K Number
K051749
Manufacturer
Date Cleared
2005-09-13

(76 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Estech Cobra Surgical System is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during general surgery to coagulate soft tissues. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

Device Description

The Estech Cobra Surgical System is a family of devices including RF Ablation Probes, with either a malleable or flexible shafts, used in conjunction with the Cobra Electrosurgical Unit (ESU) and other accessories, including the Estech Cobra Bipolar Recording Electrode. The System is intended for use by surgeons for the coagulation of cardiac and soft tissues during open surgical procedures.

AI/ML Overview

This device (Estech Cobra Surgical System) is an electrosurgical device, and the submission is a 510(k) premarket notification for an amendment to the existing device. The information provided heavily focuses on regulatory comparisons to predicate devices rather than a detailed clinical study for acceptance criteria.

Based on the provided text, the "acceptance criteria" for this amendment are primarily based on substantial equivalence to existing predicate devices.

Here's an analysis of the provided text in relation to your questions, noting where information is not available:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied to be substantial equivalence to predicate devices in terms of intended use, technology, design, and materials, and satisfactory preclinical performance data for any modifications. Specific numerical performance metrics (e.g., coagulation depth, lesion size) are not explicitly stated as "acceptance criteria" in a table format with corresponding reported performance, as would be typical for a new device's clinical study.

Acceptance Criteria CategoryReported Device Performance Assessment in Submission
Substantial Equivalence (General)The XL version of the Estech Cobra Surgical System is substantially equivalent in intended use, technology, design and materials to the predicate device.
The Estech Cobra Surgical Probe is substantially equivalent in intended use, technology, design and materials to the predicate system.
The Estech Cobra Bipolar Recording Electrode is substantially equivalent to the surgical console.
BiocompatibilityNo biocompatibility data is necessary since the materials in the modified ablation probes (XL and Surgical Probe) are identical to those in the ablation probes described in the predicate 510(k)s.
Preclinical Performance - XL VersionPreclinical performance data is supplied to demonstrate that the XL version of the cooled probe meets pertinent specifications as the predicate version.
Preclinical Performance - Surgical ProbeNo performance data is necessary for the Estech Cobra Surgical Probe since it is identical in design and manufacture to the predicate device.
Preclinical Performance - Bipolar Recording ElectrodePreclinical performance data is supplied to demonstrate that the Estech Cobra Bipolar Recording Electrode operates as intended.
Compliance with StandardsAdditionally, Estech certifies that the accessories are in compliance with the pertinent standard for electrosurgical devices, AAMI HF-18.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a clinical "test set" with human subjects or a specific sample size for a study proving acceptance criteria. The performance data mentioned is "preclinical performance data." This usually refers to bench testing or animal studies.

  • Sample Size: Not specified for preclinical performance tests.
  • Data Provenance: "Preclinical performance data" implies laboratory or animal testing, not human clinical trials. No country of origin for such data is specified, but it would typically be conducted by the manufacturer or a contract research organization. The context suggests that the submission is from the US, implying data generated or submitted to the US FDA.
  • Retrospective or Prospective: Not applicable as it's not a clinical study involving patients. Preclinical data would be generated prospectively as part of the validation process.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. Since the described data is "preclinical performance data" and not a clinical study with a human "test set" requiring ground truth established by experts, this section is not applicable.

4. Adjudication Method for the Test Set

Not applicable, as there is no described clinical "test set" requiring expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study is mentioned. The device is a surgical tool, not an imaging or diagnostic device that typically undergoes MRMC studies.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is a physical surgical device, not an algorithm.

7. Type of Ground Truth Used

For the preclinical data provided, the "ground truth" would be established by:

  • Engineering specifications: Performance of the device (e.g., power output, temperature control, lesion size) compared against predefined engineering targets derived from the predicate device's performance.
  • Compliance with standards: Meeting the requirements of AAMI HF-18.
  • Predicate device performance: The "ground truth" for the new XL version is the established performance characteristics and safety profile of the predicate device.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML algorithm. There's no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical electrosurgical device.

{0}------------------------------------------------

Endoscopic Technologies, Inc.

ESTECH Cobra System Model Amendma Premarket Notification

SECTION 1 - GENERAL DESCRIPTION 510(k) Notification Information Collection

Sponsor Endoscopic Technologies, Inc.; d/b/a Estech

K 051749

Person Responsible for File Art Bertolero Chief Executive Officer Estech Technologies

Device Trade Name Estech Cobra Surgical System™ Accessories presented in this premarket notification: Estech Cobra Adhere XL Surgical System Estech Cobra Surgical Probe Estech Cobra Bipolar Recording Electrode

Common Name Electrosurgical Probes and accessories

Classification Name

Classification Name
Class II, Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400);

Product Code GEI

Establishment Registration Number 2953686

Address of Manufacturing Facility and Sterilization Site:

