K Number
K042692

Validate with FDA (Live)

Device Name
CLARITY FUSION
Date Cleared
2004-10-07

(7 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To detect or image the distribution of radionuclides in the body or organ, using the following techniques:

  • Multiplanar Reconstruction (MPR)
  • Maximum/Minimum Intensity Projection (MIP) .
  • . Image Contrast Manipulation
  • Image Zoom Manipulation .
  • Automatic registration with Mutual Information Technique .
Device Description

Clarity Fusion is an option within the Clarity PET system so the end-user may differentiate between the two different applications within the same software program. This viewing of fused images is the ability to see the same datasets, as if they were being viewed independently, overlapped with each others. These fused images will allow a physician to look at two different datasets from different modalities. This is common medical practice that has now been computerized.

AI/ML Overview

The provided text is a 510(k) summary for the Clarity Fusion device. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the specific information required to answer your request about acceptance criteria, device performance, study details (sample sizes, ground truth establishment, expert qualifications, adjudication methods), or multi-reader multi-case studies.

This submission is a "Special 510(k)", which typically streamlines the review process for modifications to a legally marketed device if the changes do not affect the device's fundamental scientific technology or intended use. In such cases, extensive new clinical performance studies are often not required, and instead, the focus is on verification and validation that the modifications meet the original safety and effectiveness requirements and maintain substantial equivalence.

Therefore, many of the details you're asking for (like specific acceptance criteria performance data, sample sizes of test/training sets, expert qualifications, or MRMC studies) are not present in this type of 510(k) summary. These types of details would typically be found in a Traditional 510(k) submission where a new device is being introduced.

Here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Cannot provide. This information is not present in the provided text. The submission relies on demonstrating substantial equivalence to a predicate device (Clarity PET) and states that "software validation/verification and a Declaration of Conformity were submitted to demonstrate substantial equivalence." It does not provide specific acceptance criteria or performance metrics for the Clarity Fusion itself beyond its functionalities listed.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Cannot provide. This information is not present.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Cannot provide. This information is not present. Ground truth methodology and expert involvement are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Cannot provide. Adjudication methods are not mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was mentioned. The device described is an image processing system that fuses images from different modalities and provides tools like MPR, MIP, contrast manipulation, and automatic registration. It's not presented as an AI-enabled diagnostic aid that would typically be evaluated with MRMC studies comparing human readers with and without AI assistance. It "serv[es] primarily as an aid in image display" and is "not intended to replace the final interpretation of physicians."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable/Cannot provide. The device is an image display and processing tool, not an AI algorithm performing a diagnostic task independently. Its function is to enhance diagnostic image quality and aid in image display for a physician, not to operate as a standalone diagnostic algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Cannot provide. The method for establishing ground truth is not described in this document.

8. The sample size for the training set

  • Cannot provide. This information is not present.

9. How the ground truth for the training set was established

  • Cannot provide. This information is not present.

{0}------------------------------------------------

PREMARKET NOTIFICATION SPECIAL 510(K) SUMMARY 8

Sponsor:

Shared P.E.T. Imaging, LLC 4912 Higbee Avenue NW. Ste 100 Canton OH 44718 Telephone: (330) 491-0480 Fax: (330) 491-0488 Contact: Marc A. Simms, Director of Information Technology

Proprietary Name:Clarity Fusion
Common Name:Clarity Fusion
Classification Name:System, Image Processing, Radiological
Reference Number:§892.2050
Owner/Operator Number:9056369
Establishment RegistrationNumber:3004082056
Listing Status:Active
Proposed Regulatory Class:Class II
Device Product Code:LLZ
Panel / Code:Radiology
Reason for Special 510(k):Establish Substantial Equivalence to a previously marketed predicatedevice, Clarity PET. The Clarity PET (K032866) is alsomanufactured by Shared PET Imaging. This SPECIAL 510(k)merely adds a module to Clarity PET for enhanced diagnostic imagequality. The intended use and indications for are not changed fromthe original 510(k)

{1}------------------------------------------------

. Device Description and Intended Use:

Clarity Fusion is an option within the Clarity PET system so the end-user may differentiate between the two different applications within the same software program. This viewing of fused images is the ability to see the same datasets, as if they were being viewed independently, overlapped with each others. These fused images will allow a physician to look at two different datasets from different modalities. This is common medical practice that has now been computerized.

. General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warning, to provide for safe and effective use of the device. Serving primarily as an aid in image display, this device has no direct adverse effect on health as the results are to be integrated into all of the information a physician will use to form a final interpretation.

Substantial Equivalence: .

As required for a special 510(k), the Clarity Fusion has the identical indications as the Clarity PET and software validation/verification and a Declaration of Conformity were submitted to demonstrate substantial equivalence.

. 510(k) Cleared Indications for Use:

To detect or image the distribution of radionuclides in the body or organ, using the following techniques:

  • . Multiplanar Reconstruction (MPR)
  • Maximum/Minimum Intensity Projection (MIP) .
  • . Image Contrast Manipulation
  • Automatic registration with Mutual Information Technique .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The bird is positioned to the left of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 7 - 2004

'Shared P.E.T. Imaging, LLC % Mr. Russ Pagano Official Correspondent M Squared Associates, Inc. 719 A Street, NE WASHINGTON DC 20002 Re: K042692

Trade/Device Name: Clarity Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system

Regulatory Class: II Product Code: 90 LLZ Dated: September 29, 2004 Received: September 30, 2004

Dear Mr. Pagano:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we nave reviewed four courte is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the encreated of the enactment date of the Medical Device Amendments, or to devices that have been rear 20, 1770, accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approval controls provisions of the Act. The general controls provisions of the Act include reguirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket " 11 your do roo to subject to such additional controls. Existing major regulations affecting your Apploral), It they of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statues and regulations administered by other Federal agencies. You must comply with all the I oc 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmadsmamain.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

K042692 510(k) Number (if known): Device name: Clarity Fusion Indications for use:

To detect or image the distribution of radionuclides in the body or organ, using the following techniques:

  • Multiplanar Reconstruction (MPR) .
  • Maximum/Minimum Intensity Projection (MIP) .
  • . Image Contrast Manipulation
  • Image Zoom Manipulation .
  • Automatic registration with Mutual Information Technique .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)

Prescription use _X (Per 21 CFR801.109) OR

Over-the Counter use (Optional format 1-2-96)

Nancy C. Boughton

510(k) Number

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).