When used as a pedicle screw fixation system in the non-cervical posterior spine in skeletally mature patients, the TSRH® Spinal System is indicated for one or more of the following: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scoliosis, (6) kyphosis, (7) spinal tumor, and/or (8) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the TSRH® Spinal System is indicated for skeletally mature patients: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) joint; (2) who are receiving fusions using autogenous bone graft only; (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the TSRH® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spondylolisthesis, (3) fracture, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) spinal stenosis, (6) pseudarthrosis, (7) tumor resection, and/or (8) unsuccessful previous attempts at spinal fusion.
For anterior use only the TSRH® Spinal System has the additional indications of: (1) spinal stenosis and/or, (2) spondylolysis.
The purpose of this 510(k) submission is to add modified TSRH® components along with previously cleared ARROWHEAD® Spinal System, CD HORIZON® Spinal System and COLORADO II® Spinal System components to the existing TSRH® Spinal System.
The TSRH® Spinal System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.
The TSRH® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, cross connectors, and connecting components. In addition, GDLH® rods, DYNALOK PLUS™ bolts, CD HORIZON® Low Profile MULTI-SPAN® CROSSLINK® Plates, GDLH® rod/bolt connectors, GDLH® Variable Angle T-Bolts, and GDLH® and CD HORIZON® set screws and locking screws may be used with the TSRH® Spinal System.
The TSRH® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The hooks are intended for posterior use only and the staples are for anterior use only. The TSRH-3D® connectors and TSRH-3D® screws are intended for posterior use only.
The TSRH® Spinal System components are fabricated from stainless steel. Alternatively, they may be fabricated from medical grade titanium alloy or medical grade titanium. The TSRH® Spinal System may be sold sterile or non-sterile.
This document pertains to the Medtronic Sofamor Danek TSRH® Spinal System and is a 510(k) summary for regulatory clearance, not a study evaluating device performance against acceptance criteria in the typical sense of a clinical or AI-based study. It focuses on demonstrating substantial equivalence to previously cleared devices through mechanical testing and comparison of indications for use.
Therefore, many of the requested categories for acceptance criteria and study details are not directly applicable to this type of regulatory submission. However, I can extract the relevant information where possible and indicate when a category is not relevant or not provided in the document.
1. Table of Acceptance Criteria and Reported Device Performance
For this 510(k) submission, the primary "acceptance criterion" is demonstrating substantial equivalence to predicate devices, which is largely met through mechanical testing and comparison of indications for use. The document refers to "mechanical test results" as the primary evidence. Specific quantitative acceptance criteria or detailed performance metrics are not given in this summary.
| Acceptance Criterion (Regulatory) | Reported Device Performance |
|---|---|
| Substantial Equivalence to Predicate Devices (K050282, K042025, K991091, K032560) | Met: "The subject components were demonstrated to be substantially equivalent..." "Mechanical test results were provided in support of this application." The labeling is also identical to a cleared predicate. |
| Safety and Effectiveness for Stated Indications | Assumed Met: By demonstrating substantial equivalence to devices already cleared for safety and effectiveness for similar indications. |
| Material Compatibility | Met: Fabricated from stainless steel, medical grade titanium alloy, or medical grade titanium, consistent with standards for implants. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size: Not specified in terms of the number of specific test articles or components. The document refers to "mechanical test results" in general.
- Data Provenance: The document does not specify the country of origin for the mechanical test data. Given it's a US FDA submission, it's typically assumed such testing is conducted in a compliant manner, often in the US or in accredited labs elsewhere. The testing is for pre-market notification, so it is, in effect, prospective for the purposes of this submission, though the testing itself would have been conducted prior to submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This concept is not applicable to a mechanical testing submission for substantial equivalence. "Ground truth" in this context refers to the defined engineering specifications, material properties, and mechanical load requirements that the device must meet, which are established by engineering standards and regulatory guidance for spinal implants.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods are typically relevant for human interpretation tasks or clinical outcome assessments, not for mechanical testing of an engineering device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This is a medical device (spinal implant) submission, not an AI or imaging device submission that would involve multi-reader studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a spinal implant, not an algorithm.
7. The Type of Ground Truth Used
For mechanical testing, the "ground truth" implicitly refers to engineering specifications, material standards, and biomechanical performance metrics established for spinal instrumentation. These are often derived from industry standards (e.g., ASTM, ISO) and biomechanical literature for spinal stability and fixation.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" in the context of a mechanical device submission. This term applies to machine learning/AI models.
9. How the Ground Truth for the Training Set Was Established
Not applicable (see point 8).
{0}------------------------------------------------
AUG 16 2005
TSRH® Spinal System 510(k) Summary July 2005
- I. Medtronic Sofamor Danek USA, Inc. Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133
Contact: Richard. W. Treharne
Sr. Vice President Regulatory Affairs
Proposed Proprietary Trade Name: TSRH® Spinal System II.
