K Number
K050744

Validate with FDA (Live)

Date Cleared
2005-04-21

(30 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For acquiring, viewing, handling, and storage of dental color images and radiographs.

Device Description

KaVo ERGOcom®/ERGOcam® system consists of four main components:

  • The ERGOcom® Video control box.
  • One of two versions of the ERGOcam® Cameras: The ERGOcam® 3 of the ERGOcam® 4 camera.
  • A video display, flat panel color from 15 to 19" with touch screen control.
  • A PC compatible computer with the KaVo Device Interface System Software installed. The Device Interface System has the following components:
      1. CCCDIS. The CCCDIS software is used to administer and configure the software components in the KaVo Device Interface System. The CCCDIS acts as the interface of the KaVo Device Interface System with third party applications
    • ERGOcam Control The ERGOcam Control software is used to configure individual parameters of the ERGOcam 4
    • ERGOcom Control The ERGOcom Control software is used for the PC-assisted status display, configuration and service support of the ERGOcom 3.
    • ERGOvideo Control The ERGOvideo Control software is used for generating and displaying video imaging data created with ERGOcom 2 / ERGOcom 3 andERGOcam 3 / ERGOcam 4
    • ERGOcam Viewer The ERGOcam Viewer is used for demonstrating the use of the ERGOvideo Control, ERGOcam Control and ERGOcom Control. The ERGOcam Viewer is used to store and administer imaging data that have been generated with the software components referred to above.
    • VirtualCom2 VirtualCom2 is used to convert the foot starter and camera cradle signals from ERGOcom 3 to a format that is compatible with ERGOcom 2. These signals are only used to control program modules within a practice management software that allows the recording of images with a KaVo intraoral camera, thereby meeting the interface specification for the serial interface of ERGOcom 2

The intraoral camera emits video signals which generate an image of the area under inspection (the patient's oral cavity for instance) on a screen or other visual display medium. The generated image is used for the patient's information and to support the diagnosis.

AI/ML Overview

The provided text is a 510(k) summary for the KaVo ERGOcom®/ERGOcam® Dental Image Management System. It states that the device is substantially equivalent to a predicate device (DentalEye 2, K012439) and does not contain detailed information about specific acceptance criteria, performance studies, or efficacy claims in the way a clinical study report would.

Therefore, the following information is not available in the provided document:

  • A table of acceptance criteria and reported device performance (beyond basic system specifications and functional equivalence).
  • Sample sizes used for test sets.
  • Data provenance (country of origin, retrospective/prospective).
  • Number and qualifications of experts for ground truth establishment.
  • Adjudication method for the test set.
  • Results of a multi-reader multi-case (MRMC) comparative effectiveness study.
  • Results of a standalone (algorithm-only) performance study.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for performance evaluation of clinical effectiveness.
  • Sample size for the training set.
  • How ground truth for the training set was established.

The document focuses on demonstrating substantial equivalence to a previously cleared device (DentalEye 2, K012439) based on intended use and technological characteristics, rather than presenting performance data from a clinical efficacy study.

Here's what can be extracted regarding how the device meets the acceptance criteria (which, in this context, are primarily about substantial equivalence and safety standards):

1. Acceptance Criteria and Reported Device Performance (Summary based on Substantial Equivalence Comparison):

The "acceptance criteria" here are implicitly defined by the characteristics and performance of the predicate device (DentalEye 2, K012439) and compliance with safety standards. The study/justification provided is a comparison table and a statement of compliance.

