K Number
K050910
Date Cleared
2005-06-16

(66 days)

Product Code
Regulation Number
876.4300
Panel
GU
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Roei Working Element and Roei Cutting Loops are endoscopic electrosurgical accessories intended for use with monopolar resectoscope systems for the resection of soft tissue in transuretheral bladder tumor resection (TURBT), transurethral prostatic and bladder biopsy, and transurethral resection of the prostate (TURP) procedures.

Device Description

The Roei Working Element and Cutting Loops use the same fundamental operating characteristics as other resectoscope working elements and electrode cutting loops. The Roei Working Element includes a handle and rod to which the disposable Roei Cutting Loops are attached. One end of the handle has a connection port for connection of a resectoscope optics element. The opposite end of the Roei Working Element handle has a connection port that accommodates a resectoscope sheath used to encase a Roei Cutting Loop. Additionally, the Roei Working Element handle provides an electrical connection box with an inlet receptacle for connection of a resectoscope's electrosurgical cable that is then attached to a monopolar electrosurgical generator. Once the Roei Working Element and Cutting Loop are attached to a resectoscope system, the Roei Cutting Loop receives electrical current from the electrosurgical generator. The electrical current transmitted to the Cutting Loop is then used to resect the targeted soft tissue.

In addition to serving as the housing for the Roei Cutting Loops, optics and electrical connections, the Roei Working Element permits translation of the axial motion of the handgrip into a circular axis movement of the Cutting Loop independent of the resectoscope sheath. This circular axis movement provides the ability for the surgeon to move the Cutting Loop in a side-to-side clockwise or counterclockwise direction through the soft tissue, according to the operating handle position.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Roei Working Element and Roei Cutting Loops:

Acceptance Criteria and Device Performance Study

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Safe and effective resection of soft tissue (functional performance)Animal testing demonstrated that the device "safely and effectively resects soft tissue."
Compliance with IEC 6061-2-2, Third Ed. (1998-09): Medical Electrical Equipment - Part 2: Particular Requirements for the Safety of High Frequency Surgical Equipment (Electrical Safety)The device "will be in compliance with the following recognized consensus standards prior to marketing the device." This implies an intention to comply, rather than direct reporting of achieved compliance in this document.
Compliance with AAMI/ANSI HF 18 (2001): Electrosurgical devices (Electrosurgical Performance)The device "will be in compliance with the following recognized consensus standards prior to marketing the device." This implies an intention to comply, rather than direct reporting of achieved compliance in this document.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Animal testing" as the basis for performance data.

  • Sample Size: Not specified. The document only states "Animal testing conducted...".
  • Data Provenance: The country of origin of the animal testing data is not specified. The study is prospective as it involves active animal testing to evaluate the device.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

  • Number of Experts: Not specified. The document does not provide details on who assessed the animal testing results or how "safely and effectively resects soft tissue" was determined.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

Not specified. There is no mention of an adjudication method as the evaluation appears to be based on observed outcomes from animal testing and compliance with standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission is for a medical device (surgical tools) and does not involve image-based diagnosis or human interpretation in the context of an MRMC study. The comparison made is against predicate devices in terms of intended use, technological characteristics, and principles of operation, rather than a comparative effectiveness study with human readers assisted by AI.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, in a way. The "performance data" section focuses on the device itself (the Roei Working Element and Roei Cutting Loops) functioning independently to resect soft tissue in animal models and its compliance with safety and electrosurgical standards. This can be considered a standalone evaluation of the device's functional and safety characteristics, without human intervention for interpreting results in the way an AI algorithm would be evaluated. It's a standalone test of the physical device's performance.

7. The Type of Ground Truth Used

The ground truth used for the animal testing was the observed outcome of soft tissue resection in an animal model. The statement "safely and effectively resects soft tissue" implies direct observation of the surgical outcome (e.g., successful removal of targeted tissue, lack of excessive collateral damage, appropriate healing). For the standards compliance, the ground truth would be the parameters and requirements outlined in IEC 6061-2-2 and AAMI/ANSI HF 18.

8. The Sample Size for the Training Set

Not applicable. This device is a surgical instrument, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).