K Number
K052232
Date Cleared
2005-09-28

(42 days)

Product Code
Regulation Number
870.5150
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pronto V3 Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels of the arterial systems.

Device Description

The Pronto V3 extraction catheter is a dual lumen catheter with related accessories. The extraction lumen allows for the aspiration and removal of embolic material (thrombus/debris) using the included syringe, extension line and stopcock. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate atraumatic advancement of the catheter into the blood vessel and maximize extraction of emboli/thrombi through the extraction lumen. Incorporated within the catheter distal tip is a non-blood contacting radiopaque marker for fluoroscopic visualization. The catheter is a monorail design with a distal flexible region with stiffness along the shaft tapering to a stiff proximal region. The distal segment of the catheter is coated with a hydrophilic coating to lubricate the catheter for ease of insertion. The catheter has an approximate outer diameter of 0.065 inches, allowing delivery through standard 6Fr. guide catheters. The smaller wire lumen of the catheter is able to accommodate guide wires that are ≤ 0.014" in diameter. The catheter will be available in working lengths of 40 to 145 cm with the length changes being made in the proximal rail section. The proximal end of the catheter incorporates a standard luer adapter to facilitate the attachment of the catheter to the included extension line, stopcock, and syringe. A 74 micrometer filter basket (not identified in the schematic below) is included for assistance in filtering the blood removed during the procedure for laboratory analysis of thrombus.

AI/ML Overview

The provided text describes a medical device, the Vascular Solutions Pronto V3™ Extraction Catheter, and its FDA 510(k) clearance (K052232).

Here's an analysis to extract the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (typically based on predicate device performance or relevant consensus standards)Reported Device Performance (from non-clinical testing)
Suitable for intended use: removal of fresh, soft emboli and thrombi from arterial systems.Confirmed suitability for intended use through "assessments of the design verification" and "biocompatibility assessments."
Design verification (e.g., catheter dimensions, material properties, functionality)"Testing conducted included assessments of the design verification of the Pronto V3 Extraction Catheter."
Biocompatibility"Biocompatibility assessments" were conducted.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. The submission indicates "No clinical evaluations of this product have been conducted."
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable, as no clinical testing was performed and therefore no ground truth was established by experts for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable, as no clinical testing was performed and therefore no adjudication was necessary.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a catheter, not an AI or imaging diagnostic device. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This device is a catheter, not an AI or imaging diagnostic device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

  • For the non-clinical testing, the "ground truth" would be established by engineering specifications, material standards, and in-vitro test results confirming the physical and functional properties of the device met predetermined design requirements. No clinical ground truth (e.g., expert consensus, pathology, or outcomes data) was used.

8. The sample size for the training set:

  • Not applicable. This is a physical medical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:

  • Not applicable, as no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 15, 2021

Vascular Solutions, Inc. Sara Coon Sr. R.A. Associate 6464 Sycamore Court Minneapolis, Minnesota 55369

Re: K052232

Trade/Device Name: Pronto V3 Extraction Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: OEZ

Dear Sara Coon:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 28, 2005. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory. Oconnell@FDA.HHS.gov.

Sincerely,

Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2021.09.15

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a bird in flight, composed of three curved lines.

Public Health Service

SEP 2 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vascular Solutions, Inc. c/o Ms. Sara L. Coon Senior Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, MN 55369

Re: K052232 Pronto V3TM Extraction Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: II Product Code: DXE Dated: August 16, 2005 Received: August 17, 2005

Dear Ms. Coon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Sara L. Coon

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Duna R. Vochner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

2

Enclosure

{3}------------------------------------------------

Indications For Use Statement

  • K052232 - - - - - - - - - - - - - - - -510(k) Number:

Vascular Solutions Pronto V3™ Extraction Catheter Device Name:

Indications for Use:

The Pronto V3 Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels of the arterial systems.

x Prescription Use (Part 21 CFR 801 Subpart D)

:

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

DARINA R. lochner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number Kos 222232

Page 1 of _ | _ | _ _

{4}------------------------------------------------

K052232

okl. 2 8 2005

Appendix A: Summary of Safety and Effectiveness

Common/Usual Name:Embolectomy Catheter
Product Trade Name:Pronto V3™ Extraction Catheter
Classification Name:UnclassifiedProduct Code, DXE
Manufacturer:Vascular Solutions, Inc.6464 Sycamore CourtMinneapolis, Minnesota 55369
Establishment Registration:2134812
Contact:Sara L. CoonSenior Regulatory Affairs Associate(763) 656-4300 phone(763) 656-4250 fax
Performance Standards:No performance standards have been developedunder section 514 for this device.

Device Description:

The Pronto V3 extraction catheter is a dual lumen catheter with related accessories. The extraction lumen allows for the aspiration and removal of embolic material (thrombus/debris) using the included syringe, extension line and stopcock. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate atraumatic advancement of the catheter into the blood vessel and maximize extraction of emboli/thrombi through the extraction lumen. Incorporated within the catheter distal tip is a non-blood contacting radiopaque marker for fluoroscopic visualization. The catheter is a monorail design with a distal flexible region with stiffness along the shaft tapering to a stiff proximal region. The distal segment of the catheter is coated with a hydrophilic coating to lubricate the catheter for ease of insertion. The catheter has an approximate outer diameter of 0.065 inches, allowing delivery through standard 6Fr. guide catheters. The smaller wire lumen of the catheter is able to accommodate guide wires that are ≤ 0.014" in diameter. The catheter will be available in working lengths of 40 to 145 cm with the length changes being made in the proximal rail section. The proximal end of the

{5}------------------------------------------------

catheter incorporates a standard luer adapter to facilitate the attachment of the catheter to the included extension line, stopcock, and syringe. A 74 micrometer filter basket (not identified in the schematic below) is included for assistance in filtering the blood removed during the procedure for laboratory analysis of thrombus.

Intended Use:

The Pronto V3 Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels of the arterial systems.

Summary of Non-Clinical Testing:

Testing conducted included assessments of the design verification of the Pronto V3 Extraction Catheter along with biocompatibility assessments. The results of this battery of tests confirmed the suitability of the Pronto V3 Extraction Catheter for its intended use.

Summary of Clinical Testing:

No clinical evaluations of this product have been conducted.

Predicate Devices:

The Pronto V3 Extraction Catheter is similar in intended use to the Vascular Solutions, Inc. Pronto Extraction Catheter and the Medtronic Export™ Aspiration Catheter.

Conclusions:

The Pronto V3 Extraction Catheter is substantially equivalent to the Pronto Extraction Catheter and the Export Aspiration Catheter. The testing performed confirms that the Pronto V3 Extraction Catheter will perform as intended.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

SEP 2 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vascular Solutions, Inc. c/o Ms. Sara L. Coon Senior Regulatory Affairs Associate 6464 Sycamore Court Minneapolis. MN 55369

Re: K052232 Pronto V3TM Extraction Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: II Product Code: DXF Dated: August 16, 2005 Received: August 17, 2005

Dear Ms. Coon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Ms. Sara L. Coon

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincercly yours.

Duna R. Vochner

☑ Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{8}------------------------------------------------

Indications For Use Statement

K052232 510(k) Number:

Vascular Solutions Pronto V3™ Extraction Catheter Device Name:

Indications for Use:

The Pronto V3 Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels of the arterial systems.

× Prescription Use (Part 21 CFR 801 Subpart D)

::

の 2007年の1000年の1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には1000年には1000年に10000000000000000000000000000000000000000000000000

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dauna R. Lochner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number Ko 5 2 2 3 2

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).