K Number
K060102

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-03-15

(61 days)

Product Code
Regulation Number
886.5916
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use in the correction of eyes with refractive errors resulting from corneal surgery or trauma including hyperopia and myopia, astigmatism and irregular astigmatism in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 in eyes with astigmatism or irregular astigmatism up to 6.00D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

Device Description

The SynergEves™ PS (paflufocon D hem-iberfilcon A) Hybrid Daily Wear Contact Lens is a combination rigid gas permeable contact lens corneal optic portion surrounded by a soft hydrophilic skirt that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere .
  • . in the range of 0.25 to 6.00 diopters cylinder
  • with center thickness from 0.18mm to 0.30mm .
  • with base curves of 7.10mm to 9.60mm .
  • with diameter of 14.50mm .
    The lens material (paflufocon D hem-iberfilcon A) is identical to the predicate current material cleared under K051035 and K052675. There are no differences to the chemical composition, formulation, manufacturing process, packaging and sterilization as described in the referenced 510(k).
    This lens material for the rigid portion is paflufocon D lathe cut, surrounded by soft hydrophilic copolymer (hem-iberfilcon A), sterilized by means of e-beam sterilization. When placed on the human cornea, the SynergEyes™Hybrid Contact Lens acts as a refracting medium to focus light rays onto the retina. The device is available as a lathe cut contact lens in the following design for post surgical refractive errors and trauma: multi concentric zones in blue visibility tinted material. This device is equivalent to the SynergEyes™ A and M Hybrid Contact Lens is material design and composition, and the SoftPerm® hybrid RGP manufactured by Ciba Vision Corporation.
    The SynergEyes™ PS Hybrid Daily Wear Contact Lens center optical portion is a rigid gas permeable material of (paflulofon-D) polymer. TThe soft skirt is comprised of HEMA (hydroxyethylmethacrylate) of 27% water and 73% polymer.
    The junction between the rigid material and soft material is bound by a proprietary chemical bonding method.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the SynergEyes™ PS Hybrid Contact Lens, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" with numerical thresholds in the typical sense of a medical device submission beyond showing substantial equivalence to predicates. However, it evaluates safety and efficacy by tracking specific outcomes. I'll frame the reported device performance against generally understood good outcomes for contact lenses.

Performance MetricAcceptance Criteria (Implied/Good Outcome)Reported Device Performance (SynergEyes™ PS)
SafetyLow incidence of adverse events.Adverse Events: 8 adverse events reported across 7 subjects (4 completed, 3 discontinued). Events included 1 painful light sensitivity, 1 loose interrupted suture and infiltrate, 1 corneal abrasion upon lens removal, 1 superficial abrasion, 1 subject (2 eyes) with corneal swelling due to inadequate rinsing of hydrogen peroxide solution, 1 allergic conjunctivitis, and 1 keratitis. Discontinuation due to Safety: Adverse events were a small portion of discontinuations. Specific adverse events like corneal swelling from disinfection solution indicate user error rather than intrinsic device failure.
Efficacy: Visual AcuityHigh percentage of subjects achieving excellent or good visual acuity (e.g., 20/20, 20/25).Completed Subjects: - 20/20 or better: 20.0% - 20/25 or better: 45.0% - 20/30 or better: 68.8% - 20/40 or better: 81.3% Discontinued Subjects: - 20/20 or better: 8.6% - 20/25 or better: 39.7% - 20/30 or better: 53.4% - 20/40 or better: 70.7% 3 completed eyes and 8 discontinued eyes had VA decreases of >2 lines (expected in this population).
Efficacy: Wearing TimeAcceptable daily wearing time for daily wear lenses (e.g., >8-10 hours).Average daily wearing time: 10.6 hours per day for completed patients.
ComfortLow rates of discomfort/awareness leading to discontinuation.Discontinuation due to poor comfort: 27.5% of discontinuations. Reported Symptoms (Completed Subjects): - Discomfort and awareness: 28.4% - Dryness and scratchiness: 23.4% - Itchiness and burning: 9.3% - Variable vision: 8.2% Symptoms decreased over time for completed eyes.
Overall PerformanceSatisfactory performance and safe/effective use for intended purpose."The SynergEyes™ PS Hybrid Contact Lens for Post Surgical refractive error and cases involving trauma provided satisfactory performance as expected... Overall, the lens performance demonstrated safe and effective use of the device for its intended use." It notes the higher than estimated discontinuation rate was anticipated given the challenging patient population.

