(119 days)
The DakoCytomation ER/PR pharmDx™ assay is an immunohistochemical (IHC) kit system to identify human estrogen receptor (ER) and human progesterone receptor (PR) expression in breast cancer tissues routinely processed and paraffin-embedded for histological evaluation. ER/PR pharmDx specifically detects the ER alpha protein as well as the PR located in the cell nucleus of ER and/or PR-expressing cells.
ER/PR pharmDx is indicated as an aid in identifying patients eligible for treatment with antihormonal or aromatase inhibitor therapies, as well as an aid in the prognosis and management of breast cancer.
The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is intended for laboratory use as a semi-quantitative detection of progesterone receptor by light microscopy in routinely processed normal and pathological human paraffinembedded tissue. This antibody is indicated for use as an aid in the management, prognosis and prediction of outcome of breast cancer. The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient's clinical history and other diagnostic tests by a qualified individual
The DakoCytomation ER/PR pharmDx™ assay is a semi-quantitative immunohistochemical (IHC) kit system to identify estrogen receptor (ER) and progesterone receptor (PR) expression in normal and neoplastic tissues routinely processed and paraffin-embedded for histological expressing cells respectively.
ER/PR pharmDx assay is available in two configurations, both manual and automated and is optimized for use with DakoCytomation detection systems.
The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is a component of the ER/PR pharmDx kit, which will also be commercially available as a concentrate.
Here's an analysis of the provided 510(k) summary, specifically focusing on the acceptance criteria and study proving device performance for the DakoCytomation ER/PR pharmDx™ Kit and the Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294:
Acceptance Criteria and Device Performance
The document does not explicitly present a table of numerical acceptance criteria with corresponding reported device performance values. Instead, it describes performance characteristics that were evaluated and states that the results demonstrated "a substantial degree of equivalency to the predicate devices" and "substantially equivalent staining results" to a reference method.
The primary acceptance criteria appear to be substantial equivalence to predicate devices and high concordance with a recognized reference method.
Table of Acceptance Criteria and Reported Device Performance (Inferred)
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Concordance | Substantial equivalency (e.g., >95% or similar) with predicate devices and reference methods. | 99% concordance with the reference Allred method. |
| Specificity | Demonstrates appropriate specificity compared to predicate devices. | Demonstrated specificity; results deemed substantially equivalent to predicate devices. |
| Sensitivity | Demonstrates appropriate sensitivity compared to predicate devices. | Demonstrated sensitivity; results deemed substantially equivalent to predicate devices. |
| Reproducibility | Demonstrates appropriate reproducibility compared to predicate devices. | Demonstrated reproducibility; results deemed substantially equivalent to predicate devices. |
Study Description and Details:
The summary refers to a single study that encompassed evaluations of specificity, sensitivity, reproducibility, and concordance testing.
-
Sample Size and Data Provenance:
- Test Set Sample Size: Not explicitly stated for any of the performance characteristics.
- Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
-
Number of Experts and Qualifications for Ground Truth:
- The document does not provide information on the number of experts used to establish ground truth or their specific qualifications.
-
Adjudication Method:
- The document does not specify any adjudication method used for the test set.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study is mentioned. The comparison is between the new device and predicate devices/reference methods, not an assessment of human reader improvement with or without AI assistance.
-
Standalone Performance (Algorithm Only):
- Yes, the performance characteristics (specificity, sensitivity, reproducibility, and concordance) are described for the device itself (the IHC kit and its components) when used according to its intended purpose. This represents a standalone assessment of the assay's performance.
-
Type of Ground Truth Used:
- The ground truth for the concordance testing was established using the "Allred method." The Allred score is a semi-quantitative scoring system for ER/PR expression in breast cancer, combining the proportion of positive cells and the staining intensity. This can be considered a type of expert consensus/established semi-quantitative scoring method based on histopathology. For specificity, sensitivity, and reproducibility, the ground truth would also be based on established histopathological evaluation or comparison to the predicate devices, which are already considered validated.
-
Training Set Sample Size:
- Not specified. The document primarily focuses on the validation study demonstrating substantial equivalence. It does not provide details about any internal development or training data used in the creation of the assay itself, as this would typically be a chemical/biological assay rather than a machine learning algorithm requiring a separate "training set" in the computational sense.
