(332 days)
The RefStar™ RMT External Reference Patch is indicated for use with the NAVI-STAR™ RMT catheter and the CARTO™ RMT system to provide catheter tip location as well as electrogram information.
The RefStar RMT is an external reference catheter designed to be placed on the patient's back in order to compensate for movement during electrophysiology mapping of the heart. The back of the distal end of the RefStar RMT External Reference Patch has an adhesive backing, designed to be placed externally on the patient's back. The catheter contains a location sensor, that, when used together with the CARTO RMT system and the , NAVISTAR RMT Catheter provides catheter tip location information to construct a 3D electrophysiological maps of the human heart in the Stereotaxis Magnetic Navigation System magnetic field (K013484 and K021555) in real-time.
The provided text describes a 510(k) premarket notification for the RefStar RMT External Reference Patch. This device is seeking clearance based on substantial equivalence to a predicate device, not by proving efficacy via clinical trials with specific acceptance criteria. This means the submission doesn't contain a detailed study with acceptance criteria and device performance results in the way a novel device might.
Here's a breakdown of the requested information based on the provided text:
1. Table of acceptance criteria and the reported device performance
Based on the provided text, no specific quantitative acceptance criteria or detailed performance results (e.g., accuracy, precision) are explicitly stated for the RefStar RMT. The submission focuses on demonstrating substantial equivalence.
Acceptance Criteria | Reported Device Performance |
---|---|
Not explicitly stated as quantitative acceptance criteria for the RefStar RMT. The primary criteria is "substantial equivalence" to the predicate device. | Bench testing demonstrates that the devices are functionally equivalent. RefStar RMT complies with multiple external electrical and performance standards. |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of a clinical performance study. The testing mentioned is "bench testing under simulated use conditions." Therefore, no information on sample size for a test set or data provenance (country of origin, retrospective/prospective) is provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The document does not describe a clinical study requiring expert-established ground truth.
4. Adjudication method for the test set
Not applicable. The document does not describe a clinical study requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a surface reference patch for an electrophysiology system, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant to its clearance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a hardware component of a larger electrophysiology system and does not operate as a standalone algorithm.
7. The type of ground truth used
Not applicable in the context of a clinical performance study with "ground truth." The "truth" for this submission is the established performance and safety of the predicate device. The performance of the RefStar RMT was likely assessed against established engineering specifications and functional requirements to ensure it matched the predicate.
8. The sample size for the training set
Not applicable. The document does not describe a training set as this is not an AI/machine learning model.
9. How the ground truth for the training set was established
Not applicable. The document does not describe a training set or its associated ground truth.
Summary of the Study (Bench Testing and Substantial Equivalence):
The RefStar RMT was cleared based on proving substantial equivalence to a predicate device, the Cordis-Webster RefStar™ External Surface Reference Device (K980961). The study described is bench testing under simulated use conditions. The core argument for safety and efficacy relies on:
- Identical Indication for Use: Both the RefStar RMT and the predicate device have the same indication for use: "to provide catheter tip location as well as electrogram information" when used with specific catheter and system components.
- Similar Design Requirements and Technological Characteristics: The new device meets the same design requirements and shares similar technological characteristics with the predicate.
- Functional Equivalence: Bench testing demonstrated that the devices are functionally equivalent. This would typically involve verifying that the new device performs its intended functions (e.g., providing location signals, electrical properties) within acceptable tolerances and in a manner consistent with the predicate.
- Compliance with Standards: The RefStar RMT complies with multiple external electrical and performance standards, although the specific standards are not listed.
Essentially, the "study" for this 510(k) was a comparison to an already cleared device, demonstrating that the new device performs its intended function similarly and does not raise new questions of safety or effectiveness.
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).