K Number
K050865
Date Cleared
2005-05-20

(45 days)

Product Code
Regulation Number
866.1645
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD PhoenixTM Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Emero Cram-Regarter accedes and and most Gram-positive bacteria isolates from pure culture belonging to the genera Stanylococcuses , Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent trimethoprim-sulfamethoxazole at concentrations of 0.0625/1.2-16/304 ug/mL to Streptococcus ID/AST or AST only Phoenix panels. Trimethoprimsulfamethoxazole has been shown to be active in vitro against most strains of microrganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against: Streptococcus pneumoniae

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components:

  • BD Phoenix instrument and software. ●
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
  • of TLS Forethinations: .
  • BD Phoenix AST-S Broth used for performing AST tests only. .
  • BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination.
    The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.
    The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
    The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).
AI/ML Overview

Acceptance Criteria and Device Performance for BD Phoenix™ Automated Microbiology System - Trimethoprim-sulfamethoxazole

This document describes the acceptance criteria and the study that demonstrates the BD Phoenix™ Automated Microbiology System, specifically with Trimethoprim-sulfamethoxazole (0.0625/1.2-16/304 µg/mL) and Streptococcus ID/AST or AST only Phoenix Panels, meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are derived from the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003, for establishing substantial equivalence to a reference method (CLSI broth microdilution). While specific numerical acceptance thresholds are not explicitly stated in the provided text, the criteria are implicitly demonstrated by achieving "substantially equivalent performance" and high agreement rates.

Acceptance Criteria CategorySpecific MetricStated Threshold (Implicit/Explicit)Reported Device Performance (Trimethoprim-sulfamethoxazole)
Accuracy (Clinical Isolates)Essential Agreement (EA) with CLSI reference method"Substantially equivalent performance" (implied high EA)95.8% (n=906)
Accuracy (Clinical Isolates)Category Agreement (CA) with CLSI reference method"Substantially equivalent performance" (implied high CA)95.3% (n=906)
Reproducibility (Intra-site)Overall Intra-site ReproducibilityGreater than 90%> 90%
Reproducibility (Inter-site)Overall Inter-site ReproducibilityGreater than or equal to 95%≥ 95%
Reproducibility (General)Reproducibility within ± 1 dilution (across clinical sites)95% or greater95% or greater

2. Sample Size and Data Provenance for Test Set

  • Sample Size: The test set for accuracy (Essential Agreement and Category Agreement) included 906 isolates for Trimethoprim-sulfamethoxazole. This combined "clinical, stock and challenge isolates."
  • Data Provenance: The isolates were tested across "multiple geographically diverse sites across the United States." The study included "clinical, stock and challenge isolates." The clinical isolates were compared to the CLSI reference broth microdilution method, while challenge isolates "were compared to the expected results." This indicates a mix of retrospective and prospective data, with clinical isolates likely representing prospective collection for the study, and stock/challenge isolates being pre-existing.

3. Number and Qualifications of Experts for Ground Truth (Test Set)

The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth for the test set. It refers to the CLSI reference broth microdilution method as the "reference method" for clinical isolates and "expected results" for challenge isolates. This implies that the CLSI method itself, including its standard operating procedures and trained personnel, served as the "expert" ground truth.

4. Adjudication Method (Test Set)

The document does not describe an explicit adjudication method (e.g., 2+1, 3+1). The comparison is directly made between the BD Phoenix System results and the CLSI reference broth microdilution method or "expected results" for challenge isolates. Discrepancies would likely be investigated, but a formal multi-reader adjudication process as seen in imaging studies is not applicable here.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typical for diagnostic imaging devices where human readers interpret results, and the AI's role is to assist these readers. This device is an automated microbiology system that performs the test and provides results automatically; it does not involve human interpretation that is then assisted by AI in the traditional sense.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire study described is a standalone performance evaluation of the BD Phoenix™ Automated Microbiology System. The device, without human-in-the-loop interpretation of the primary reading, directly produces Minimum Inhibitory Concentration (MIC) values and categorical interpretations (S, I, R, N). Its performance (Essential Agreement and Category Agreement) is compared against a reference standard (CLSI broth microdilution method), demonstrating its accuracy as a standalone system.

7. Type of Ground Truth Used

The primary ground truth used for clinical isolates was the CLSI reference broth microdilution method. For challenge isolates, "expected results" were used. The CLSI method is a well-established and accepted laboratory gold standard for antimicrobial susceptibility testing, involving the precise measurement of bacterial growth inhibition at various antimicrobial concentrations. This is a form of objective, highly standardized laboratory measurement.

