K Number
K052832

Validate with FDA (Live)

Date Cleared
2006-05-15

(221 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sauflon New Day® (methafilcon A) Soft (hydrophilic) Visibility tinted contact lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

Device Description

The SAUFLON New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lenses are available in an aqua visibility tint. The lens material (methafilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid, which is cross-linked with ethyleneglycol dimethacrylate. When hydrated the lens consists of 45% HEMA and 55% water by weight of saline immersed in normal saline. The lens is visibility tinted aqua with weight of summe inimelod in his meller on the well and one UV absorbing monomer is used to block UV radiation. Sauflon New Day lenses are daily disposable lenses. The average transmittance characteristics are less than 10% in the UVB range of 280 to 315nm and less than 40% in the UVA range of 315-380nm. The SAUFLON New Day Daily Disposable Lens is a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera.

AI/ML Overview

This 510(k) summary describes a contact lens, the SAUFLON New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens, and aims to establish its substantial equivalence to a predicate device, the SAUFLON 55UV (methafilcon A) Soft (hydrophilic) Visibility Tinted Contact Lens for Daily Wear (K013649).

The acceptance criteria relate to demonstrating that the new device is as safe and effective as the predicate device. The study presented is a comparison of physical, optical, and chemical properties, as well as manufacturing processes and safety tests, between the two contact lenses. No clinical effectiveness study with human readers or AI assistance is mentioned as this is a medical device approval for a contact lens.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and the Reported Device Performance

The acceptance criteria are implicitly that the new device (SAUFLON New Day) demonstrates properties, materials, and manufacturing processes that are identical or equivalent to the legally marketed predicate device (SAUFLON 55UV), and where differences exist (e.g., lens design or modality), these differences do not affect safety or efficacy.

Acceptance Criteria (Implicit from Predicate Comparison)Reported Device Performance (SAUFLON New Day)
Material Composition: Methafilcon AIdentical (Methafilcon A)
Physical/Optical Properties (e.g., Water Content, Transmittance, Dk):Identical across a range of specified measurements (Water Content: 55%, %Transmittance @590nm: 94.61%, %Transmittance @280-315nm: 9.41%, %Transmittance @316-380nm: 36.00%, Dk @35°C: 22.0 x 10^-11, Refractive Index: 1.4020, Tensile Strength: 1.47, Modulus: 0.52, Elongation at Break: 280, Toughness: 1.39)
Powers: -20.00 to +20.00 DIdentical
Color: Aquamarine VisibilityIdentical
Chemical Composition of Finished Lens: Co-polymer of 2-hydroxyethylomethacrylate and methacrylic acid cross-linked with ethyleneglycol dimethacrylate, with benzophenone type UV absorbing monomer, CI Reactive Blue Dye 4, CI Reactive Yellow Dye 86.Identical
Raw Material Specifications: Approved in K013649Identical
Manufacturing Process: Cast MouldingIdentical
Tinting Process: 'In monomer' tintingIdentical
Packaging Materials: Injected molded polypropylene blisters, aluminum foil laminate, printed cartonsIdentical
Lens Packing Solution: 0.9%w/v sodium chloride solution buffered at pH 7.5 to 7.8 with sodium hydrogen carbonate and containing 0.005% w/v poloxamer 407Identical
Packaging Method: Hermetically sealed blister packIdentical
Sterilization Method: Steam SterilizationIdentical
Shelf-life: 5 years (based on stability studies of predicate)Identical (Applicable from predicate studies due to identical materials and processes)
Toxicology (Lens Material Methafilcon A): Non-toxic, non-irritating (cytotoxicity, systemic toxicity, ocular irritation tests)Identical (Toxicity data applicable due to identical material)
Toxicology (Blister Packs): Non-toxic (cytotoxicity, systemic toxicity, ocular irritation tests)Identical (Toxicity data applicable due to identical materials)
Residual Monomers: Below detection limitsData applicable due to identical material
Leachable Color Additives: No detectable amountsData applicable due to identical dyes
Indications for Use: Myopia, Hyperopia, Astigmatism in aphakic and non-aphakic persons with non-diseased eyes, astigmatism up to 2.00 Dioptres not interfering with visual acuity.Identical
Difference in Lens Design: Spherical (predicate)Aspherical (subject device) - "This design alteration does not affect the safety or the efficacy of the lens."
Difference in Modality: Daily Wear (predicate)Daily Disposable wear (subject device) - Labeling adapted accordingly; noted as a "convenient alternative" with reduced risk due to limited exposure.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as a number of tested units for each property. The "test set" here refers to the SAUFLON New Day lens, for which various properties were measured and compared. The data provenance is from Sauflon Pharmaceuticals Ltd., United Kingdom. The studies appear to be prospective in that new measurements were taken for the subject device to compare against the known predicate device. For some aspects (e.g., stability, leachability, toxicology), the data from the predicate device's prior approval (K013649) was deemed applicable due to identical materials and processes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information is not applicable for this submission. The "ground truth" for contact lens properties like water content, transmittance, Dk, etc., is established through standardized physical, optical, and chemical testing methods, not through expert consensus on interpretation. Toxicology and leachability tests also follow established laboratory protocols.

