K Number
K043599
Manufacturer
Date Cleared
2005-02-25

(58 days)

Product Code
Regulation Number
862.1300
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Moments Menopause Check is an in-vitro diagnostic screen for the detection of FSH (follicle stimulating hormone) in urine. Change in FSH levels may be associated with changes in menopause. This kit provides a preliminary screening result. This kit is intended for over-the-counter sales.

Device Description

Immunoassay for the qualitative detection of FSH in urine.

AI/ML Overview

Here's an analysis of the provided information regarding the Moments Menopause Check (Model 9113), focusing on its acceptance criteria and the supporting study:

Acceptance Criteria and Device Performance for Moments Menopause Check (Model 9113)

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Qualitative detection of FSH in urine.The device performs qualitative detection of FSH in urine.
Minimum detectable concentration for FSH is 25 mIU/ml.The minimum detectable concentration for this test is 25 mIU/ml.
Substantially equivalent to predicate devices for qualitative detection of FSH."Laboratory studies, using clinical specimens, produced a >98% correlation when compared to the predicate devices."
Excellent overall accuracy for consumer interpretation."Consumer interpretation of the FSH test showed accuracy to be greater than 98%."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the sample size used for the clinical sample correlation study or the blind labeled spiked study.
  • Data Provenance: The studies used "clinical specimens." The country of origin for the data is not specified, but the manufacturer is Phamatech, located in San Diego, California, USA. The studies appear to be prospective, as they were conducted to evaluate the performance of the Moments Menopause Check.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the performance studies.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study being performed to assess how much human readers improve with AI vs. without AI assistance. This device is a rapid, qualitative immunoassay for over-the-counter use, and such a study design is generally not applicable to this type of diagnostic.

6. Standalone Performance Study

A standalone performance study was performed. The "clinical sample correlation study" and "blind labeled spiked study" evaluated the device's performance (qualitative detection of FSH in urine) independently. The "consumer study" also assessed the device in a standalone context, focusing on user interpretation.

7. Type of Ground Truth Used

  • Clinical Sample Correlation Study: The ground truth for this study was likely established by the "predicate devices" to which the Moments Menopause Check was compared. This implies a comparison to existing, legally marketed FSH tests which themselves have established accuracy.
  • Blind Labeled Spiked Study: For this study, the ground truth would have been based on known, controlled concentrations of FSH that were "spiked" into samples, allowing for a direct assessment of the device's ability to detect FSH at the specified threshold (25 mIU/ml).
  • Consumer Study: The ground truth for the consumer interpretation was likely derived from the objective results of the FSH test (e.g., whether FSH was actually present at the detectable level) and then compared to the consumer's visual interpretation.

8. Sample Size for the Training Set

The document does not provide information about a training set since this is a chemical test (immunoassay) and not a machine learning or AI-based device. Therefore, the concept of a "training set" as understood in AI/ML is not applicable here. The device's performance is based on its inherent chemical properties and design.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, the concept of a training set and its ground truth is not applicable to this device, which relies on immunoassay technology rather than machine learning.

{0}------------------------------------------------

K 043599

product for over the counter use. Such studies are a critical element in establishing the fundamental safety and effectiveness of the products and their appropriateness for commercial distribution.

FEB 2 5 2005

510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(C)

Identification: Moments Menopause Check Model 9113

Description: Immunoassay for the qualitative detection of FSH in urine.

Name Of Manufacturer:

Phamatech 10151 Barnes Canyon Road San Diego, California 92121, USA

Intended Use: The Moments Menopause Check is a rapid, qualitative immunoassay for the detection of follicle stimulating hormone (FSH) in urine. The minimum detectable concentration for this test is 25 mlU/ml. This assay is intended for use in the home to assist in the early detection of menopause.

The Moments Menopause Check, like many commercially available FSH Technology: screening tests, qualitatively measures the presence of follicle stimulating hormone by visual color sandwich one step immunoassay technology. Examples of such predicate devices include the Phamatech (San Diego, CA) Moments Menopause Check (Models 9111 and 9112) and the Estroven Menopause Monitor (Distributed by Amerifit Nutrition, Inc. Bloomfield, CT). All of the above devices rely on the basic immunochemical sandwich assay principle of recognition and formation of specific antibody / FSH / colored (labeled) antibody complexes.

The product performance characteristics of the Moments Menopause Performance: Check were evaluated in a clinical sample correlation study and a blind labeled spiked study. The results of these studies demonstrate the Moments Menopause Check to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of follicle stimulating hormone in urine. Laboratory studies, using clinical specimens, produced a >98% correlation when compared to the predicate devices.

A consumer study was also performed, in it the Moments Menopause Check exhibited excellent overall accuracy. Consumer interpretation of the FSH test showed accuracy to be greater than 98%

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FEB 2 5 2005

Mr. Carl Mongiovi Vice President Phamatech, Inc. 10151 Barnes Canyon Road San Diego, CA 92121

K043599 Re:

Trade/Device Name: Moments Menopause Check (Model 9113) Regulation Number: 21 CFR 862.1300 Regulation Name: Follicle-stimulating hormone test system Regulatory Class: Class I Product Code: CGJ Dated: February 14, 2005 Received: February 17, 2005

Dear Mr. Mongiovi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commercc prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Jean M. Cooper MS, DUM

Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR USE

Applicant: Phamatech

510 (k) Number (if known): _K043599

Device Name: Moments Menopause Check (Model 9113)

Indications for Use:

The Moments Menopause Check is an in-vitro diagnostic screen for the M The Moments Michopause Shoulating hormone) in urine. Change in FSH levels aelection of For (follicio shimalains) nonopause. This kit provides a preliminary may be dssociated will relagos in money club in is intended for over-thecounter sales.

Prescription Use:_____________________________________________________________________________________________________________________________________________________________ Part 21 CFR 801 Subpart D

Over the Counter_ AND/OR 21 CFR 807 Subpart C

Concurrence of CDRH Office of In Vitro Diagnostic Devices (OIVD)

_
Division Sign-off

Division Sign-off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k): K043599

§ 862.1300 Follicle-stimulating hormone test system.

(a)
Identification. A follicle-stimulating hormone test system is a device intended to measure follicle-stimulating hormone (FSH) in plasma, serum, and urine. FSH measurements are used in the diagnosis and treatment of pituitary gland and gonadal disorders.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.