(72 days)
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent cefoxitin at concentrations of 0.5-64 ug/mL to gram-negative ID/AST or AST only Phoenix panels. Cefoxitin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli Klebsiella spp. (including K. pneumoniae) Morganella morqanii
Proteus mirabilis Proteus vulgaris Providencia spp. (including P. rettgeri)
Active In Vitro Against:
Klebsiella oxytoca
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth t determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
Here's an analysis of the provided text regarding the acceptance criteria and study for the BD Phoenix™ Automated Microbiology System - Cefoxitin – 0.5-64 µg/mL.
Acceptance Criteria and Device Performance Study
The study evaluates the performance of the BD Phoenix™ Automated Microbiology System for antimicrobial susceptibility testing of Cefoxitin against specific Gram-negative bacteria. The performance is assessed by comparing its results to a CLSI reference broth microdilution method.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the "Essential Agreement (EA)" and "Category Agreement (CA)" metrics, which are standard for AST systems. While the document states that the system met these criteria (demonstrated "substantially equivalent performance"), the actual numerical acceptance limits for EA and CA are not explicitly defined in the provided text. However, typical FDA guidance for AST devices requires high percentage agreement. The table below presents the reported performance based on the general statements.
| Acceptance Criteria Metric | Reported Device Performance | Comments |
|---|---|---|
| Essential Agreement (EA) | Not explicitly quantified, but stated to demonstrate "substantially equivalent performance" compared to the CLSI reference method. EA occurs when the BD Phoenix™ System agrees exactly or within ± one two-fold dilution to the reference result. | The document implies the EA met an acceptable threshold, but no specific percentage is given in the readable text of "Table 1". |
| Category Agreement (CA) | Not explicitly quantified, but stated to demonstrate "substantially equivalent performance" compared to the CLSI reference method. CA occurs when the BD Phoenix™ System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant). | Similar to EA, CA met an implied acceptable threshold. The readable text of "Table 1" does not provide specific percentages. |
| Intra-site Reproducibility | Greater than 90% | Achieved at three sites, tested in triplicate on three different days. |
| Inter-site Reproducibility | Greater than 95% | Across three sites. |
Note on Table 1: The content of "Table 1" in the provided text is highly corrupted and unreadable. Therefore, specific numerical performance metrics for EA and CA cannot be extracted directly from it. The description focuses on the definition of EA and CA and the overall conclusion of substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document mentions "clinical, stock and challenge isolates" were tested. While it states "Table 1 summarizes the performance for the isolates tested in this study," the corrupted nature of Table 1 prevents extraction of the exact number of isolates for the Cefoxitin-specific study. However, for a 510(k) submission for AST, this would typically involve hundreds to thousands of isolates across various species.
- Data Provenance: The isolates were tested "across multiple geographically diverse sites across the United States." This indicates a prospective collection or a mix of prospective (clinical) and well-characterized retrospective (stock/challenge) isolates. The "clinical isolates" would be prospectively collected from patient samples, while "stock and challenge isolates" are typically well-characterized laboratory strains used for specific testing purposes.
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the number or qualifications of experts used to establish the ground truth. However, the ground truth for antimicrobial susceptibility testing is established by the "CLSI reference broth microdilution method," which is a highly standardized laboratory procedure. This method is performed by trained laboratorians according to rigorous protocols, rather than relying on subjective expert interpretation like in image-based diagnostics. The "experts" in this context are the trained laboratory personnel performing the reference method.
4. Adjudication Method for the Test Set
There is no mention of an adjudication method in the context of expert review. This is expected because the ground truth for AST is determined by a standardized reference laboratory method (CLSI broth microdilution), which yields an objective result (MIC value and categorical interpretation). Discrepancies between the device and the reference method are analyzed for agreement, not adjudicated by a panel of human experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This type of study is typically performed for diagnostic devices where human interpretation of results is involved (e.g., radiology, pathology). The BD Phoenix™ System is an automated system for objective measurement of microbial growth inhibition, and its "effectiveness" is compared against a gold standard method (CLSI reference) rather than against human readers' performance with or without AI assistance.
