(27 days)
The IRMA TRUpoint™ Creatinine Control Kit is for use on the IRMA TRUpoint™ Blood Analysis System to perform Quality Control assays for Creatinine on the IRMA TRUpoint™ Blood Analysis System. For in vitro Diagnostic Use Only
The Creatinine Control Kit includes creatinine control materials at two levels for monitoring the IRMA analyzer performance at different points in the range of intended clinical utility. The control materials are packaged in capped luer lock syringes, each containing 1.5 ml of solution. The syringes are in turn packaged in pouches. Three (3) pouches/syringes for low level and three (3) pouches/syringes for high level are packaged in a box. The Cr control is an aqueous based solution of creatinine and sucrose. This control contains no human or biological materials.
The provided document is a 510(k) summary for the IRMA TRUpoint™ Creatinine Control Kit. This submission is for a quality control material intended to monitor the performance of an analyzer, not for a diagnostic device that measures patient analytes directly. Therefore, the concepts of sensitivity, specificity, or diagnostic accuracy as typically applied to medical devices for patient diagnosis or screening do not apply in the same way.
The "acceptance criteria" for a quality control material generally relate to its stability, consistency, and ability to generate expected values when tested on the intended analyzer. The study presented here is a demonstration of substantial equivalence to existing legally marketed predicate devices, rather than a performance study with traditional clinical acceptance criteria.
Here's an interpretation based on the provided text, aligning with the spirit of the request by describing the implicit "acceptance criteria" for a QC device and how the submission addresses them:
Acceptance Criteria and Device Performance for a Quality Control Material (Derived from Substantial Equivalence)
For a Quality Control material like the IRMA TRUpoint™ Creatinine Control Kit, the "acceptance criteria" are implicitly met if the device demonstrates substantial equivalence to predicate devices. This means it performs its intended function (monitoring analyzer performance) in a comparable manner, considering its form, matrix, intended use, and storage conditions.
Acceptance Criteria (Implied for QC Material Substantial Equivalence) | Reported Device Performance |
---|---|
Similar Intended Use: | The IRMA TRUpoint™ Creatinine Control Kit is "intended to be used to perform Quality Control assays for Creatinine on the IRMA TRUpoint™ Blood Analysis System." This intended use is comparable to predicate devices which also "perform quality control assays" for analytes. |
Similar Form and Matrix: | The device is "comprised of materials in the same form and aqueous matrix" as the Bionostics Inc. RNA 823 Controls (K943754). It is also in "the same form" as Medical Analysis Systems Moni-Trol H (K030942), though the latter is human serum based, while the IRMA TRUpoint™ Creatinine Control is aqueous. |
Similar Analyte Control: | The IRMA TRUpoint™ Creatinine Control contains only creatinine. One predicate, Medical Analysis Systems Moni-Trol H, is explicitly stated to be "for the control of a similar analyte, creatinine." While the other predicate (Bionostics RNA QC 823) is a multi-analyte control without creatinine, the overall substantial equivalence claim encompasses the specific analyte being controlled. |
Comparable Storage Conditions: | The IRMA TRUpoint™ Creatinine Control and the Bionostics Inc. RNA 823 Controls are both stored at 2°C - 8°C. The Medical Analysis Systems Moni-Trol H has different storage conditions (-20°C). However, the overall substantial equivalence argument considers the composite similarities. |
Safety and Effectiveness (implicitly similar): | By demonstrating substantial equivalence to legally marketed devices, it is inferred that the new device shares similar safety and effectiveness profiles for its intended function as a quality control material. Unlike patient-contact devices, the primary "safety" here relates to stable, reliable control values and lack of biohazards (explicitly stated: "This control contains no human or biological materials"). "Effectiveness" relates to its ability to perform the QC function. |
Study Details (Based on the 510(k) Summary)
The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicates rather than presenting de novo clinical study data with detailed performance metrics. Therefore, many of the requested details about specific study design (e.g., sample sizes, ground truth establishment, expert adjudication) are not applicable or not explicitly provided for this type of submission.
Here's an attempt to address the points based solely on the provided text, indicating when information is not available:
-
A table of acceptance criteria and the reported device performance:
- Addressed in the table above. The "acceptance criteria" are implied by the requirements for substantial equivalence. The "reported device performance" is the claim of substantial equivalence based on the comparisons made.
-
Sample size used for the test set and the data provenance:
- Not explicitly provided. This submission describes the device and compares its technical characteristics to predicates to establish substantial equivalence. It does not detail specific performance testing data (e.g., precision, accuracy, stability studies) that would typically involve a "test set" in the context of diagnostic accuracy. While such studies would have been performed by the manufacturer, the 510(k) summary focuses on the comparative argument.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. For a quality control material, "ground truth" isn't established by experts in the same way it would be for a diagnostic image or pathology slide. The "truth" for a QC material is its known concentration of the analyte, which is verified through analytical methods. The submission doesn't detail the analytical validation process for the control itself, only its comparison to predicates.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. Adjudication is relevant for subjective assessments (e.g., human reader interpretations). This is a technical comparison for a quality control material, not a diagnostic interpretation study.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This submission is for a quality control material, not an AI or imaging diagnostic device that would involve human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a quality control material, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not explicitly stated/Not applicable in the human-centric sense. For a quality control material, the "ground truth" would be the assigned value or target value for creatinine, determined through highly accurate reference methods or inter-laboratory consensus, which the control material is manufactured to. The document does not detail how these target values were established for the IRMA TRUpoint™ Creatinine Control Kit, but focuses on its equivalence in form and function to other marketed QC materials.
-
The sample size for the training set:
- Not applicable. This is a physical quality control material, not a machine learning model, so there is no "training set."
-
How the ground truth for the training set was established:
- Not applicable. See point 8.
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.