K Number
K051035
Manufacturer
Date Cleared
2005-09-02

(133 days)

Product Code
Regulation Number
886.5916
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SynergEyes® A and M (paflufocon D hem-iberfilcon A) Hybrid Contact Lenses for daily wear are indicated for use in the concession of inyporopic myspice eyes. The lenses are indicated for daily presuyopla, in aphakic and not aprilike, nor alocused byest with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00D. The lenses may be disinfected using only a presoyopia, add powers between · 1.00 and · 1.00 and · 1.00 lense · Provinsible lenses.

Device Description

The SynergEyes™ A and M (paflufocon D hem-iberfilcon A) Hybrid Daily Wear Contact Lens is a The Syreigeyes ... A and M (palificon B non ibernison) , hydrion surrounded by a soft hydrophilic skirt that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere, ●
  • with center thickness from 0.12mm to 0.30mm .
  • with base curves of 7.10mm to 9.00mm .
  • with diameter of 14.50mm .
  • .
    This lens material for the rigid portion is paflufocon D lathe cut, surrounded by soft hydrophilic This lefts material for the ngla portion is panaloom a largerilization. When placed on the copolymer (nemi-benicon A), stemized by means of orefracting medium to focus light numan contea, the Gynergeyos - Trybha black as a lathe cut contact lens in the following designs: rays offe the relific. "The device is than be on and as a lable visibility tinted material. This device is spielical, tone, mamocal, and asphone ouride in the soft lens skirt manufactured by Ciba Vision Corporation.
    The SynergEyes™ A and M Hybrid Daily Wear Contact Lens is a rigid gas permeable material of The soft skirt is comprised of HEMA (paflulofon-D) rigid gas permeable polymer. (hydroxyethylmethacrylate) of 27% water and 73% polymer.
    The junction between the rigid material and soft material is bound by a proprietary chemical bonding method.
AI/ML Overview

The provided text describes the 510(k) summary for the SynergEyes™ A and M Hybrid Daily Wear Contact Lens. It outlines the device description, intended use, predicate devices, characteristics, and non-clinical and clinical studies conducted to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" for the clinical study in a quantifiable manner (e.g., "comfort must be >X%"). Instead, it describes the outcomes and compares them to predicate devices, focusing on demonstrating safety and efficacy.

CategoryAcceptance Criteria (Implied / Comparison)Reported Device Performance (SynergEyes™ Hybrid Contact Lens)
SafetyNo significant adverse events, minimal complaints, stable keratometric readings. Comparative to predicate devices' historical experience.- No adverse events reported. - Complaints were minimal, with comfort being the most prevalent. - One complication, "Poor outcome with lenses," reported; subject treated and discontinued. - No other significant safety findings, though the comfort complaint was "lost during the course of the study" (possibly meaning resolved or no longer primary).
EfficacyNo negative changes in refractive correction or visual acuity. Comparative to predicate devices' historical experience.- No negative changes in efficacy upon evaluation of the data (no details on specific metrics provided beyond "changes in refractive correction, lens visual acuity").
Material PropertiesSubstantial equivalence to predicate materials (Paragon HDS 100, Synergicon A).- Material property data for SynergEyes™ and Paragon HDS 100 were "substantially equivalent." - Listed Dk permeability for RGP Center (FATT: 145, ISO: 100) and Soft Skirt (9.1) compared favorably to SoftPerm® (FATT: 14, Soft Skirt: 5.5). - Dk/L transmissibility (FATT: 66-77, ISO: 46-53) compared to SoftPerm® (ISO: 17.5). - Light transmittance (>90%) compared to SoftPerm® (88-92%).
ToxicologyMeet ISO 10993 standards (no cytotoxicity, no systemic toxicity, no ocular irritation).- Cytotoxicity Test (ISO 10993-5): Test articles were not cytotoxic. - Acute Systemic Injection Test in mouse (ISO 10993-11): Each test article met the test requirements (no systemic toxicity). - Ocular Eye Irritation Test in rabbit (ISO 10993-10): No evidence of ocular irritation; extracts are not considered irritants to ocular tissue.
SterilizationMaintain sterility and stability.- Sterilized using e-Beam sterilization. - Established 1-year sterility and stability with an approved protocol.