4. Adjudication method for the test set

Not Applicable. As mentioned above, the "ground truth" is derived from objective physical, optical, and chemical measurements and standard laboratory tests, not from subjective expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is a submission for a medical device (contact lens), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable and was not performed.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the comparison is based on:

  • Standardized physical, optical, and chemical measurements: For properties like water content, transmittance, Dk, refractive index, tensile strength, etc.
  • Laboratory test results: For toxicology (cytotoxicity, systemic toxicity, ocular irritation), residual monomers, and color leachability studies.
  • Manufacturing process documentation: To confirm identical processes.
  • Reference to prior FDA clearance (K013649): For established safety and efficacy of the predicate device's materials and processes.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device. There is no concept of a "training set" in this context.

9. How the ground truth for the training set was established

Not applicable. As above, there is no training set for this type of device. The ground truth for the comparison was established through objective, standardized testing and the documented equivalence to a previously cleared device.

{0}------------------------------------------------

MIY 15 2005

K052832

510(K) SUMMARY AS REQUIRED BY SECTION 807.92(c)

SUBMITTER INFORMATION:

Sauflon Pharmaceuticals Ltd. Company Name:

49 - 53 York Street Address: Twickenham Middlesex TW1 3LP 020 8322 4231 Phone:

020 8891 2833 Fax:

Ms Tanya Fair Contact Person:

DATE SUMMARY PREPARED: 22nd September 2005

DATE SUMMARY REVISED: 11th May 2006

DEVICE NAME:

(methafilcon Soft Day A) SAUFLON New Trade Name: (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens Contact Lens Common Name: CLASS II (21 CFR 886.5925) CODE - MVN;LPL Classification: SOFT (HYDROPHILIC) CONTACT LENS

SUBSTANTIAL EQUIVALENCE:

SAUFLON New Day® (methafilcon A) Soft (hydrophilic) Visibility Tinted Daily Disposable Contact Lens are substantially equivalent to SAUFLON 55UV (methafilcon A) Soft (hydrophilic) Visibility Tinted Contact Lenses for Daily Wear that received market clearance pursuant to K013649 and are currently marketed in the USA.

DESCRIPTION of the DEVICE:

.

The SAUFLON New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lenses are available in an aqua visibility tint. The lens material (methafilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate

{1}------------------------------------------------

(HEMA) and methacrylic acid, which is cross-linked with ethyleneglycol dimethacrylate. When hydrated the lens consists of 45% HEMA and 55% water by weight of saline immersed in normal saline. The lens is visibility tinted aqua with weight of summe inimelod in his meller on the well and one UV absorbing monomer is used to block UV radiation. Sauflon New Day lenses are daily disposable lenses.

The average transmittance characteristics are less than 10% in the UVB range of 280 to 315nm and less than 40% in the UVA range of 315-380nm.

The SAUFLON New Day Daily Disposable Lens is a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

Chord Diameter:14.0mm to 15.0mm
Centre Thickness @ -3.00DS:0.07mm
Base Curve:8.70mm
Powers:-20.00 to +20.00Presently available:-0.50 to -6.00 (0.25DS steps)-6.00 to -10.00 (0.50DS steps)+0.50 to +6.00 (0.25DS steps)
The physical/optical properties of the lens are:
Refractive Index:1.40 +/- 0.32
%Transmittance @ 590nm94.61
% Transmittance @ 280-315nm9.41
% Transmittance @ 316-380nm36.00
Surface Character:Hydrophilic
Water Content:55%
Specific Gravity1.09
Oxygen Permeability (Dk):$22.0 x 10^{-11} (cm^2 / sec)$
(Fatt Method for determination of oxygen permeability)(ml O2 / ml x mm Hg) at 35°C

{2}------------------------------------------------

The table overleaf Table 7.1 Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft (Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft Disposable + Inted Contact Lens for Daily Wear shows the rationale for (11) and the Saufloon the predicate device Sauflon 55UV and the Sauflon New Day Lens.

