When intended to promote fusion of the cervical spine and the thoracic spine, (CI-T3), the VERTEX® Reconstruction System is indicated for the following:
DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
The use of screws (3.5mm, 4.0mm and 4.5mm cancellous and 4.0mm cortical) is limited to placement in T1-T3 in treating thoracic conditions only. The screws are not intended to be placed in the cervical spine.
Titanium ATLAS™ Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
The VERTEX® Reconstruction System is a posterior system, which consists of a variety of shapes and sizes of rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailormade for the individual case. Titanium ATLAS™ cable may be used with this system at the surgeon's discretion.
The VERTEX® Reconstruction System is fabricated from medical grade titanium alloy. The VERTEX® Reconstruction System also includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol – NiTi). Shape Memory Alloy is compatible with titanium or titanium alloy implants only. Do not use with stainless steel.
The purpose of this submission was to add modified setscrews to the existing system.
The provided text is a 510(k) summary for the VERTEX® Reconstruction System, a spinal fixation device. It details the device's description, indications for use, and a claim of substantial equivalence to previously cleared devices.
Crucially, this document does not contain information about acceptance criteria or a study demonstrating the device meets such criteria in terms of performance metrics like sensitivity, specificity, accuracy, or similar quantitative measures. The approval is based on "substantial equivalence" to existing cleared devices, primarily supported by mechanical test results.
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor can I provide answers to the questions regarding sample size, data provenance, expert involvement, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details. These types of details are not typically found in a 510(k) summary for a spinal implant device, as the regulatory pathway relies on demonstrating equivalence rather than establishing new performance benchmarks through clinical or AI-specific studies.
The document states:
- Substantial Equivalence: "Documentation, including mechanical test results, has been provided which demonstrates that the subject VERTEX™ Reconstruction System components are substantially equivalent to VERTEX® Reconstruction System components previously cleared in K042498 (SE 10/07/04) and K042789 (SE 12/21/04)."
This indicates that the "study" proving the device met the criteria was a demonstration of its mechanical and material properties being comparable to already approved devices, rather than a clinical trial or performance study against specific acceptance criteria for diagnostic or algorithmic performance.
{0}------------------------------------------------
AUG 2 3 2005
KOS2180
VERTEX® Reconstruction System 510(k) Summary July 2005
| I. | Company: | Medtronic Sofamor Danek USA, Inc.1800 Pyramid PlaceMemphis, Tennessee 38132(901) 396-3133 |
|---|---|---|
| Contact: | Richard W. Treharne, Ph.D.Senior Vice President, Regulatory Affairs | |
| II. | Product Name: | VERTEX Reconstruction System |
| Classification Name: | Spinal Interlaminal Fixation Orthosis | |
| Regulation Number: | 888.3050 | |
| Code: | KWP |
III. Description:
The VERTEX® Reconstruction System is a posterior system, which consists of a variety of shapes and sizes of rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailormade for the individual case. Titanium ATLAS™ cable may be used with this system at the surgeon's discretion.
The VERTEX® Reconstruction System is fabricated from medical grade titanium alloy. The VERTEX® Reconstruction System also includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol – NiTi). Shape Memory Alloy is compatible with titanium or titanium alloy implants only. Do not use with stainless steel.
The purpose of this submission was to add modified setscrews to the existing system.
IV Indications
When intended to promote fusion of the cervical spine and the thoracic spine, (CI-T3), the VERTEX® Reconstruction System is indicated for the following:
DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.
$$
\begin{array}{ccc}
\heartsuit & \heartsuit & \heartsuit & \heartsuit \
\end{array}
$$
{1}------------------------------------------------
KOSZIYO
Hooks and Rods
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
Screws/Connectors
The use of screws (3.5mm, 4.0mm and 4.5mm cancellous and 4.0mm cortical) is limited to placement in T1-T3 in treating thoracic conditions only. The screws are not intended to be placed in the cervical spine.
Titanium ATLAS™ Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
Substantial Equivalence: V.
Documentation, including mechanical test results, has been provided which demonstrates that the subject VERTEX™ Reconstruction System components are substantially equivalent to VERTEX® Reconstruction System components previously cleared in K042498 (SE 10/07/04) and K042789 (SE 12/21/04).
ﻢ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ 1
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 3 2005
Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek, Inc. 1800 Pyramid Place Memphis, Tennessee 38132
Re: K052180
Trade/Device Name: VERTEX® Reconstruction System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: August 9, 2005 Received: August 10, 2005
Dear Dr. Treharne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1 July 2005
510(k) Number (if known): ≤052183
Device Name:
Indications for Use
When intended to promote fusion of the cervical spine and the thoracic spine, (C1-T3), the VERTEX® Reconstruction System is indicated for the following:
DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.
Hooks and Rods
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
Screws/Connectors
The use of screws (3.5mm and 4.0mm cancellous, and 4.0mm cortical) is limited to placement in TI-T3 in treating thoracic conditions only. Screws are not intended to be placed in the cervical spine.
Titanium ATLAS™ Cable used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
AND/OR Prescription Use ※ Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of General, Restorative, and Neurological Devices
-tth) Mmm K052180
N/A