2. Sample Size for the Test Set and Data Provenance:

  • Sample Size (Clinical Study Subjects): 107 subjects (214 eyes)
  • Data Provenance: The document does not explicitly state the country of origin. Given the context of a 510(k) summary for the FDA in the US, it is highly probable the study was conducted in the United States, or at least under US regulatory standards. The study was prospective, as indicated by "A Clinical Study was performed to establish the safety and efficacy..." and "The clinical investigation...was conducted on 107 subjects for a period of 3 months."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • This information is not provided. The clinical study assessed safety and efficacy in human subjects, relying on clinical observations and measurements rather than expert consensus on a test set of pre-existing data. For regulatory applications like this, the "ground truth" for clinical performance is derived from the direct observation of patient outcomes and measurements during the study.

4. Adjudication Method for the Test Set:

  • This information is not explicitly stated. For clinical studies, "adjudication" typically refers to independent review of adverse events or specific endpoints by experts to ensure consistency and objectivity. While adverse events were reported and assessed, the specific adjudication method is not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is specifically designed to evaluate the impact of an AI system on human reader performance, usually in image interpretation tasks. This document pertains to a physical medical device (contact lens) and its clinical performance, not an AI software.

6. Standalone (Algorithm Only) Performance Study:

  • No, a standalone performance study was not done. This concept is relevant for AI algorithms that operate without human intervention. The device in question is a contact lens, which is physically worn and interacts with the human eye.

7. Type of Ground Truth Used:

  • The ground truth for the clinical study was derived from direct clinical observation, measurements from human subjects, and patient self-reported symptoms/satisfaction (e.g., adverse events, slit lamp findings, refractive correction changes, visual acuity, comfort complaints). For material properties, the ground truth was laboratory testing results.

8. Sample Size for the Training Set:

  • This is not applicable as the document describes a physical medical device. There is no "training set" in the context of an AI algorithm learning from data. The clinical study results are used to demonstrate the device's performance, not to train an algorithm.

9. How Ground Truth for the Training Set Was Established:

  • Not applicable (see point 8).

{0}------------------------------------------------

K 05/035 510 (K) SUMMARY

Applicant's Name and Address

SynergEyes, Inc.1926 Kellogg Ave.Carlsbad, CA 92008(760) 476 9410
Contact PersonRichard Lippman, OD FAAOVice President Ophthalmic Regulatory ProductsR.P. Chiacchierini & Associates, LLC Suite 3015825 Shady Grove RoadRockville, Maryland 20850(240) 683-3738
1. Identification of device

Contact Lens Common Name: SynergEyes™(paflufocon D hem-iberfilcon A) Hybrid Trade Name: Daily Wear Lens Daily Wear Soft (hydrophobic) Contact Lens Classification: Class II (21 CFR 886.5916) Device classification:

2. Description of device

The SynergEyes™ A and M (paflufocon D hem-iberfilcon A) Hybrid Daily Wear Contact Lens is a The Syreigeyes ... A and M (palificon B non ibernison) , hydrion surrounded by a soft hydrophilic skirt that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere, ●
  • with center thickness from 0.12mm to 0.30mm .
  • with base curves of 7.10mm to 9.00mm .
  • with diameter of 14.50mm .
  • .

This lens material for the rigid portion is paflufocon D lathe cut, surrounded by soft hydrophilic This lefts material for the ngla portion is panaloom a largerilization. When placed on the copolymer (nemi-benicon A), stemized by means of orefracting medium to focus light numan contea, the Gynergeyos - Trybha black as a lathe cut contact lens in the following designs: rays offe the relific. "The device is than be on and as a lable visibility tinted material. This device is spielical, tone, mamocal, and asphone ouride in the soft lens skirt manufactured by Ciba Vision Corporation.

The SynergEyes™ A and M Hybrid Daily Wear Contact Lens is a rigid gas permeable material of The soft skirt is comprised of HEMA (paflulofon-D) rigid gas permeable polymer. (hydroxyethylmethacrylate) of 27% water and 73% polymer.

The junction between the rigid material and soft material is bound by a proprietary chemical bonding method.

