K Number
K050536
Manufacturer
Date Cleared
2005-03-24

(22 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quest Diagnostics Immunoassay/TDM Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Device Description

Quest Diagnostics Immunoassay/TDM Control is prepared from human serum, with added constituents of human and animal origin, chemicals, and therapeutic drugs. The control is provided in lyophilized form for increased stability.

AI/ML Overview

This is a premarket notification for the Quest Diagnostics Immunoassay/TDM Control. Based on the provided document, here's an analysis of the acceptance criteria and supporting study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document is a 510(k) premarket notification claiming substantial equivalence to a predicate device. For such devices, the "acceptance criteria" are generally related to demonstrating comparable performance (precision, stability) to the predicate device. The performance of the new device is typically evaluated against established ranges or specifications for each analyte, which are implicitly derived from the predicate device's expected performance and clinical utility.

The document primarily focuses on stability claims as explicit performance criteria for the new device.

Acceptance Criteria (Stability Claims)Reported Device Performance (Quest Diagnostics Immunoassay/TDM Control)
Open Vial Stability: Stability when stored tightly capped at 2 to 8°C.7 days at 2 to 8°C, with the following exceptions:
- C-Peptide, Folate, and PSA are stable for 3 days.
- Gastrin, Free PSA, and Intact PTH should be assayed immediately after reconstitution.
Shelf Life: Stability when stored at 2 to 8°C.36 months when stored at 2 to 8°C.

Note: The document does not explicitly state numerical acceptance criteria for assay performance (e.g., specific coefficients of variation or bias limits). Instead, it relies on demonstrating that the new control performs equivalently to the predicate and meets its intended use as a quality control serum to monitor the precision of laboratory testing procedures for listed analytes. This equivalence is implicitly accepted by the FDA's 510(k) clearance based on the provided data.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Immunoassay/TDM Control. Product claims are as follows:..."

  • Sample Size for Test Set: The specific sample sizes (e.g., number of control lots, number of replicates, number of instruments used) for the stability studies are not provided in this summary. It simply states "Stability studies have been performed."
  • Data Provenance: The document does not explicitly state the country of origin of the data. However, the submitter is Bio-Rad Laboratories, located in Irvine, California, USA. The testing would presumably have been conducted in a facility under their control, likely in the USA. The studies are prospective as they were specifically performed to determine the stability claims for this new device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable to a quality control material where the "ground truth" relates to the assigned values or reference ranges of the analytes within the control. For quality control materials, the "ground truth" (the expected range of values) is established through rigorous internal validation processes by the manufacturer, often using reference methods and statistical analysis rather than expert consensus on individual "cases." The intended use is to monitor the precision of laboratory tests, meaning the control's own stability and performance consistency are critical.

4. Adjudication Method for the Test Set

This is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or diagnostic results where expert consensus is needed to determine the "true" clinical outcome or diagnosis for a case. For a quality control material, the performance assessment is based on analytical measurements and statistical process control, not subjective interpretation requiring adjudication among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. An MRMC study is designed to evaluate the diagnostic performance of a device (often imaging-related) in an clinical setting, often comparing human interpretation with and without AI assistance across multiple readers and cases. The Quest Diagnostics Immunoassay/TDM Control is a quality control material, not a diagnostic device that requires human interpretation or AI assistance in a diagnostic workflow.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical quality control serum, not an algorithm, and does not operate "standalone" in the sense of a software-only diagnostic tool. Its performance is demonstrated through its own chemical and physical stability when used within laboratory testing procedures.

7. The Type of Ground Truth Used

For a quality control material, the "ground truth" for its performance is typically:

  • Analytical Measurement: The measured concentrations or activities of the analytes within the control using calibrated reference methods.
  • Stability over Time: Demonstrated by empirical testing that the analyte values remain within acceptable ranges over the claimed shelf life and open-vial stability periods.
  • Material Homogeneity and Consistency: Ensuring that each vial performs consistently within a lot.

The document states: "Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Immunoassay/TDM Control." This implies that the ground truth for these claims was established through prospective analytical testing and measurement of the control material over specified timeframes and storage conditions.

