(35 days)
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies of clinically significant aerobic gram positive bacteria. VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole is designed for Trimethoprim/sulfamethoxazole susceptibility testing of Staphylococcus aureus, Enterococcus species, and Streptococcus species including Streptococcus agalactiae, and S. pneumoniae. VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole is a qualitative test. It is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole is designed for antimicrobial susceptbility since with the VITEK® 2 and VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to r he unlinerebial procentrations in mcg/mi. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study details for the VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole device:
Acceptance Criteria and Device Performance for VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Overall Category Agreement with NCCLS reference macrobroth dilution method | 99.6% |
| Reproducibility | Acceptable results |
| Quality Control | Acceptable results |
Note: The document implicitly defines acceptable results for reproducibility and quality control by stating they were "acceptable." Specific quantitative thresholds for these metrics are not provided in the excerpt.
2. Sample Size Used for the Test Set and Data Provenance
The document states that the evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. However, it does not provide a specific sample size (number of isolates/strains) used for the external evaluations (test set).
The data provenance is retrospective and prospective, as it includes both clinical isolates (which can be retrospective or freshly collected) and stock challenge strains. The country of origin of the data is not explicitly stated, but the submitter (bioMérieux, Inc.) is based in the USA, and the FDA is the regulatory body, suggesting the data is likely from the USA or was submitted for approval in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not explicitly state the number of experts used or their qualifications. The "NCCLS reference macrobroth dilution method" is the ground truth, which is a standardized laboratory procedure, not typically established by individual experts but rather by a consensus of scientific bodies (e.g., Clinical and Laboratory Standards Institute, formerly NCCLS). The performance of this reference method would be carried out by trained laboratory personnel, but they are not categorized as "experts" in the context of establishing ground truth in the same way, for example, a radiologist interprets an image.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set. The ground truth is established by the "NCCLS reference macrobroth dilution method," which is a direct comparative method rather than an interpretive classification requiring adjudication of multiple readings.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done as described in the provided text. This study is focused on the device's performance against a reference method, not on how human readers improve with or without AI assistance.
6. Standalone (Algorithm Only) Performance Study
Yes, a standalone performance study was done. The entire evaluation described is a standalone study, comparing the device (VITEK® 2 system) directly to the NCCLS reference macrobroth dilution method. The VITEK® 2 system's performance is reported (99.6% overall Category Agreement) without human interpretation in the results. The system automatically processes the cards and generates a report.
7. Type of Ground Truth Used
The type of ground truth used is the NCCLS reference macrobroth dilution method. This is a laboratory-based, gold-standard method for determining antimicrobial susceptibility, providing quantitative Minimum Inhibitory Concentration (MIC) values and interpretive categories.
8. Sample Size for the Training Set
The document does not specify a sample size for the training set. The provided text focuses on the performance evaluation (validation) of the device against a reference method. It's common for such submissions to describe the validation data rather than the internal development/training data, especially for hardware-based systems like the VITEK 2.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. This detail would typically be part of the device's internal development process and is not included in this 510(k) summary, which focuses on the validation of the final product. Assuming a similar development process, it is likely that reference methods similar to the NCCLS macrobroth dilution method were used during training and development phases.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for BioMerieux. The logo consists of the company name in a stylized font, with a circular graphic above it. A thin line extends vertically through the center of the circle and the text.
MAY 2 3 2005
VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole
510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|---|---|
| Address: | 595 Anglum RoadHazelwood, MO 63042 |
| Contact Person: | Nancy WeaverStaff Regulatory Affairs Specialist |
| Phone Number. | 314-731-8695 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | April 12, 2005 |
| Device Name: | |
| Formal/Trade Name: | VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole |
| Classification Name: | Fully Automated Short-Term Incubation CycleAntimicrobial Susceptibility Device, 21 CFR 866.1645 |
| Common Name: | VITEK 2 AST-GP SXT |
| Prodicate Device: | VITEK 2 Gram Positive AST for Sparfloxacin |
D. 510(k) Summary:
C. Predicate Device:
B.
VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole is designed for antimicrobial susceptbility VITER 2 Oran't oolare himourophin since with the VITEK® 2 and VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to r he unlinerebial procentrations in mcg/mi. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
(N50510/S141)
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK 2 Gram Positive Trimethoprim/sulfamethoxazole demonstrated substantially equivalent performance when compared with the NCCLS reference macrobroth dilution method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003.
bioMérieux, Inc.
800/638-4835 Fax: 314/731-Phone: 314/731-8500 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA 8700
{1}------------------------------------------------
The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Positive The Premiation (o ro[it]) procente levaluation was conducted with fresh and stock Trined isolates and stock challenge strains. The external evaluations were designed to confirm the cillitical solates and stock onailer Trimethoprim/ suffamethoxazole by comparing its acceptablily of VTLR 2 Oran Positive minorition reference method. The data is representative of performance with the NOOLS macrosition relevel instrument platforms, as evidenced in the periormance on bout the VTTEK 2 and VTER 2 Compact file, K050002 . VITEK 2 Gram Positive AST equivalency study presented in the VITER a Company of 99.6% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY 2 3 2005
Ms. Nancy Weaver Staff Regulatory Affairs Specialist BioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320
K050976 Re:
K02070
Trade/Device Name: VITEK® 2 Gram Positive Trimethoprim/Sulfamethoxazole (< 0.5/9.5 - ≥ 16/304 µg/mlT Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 12, 2005 Received: April 18, 2005
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your bector of the device is substantially equivalent (for the indications felerenced above and nave determined a vices and and one ices marketed in interstate for use stated in the encreate) to regarry the enactment date of the Medical Device Amendments, or to contineree prior to May 20, 1978, the encordance with the provisions of the Federal Food, DINE, devices that have been recuire approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval application (PM and Cosmette Act (110) that do noves, subject to the general controls provisions of the Act. The 1 ou may, therefore, market the device, equirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is Classified (300 a00 v) incl. Existing major regulations affecting your device it may be subject to such additional democrations (CFR), Parts 800 to 895. In addition, FDA can be found in The 21, Occareents concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I Dri s issuated of a bustical with other requirements of the Act that FDA has made a decemination administered by other Federal agencies. You must of any Federal statutes and regulations and limited to: registration and listing (21 comply with an the Act 3 requirements, into and 809); and good manufacturing practice CFR Part 807), labeling (21 CF RT 11 in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter whi anow you to ocgin manceang your and equivalence of your device to a legally prematication. "The I DTV intentig gour device and thus, permits your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, of k If you desire specific information as advertising of your device, please contact the Office of In of questions on the promotion and Safety at (240)276-0484. Also, please note the Viro Diagnostic Device Device Liver and to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the 1 ou may of and Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sales, a Form
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): $05097 (e
Device Name: VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole (< 0.5/9.5 – ≥ 16/304 µg/ml)
Indications For Use:
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the The VTER 2 Antimioronal Systems for the automated quantitative or qualitative or qualitative or qualitative or qualitative or qualitative or qualitative or chiliante or VITER 2 and VTTEN 2 Oompact Uyically significally significant aerobic gramsusceptibility testing of isolatou contrile in the more of streptococcus agalactiae, and S. pneumoniae.
S. priedmoniae.
VITEK® 2 Gram Positive Trimethoprim/sulfamethoxazole is designed for Trimethoprim VITEK 2 Gram Postive Thinethophiniounciller. VITEK 2 Gram Positive Trimethopim/
susceptibility testing of Staphylococcus aureus. VITEK 2 Gram Positive NITEK 2 only!TE Susceptiblity testing of Oraphyloooous and out of the VITEK 2 and VITEK 2 and VITEK
sulfamethoxazole is a qualitative test. It is intended for use with the virus and VITEK Sullamethoxazole is a qualitative tocain in the determination of in vitro susceptibility to antimicrobial agents.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Luddi L. Pool
Division Sign-Off
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
p. 10
K650976 510(k)________________________________________________________________________________________________________________________________________________________________________
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”