3. Intended use

SynergEyes® A and M (paflufocon D hem-iberfilcon A) Hybrid Contact Lenses for daily wear are Sylelgcyes® A and M (pundloodn's honeropic, myopic and astigmatic refractive error including indicated for use in the concession of inyporopic myspice eyes. The lenses are indicated for daily presuyopla, in aphakic and not aprilike, nor alocused byest with astigmatism up to 6.00 D. For

{1}------------------------------------------------

presbyopia, add powers between +1.00 and +4.00D. The lenses may be disinfected using only a presoyopia, add powers between · 1.00 and · 1.00 and · 1.00 lense · Provinsible lenses.

4. Predicate devices

  1. Predicate devices
    The predicate lenses were selected to address: material (silicone acrylate and many of the lange - Additionally for The predicate lenses were selected to address. matchan (the lens. Additionally for optic portion of the lens and (syllergion) A) for the Soll only Portand Or and Group I low water non-innic
    lens design (hybrid RGP/soft), lens group (silicone acrylate, data Tens design (nybrid RGP/Soft), lefls group (silleone devices for indication for use and parameter comparison.

Predicate Devices:

  • For the RGP Center Portion: .
Paragon Vision SciencesP870024/S39 and S43K940277K002845
Polymer Technology (B&L) Boston XO:K000795
  • For the Hybrid (RGP-soft skirt concept) .
    P840004 Saturn II (synergicon A) Mixed use Lens

5. Characteristics

  1. Characteristics
    The physical and dimensional characteristics of the SynergEyes™ Hybrid Contact Lens are The physical and differistial characteristics of the Synorgeyos - Trycher Contact Lens in the following table.
SynergEyes™(paflufocon D-iberfilcon A) HybridContact LensSoftPerm®(synergicon A)Contact Lens(Predicate Lens)
Lens Characteristics
ManufacturerSynergEyes™ Inc.Ciba Vision Care
Base Curves7.10-9.00mm7.10-8.10mm
RGP Center8.40mm8.00mm
Posterior Optic ZoneDiameter7.80mm7.00mm
Lens DesignsSphere, Aspheric,Front Surface Toric,MultifocalSphere, FrontSurface Toric
Diameters:14.5mm14.3mm
Power Range-20.00 to + 20.00D-13.00 to +6.00D
Center Thickness0.12 to 0.30mm0.08mm to 0.28mm
Refractive Index (RGP)1.4421.53
Wetting Angle (RGP)42°21° CLMA
Specific Gravity (RGP)1.101.015
Hardness79
Indications for UseDaily WearDaily Wear
UV BlockingNoNo
MaterialPaflufocon D centerSynergicon A

{2}------------------------------------------------

RGP hem-iberfilcon A(HEMA, MEMA)RGP and HEMA Skirt
TintVisibility BlueClear
Soft Skirt Water Content27%25%
Core (RGP) Water Content< 1%< 0.2%

6. Non clinical studies

Non-clinical studies are summarized below:

Chemistry and leachability

  • Material property data were generated on the SynergEyes™ and the Paragon HDS 100 . materials. The material properties were substantially equivalent.
  • The lens care product manufacturers have been previously shown compatible for RGP and . soft (Hydrophilic) lenses with their products.
  • Son (Tyarophilib) for the SynergEyes™ Hybrid Contact Lens is based upon stability . protocols included with this notification.
  • protocolories were conducted to determine the residual monomers on the subject device. .

Toxicology, lenses materials

I Uncology, lefises materials
In accordance with the May 1994 Guidance Document for Daily Wear contact lenses, toxicology in accordance with the May 100 - Caractives M Hybrid Contact Lens. The results are summarized below:

Cvtotoxicity Test: .

Cytotoxicity Tests have been conducted on the subject device according to ISO 10993-5. Gylotoxicily Tests Trave Devices- Tests for Cytotoxicity: In vitro Methods guidelines, was conducted on the test articles, to determine the potential for Cytotoxicity.

conduction the tool and the positive controls performed as anticipated. Under the conditions of the study, the test articles were not cytotoxic.

Acute Systemic Injection Test in the mouse: .

An evaluation of the test articles for systemic toxicity in mice after a single intravenous An Evaluation or a single intraperitoneal administration has been conducted according to the ISO 10993-11: Biological Evaluation of Medical Devices. Tests for Systemic Toxicity.