8. The Sample Size for the Training Set

This is not applicable. Quality control materials are manufactured and validated, not "trained" in the way an AI algorithm or machine learning model would be. There is no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set was Established

As there is no "training set" for a quality control material, this question is not applicable. The underlying properties of the control material (analyte concentrations, matrix characteristics) are established during the manufacturing and formulation process, and then its performance (stability) is verified through specific studies as described above.

{0}------------------------------------------------

MAR 2 4 2005

Bio-Rad Laboratories Premarket Notification Section 510(k) for Quest Diagnostics Immunoassay/TDM Control Summary of Safety and Effectiveness

Page 1 of 3

Ko 50536

Summary of Safety and Effectiveness Quest Diagnostics Immunoassay/TDM Control

1.0 Submitter

Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 (949) 598-1200 Telephone: (949) 598-1555 Fax:

Contact Person

Maria Zeballos Regulatory Affairs Specialist Telephone: (949) 598-1367

Date of Summary Preparation

February 24, 2005

2.0 Device Identification

Quest Diagnostics Immunoassay/TDM Control Product Trade Name: Multi-Analyte Controls, (Assayed and Unassayed) Common Name:

Classifications: Product Code: Requiation Number:

Class I JJY CFR 862.1660

Device to Which Substantial Equivalence is Claimed 3.0

Lyphochek Immunoassay Plus Control Bio-Rad Laboratories Irvine, CA 92618 Docket Number: K981532

4.0 Description of Device

Quest Diagnostics Immunoassay/TDM Control is prepared from human serum, with added constituents of human and animal origin, chemicals, and therapeutic drugs. The control is provided in lyophilized form for increased stability.

5.0 Statement of Intended Use

Quest Diagnostics ImmunoassayTDM Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

008_SSE_Quest_IA/TDM_930 2/05

{1}------------------------------------------------

6.0 Comparison of the new device with the Predicate Device

Quest Diagnostics Immunoassay/TDM Control claims substantial equivalence to the Lyphochek Immunoassay Plus Control currently in commercial distribution (K981532).

CharacteristicsQuest DiagnosticsImmunoassay/TDM Control(New Device)Bio-Rad LaboratoriesLyphochek Immunoassay Plus Control(Predicate Device K981532)
Similarities
Intended UseQuest Diagnostics Immunoassay/TDM Control is intended for useas a quality control serum to monitor the precision of laboratorytesting procedures for the analytes listed in the package insert.Lyphochek Immunoassay Plus Control is intended for use as anassayed quality control serum to monitor the precision oflaboratory testing procedures for the analytes listed in thepackage Insert.
FormLyophilizedLyophilized
MatrixHuman SerumHuman Serum
PreservativesDoes not Contains preservativesDoes not Contains preservatives
Storage (Unopened)2°C to 8°CUntil expiration date2°C to 8°CUntil expiration date
Differences
Reconstituted VialClaim7 days at 2 to 8°Cwith the following exceptions:(1) C-Peptide, Folate and PSA 3 days(2) Gastrin, Free PSA and Intact PTH assay immediately7 days at 2°C to 8°Cwith the following exceptions:(1) Folate and PSA 3 days,(2) C-Peptide 1 day,(3) Intact PTH 16 hours,(4) ACTH, Calcitonin, Gastrin and Free PSA assay immediately
After Reconstitutingand FreezingNo claimsAll analytes 30 days at -10 to -20°C
AnalytesContains the following analytes:• 17-a-Hydroxyprogesterone• Acetaminophen• Aldosterone• Alpha Fetoprotein (AFP)• Amikacin• Androstenedione• Caffeine• Carbamazepine• CEA• Chloramphenicol• Cortisol• C -Peptide• DHEA Sulfate• Digoxin• Disopyramide• Estradiol• Estriol, Free• Ethosuximide• Ferritin• Folate• FSH• Gastrin• Gentamicin• Homocysteine• HCG• HGH• Immunoglobulin E (IgE)Contains the following analytes:• Lidocaine• Lithium• LH• NAPA• PTH, Intact• Phenobarbital• Phenytoin• Primidone• Procainamide• Progesterone• Prolactin• PAP• PSA• PSA, Free• Quinidine• Salicylate• T3, Free• T3, Total• T4, Free• T4, Total• Testosterone• Theophylline• Tobramycin• TSH• Valproic Acid• Vancomycin• Vitamin B12• 11-Deoxycortisol• 17-a-Hydroxyprogesterone• Acetaminophen• ACTH• Alpha Fetoprotein (AFP)• Amikacin• Aldosterone• Amitriptyline• Androstenedione• Caffeine• Calcitonin• Carbamazepine, Free• Carbamazepine• CEA• Chloramphenicol• Cortisol• C-Peptide• DHEA Sulfate• Digoxin• Disopyramide• Cyclosporine• Desipramine• DHEA• Estradiol• Estriol, Free• Estriol, Total• Estrogens, Total• Immunoglobulin E (IgE)• Insulin• Lidocaine• Lithium• LH• NAPA• Netilmicin• Nortriptyline• PTH• Phenobarbital• Phenytoin• Phenytoin, Free• Primidone• Procainamide• Progesterone• Prolactin• PAP• Propanolol• PSA• PSA, Free• Quinidine• Salicylate• T3 Free• T3 Total• T3 Uptake• T4 Free• T4 Total