6. Standalone Performance Study (Algorithm only without human-in-the-loop performance)
Yes, a standalone performance study was done. The entire study described is a standalone performance evaluation of the BD Phoenix™ Automated Microbiology System. The device directly provides Minimum Inhibitory Concentration (MIC) values and categorical interpretations (S, I, R) without human intervention in the primary interpretation loop. The Phoenix System's results are then compared to the CLSI reference method.
7. Type of Ground Truth Used
The ground truth used is the CLSI reference broth microdilution method. This is a highly standardized and accepted laboratory method for determining antimicrobial susceptibility, widely considered the "gold standard" for this type of testing. For challenge isolates, "expected results" were used, which are also derived from established reference methods and historical characterization.
8. Sample Size for the Training Set
The document does not provide information about a separate "training set" or its sample size. Antomicrobial Susceptibility Testing (AST) systems like the BD Phoenix are developed and validated using a different paradigm than typical AI/ML algorithms that require explicit training sets. The development typically involves:
- Algorithm development: Based on principles of microbiology, redox indicators, and bacterial growth kinetics.
- Calibration: Using a diverse set of characterized isolates to calibrate the system's readings and interpretation rules against the reference method. This is more akin to a calibration/optimization set than a traditional ML "training set."
The isolates mentioned (clinical, stock, challenge) are part of the performance evaluation or test set that demonstrates the device's accuracy against the reference method, not a separate training set for an AI/ML model.
9. How the Ground Truth for the Training Set Was Established
As noted above, a traditional "training set" for an AI/ML model is not explicitly mentioned or implied for this device's validation. If any internal calibration or optimization involved a set of isolates, their ground truth would have been established using the CLSI reference broth microdilution method, consistent with the test set's ground truth. However, the document provided focuses solely on the performance evaluation against the reference standard for regulatory submission.
{0}------------------------------------------------
510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company APR 14 20067 Loveton CircleSparks, MD 21152Phone: (410) 316 - 4278Fax: 410-316-4499 |
|---|---|
| CONTACT NAME: | Monica E GiguereRegulatory Affairs Specialist |
| DATE PREPARED: | April 5, 2006 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -Cefoxitin – 0.5-64 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), and Cefoxitin - 0.5-64 µg/ml (K032567,September 17, 2003). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin. |
DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth t determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional organism groups with Cefoxitin - 0.5-64 ug/mL and the BD PhoenixTM Automated Microbiology System.
SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.
Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and associated reagents.
The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.
{2}------------------------------------------------
Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
Performance of BD Phoenix System by Drug Table 1:
| on and in American in sunderLA But Chical ConcessionAntimicrobial | SHOW OF THE BLUE OF THE F. F.oncentration | -------------------------------------------------------------------------------11 | Cabon AV 110 1100 1 11 11 | 1 | And Analy Property Contract Company Comments of Children Comments of ChildrenC |
|---|---|---|---|---|---|
| STATE LABOR HALL Walnutrefoxitin | money may consider of a concessful an42/mi | 1400CARLES LAND LA CLASS & LEASE SELLE | 40CALLER LEASE CHANNE AN CO | 1900- And - | C CA has an and |
Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), and Cefoxitin -- 0.5-64 ug/ml (K032567, September 17, 2003).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a black and white logo. The logo is circular in shape and contains an emblem in the center. The emblem is a stylized depiction of an eagle or bird-like figure. The text "U.S. Public Health Service" is arranged around the perimeter of the circle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostic Systems 7 Loveton Circle Sparks, MD 21152
APR 1 4 2006
Re: K060257/ S1
Trade/Device Name: BD Phoenix™ Automated Microbiology System Cefoxitin (0.5-64 ug/mL) Gram-negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: January 31, 2006 Received: February 1, 2006
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 --
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to procced to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sally a Hong
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: K060257/S1
BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Device Name: Cefoxitin - 0.5-64 ug/mL on Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent cefoxitin at concentrations of 0.5-64 ug/mL to gram-negative ID/AST or AST only Phoenix panels. Cefoxitin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli Klebsiella spp. (including K. pneumoniae) Morganella morqanii
Proteus mirabilis Proteus vulgaris Providencia spp. (including P. rettgeri)
Active In Vitro Against:
Klebsiella oxytoca
Prescription Use _ (Per 21 CFR 801.109) Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11013
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”