Regulation Numbers: 888.3050, 888.3060 and 888.3070 Regulation Names: Spinal Interlaminal Fixation Orthosis, Spinal Intervertebral Body Fixation Orthosis, Spondylolisthesis Spinal Fixation Device System, and Pedicle Screw Spinal System. Codes: NKB, KWP, MNH and MNI Class III - Pre-amendment for indication of degenerative disc disease
III. Description
The purpose of this 510(k) submission is to add modified TSRH® components along with previously cleared ARROWHEAD® Spinal System, CD HORIZON® Spinal System and COLORADO II® Spinal System components to the existing TSRH® Spinal System.
The TSRH® Spinal System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.
The TSRH® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, cross connectors, and connecting components. In addition, GDLH® rods, DYNALOK PLUS™ bolts, CD HORIZON® Low Profile MULTI-SPAN® CROSSLINK® Plates, GDLH® rod/bolt connectors, GDLH® Variable Angle T-Bolts, and GDLH® and CD HORIZON® set screws and locking screws may be used with the TSRH® Spinal System.
The TSRH® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The hooks are intended for posterior use only and the staples are for anterior use only. The TSRH-3D® connectors and TSRH-3D® screws are intended for posterior use only.
000007
{1}------------------------------------------------
The TSRH® Spinal System components are fabricated from stainless steel. Alternatively, they may be fabricated from medical grade titanium alloy or medical grade titanium. The TSRH® Spinal System may be sold sterile or non-sterile.
Indications for Use: IV.
When used as a pedicle screw fixation system in the non-cervical posterior spine in skeletally mature patients, the TSRH® System is indicated for one or more of the following: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scollosis, (6) kyphosis, (7) spinal tumor, and/or (8) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the TSRH® Spinal System is indicated for skeletally mature patients: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) joint; (2) who are receiving fusions using autogenous bone graft only; (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the TSRH® Spinal System is intended for the following indications: (1) degenerative disc discase (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spondylolisthesis, (3) fracture, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) spinal stenosis, (6) pseudarthrosis, (7) tumor resection, and/or (8) unsuccessful previous attempts at spinal fusion.
For anterior use only the TSRH® Spinal System has the additional indications of: (1) spinal stenosis and/or, (2) spondylolysis.
V. Substantial Equivalence:
The subject components were demonstrated to be substantially equivalent to TSRH® Spinal System components manufactured by Medtronic Sofamor Danek and cleared by the FDA in K050282 (02/18/05), as well as to CD HORIZON® Spinal System components cleared in K042025 (SE 08/25/04), COLORADO II® Spinal System components cleared in K991091 (SE 05/10/00) and ARROWHEAD® Spinal System components cleared in K032560 (SE 03/09/04). Mechanical test results were provided in support of this application. The labeling for this device is identical to that cleared by the agency in K050282.
© 2005 Medtronic Sofamor Danek
000008
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three stripes representing the three branches of government. The eagle is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular pattern. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 6 2005
Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132
Re: K052054
Trade/Device Name: TSRH® Spinal System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3060, 21 CFR 888.3070 Regulation Name: Spinal interlaminal fixation orthosis , Spinal intervertebral body fixation orthosis, Pedicle screw spinal system Regulatory Class: III Product Code: KWP, MNH, MNI, NKB Dated: July 28, 2005 Received: July 29, 2005
Dear Dr. Treharne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
80
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 510(k) Number (if known): | KOS2054 |
|---|---|
| Device Name: | TSRH ® Spinal System |
Indications for Use:
. .
When used as a pedicle screw fixation system in the non-cervical posterior spine in skeletally mature patients, the TSRH® Spinal System is indicated for one or more of the following: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scoliosis, (6) kyphosis, (7) spinal tumor, and/or (8) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the TSRH® Spinal System is indicated for skeletally mature patients: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (LS-S1) joint; (2) who are receiving fusions using autogenous bone graft only; (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the TSRH® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spondylolisthesis, (3) fracture, (4) spinal deformities (i.e., scoliosis, and/or lordosis), (5) spinal stenosis, (6) pseudarthrosis, (7) tumor resection, and/or (8) unsuccessful previous attempts at spinal fusion.
For anterior use only the TSRH® Spinal System has the additional indications of: (1) spinal stenosis and/or, (2) spondylolysis.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) (Optional 1-2-96) | X |
|---|---|
| --------------------------------------------------------- | --- |
OR
| Over-the-counter Use | _________________ |
|---|---|
| ---------------------- | ------------------- |
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K052054 |
|---|---|
| --------------- | --------- |
000006
N/A