Table 1. Similarities and Differences between new and predicate device.

008_SSE_Quest_IA/TDM_930 2/05



{2}------------------------------------------------

Bio-Rad Laboratories Dio-Nab Laboratones

Premarket Notification Section 510(k) for Quest Diagnostics Immunoassay/TDM Control Summary of Safety and Effectiveness

CharacteristicsQuest DiagnosticsImmunoassay/TDM Control(New Device)Bio-Rad LaboratoriesLyphochek Immunoassay Plus Control(Predicate Device K981532)
Does not Contain the following analytes:• 11-Deoxycortisol• Aldosterone• Amitriptyline• Calcitonin• Carbamazepine, Free• Cyclosporine *• Desipramine• DHEA• Estriol, Total• Estrogens, Total• Flecainide• 25-Hydroxy Vitamin D• Angiotensin I• Fructosamine• Glucagon• Iron• TIBC• HCG-Beta Subunit• Imipramine• Netilmicin• Nortriptyline• Phenytoin, Free• Propanolol• T3 Uptake• TCA Screen• Testosterone, Free• Valproic Acid, Free• Immunoglobulin A (IgA)• Immunoglobulin G (IgG)• Immunoglobulin M (IgM)• Somatomedin-C• TBG• Thyroglobulin• Folate• Flecainide• FSH• Gastrin• Gentamicin• hCG and hCG-Beta Subunit• hGH• Imipramine• 25-Hydroxy Vitamin D• Angiotensin I• Fructosamine• Glucagon• Iron• TIBCDoes not Contain the following analytes:• Homocysteine• Testosterone, Free• Theophylline• Tobramycin• TSH• Valproic Acid• Valproic Acid, Free• Vancomycin• Vitamin B12• Immunoglobulin A (IgA)• Immunoglobulin G (IgG)• Immunoglobulin M (IgM)• Somatomedin-C• TBG• Thyroglobulin

1.0 STATEMENT OF SUPPORTING DATA

Stability studies have been performed to determine the open vial stability and shelf life for the Quest Diagnostics Immunoassay/TDM Control. Product claims are as follows:

  • Open vial Stability: 7days when stored tightly capped at 2 to 8℃ with the following exceptions: ● C-Peptide, Folate and PSA are stable for 3 days and Gastrin, Free PSA and Intact PTH should be assayed immediately after reconstitution.
  • . Shelf Life: 36 months when stored at 2 to 8 °C

All supporting data is retained on file at Bio-Rad Laboratories.

008_SSE_Quest_IA/TDM_930 2/05

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

MAR 2 4 2005

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Elizabeth Platt Regulatory Affairs Manager/ Quality Assurance Bio-Rad Laboratories, QSD 9500 Jeronimo Road Irvine, CA 92618-2017

K050536 Re:

Trade/Device Name: Ouest Diagnostics Immunoassay/TDM Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: February 24, 2005 Received: March 2, 2005

Dear Ms. Platt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Sincerely yours,

Jean M. Cooper MS, DVM

Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

CHClosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Quest Diagnostics Immunoassay/TDM Control

Indications For Use:

Quest Diagnostics Immunoassay/TDM Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Alute Sutt

(Division Sign-Off) Division of Clinical Laboratory Device 510(k) Number


Page 1 of

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.