10355-11. Diological Evaluation of Moulous or not the test article extracts. Each test article met the test requirements.

Ocular Eye Irritation Test in the rabbit: .

An evaluation of the ocular irritation of 0.9% NaCl of the subject article after a single instillation in An evaluation of the outling to 1010 has in 10993-10: Biological Evaluation of Medical Devices. Tests for the Irritation and Sensitization.

Devidence of ocular irritation was observed in the rabbits. The test article extracts are not considered irritants to the ocular tissue of rabbits.

Validity of Scientific Data:

A contract laboratory using Good Laboratory Practices conducted the Toxicology studies. Chemistry leachables studies were conducted in-house. Certification as to implementation and conduct under Good Laboratory Practices may be found with each report.

Solution Compatibility

Lens materials used in this device are already recognized materials previously approved for use with all approved lens care products. No further testing was conducted.

Microbiology and sterilization

{3}------------------------------------------------

The SynergEyes™ Hybrid Contact Lenses are sterilized using e-Beam sterilization. Additional autonoions i he SynergEyes ™ Tryond Outlact Echsed as 1 year as of the current date. Additional extensions stenlity and stablity has boon ectablished with an approved protocol.

7. Packaging

  1. Packaging
    The primary lens container for shipping is a sterile enclosed medical grade glass vial capped with The primary fens containor for enipping of in a sterile buffered normal saline.

8. Clinical data

  • A Clinical Study was performed to establish the safety and efficacy of the SyncrgEyes of the . A Clinical Study Was performed to occubition the clinical investigation of the SynergEyes™Hybrid Contact Lens was conducted on 107 subjects for a period of 3 months. Synergeyes "Hybru Contact Echs was consent document regarding participation in the All subjects read and signed an monthed outhorn document of and international ethical Investigation Which was conducted in ochilemanse to completed the protocol, 46 (43%) Standards und Togalations - Com for discontinuation were as follows:
Over All Visits
Reason For Discontinuation# Subjects%
Poor Comfort1532.6%
Loss of Interest1328.3%
Poor Vision613.0%
Difficulty Handling12.2%
Other36.5%
Medical Condition - not study related24.3%
Non-Compliance48.7%
Poor Outcome with Lenses24.3%
Total Eyes46

The "other" reasons included travel conflicts (1), use of reading glasses inconvenient (1), and subject not happy with lenses (1).

  • There was one complication reported as "Poor outcome with lenses" during the study, a . There was one complication Topened as The subject was treated and was discontinued.
  • Safety was assessed by adverse events, positive slit lamp findings, symptoms problems and . Safely was assessed by adverse events, positive on laims him adverse events reported.
    complaints and changes in keratometric readings. There were no adverse event provided complaints and complaints were minimal with comfort being the most prevalent Symploms problems and complaints were minimal with time. There were no other significant safety finding although the complaint alminioned mas lost during the course of the study.
  • Efficacy was evaluated by changes in refractive correction, lens visual actury, average . Elicacy was Evanuatou by changes were noted in efficacy upon evaluation of the data.

In conclusion, the SynergEyes™ Hybrid Contact Lens material performance is equivalent in PCP In Collidation, the Oyfergeyes - Trybrid - Ochace for this class of materials, as well as for RGP daily and soft daily wear lenses in general.

  1. Conclusions drawn from studies Substantial Equivalence:

{4}------------------------------------------------

Information provided in this 510(k) establishes that the SynergEyes™Hybrid Contact Lens are reise Information provided in this 210(k) establishes that the others and do not raises and do not raise equivalent in optical, chemical and prysical propertion de monstrated safe and any questions of salery and where possible equivalence to historical experience with effective lens peromance, and where possible oquivalent to the predicate devices material, predicate devices. The device is substantally Cultivation (synergicon A) lense Paragon HDS under For 0024/000 and CTO, and To 10-11-14 and Comprised of a rigid center optic portion and a soft skirt portion.