2. Sample Size for Test Set and Data Provenance

  • Sample Size for Test Set:
    • Subjects: 76 subjects were dispensed into the study.
    • Eyes: 80 eyes (from 44 completed subjects) reached study completion. The total number of eyes enrolled isn't explicitly stated but would be at least 76 (assuming one eye per subject) and up to 152. The "58 eyes" mentioned for discontinued subjects suggests that some subjects wore lenses in both eyes.
  • Data Provenance: The study was a "three month clinical study," implying a prospective design within a controlled clinical environment (likely in the US, given the FDA submission). No specific country of origin is mentioned beyond the applicant's US address.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This is not applicable. The study is a clinical trial assessing the direct performance of the contact lens on patients. The "ground truth" is the actual patient experience and objective clinical measurements (visual acuity, adverse events) as documented by clinical investigators and optometrists/ophthalmologists involved in the study. There's no mention of independent experts establishing a "ground truth" for a diagnostic performance metric.

4. Adjudication Method for the Test Set

This is not applicable, as it's a clinical trial for a contact lens, not a diagnostic device requiring adjudicated interpretations of images or data. Clinical investigators and study coordinators would have managed data collection and reported outcomes.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where multiple readers interpret cases to assess diagnostic accuracy and inter-reader variability. This document describes a clinical trial for a contact lens.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

No, a standalone algorithm performance study was not done. This document describes a clinical trial for a physical medical device (contact lens) worn by human patients, inherently involving human interaction (wearing, fitting, clinical evaluation).

7. The Type of Ground Truth Used

The "ground truth" in this context is based on clinical observation, patient-reported outcomes, and objective ophthalmological measurements collected during the prospective clinical study. This includes:

  • Measurement of visual acuity (Snellen VA).
  • Documentation of adverse events.
  • Patient-reported symptoms and complaints (comfort, dryness, variable vision).
  • Measured wearing time.
  • Clinical assessments by practitioners (e.g., corneal health).

8. The Sample Size for the Training Set

This information is not provided and is not applicable in the context of this device submission. This device (contact lens) is a physical product, not an AI/ML-based algorithm that requires a "training set" in the computational sense. The document describes a clinical study to evaluate the device's performance, not to develop or train an algorithm.

9. How the Ground Truth for the Training Set was Established

This information is not provided and is not applicable, as there is no "training set" for an AI/ML algorithm.

{0}------------------------------------------------

SynergEyes
Carlsbad, CA 92008

Kobo102

MAR 1 5 2006

510(K) SUMMARY

Applicant's Name and Address

SynergEyes ™, Inc.1926 Kellogg Ave. Suite 100Carlsbad, CA 92008(760) 476-9410
Contact PersonRichard Lippman, OD FAAOVice President Ophthalmic Regulatory ProductsR.P. Chiacchierini & Associates, LLC
15825 Shady Grove Road Suite 30
Rockville, Maryland 20850
(240) 683-3738
1. Identification of device
Common Name:Contact Lens
Trade Name:SynergEyes™ PS (paflufocon D hem-iberfilcon A)Hybrid Daily Wear Lens For Post Surgical and Trauma
Classification:Daily Wear Contact Lens (other material)
Device classification:Class II (21 CFR 886.5916)

2. Description of device

The SynergEves™ PS (paflufocon D hem-iberfilcon A) Hybrid Daily Wear Contact Lens is a combination rigid gas permeable contact lens corneal optic portion surrounded by a soft hydrophilic skirt that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere .
  • . in the range of 0.25 to 6.00 diopters cylinder
  • with center thickness from 0.18mm to 0.30mm .
  • with base curves of 7.10mm to 9.60mm .
  • with diameter of 14.50mm .

The lens material (paflufocon D hem-iberfilcon A) is identical to the predicate current material cleared under K051035 and K052675. There are no differences to the chemical composition, formulation, manufacturing process, packaging and sterilization as described in the referenced 510(k).

This lens material for the rigid portion is paflufocon D lathe cut, surrounded by soft hydrophilic copolymer (hem-iberfilcon A), sterilized by means of e-beam sterilization. When placed on the human cornea, the SynergEyes™Hybrid Contact Lens acts as a refracting medium to focus light rays onto the retina. The device is available as a lathe cut contact lens in the following design for post surgical refractive errors and trauma: multi concentric zones in blue visibility tinted material. This device is equivalent to the SynergEyes™ A and M Hybrid Contact Lens is material design and composition, and the SoftPerm® hybrid RGP manufactured by Ciba Vision Corporation.