Acceptance Criterion (Implicit)Reported Device Performance (KaVo ERGOcom®/ERGOcam®)
Intended Use: Acquiring, viewing, handling, and storage of dental color images and radiographs.Meets: "For acquiring, viewing, handling, and storage of dental color images and radiographs." (Identical to predicate)
System Implementation: Software or Hardware/SoftwareMeets: Hardware and Software supplied (Predicate was software-only, but this is an enhancement, not a deviation from intended use)
Host Platform: Intel or AMD PCMeets: PC Pentium, 1800 MHz or higher.
Operating System: Windows 95/98/ME/2000/NTMeets: Windows 2000, Windows XP. And Microsoft .NET Framework.
Host RAM: 32 MB minimumMeets/Exceeds: 256 MB
Host Magnetic Storage: At least 500 MBMeets/Exceeds: Sufficient hard disk space to store imaging data., Usually 20 GB
CD ROM: For installationMeets: Yes (for installation)
Host Processor Speed: Pentium II 233 MHz or betterMeets/Exceeds: PC Pentium, 1800 MHz or higher.
Host Monitor Size: Any VGA or better PC color monitorMeets/Exceeds: XGA or better, 15" or 19" Touch Display.
Display Resolution: Minimum 800 X 600, Recommended 1024 X 768Meets/Exceeds: Screen resolution 1024 X 768, with min. true color depth (32 bit).
User Display Preferences: YesMeets: Yes
Receive Images from other Systems: YesMeets: Yes
Images Displayed: Dental X-Rays, Intraoral Images, Extraoral ImagesMeets/Exceeds: Dental X-Rays, Intraoral Images, Extraoral Images (face, etc), Entertainment (TV) for patients, Instructional Video, Patient Administration.
Safety Standards: (Predicate listed "Not applicable. Software only supplied.")Meets: UL/CSA standards for safety met. System has been thoroughly tested to IEC/UL/CSA standards including electrical safety and electromagnetic compatibility and has been found to comply with those standards.

2. Study/Evidence Proving Acceptance:

The "study" described is a comparative analysis for substantial equivalence to a predicate device (DentalEye 2, K012439) and testing for compliance with relevant safety standards.

Key evidence provided:

  • Comparison Table: A detailed side-by-side comparison of the KaVo ERGOcom®/ERGOcam® system with the predicate device (DentalEye 2) across various technical and functional specifications.
  • Statement of Compliance: A declaration that the system has been "thoroughly tested to IEC/UL/CSA standards including electrical safety and electromagnetic compatibility and has been found to comply with those standards."
  • Conclusion of Substantial Equivalence: A statement that "In all important respects, the KaVo ERGOcom®/ERGOcam® Dental Image System is substantially equivalent to one or more predicate systems, including the one named above."

Since this is a 510(k) submission primarily demonstrating substantial equivalence for an image management system without claims of diagnostic aid or advanced clinical performance, the following information is not applicable or not provided in the document:

  • Sample size for the test set and data provenance: No clinical test set data is presented for efficacy. The "test set" implicitly refers to the physical device undergoing safety and compatibility testing.
  • Number of experts and qualifications for ground truth: Not applicable for this type of submission focused on substantial equivalence of an image management system.
  • Adjudication method for the test set: Not applicable.
  • MRMC comparative effectiveness study: Not applicable, as no claims are made about human reader improvement with AI assistance.
  • Standalone (algorithm only) performance: Not applicable, as this is an image management system with human-in-the-loop operation.
  • Type of ground truth used: Not applicable for establishing clinical efficacy, as none is measured or reported.
  • Sample size for the training set: Not applicable, as this is not an AI/machine learning device with a "training set" in the typical sense.
  • How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

APR 2 1 2005

K050744

EXHIBIT 2 510(k) Summary

KaVo America Corporation 340 East Main Street Lake Zurich, Illinois 60047 Tel: 847 / 550 - 6800 Fax: 847 / 550 - 6825 Contact: John Miller, Director of RA/QA March 7, 2005

  1. Identification of the Device:

Proprietary-Trade Name: KaVo ERGOcom®/ERGOcam® Classification Name: Dental Hand Instrument, Laser Fluorescence Caries Detection Device, Product Codes Picture archiving and communications system, 90 LLZ or System, x-ray, extraoral source, digital Product Code MUH Common/Usual Name: Dental Image Management System