Clinical Equivalence to Predicate Devices Historical Experience

SynergEyes™Hybrid ContactLensJurkus(1998)Chung(2001)
No. of Lenses (Eyes)2142135
No. Patients1071128
Gender72F/32M5F/6M11F/17M
Mean Age43F/39.9M41 ± 19
LensSynergEyes™HybridSaturn IISoftPerm-PriorRGP users
IndicationMyopia,Hyperopia,AstigmatismAstigmatismKeratoconus 22/35(62.9%)PK 10/35(28.6%)
Completions61 (57%)4 (36%)17 (66.7%)
Discontinuations46 (43%)7 (64%)11 (33.3%)
Comfort32%28%45.5%
Vision13%28%9.1%

Material Equivalence Table

Material Comparison
SynergEyes™ Hybrid Contact LensSoftPerm® Contact Lens
PRODUCTION METHODLathingLathing
INTENDED USEDaily WearDaily Wear
MATERIALPaflufocon D Centerhem-iberfilcon A skirtSynergicon-A
TypeGroup 1 Low WaterGroup 1 Low Water
Surface ChargeNon-ionicNon-ionic
Color additive (Scientific name)D&C Green 6
UV additiveNoNo
Dk permeability:RGP Center:RGP Center:
1. Revised FATT Polarimetricmethod with edge correction@ 35°C x $10^{11}$ (cm²/sec) (mlO2/ml x mm Hg)FATT: 145ISO: 100FATT: 14
2. ISO 9913-1Soft Skirt: 9.1Soft Skirt: 5.5

{5}------------------------------------------------

Dk/L- Lens transmissibility:FATT:66-77
1. Revised FATT(Through power range -20 to +20D)ISO:46-53ISO:17.5
2. ISO 9913-1 Polarimetricmethod :
Light transmittance:(380nm to 780nm)>90%88-92%

Risk and Benefits:

The risks of the subject device are the same as those normally attributed to the wearing of RGP The nate of the oublice areas on a daily wear base. The benefits to the patient are the and soft (hydrophillo) vontact lenses on drophilic) contact lenses. Overall, the risks and benefits Same as those for other NOT and out (mychophilis) contact lenses and raise no additional concerns for safety or effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The caduceus is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 _ 2005

SynergEves, Inc. C/O Richard E. Lippman, O.D., F.A.A.O. Vice President for Ophthalmic Product Regulatory Affairs P. Chiacchierini & Associates, LLC 15825 Shady Grove Rd., Suite 30 Rockville, MD 20850

K051035 Re:

Trade/Device Name: SynergEyes™ (paflufocon D hem-iberfilcon A) Hybrid Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HQD Dated: September 1, 2005 Received: September 1, 2005

Dear Dr. Lippman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, atores, while Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device to such additional controls. Existing major regulations affecting your device can may or subject to December of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean i lease oc acrised that I Drivision that your device complies with other requirements of the Act that I Dr has intatutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 - Richard E. Lippman, O.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter with and wy of substantial equivalence of your device of your device to a legally promated predicate device results in a classification for your dovice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please it you dobite of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David M. Whipple

David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure 1

{8}------------------------------------------------

SYNERGEYES, INC.SECTION 2
510(k) Premarket NotificationINDICATION FOR USE STATEMENT
SYNERGEYES (paflufocon D hem-iberfilcon A) HYBRID CONTACT LENSDAILY WEAR CONTACT LENS

SECTION 2 : INDICATIONS FOR USE STATEMENT

510(k) Number (if known) _K651035 (5001

Device Name: SynergEyes@ A and M (paflufocon D hem-iberfilcon A) Hybrid Contact Lens

Indication for Use

SynergEyes® A and M (paflufocon D hem-iberfilcon A) Hybrid Contact Lenses for daily wear are SynelgEyes® A and M (pandroon D fren ibenneon A), and astigmatic refractive erroringling and Indicated tof the concection of hyperoped my of the enses are indicated for daily presuyopla, in aphakis and not apnake, non dious of of of ever with astigmatism up to 6.00 D. For weal for the correction of up to 120.00 and +4.00D. The lenses may be disinfected using only a
presbyopia, add powers between +1.00 and +4.00D. The lenses nay be disinfected presoyopla, add powers between + 1.00 and + 1.000 - we + and rigid gas permeable lenses.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------
Prescription Use
--------------------
OR Over-the -counter-use
----------------------------
Karer Wabanten (Division Sign-off) Division of Ophthalmic Devices
--------------------------------------------------------------------------------------------------------------
510(k) NumberK051035/5001
-----------------------------

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.