Dick's Formalwear® Male Latex Condoms is to be used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases, including human immunodeficiency virus (HIV)).
These condoms are made of a high quality natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These three types of condoms (plain, ribbed, and dotted), have a nominal length of 180 mm, nominal width of 54 mm, and a nominal thickness of 0.05 mm. They are all lubricated with a silicone lubricant and have a reservoir tip for extra safety. The condoms are also printed on the exterior of the latex with a graphic design per section 11.1 of ISO 4074:2002 and have passed toxicity and biocompatibility testing.
The provided text describes a 510(k) summary for Dick's Formalwear® Male Latex Condoms, not a study involving AI or complex medical imaging. Therefore, most of the requested information (like sample size for test/training sets, expert qualifications, MRMC studies, standalone algorithm performance, etc.) is not applicable or available in the document.
However, based on the information provided, here's what can be extracted and formatted:
Acceptance Criteria and Device Performance for Dick's Formalwear® Male Latex Condom
The device performance is primarily evaluated against established industry standards and AQL (Acceptable Quality Limit) levels for manufacturing quality.
-
Table of Acceptance Criteria and Reported Device Performance:
Criteria Acceptance Standard Reported Device Performance Conformance to ASTM Latex Condom Standard ASTM D3492 Designed in conformance AQL for ASTM D3492 0.25 (specifically for the standard) Met (implied by conformance statement) Conformance to ISO (manufacturing) ISO 4074:2002 Manufactured in "recognition of" (implied conformance) Air Burst Requirements (part of GMP) Established and implemented, per ISO 4074:2002 Air burst testing yielded an AQL of 1.0 Electronic Testing Not explicitly stated (implied "pass") Each condom is electronically tested Graphic Design Per section 11.1 of ISO 4074:2002 Printed on the exterior of the latex with a graphic design Toxicity and Biocompatibility Passed Passed -
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated. The AQL (Acceptable Quality Limit) values (0.25 for ASTM D3492 and 1.0 for air burst testing) indicate statistical sampling was used during manufacturing quality control, but the specific sample sizes are not provided in this summary.
- Data Provenance: The testing is internal to the manufacturing process, likely conducted at the manufacturing facility. There is no mention of country of origin of data beyond the manufacturing location. The testing would be considered prospective as it's part of the ongoing quality control during production.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
- This question is not applicable. The "ground truth" for condom quality is established by objective physical and material standards (ASTM, ISO), along with internal quality control procedures and testing. There isn't a subjective "expert review" of individual condoms in the way there would be for image analysis.
-
Adjudication Method for the Test Set:
- This question is not applicable. Quality control for manufacturing physical devices like condoms typically involves objective measurements and adherence to acceptance ranges, not expert adjudication in the clinical sense.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- This question is not applicable. This device is a physical medical device (condom), not an AI-assisted diagnostic tool.
-
Standalone (Algorithm Only) Performance Study:
- This question is not applicable. This device has no embedded algorithm.
-
Type of Ground Truth Used:
- The ground truth is based on objective, established industry standards and specifications for latex condoms, including:
- ASTM D3492 (Latex Condom Standard)
- ISO 4074:2002 (Natural rubber latex condoms – Requirements and test methods)
- Internal Good Manufacturing Practice (GMP) air burst requirements
- Electronic testing for defects.
- The ground truth is based on objective, established industry standards and specifications for latex condoms, including:
-
Sample Size for the Training Set:
- This question is not applicable. There is no "training set" in the context of this device. The manufacturing process and quality control learn from adherence to established standards and continuous process improvement, not a data-driven training set for an algorithm.
-
How the Ground Truth for the Training Set Was Established:
- This question is not applicable for the reasons stated above.
{0}------------------------------------------------
K050913
| III. 510(k) SUMMARY | APR | 4 2006 | |
|---|---|---|---|
| Submitted by: | Dick's Formalwear, LLC3507 W. Bethany Home RoadPhoenix, AZ 85019480-659-1146 | ||
| Contact Person: | Curt Strenk | ||
| Date Prepared: | April 7, 2005 | ||
| Proprietary Name: | Dick's Formalwear® Male Latex Condom | ||
| Common Name: | Male Latex Condom | ||
| Classification Name: | Condom |
- These condoms are made of a high quality natural Description of the Device: rubber latex sheath, which completely covers the penis with a closely fitted membrane. These three types of condoms (plain, ribbed, and dotted), have a nominal length of 180 mm, nominal width of 54 mm, and a nominal thickness of 0.05 mm. They are all lubricated with a silicone lubricant and have a reservoir tip for extra safety. The condoms are also printed on the exterior of the latex with a graphic design per section 11.1 of ISO 4074:2002 and have passed toxicity and biocompatibility testing.
- Intended Uses of the Device: These condoms are to be used for contraception and for prophylactic purposes. they are also intended to help prevent pregnancy and the transmission of sexually transmitted diseases, including human immunodeficiency virus (HIV).
- Technological Characteristics: These condoms have the same technological characteristics as condoms that are currently distributed throughout the USA. They are designed in conformance with ASTM Latex Condom Standard D3492 with an AQL of 0.25. These condoms are made of high quality natural rubber latex. Each condom is electronically tested. They are manufactured, in recognition of ISO 4074:2002, under established and implemented air burst requirements as part of good manufacturing practice procedures. The results of air burst testing yielded an AQL of 1.0. These condoms conform to domestic and international standards.
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows a circular seal with the Department of Health & Human Services USA text around the border. Inside the circle is an abstract image of an eagle. The eagle is facing to the right.
4 2006 APR
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Curt Strenk Managing Member Dick's Formalwear, Inc. 3507 W. Bethany Home Road PHOENIX AZ 85019
Re: K050913
Trade Name: Dick's Formalwear® Male Latex Condoms Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: March 22, 2006 Received: March 23, 2006
Dear Mr. Strenk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K050913
VIII. INDICATIONS FOR USE STATEMENT
| Device Name : | Dick's Formalwear® Male Latex Condoms |
|---|---|
| Indications for Use: | Dick's Formalwear® Male Latex Condoms is to be used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases, including human immunodeficiency virus (HIV)). |
Prescription Use (21 CFR 801 Subpart D) OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Taint K. Sloggrm
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number Kc_569113/5002
§ 884.5300 Condom.
(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.