  1. Equivalent legally marketed device: DentalEye 2, K012439.
  • Indications for Use (intended use): For acquiring, viewing, handling, and storage of dental 3. color images and radiographs.
  • Description of the Device: KaVo ERGOcom®/ERGOcam® system consists of four main ধ components:
    • The ERGOcom® Video control box. .
    • One of two versions of the ERGOcam® Cameras: The ERGOcam® 3 of the ERGOcam® 4 . camera.
    • A video display, flat panel color from 15 to 19" with touch screen control .
    • A PC compatible computer with the KaVo Device Interface System Software installed. The . Device Interface System has the following components:
        1. CCCDIS. The CCCDIS software is used to administer and configure the software components in the KaVo Device Interface System . The CCCDIS acts as the interface of the KaVo Device Interface System with third party applications
      • ERGOcam Control The ERGOcam Control software is used to configure individual 2. parameters of the ERGOcam 4
      • ERGOcom Control The ERGOcom Control software is used for the PC-assisted status 3. display, configuration and service support of the ERGOcom 3.
      • ERGOvideo Control The ERGOvideo Control software is used for generating and 4. displaying video imaging data created with ERGOcom 2 / ERGOcom 3 andERGOcam 3 / ERGOcam 4
      • ERGOcam Viewer The ERGOcam Viewer is used for demonstrating the use of the 5. ERGOvideo Control, ERGOcam Control and ERGOcom Control. The ERGOcam Viewer is used to store and administer imaging data that have been generated with the software components referred to above.
      • VirtualCom2 VirtualCom2 is used to convert the foot starter and camera cradle signals 6. from ERGOcom 3 to a format that is compatible with ERGOcom 2 .These signals are only used to control program modules within a practice management software that

{1}------------------------------------------------

allows the recording of images with a KaVo intraoral camera, thereby meeting the interface specification for the serial interface of ERGOcom 2

The intraoral camera emits video signals which generate an image of the area under inspection (the patient's oral cavity for instance) on a screen or other visual display medium. The generated image is used for the patient's information and to support the diagnosis.

DescriptionDentalEye 2 K012439KaVo ERGOcom®/ERGOcam®
ImplementationSoftware onlyHardware and Software supplied
Host platformIntel or AMD based Personal ComputerPC Pentium, 1800 MHz or higher.
Operating systemWindows 95 Windows 98Windows MEWindows 2000Windows NTWindows 2000, Windows XP. And Microsoft .NET Framework.
Host RAM32 MB minimum256 MB
Host Magnetic StorageAt least 500 MBSufficient hard disk space to store imaging data., Usually 20 GB
Host Floppy Drives:Not requiredNot required
CD ROMYes (for installation)Yes (for installation)
Host Processor SpeedPentium II 233 MHz or betterPC Pentium, 1800 MHz or higher.
Host Monitor SizeAny VGA or better PC color monitorXGA or better, 15" or 19" Touch Display.
Display ResolutionMinimum 800 X 600Recommended1024 X 768Screen resolution 1024 X 768, with min. true color depth (32 bit).
User Display PreferencesYesYes
USB and S Video supportNOYes
Receive Images from other SystemsYesYes
Images DisplayedDental X-Rays, Intraoral Images, Extraoral Images (face, etc)Dental X-Rays, Intraoral Images, Extraoral Images (face, etc), Entertainment (TV) for patients, Instructional Video, Patient Administration.
Safety StandardsNot applicable. Software only supplied.UL/CSA standards for safety met.
    1. Safety and Effectiveness, comparison to predicate device:
    1. Conclusion: In all important respects, the KaVo ERGOcom®/ERGOcam® Dental Image System is substantially equivalent to one or more predicate systems, including the one named above. The system has been thoroughly tested to IEC/UL/CSA standards including electrical safety and electromagnetic compatibility and has been found to comply with those standards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form, with three figures connected in a flowing design.

Public Health Service

APR 2 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

KaVo America Corporation % Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates PO Box 7007 DEERFIELD IL 60015

Re: K050744

Trade/Device Name: KaVo ERGOcom® / ERGOcam® Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system ﺎ Regulatory Class: II Product Code: MUH and LLZ Dated: March 17, 2005 Received: March 25, 2005

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced abone and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been rear 20, 1770, and coance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the 1 oderal building, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will anow you to begin manusing of substantial equivalence of your device to a legally prematRet notification: "The PDF maning of our device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de received in the regulation the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other-240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Alls, please note the regulation on your responsibilities under the Act from the 807.97). Tou may outlar other general and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

ﯿﮟ

050744 510(k) Number (if known):

Device Name:_________________________________________________________________________________________________

Indications For Use:

For acquiring, viewing, handling, and storage of dental color images and radiographs.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

్లో

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

  • Nancy C Brogdon
    (Division Sign-Off)
    Division of Reprodu
    and Radiological
    510(k) Number
    K050744

Page 1 of 1

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).