This substantial equivalence is based on the lens having the same indications for use, being equivalent in terms of performance and sharing the same chemical composition while utilising the same manufacturing method, sterilisation and packaging processes. Wheelers material Methafilcon A utilised by both predicate and test device was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests as well as no residual monomers or dyes being present during leachability studies. as no reciente device Sauflon 55UV has an established shelf-life of 5 years based on both accelerated and real time stability studies presented in K013649 The rationale for these studies being applicable to the Sauflon New Day is that the same materials and processes are used for manufacture and packaging and there is no reason to believe that this lens cannot exceed the already established shelf-life. The lens do vary in design in that the Sauflon 55UV lens is of spherical design while the Sauflon New Day lens is aspheric. This design alteration does not affect the safety or the efficacy of the lens. The design engineering drawing for the aspheric design is included in Section 8 of this submission.

Due to the varying modality of the predicate i.e. a daily wear lens and subject lens i.e. a daily disposable lens the labelling of the package insert does vary in that the package insert for a daily disposable is not required to contain as much detail as a daily wear lens requires. Items such as a cleaning and care regime for the lenses are not required as the New Day lens is disposed of following use.

The package insert does however contain all necessary information for the safe and effective use of a daily disposable lens including lens handling, insertion, removal and precautions.

The final draft labelling for the Sauflon New Day lens blister pack, carton, patient instructions and professional fitting and information guide are included in Section 5 of this submission. The patient instructions for the predicate device are included in Section 6 of this submission.

{3}------------------------------------------------

Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) VisibilityTinted Contact Lens for Daily Wear.
PREDICATE DEVICE - SAUFLON 55UV (K013649)SUBJECT DEVICE - SAUFLON NEW DAY
LENS DESCRIPTION(Methafilcon A) Soft (Hydrophilic) Visibility TintedContact LensIdentical
INDICATIONS FOR USEMyopia, Hyperopia and AstigmatismIdentical
MODALITYDaily WearDaily Disposable wear
LABELLINGFollows the recommendations given in the guidancedocument for Daily Wear Soft Contact lenses issued byFDA in May 1994.Due to the fact that these are daily disposable lenses and notdaily wear lenses the labelling has been tailored for this andareas such as care regime or replacement programmes of thelenses are not relevant and therefore have not been included.
Labelling has been provided for lens blister, carton, PatientInstructions and Professional Fitting and Information guidesuitable for a daily wear lens.Labelling has been provided for lens blister, carton, PatientInstructions and Professional Fitting and Information guidesuitable for a daily disposable lens.
LENS MATERIALMethafilcon AIdentical
WATER CONTENT55%Identical
%TRANSMITTANCE @590NM94.61%Identical
%TRANSMITTANCE @280-315NM9.41%Identical
% TRANSMITTANCE @31680NM36.00%Identical
DK @35°C (EDGECORRECTED)22.0 x 10-11Identical
POWERS-20.00 to +20.00 DIdentical
Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) VisibilityTinted Contact Lens for Daily Wear.
PREDICATE DEVICE - SAUFLON 55UV (K013649)SUBJECT DEVICE -SAUFLON NEW DAY
COLOURAquamarine VisibilityIdentical
REFRACTIVE INDEX1.4020Identical
TENSILE STRENGTH1.47Identical
MODULUS0.52identical
ELONGATION AT BREAK280Identical
TOUGHNESS1.39Identical
LENS DESIGNSphericalAspherical
CHEMICAL COMPOSITIONOF FINISHED LENSMethafilcon A which is a copolymer of 2-hydroxyethylomethacrylate and methacrylic acid crosslinked with ethyleneglycol dimethacrylate.with benzophenone type UV absorbing monomer, 2-(4-benzoyl-3-hydroxyphenoxy) ethyl acrylate.Identical
RAWMATERIALSPECIFICATIONSCopies of all the raw material specifications were includedand approved in K013649Identical
MANUFACTURING PROCESSCast MouldingIdentical
Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) VisibilityTinted Contact Lens for Daily Wear.
PREDICATE DEVICE - SAUFLON 55UV (K013649)SUBJECT DEVICE - SAUFLON NEW DAY
TINTING PROCESS'In monomer' tintingIdentical
TINT USEDCI Reactive Blue Dye 4CI Reactive Yellow Dye 86Identical
PACKAGING MATERIALSInjected molded polypropylene blisters covered byaluminium foil laminate and blister strips are packed intoprinted cartonsIdentical
LENS PACKING SOLUTION0.9%w/v sodium chloride solution buffered at pH 7.5 to7.8 with sodium hydrogen carbonate and containing0.005% w/v poloxamer 407Identical
PACKAGING METHODHermetically sealed blister packIdentical
STABILITY TESTSReal time studies (5 years at 20°C) and accelerated stabilitystudies (45°C) carried out.All parameters including sterility remained withinspecification.These studies are applicable to the New Day lens.
LEACHABILITYSTUDIES-RESIDUALMONOMERSResidual Monomers were analysed by HPLC. The levels ofmonomers were below detection limitsAs the material used in the two lenses i.e. Methafilcon A andthe composition of the lenses are identical this leachabilitydata is therefore applicable.
Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft (Hydrophilic) Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear.
PREDICATE DEVICE - SAUFLON 55UV (K013649)SUBJECT DEVICE - SAUFLON NEW DAY
LEACHABILITY STUDIES -COLOUR ADDITIVESNo detectable amounts of the dyes were found in the extract tested.As the dyes used in the New Day lens is identical to those used for Sauflon 55UV lens the leachability data is therefore applicable.
STERILISATION METHODSterilised in an autoclave by Steam SterilisationIdentical
SHELF-LIFEShelf-life is 5 years based on stability studiesIdentical
TOXIOCOLOGY OF THE LENSThe lens material Methafilcon A was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests.The identical lens material is utilised therefore the toxicity data is applicable.
TOXICOLOGY OF THEBLISTER PACKSThe moulded polypropylene blister material was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests.The identical materials are utilised in the packaging of the Sauflon New Day lens therefore the toxicity data is applicable.