8. Sample Size for the Training Set

The document does not specify the sample size for the training set. The provided text focuses on the performance evaluation (test set) for the specific antimicrobial agent and organism. Automated systems like the BD Phoenix™ are often developed using large, diverse internal datasets over time, but details of the training data for this specific component are not given.

9. How the Ground Truth for the Training Set Was Established

The document does not describe how the ground truth for the training set was established. Similar to the sample size, the specifics of the training data development and ground truth establishment for the underlying algorithm of the BD Phoenix System are not provided in this 510(k) summary. It can be inferred that similar reference methods (like CLSI broth microdilution) would have been used during the development and training phases of the system to establish the accuracy and reliability of its automated interpretations.

{0}------------------------------------------------

CONFIDENTIAL AND PROPRIETARY MAY 20 2005

510(k) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 410-316-4287Fax: 410-316-4499
CONTACT NAME:Monica Evelyn Giguere,RA Specialist
DATE PREPARED:April 1, 2005
DEVICE TRADE NAME:BD PhoenixTM Automated Microbiology System -Trimethoprim-sulfamethoxazole 0.0625/1.2-16/304 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD PhoenixTMAutomated Microbiology System with Gatifloxacin (K020323May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD PhoenixTM Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components:

  • BD Phoenix instrument and software. ●
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
  • of TLS Forethinations: .
  • BD Phoenix AST-S Broth used for performing AST tests only. .
  • BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination.

{1}------------------------------------------------

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Streptococcus ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of streptococcal isolates. Each site tested the isolates in triplicate on three different days using one lot of Streptococcus Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than or equal to 95% for the streptococcal isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Streptococcus Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or nonsusceptible).

Table 1 summarizes the performance for the isolates tested in this study.

Performance of BD Phoenix System for Streptococcal Organisms by Drug Table 1:

AntimicrobialConcentrationEA (n)EA (%) CA (n)I CA (%)
Trimethoprim-' 0.0625/1.2-16/304 μg/mL906તેરી જેવી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી90695.3
sulfamethoxazole

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

{3}------------------------------------------------

SUMMARY INFORMATION FOR TRIMETHOPRIM-SULFAMETHOXAZOLE Available Range 0.0625/1.2-16/304 µg/mL

Performance

Accuracy

AntimicrobialConcentrationEA (n)EA (%)CA (n)CA (%)
Trimethoprim-sulfamethoxazole0.0625/1.2-16/304 μg/mL90695.890695.3

Reproducibility

Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution.

Breakpoints - CLSI, FDA

THE THE THE LABOUR LA S1And Address of AcresJrganismAA 10 10 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0-----
1Streptococcus pneumoniae------------------------------------------------------------------------------------------------------------------------------------------------------------------------------000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000ViviCanadian Parcel Products of the Property of Children0190≥4/7Acres of the Art

Recommended Quality Control Organisms

Quality Control StrainMIC RangeSource
Streptococcus pneumoniae ATCC 496190.12/2.4 – 1/19 µg/mLCLSI

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by a series of three curved lines that resemble a stylized staff with intertwined snakes, a symbol often associated with medicine and healthcare.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 2 0 2005

Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

K050865 Re:

Trade/Device Name: BD Phoenix™ Automated Microbiology System Trimethoprim-sulfamethoxazole (0.0625/1.2-16/304 ug/mL) -Streptococcus ID/AST or AST only Phoenix Panels Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 1, 2005 Received: April 6, 2005

Dear Ms. Giguere:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass surfal in the enactment date of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may; aroney manage of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is classified dditional controls. Existing major regulations affecting your device it may be subject to Sach as Federal Regulations (CFR), Parts 800 to 895. In addition, FDA ound be louits in The Enouncements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean i trass be actived that I bermination that your device complies with other requirements of the Act that I Dr Has intatutes and regulations administered by other Federal agencies. You must or any I edital statutes and equirements, including, but not limited to: registration and listing (21 compry with an and the Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to Iogally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sale, a Hogg

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k) Number: K050865

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent trimethoprim-sulfamethoxazole (0.0625/1.2-16/304 ug/mL) - Streptococcus ID/AST or AST only Phoenix Panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Emero Cram-Regarter accedes and and most Gram-positive bacteria isolates from pure culture belonging to the genera Stanylococcuses , Enterococcus, and Streptococcus.

This premarket notification is for the addition of the antimicrobial agent trimethoprim-sulfamethoxazole at concentrations of

0.0625/1.2-16/304 ug/mL to Streptococcus ID/AST or AST only Phoenix panels. Trimethoprimsulfamethoxazole has been shown to be active in vitro against most strains of microrganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Streptococcus pneumoniae

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Ludden. Poole

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K050865

BD Diagnostic Systems Becton, Dickinson and Company

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”