K Number
K042953
Date Cleared
2004-11-22

(27 days)

Product Code
Regulation Number
888.3070
Panel
OR
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PROTEX™ Stabilization System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudoarthrosis and failed previous fusion.

In addition, the PROTEX™ Stabilization System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

When used as a posterior non-pedicle screw fixation system, the PROTEX ™ Stabilization System is intended for the treatment of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or fordosis, Scheuermann's disease), fracture, pseudarthrosis, tumor resection, and/or failed previous fusion. Overall levels of fixation are T1-sacrum/ilium.

When used as an anterolateral thoracolumbar system, the PROTEX™ Stabilization System is intended for anterolateral screw (with or without staple) fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e. scoliosis, kyphosis, and/or lordosis), fracture or dislocation of the thoracolumbar spine, pseudoarthrosis, tumor resection, and/or failed previous fusion. Levels of screw fixation are T8-L5.

Device Description

The PROTEX™ Stabilization System consists of a variety of shapes and sizes of rods, hooks, monoaxial screws, polyaxial screws, locking caps, tconnectors, staples, and associated manual surgical instruments. Implant components can be rigidly locked into a variety of configurations for the individual patient and surgical condition. Polyaxial screws, hooks, and tconnectors are intended for posterior use only. AccuRods are intended for posterior use with polyaxial and monoaxial screws only. Staples are intended for anterior use only. Rods and monoaxial screws may be used anteriorly or posteriorly. Locking caps are used to connect screws or hooks to the rod.

The implants are composed of titanium alloy as specified in ASTM F136 and F1295.

AI/ML Overview

The PROTEX(TM) Stabilization System is a spinal implant system. The provided text is a 510(k) summary for this device, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study that proves the device meets specific acceptance criteria in the way a clinical trial for a novel AI device would.

Therefore, the requested information points regarding acceptance criteria, device performance, sample sizes, expert involvement, and ground truth establishment, cannot be found in this 510(k) submission. This document is a regulatory filing for a medical device seeking market clearance based on substantial equivalence, not a clinical study report.

Here's why each point cannot be addressed with the provided text:

  1. A table of acceptance criteria and the reported device performance: This type of information would be found in a clinical study report or a detailed design verification/validation report, neither of which is present in a 510(k) summary. The 510(k) focuses on a comparison to a predicate device.
  2. Sample sized used for the test set and the data provenance: A 510(k) summary does not typically include details of clinical studies with test sets in the manner of an AI/diagnostic device. The "test set" for a spinal implant would usually involve mechanical testing and pre-clinical studies, not clinical data sets in the sense of AI. No sample sizes or data provenance are mentioned for any such 'test set'.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is irrelevant for a spinal implant 510(k) submission. "Ground truth" in this context would typically refer to gold standard clinical outcomes or pathological findings in a clinical study, which is not described.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of submission. Adjudication methods are relevant for clinical trials where expert consensus is needed to define an outcome.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This type of study is completely unrelated to a spinal implant system. MRMC studies are designed for diagnostic imaging devices and AI algorithms that assist human interpretation.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The PROTEX™ system is a physical surgical implant, not an algorithm.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): As noted, the concept of "ground truth" as typically discussed for AI or diagnostic devices does not apply here. For a spinal implant, success is measured by mechanical stability, biocompatibility, and clinical outcomes after surgery.
  8. The sample size for the training set: Not applicable. There is no "training set" for a physical spinal implant device in the context of machine learning or AI.
  9. How the ground truth for the training set was established: Not applicable for the same reason as point 8.

In summary, the provided document is a regulatory submission for a physical medical device (spinal implant) and does not contain the type of AI/diagnostic study information requested. It focuses on device description, indications for use, and comparison to predicate devices for substantial equivalence.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.