{4}------------------------------------------------

.

:

.

·

{5}------------------------------------------------

and the comments of the comments of

,

{6}------------------------------------------------

:

,

{7}------------------------------------------------

PRECLINICAL TESTING

The results of toxicology testing (cytotoxicity, acute systemic toxicity and acute ocular irritation) show the lens material methafilcon A to be non-toxic and nonirritating. Furthermore, the results of residual monomer and colour leachability testing demonstrate that the respective extracts did not contain significant levels of leachable colour or residual monomers.

The physical optical, and chemical properties of the SAUFLON New Day® Daily Disposable Lenses (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens are equivalent to those of Sauflon 55 UV (methafilcon A) Soft Hydrophilic Visibility Tinted Contact Lens for Daily Wear. This lens is in group 4, Ionic, high water content polymers as established by the FDA and located in the Guidance Document for Daily Wear Contact Lenses, Revised Edition May 1994.

The lens will be sterilised and packaged in the same manner as previously cleared in K013649. This lens will also be sterility released by parametric release, as cleared in K013649.

INDICATIONS FOR USE

The SAUFLON New Day® Daily Disposable Lenses (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

The predicate device Sauflon 55UV lens is prescribed for daily wear however the Sauflon New Day is a Daily Disposable lens.

The New Day lens is a convenient alternative to a daily wear lens as the patient can dispose of the lens at the end of the day and then insert a brand new pair the next morning.

This eliminates the need for the patient to carry out a cleaning regime for the lens and the risk of experiencing contact lens related problems is reduced due to the limited exposure time of using one particular set of lenses.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the agency's mission to promote health, well-being, and human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 15 2006

Sauflon Pharmaceuticals, Ltd. c/o Ms. Tanya Fair 49-53 York Street Twickenham, Middlesex TW1 3LP, United Kingdom

Re: K052832

. K022032
Trade/Device Name: Sauflon New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens

Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lenses Regulatory Class: II Product Code: MVN; LPL Dated: May 4, 2006 Received: May 4, 2006

Dear Ms. Fair:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

MB Egerthuis - MS

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K052832

Device Name:

SAUFLON New Day (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens

Indications For Use:

The Sauflon New Day® (methafilcon A) Soft (hydrophilic) Visibility tinted contact lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of the CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)OROver-The Counter
Ming-chner Shi
(Division Sign-Off)Division of Ophthalmic Ear,Nose and Throat Devises
510(k) NumberK052832

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.