The SynergEyes™ PS Hybrid Daily Wear Contact Lens center optical portion is a rigid gas permeable material of (paflulofon-D) polymer. TThe soft skirt is comprised of HEMA (hydroxyethylmethacrylate) of 27% water and 73% polymer.

The junction between the rigid material and soft material is bound by a proprietary chemical bonding method.

{1}------------------------------------------------

SynergEyes™, Inc. Carlsbad, CA 92008

3. Intended use

SynerqEyes™ PS (paflufocon D hem-iberfilcon A) Hybrid Contact Lenses are indicated for use in the correction of eyes with refractive errors resulting from corneal surgery or trauma including hyperopia and myopia, astigmatism, and irregular astigmatism in aphakic and not aphakic, nondiseased eves. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism or irregular astigmatism up to 6.00 D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

4. Predicate devices

The predicate lens SynergEyes™ A and M (paflufocon D hem iberfiicon A) Hybrid Contact Lenses were selected to address material use and design (rigid lens center and soft lens skirt).

The material portion is equivalent to the predicate lenses. The optic center rigid portion of the lens is a silicone acrylate- paflufocon D used in the cleared predicates K051035 and K052675 as well as referenced HDS material from Paragon Vision Sciences P870024/S36 and 43.

The soft skirt portion of the subject device, hem-iberfilcon A, a Group I low water non-ionic material, is equivalent to its predicate K051045 and K052675.

Further equivalence to predicate devices for indication for use and parameter comparison may be found in the equivalence matrix below.

Predicate Devices:

  • For the Hybrid (RGP-soft skirt concept): ●
SynergEyes™™, Inc.K051035K052675Cleared: 9/2/2005Cleared: 11/30/2005
SoftPerm (synergicon A) Mixed use LensP840004/S7App: 10/14/1993
• For the RGP Center Portion:
SynergEyes™™, Inc.K051035K052675Cleared: 9/2/2005Cleared: 11/30/2005
Paragon Vision SciencesP870024/S36 and S43(App: 11/23/1993 and 4/13/2002)
K940277Cleared: 5/9/1994
• For the Soft Skirt Portion:
SynergEyes™™, Inc.K051035K052675Cleared: 9/2/2005Cleared: 11/30/2005
SoftPerm (synergicon A) Mixed use lensP840004/S7App: 10/14/1993

5. Characteristics

The physical and dimensional characteristics of the SynergEyes™ PS Hybrid Contact Lens are compared to the characteristics of the predicate devices SynergEyes™ A and M lenses, and SoftPerm® Contact Lens in the following table.

{2}------------------------------------------------

SynergEyes™ PS(paflufocon D-iberfilcon A) HybridContact Lens(Subject Device)SynergEyes™ A andM (paflufocon D-iberfilcon A) HybridContact LensK051035(Predicate Device)SoftPerm®(synergicon A)Contact LensP840004/S7(Predicate Lens)
Lens Characteristics
ManufacturerSynergEyes™ Inc.SynergEyes™ Inc.Ciba Vision Care
Base Curves7.10-9.60mm7.10-9.00mm7.10-8.10mm
RGP Center9.00mm8.40mm8.00mm
Posterior Optic ZoneDiameter8.40mm7.80mm7.00mm
Lens DesignsSphere, Aspheric,Front Surface Toric,Sphere, Aspheric,Front Surface Toric,MultifocalSphere, FrontSurface Toric
Diameters:14.5mm14.5mm14.3mm
Power Range-20.00 to + 20.00D-20.00 to + 20.00D-13.00 to +6.00D
Cylinder Power Range0.50 to 6.00D0.50 to 10.00D
Add Power for Near0.75 to 3.00D
Center Thickness0.18 to 0.30mm0.12 to 0.30mm0.08mm to 0.28mm
Refractive Index (RGP)1.442 (Nd @ 25°C)1.442 (Nd @ 25°C)1.53
Wetting Angle (RGP)42°42°21° CLMA
Specific Gravity (RGP)1.101.101.015
Hardness7979
Indications for UseDaily WearDaily WearDaily Wear
UV BlockingNoNoNo
MaterialPaflufocon D centerRGP hem-iberfilcon A(HEMA, MEMA)Paflufocon D centerRGP hem-iberfilcon A(HEMA, MEMA)Synergicon ARGP and HEMASkirt
TintVisibility BlueVisibility BlueClear
Soft Skirt Water Content27%27%25%
Core (RGP) Water Content< 1%< 1%< 0.2%

6. Non clinical studies:

Results from the full series of physical/chemical, toxicological, and sterility tests were conducted under K051035 for the equivalent lens, the predicate lens for this device. No further non-clinical studies were performed for this submission.