-
How Ground Truth for Training Set was Established:
- Not applicable as this is not a machine learning device and no "training set" is described. The development of such IHC kits typically involves chemical formulation, antibody selection, and optimization against known positive and negative tissue samples, but these are not referred to as a "training set" in the context of this document.
{0}------------------------------------------------
FEB 1 5 2005
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______K042884
| Submitter: | DakoCytomation California, Inc.6392 Via RealCarpinteria, CA 93013PH. 805.566.6655 FX. 805.566.0866Establishment registration number: 2022180 |
|---|---|
| Contact: | Tiffany D. Almeroth, RACSr. Regulatory Affairs SpecialistPH. 805.566.3041 |
| Date Summary Prepared: | September 30, 2004 |
| Device Name(s): | DakoCytomation ER/PR pharmDx™ Kit.Immunohistochemistry kit.(Code K1903/K1904) |
| DakoCytomation Monoclonal Mouse Anti-Human ProgesteroneReceptor, clone PgR 1294.Antibody for immunoenzymatic staining.(Code M3568) | |
| Device Classification: | Class II, for prognostic immunohistochemical staining reagents21 CFR 864.1860 |
| Panel: | Hematology and Pathology Devices PanelDivision of Clinical Laboratory Devices. |
| Predicate Devices: | Estrogen Receptor• DakoCytomation Monoclonal Mouse Anti-HumanEstrogen Receptor, Clone 1D5, (K993957)• Ventana ER clone 6F11, (K984567)Progesterone Receptor• DakoCytomation Monoclonal Mouse Anti-HumanProgesterone Receptor, Clone PgR 636, (K020023) |
Device Description:
The DakoCytomation ER/PR pharmDx™ assay is a semi-quantitative immunohistochemical (IHC) kit system to identify estrogen receptor (ER) and progesterone receptor (PR) expression in normal and neoplastic tissues routinely processed and paraffin-embedded for histological expressing cells respectively.
ER/PR pharmDx assay is available in two configurations, both manual and automated and is optimized for use with DakoCytomation detection systems.
{1}------------------------------------------------
The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is a component of the ER/PR pharmDx kit, which will also be commercially available as a concentrate.
Intended Use:
For In Vitro Diagnostic Use
The DakoCytomation ER/PR pharmDx™ assay is an immunohistochemical (IHC) kit system to identify human estrogen receptor (ER) and human progesterone receptor (PR) expression in breast cancer tissues routinely processed and paraffin-embedded for histological evaluation. ER/PR pharmDx specifically detects the ER alpha protein as well as the PR located in the cell nucleus of ER and/or PR-expressing cells.
ER/PR pharmDx is indicated as an aid in identifying patients eligible for treatment with antihormonal or aromatase inhibitor therapies, as well as an aid in the prognosis and management of breast cancer.
The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is intended for laboratory use as a semi-quantitative detection of progesterone receptor by light microscopy in routinely processed normal and pathological human paraffinembedded tissue. This antibody is indicated for use as an aid in the management, prognosis and prediction of outcome of breast cancer. The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient's clinical history and other diagnostic tests by a qualified individual
Substantial Equivalence:
The DakoCytomation anti-estrogen receptor reagent cocktail. clones 1D5 and ER-2-123. is substantially equivalent to the Ventana ER clone 6F11 and DakoCytomation monoclonal mouse anti-human ER, clone 1D5 in that these products specifically bind to estrogen receptor proteins located in the nuclei of cells. These products require similar detection chemistry principles for visualization of the product, and both aid in the prognosis of breast carcinoma. The difference in visualization does not introduce new issues of safety and effectiveness.
The DakoCytomation PgR 1294 antibody is substantially equivalent to the DakoCytomation PgR 636 antibody, in that these products both bind to progesterone receptor proteins located in the nuclei of breast cancer cells, and use similar detection chemistry principles for visualization.
Performance Characteristics:
Performance characteristics evaluated in support of the ER/PR pharmDx™ IHC kit and components include results on specificity, sensitivity, reproducibility, and concordance testing. Results of all testing conducted have demonstrated a substantial degree of equivalency to the predicate devices listed above. Further, concordance testing between the DakoCytomation ER/PR pharmDx kit and the reference, Allred method, demonstrated
{2}------------------------------------------------
substantially equivalent staining results (99% concordance).