Manufacturing FacilitySterilization Site
4135 Blackhawk Plaza CircleSuite 150Danville, CA 94506Sterigenics5725 W. Harold Gatty Dr.Salt Lake City, UT 84116
Tel: 925-648-3500Tel: 801-328-9901
Contact: Art BertoleroFax: 801-328-9002
Establishment Registration:2953686Establishment Registration:1721676

{1}------------------------------------------------

SECTION 2 - 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

CategoryComments
Sponsor:Estech4135 Blackhawk Plaza Circle .Suite 150Danville, CA 94506Tel: 925-648-3500
Correspondent:Craig CoombsCoombs Medical Device Consulting1193 Sherman StreetAlameda, CA 94501
Contact Information:Tel: 510-337-0140Fax: 510-337-0416
Device Common Name:Electrosurgical Probe
Device Proprietary Name:Estech Cobra Surgical System™
Device Classification:21 CFR 878.4400

e Information:

Predicate Device Information:

Predicate Devices:Electrosurgical Probe (K981981)Cobra Adhere Surgical System (K041599)Detect Mapping and Pacing Tool (K040812)
Predicate Device Manufacturers:Boston Scientific CorporationEndoscopic TechnologiesMedtronic
Predicate Device Common Name:Electrosurgical Probe
Predicate Device Classification:21 CFR 878.4400
Predicate Device Classification Number:Class II

b. Date Summary Prepared

July 22, 2005

c. Description of Device

The Estech Cobra Surgical System is a family of devices including RF Ablation Probes, with either a malleable or flexible shafts, used in conjunction with the Cobra Electrosurgical Unit (ESU) and other accessories, including the Estech Cobra Bipolar Recording Electrode. The System is intended for use by surgeons for the coagulation of cardiac and soft tissues during open surgical procedures.

d. Intended Use

The intended use for the Estech Cobra Surgical System is as follows: The Estech Cobra Surgical System is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during

{2}------------------------------------------------

general surgery to coagulate soft tissues. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

e. Comparison to Predicate Device

e. Comparison to Predicate Device
The XL version of the Estech Cobra Surgical System is substantially equivalent in intended use, technology, design and materials to the predicate device. The XL version intended use, technology, design and materials works. The original promarket notification
has up to seventeen electrodes on the ablation world it's in size and handle has up to seventeen electrodes on the abliated process differ in size and handle
device had an upper limit of seven electrodes. The Stabilizers add hament of the device had an upper mint of seven sieet. See of cilitate the placement of the System.

The Estech Cobra Surgical Probe is substantially equivalent in intended use, technology, I he Estech Cobra Surgical I rove is substantially.
design and materials to the predicate system. The Cobra Surgical Probed it does not design and materials to the predicate systems be run through the shaft and it does not the capability for a coomig steam of water to be country works.
interface with a Stabilizer. This version of the Cobra Surgical Probe is identical to the Boston Scientific Electrosurgical Probe.

The Estech Cobra Bipolar Recording Electrode is substantially equivalent to the surgical The Estech Cobra Bipolar Recording Blocared is the predicate device. Both are handintended use, sellsing and pacing characterial in an open surgical field. They are not implantable.

f. Summary of Supporting Data

f. Summary of Supporting Data
No biocompatibility data is necessary since the materials in the modified ablation probes No biocompatibility data is necessary only and described in the predicate 510(k)s. The (XL and Surgical I root) are receited to those in the ablation probes.

Preclinical performance data is supplied to demonstrate that the XL version of the cooled Precimical performance data is supplied to usine pertinent specifications as the predicate version.

No performance data is necessary for the Estech Cobra Surgical Probe since it is identical No performance catand manufacture to the predicate device.

Preclinical performance data is supplied to demonstrate that the Estech Cobra Bipolar Recording Electrode operates as intended.

Additionally, Estech certifies that the accessories are in compliance with the pertinent Additionally, Estech certifics that the accessive are arrest HF-18: Electrosurgical devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, which is a traditional symbol of medicine, but it is rendered in a more abstract and modern design.

FEB 2 1 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Estech, Inc. c/o Mr. Craig Coombs Coombs Medical Device Consulting 1193 Sherman Street Alameda, CA 94501

Re: K051749

Trade/Device Name: Estech Cobra Surgical System™ Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II (two) Product Code: OCL, GEL Dated: June 27, 2005 Received: July 7, 2005

Dear Mr. Coombs:

This letter corrects our substantially equivalent letter of September 13, 2005.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Craig Coombs

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

ellmoll

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

051749

Estech Cobra Surgical System™ Device Name:

Indications For Use:

The Estech Cobra Surgical System is intended for the coagulation The Estech Oobra Surgiour System Systems (RF) energy during cardiac of cardiac tissue doing factor of the system can be used during general surgery to surgery. The System can be used daning goneral bangulate coagulate oft tissue to produce hemostasis.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(Please Do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Barbare concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)Page 1 of ______
Division of General, Restorative,and Neurological Devices
CONFIDENTIALPage 1-5
510(k) NumberK051749

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.