7. Packaging

The primary lens container for shipping is a sterile enclosed medical grade glass vial capped with a screw cap and sterilized with the e-beam sterilization process. The lens is immersed in a sterile buffered normal saline.

8. Clinical data:

A three month clinical study of the SynergEyes™ PS Hybrid Contact Lens was conducted to assess safety and effectiveness for vision correction in daily wear for patients suffering from post surqical refractive error including nearsightedness, astigmatism and irregular astigmatism, or trauma to the eye. The study was designed to evaluate contact lens visual acuity and wearing time; and assess contact lens adverse events and loss of visual acuity.

Overall Findings:

A total of 76 subjects were dispensed into the study of which 44 (80 eyes) [57.9%] completed the study and 32 subjects (58 eves) [42.1%] were discontinued from the study. An additional 8 subjects were not dispensed lenses.

{3}------------------------------------------------

SynergEyes™, Inc. Carlsbad, CA 92008

The population demographics were similar to other contact lens studies with a female to male gender ration of 1.3 to 1.0. The average age of the completed and discontinued subjects was 49 and 43.5 respectfully.

Safety:

Eight (8) adverse events were reported during the study for 4 completed subjects and 3 discontinued subjects. The adverse events included 1 painful light sensitivity (photophobia); 1 loose interrupted suture and infiltrate; 1 abrasion of the cornea upon lens removal; 1 small superficial abrasion; 1 subject (2 eyes) with swelling of the corneas due to inadequate rinsing of hydrogen peroxide disinfection solution from lenses; 1 allergic conjunctivitis; and 1 keratitis intis.

Thirty-two (32) subjects (58 eyes) [42.1%] were discontinued from the study. An additional 8 subjects were never dispensed lenses. The most common reasons for discontinuation were poor outcome with lenses (32.5%), poor comfort (27.5%), poor vision (12.5%), loss of interest (10.0%), non-compliance (10.0%), and handling difficulty (2.5%).

Symptoms, Problems, and Complaints: For the completed eyes, no symptoms were reported at 46.2% of the dispensing visit or follow-up visit examinations and for discontinued eves no symptoms were reported at 28.8% of the dispensing visit or follow-up visit examinations. Symotoms decreased over time for completed eyes but not for discontinued eyes. The most common symptoms reported were discomfort and awareness (28.4% completed, 45.8% discontinued), dryness and scratchiness (23.4%, 22.1%), itchiness and burning (9.3%, 8.9%), and variable vision (8.2%, 13.3%).

Efficacy:

Visual Acuity- Final visual acuity for completed subjects was 20/20 or better (20.0%), 20/25 or better (45.0%), 20/30 (68.8%), and 20/40 or better (81.3%). The visual acuity rates for discontinued subjects were 8.6%, 39.7%, 53.4%, and 70.7% respectfully. Vision correction fluctuated as expected with the instability of the corneal curvature from post surgical conditions and trauma under the contact lens contributing to the change. Three (3) completed eyes and 8 discontinued eyes were reported to have VA decreases of more than 2 lines of Snellen VA when comparing the contact lens VA with the best corrected VA. These findings are expected with this population.

For Wearing Time: Over the study period the average daily wearing time reported by completed patients was 10.6 hours per day.

Conclusion:

The SynergEyes™ PS Hybrid Contact Lens for Post Surgical refractive error and cases involving trauma provided satisfactory performance as expected. The higher than estimated discontinuation rate was anticipated due to the nature of the subject population with the inclusion of subjects who might otherwise have less successful outcomes with other lens types. Overall, the lens performance demonstrated safe and effective use of the device for its intended use.

9. Conclusions drawn from studies Substantial Equivalence:

Information provided in this 510(k) establishes that the SynergEyes™ PS Hybrid Contact Lens are equivalent in optical, chemical and physical properties of the predicate devices and do not raise any questions of safety and effectiveness. The clinical evaluation demonstrated safe and effective lens performance, and where possible equivalence to historical experience with predicate devices. The device is substantially equivalent to the predicate devices material, SynergEves A and M Hybrid Contact Lens (K051035), Paragon HDS-100 (P870024/S36 and

{4}------------------------------------------------

S43), and (K940277); and SoftPerm (synergicon A) contact lens (P840004/S7), and indication for use as a hybrid lens material comprised of a rigid center optic portion and a soft skirt portion.