Therefore, based on the information provided in this premarket notification, DakoCytomation concludes that the devices listed above are safe, effective and substantially equivalent to their respective predicate devices in their indications for use, device design, materials, operational principles, and intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is centered on the image and is the only element present.
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
FEB 1 5 2005
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Tiffany D. Almeroth, RAC Senior, Regulatory Affairs Specialist DakoCytomation California, Inc. 6392 Via Real Carpinteria, California 93013
K042884 Re:
Trade/Device Name: DakoCytomation ER/PR pharmDxTM Kit Regulation Number: 21 CFR § 864.1860 Regulation Name: Immunohistochemistry Reagents and Kits Regulatory Class: II Product Code: MYA, MXZ Dated: January 3, 2005 Received: January 6, 2005
Dear Ms. Almeroth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 -
If you desire specific information about the application of labeling requirements to your device, rr you stions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
lobatz Beckerh
Robert L. Becker, Jr., MD, PH. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): __K042884
Monoclonal Mouse Anti-Human Progesterone Receptor, Clone: Device Name: PgR 1294 (DakoCytomation Code No. M3568)
Indications For Use:
The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is intended for laboratory use as a semi-quantitative detection of progesterone receptor by light microscopy in routinely processed normal and or progotical human paraffin-embedded tissue. This antibody is indicated for use as an aid in the management, prognosis and prediction of outcome of breast cancer. The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient's clinical history and other diagnostic tests by a qualified individual.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | OR | Over-The-Counter Use | |
|---|---|---|---|
| (Per 21 CFR 801.109) | (Per 21 CFR 801.110 | ||
| IVD Use | |||
| (Per 21 CFR 801.119 | (Optional Format 1-2-96) |
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510(k) | K042884 |
|---|---|
| -------- | --------- |
{6}------------------------------------------------
510(k) Number (if known):__K042884
DakoCytomation ER/PR pharmDx™ Device Name: (DakoCytomation Code No. K1903)
Indications For Use:
The DakoCytomation ER/PR pharmDx™ assay is a semi-quantitative immunohistochemical (IHC) kit system to identify estrogen receptor (ER) and progesterone receptor (PR) expression in normal and neoplastic tissues routinely processed and paraffin-embedded for histological evaluation. ER/PR pharmDx specifically detects the ER alpha protein as well as the PR protein located in the cell nucleus of ER and PR-expressing cells respectively.
ER/PR pharmDx is indicated as an aid in identifying patients eligible for treatment with anti-hormonal or aromatase inhibitor therapies, as well as an aid in the prognosis and management of breast cancer.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | ✓ (Per 21 CFR 801.109) | OR | Over-The-Counter Use | (Per 21 CFR 801.110) |
|---|---|---|---|---|
| IVD Use | (Per 21 CFR 801.119 |
Division Sign-Off

(Optional Format 1-2-96)
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510(k) | K042884 |
|---|---|
| -------- | --------- |
§ 864.1860 Immunohistochemistry reagents and kits.
(a)
Identification. Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.(b)
Classification of immunohistochemistry devices. (1) Class I (general controls). Except as described in paragraphs (b)(2) and (b)(3) of this section, these devices are exempt from the premarket notification requirements in part 807, subpart E of this chapter. This exemption applies to IHC's that provide the pathologist with adjunctive diagnostic information that may be incorporated into the pathologist's report, but that is not ordinarily reported to the clinician as an independent finding. These IHC's are used after the primary diagnosis of tumor (neoplasm) has been made by conventional histopathology using nonimmunologic histochemical stains, such as hematoxylin and eosin. Examples of class I IHC's are differentiation markers that are used as adjunctive tests to subclassify tumors, such as keratin.(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
(3) Class III (premarket approval). IHC's intended for any use not described in paragraphs (b)(1) or (b)(2) of this section.
(c)
Date of PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the Federal Food, Drug, and Cosmetic Act is required for any device described in paragraph (b)(3) of this section before this device may be commercially distributed. See § 864.3.