchronized cardioversion is appropriate.

The advisory function should be used to confirm ventricular fibrillation and wide complex ventricular The advisory function should be about to somments meeting the three conditions indicating lack of circulation (listed above).

Intended Use - Semiautomatic Operation (AED)

The R Series products are designed for use by emergency care personnel who have completed THE IT Series produces are doolghed for ass of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a r ricy are spoolingally coorgesuscitation involving CPR, transportation, and definitive care are incorporated into a medically approved patient care protocol.

Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac 000 or there there is apparent lack of circulation as indicated by:

  • Unconsciousness .
  • Absence of breathing .
  • . Absence of pulse.

(continued on next page)

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular DevicesPage 1 of 3

{6}------------------------------------------------

Indications for Use (continued from previous page)

Specifications for the ECG rhythm analysis function are provided in the section, "ECG Rhythm Analysis Algoirhtm Accuracey" on page A-26.

When the patient is less than 8 years of age or weighs less than 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

Intended-Use - ECG Monitoring

The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the use of the R Series defibrillator, basic life and/or advanced life support, or other physician-authorized emergency medical training.

Intended-Use - CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

The CPR monitoring function is not intended for use on patients under 8 years of age.

External Pacemaker (Pacer Version Only)

Intended Use — Pacemaker

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.

The purposes of pacing include:

Resuscitation from standstill or bradycardia of any etiology:

Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.

As a standby when standstill or bradycardia might be expected:

Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.

Suppression of tachycardia:

Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.

(continued on next page)

S.M.

(DIVision Sign-Off)
Division of Cardiovascular Devices
510(k) Number K060559

{7}------------------------------------------------

Indications for Use (continued from previous page)

Pediatric Pacing

Pacing can be performed on pediatric patients weighing 33lb. (15kg) or less using ZOLL pediatric hands-free therapy electrode pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.

Intended-Use - SpO2 Monitoring

The R Series pulse oximeter, with the Masimo® SET® technology and the LNOP® series of oximeter sensors, is indicated for the continuous, noninvasive monitoring of arterial oxygen saturation (SpO₂) and pulse rate during both no-motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital environment.

QSSM

f Cardiovascular Devices 510(k) Numb

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.