SynergEyes™ PSHybrid Contact LensJurkus(1998)Chung(2001)
No. of Lenses (Eyes)802135
No. Patients751128
Gender42F/33M5F/6M11F/17M
Mean Age49 (Completed)41 + 19
LensSynergEyes PSSaturn IISoftPerm/Prior RGPusers
IndicationMyopia, Hyperopia, Irr.AstimgatismAstigmatismKeratoconus 22/35(62.9%) and PK 10/35(28.8%)
Completions44 (57.9%)4 (36%)17 (66.7%)
Discontinuations32 (42.1%)7 (64%)11 (33.3%)
For Comfort27.5%28%45.5%
For Vision12.5%28%9.1%

Clinical Experience Equivalence to Other Studies

Material Equivalence Table

SynergEyes™ PS HybridContact Lens forKeratoconus(Subject Device)SynergEyes™ A and MHybrid Contact Lens forDaily Wear(K051035)SoftPerm® ContactLensP840004/S7
PRODUCTION METHODLathingLathingLathing
INTENDED USEDaily WearDaily WearDaily Wear
MATERIALPaflufocon D Centerhem-iberfilcon A skirtPaflufocon D Centerhem-iberfilcon A skirtSynergicon-A
TypeGroup 1 Low WaterGroup 1 Low WaterGroup 1 Low Water
Surface ChargeNon-ionicNon-ionicNon-ionic
Color additiveD&C Green 6D&C Green 6
UV additiveNoNoNo
Dk permeability:1. Revised FATT Polarimetricmethod with edge correction@ 35°C x 10⁻¹¹ (cm²/sec) (mlO₂/ml x mm Hg)RGP Center:FATT: 145ISO: 100RGP Center:FATT: 145ISO: 100RGP Center:ISO: 14
2. ISO 9913-1Soft Skirt: 9.3Soft Skirt: 9.3Soft Skirt: 5.5
Dk/L- Lens transmissibility:1. Revised FATT(Through power range -20 to+ 20D)2. ISO 9913-1 Polarimetricmethod :FATT: 66-77ISO: 46-53FATT: 66-77ISO: 46-53ISO: 17.5
Light transmittance(380nm to 780nm)>90%>90%88-92%

Risk and Benefits:

The risks of the subject device are the same as those normally attributed to the wearing of RGP and soft (hydrophilic) contact lenses on a daily wear base. The benefits to the patient are the same as those for other RGP and soft (hydrophilic) contact lenses. Overall, the risks and benefits associated with daily wear contact lenses are the same as for other daily wear contact lenses and raise no additional concerns for safety or effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 5 2006

SynergEyes, Inc. C/O Richard E. Lippman, O.D., F.A.A.O. Vice President for Ophthalmic Product Regulatory Affairs P. Chiacchierini & Associates, LLC 15825 Shady Grove Rd., Suite 30 Rockville, MD 20850

K060102 Re:

K000102
Trade/Device Name: SynergEyes™ PS (paflufocon D hem-iberfilcon A) Hybrid Contact Lens for Post Surgical Refractive Error and Trauma for Daily Wear Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HQD Dated: January 13, 2006 Received: January 13, 2006

Dear Dr. Lippman:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 1976, the enaordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Act (Act) market the device, subject to the general controls provisions of the Act. The You may, ulciclore, market the de reco, include requirements for annual registration, listing of general controls provisions of the 110 labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classifica (sec above) the existing major regulations affecting your device can
may be subject to such additional controls. Existing major regals may be subject to such adultional controls. Extronic may of to 898. In addition, FDA may be found in the Code of Poderal Registing your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination are not mean Please be advised that FDA s issuative of a stobation with other requirements of the Act
that FDA has made a determination that your device complies with other requirements o that FDA has made a decemmanon and your cared by other Federal agencies. You must or any Federal Statutes and regulations annuding, but not limited to: registration and listing (21 comply with an the Act 3 requirements, nevalles on and acturing practice requirements as sett CFR Part 807), labeling (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 - Richard E. Lippman, O.D., F.A.A.O.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincercly yours,

MB Egelund - MS

Malvina B. Eydelman, M.D Acting Division Director Division of Ophthalmic and Far, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):__K060102

SynergEyes™ PS (paflufocon D hem-iberfilcon A) Hybrid Contact Device Name: Lens for Post Surgical Refractive Error and Trauma for Daily Wear

For use in the correction of eyes with refractive errors resulting Indications For Use: from corneal surgery or trauma including hyperopia and myopia, astigmatism and irregular astigmatism in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 in eyes with astigmatism or irregular astigmatism up to 6.00D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

Prescription Use / X
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Karen Wabnstr

(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises

510(k) Number K060102

Page 1 of 1

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.