K Number
K042529
Date Cleared
2004-10-06

(19 days)

Product Code
Regulation Number
882.1460
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BalanceBack VNG device is intended for recording, viewing, and analyzing eye movements in support of identifying balance disorders in human subjects. The VNG is intended for use only by trained physicians or clinicians in an appropriate doctor's office or health care facility. This device provides no diagnoses nor does it provide diagnostic recommendations.

Device Description

Eve movements have long been known to provide important clues to the identification and location of balance disorders. Physicians and other clinicians employ the VNG system to conduct a set of established, well-known and well-documented eye movement tests. The results of the VNG eye movement tests are combined with other clinical information so that the attending clinicians may render findings on the health of the patient's primary balance sensors and related balance processing system.

Using the VNG, clinicians guide patients through a series of eye movement tests. A testing battery (i.e., an examination) typically consists of all eight tests supported by the VNG: caloric, positional, Hallpike, gaze, saccades, smooth pursuit, optokinetic, and high-frequency head shake. The attending physician may select one, several, or all tests, and has full flexibility in determining the sequence of the tests to be performed.

The patient under examination wears the VNG goggles. The goggles contain one IR light source and one IR camera for each eye. The IR video cameras sense the infrared signal reflected from the eyes and create a continuous monochromatic (grayscale) video motion picture stream of the eyes. The infrared signal is used because the human eye cannot detect the presence of the infrared signal and therefore this illumination does not create perturbations in measurements of eye movement. Measurements may therefore be made both in the presence and in the absence of visible light, thus providing clinicians with maximum flexibility for making measurements in a wide variety of situations.

The video stream from the IR cameras is digitized and transferred into the memory of the digital computer where signal processing software tracks, records, analyzes and displays eye movements for the attending clinician. The VNG system makes appropriate calculations and presents reports of well-known and well-documented parameters associated with each eye movement test.

Some eye tests require the patient to follow a visual stimulus. A projector, controlled by the VNG software, displays stimuli on any light-colored surface, such as a wall or projection screen. The goggles and VNG software record the patient's eye movements while he/she follows the stimulus.

AI/ML Overview

The provided text describes the BalanceBack VNG System, a device for recording, viewing, and analyzing eye movements to assist in identifying balance disorders. However, the document primarily focuses on establishing substantial equivalence to predicate devices and adherence to recognized electrical safety and performance standards. It does not present a study with specific acceptance criteria that the device's performance was measured against in terms of clinical outcomes or diagnostic accuracy.

Here's an analysis based on the provided text, highlighting the absence of the requested information:

1. A table of acceptance criteria and the reported device performance

The document mentions "Verification procedures with pass/fail criteria were developed to ensure that the product met all the specified requirements." However, it does not explicitly state what these acceptance criteria were in terms of device performance for identifying balance disorders, nor does it report the device's performance against such criteria. The criteria mentioned are related to electrical safety and product standards.

Acceptance CriterionReported Device Performance
Not provided in the document for clinical/diagnostic performance. The document focuses on compliance with general and electrical safety standards.Not provided in the document for clinical/diagnostic performance. The document states "various tests of the hardware and software are being performed to verify system specifications" but does not detail the results.
Compliance with ANSI Z-136.1 (2000)"A certified body shall conduct tests to determine the conformance of the device to the recognized standards shown below. The device will not be marketed or sold until it has been certified to pass these tests." (Implied compliance, but no specific report of passing these for this submission).
Compliance with IEC 60601-1 (2004)(Same as above)
Compliance with IEC 60601-1-1 (2000)(Same as above)
Compliance with IEC 60601-1-2 (2001)(Same as above)
Compliance with UL 60601-1 (2003)(Same as above)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not provided. The document does not describe a clinical test set or data. The "Testing" section refers to hardware and software verification, not clinical performance studies on human subjects to assess diagnostic aid capabilities.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable/Not provided. Since no clinical test set for diagnostic accuracy is described, there's no mention of experts establishing ground truth for such a test. The device "provides no diagnoses nor does it provide diagnostic recommendations."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable/Not provided. No clinical test set warranting adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable/Not provided. This device is presented as a tool to aid clinicians by recording, viewing, and analyzing eye movements, not as an AI-assisted diagnostic tool that would typically be evaluated in an MRMC study for improved human reader performance. The device "provides no diagnoses nor does it provide diagnostic recommendations."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable/Not provided. The device is described as "intended for use only by trained physicians or clinicians" and it "provides no diagnoses nor does it provide diagnostic recommendations." Therefore, a standalone diagnostic performance evaluation would not be relevant given its stated function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable/Not provided. No clinical performance study requiring ground truth is described.

8. The sample size for the training set

Not applicable/Not provided. The document describes a medical device for VNG, and there is no indication of machine learning or AI components that would require a "training set" in the context of diagnostic algorithms.

9. How the ground truth for the training set was established

Not applicable/Not provided. As there's no mention of a training set for an AI/ML algorithm, this information is not relevant or present in the document.

In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence to predicate devices based on similar technology and intended use, and indicates compliance with recognized electrical safety and performance standards (though the actual certification is pending marketing). It does not include a study demonstrating the clinical performance or diagnostic accuracy of the BalanceBack VNG System against specific acceptance criteria for identifying balance disorders. The device is explicitly stated not to provide diagnoses or diagnostic recommendations, which simplifies the regulatory pathway by avoiding requirements for extensive clinical performance studies to prove diagnostic accuracy.

{0}------------------------------------------------

OCT 6 - 2004

510(k) Summary E

Submitter's Name:Fall Prevention Technologies, LLC
Submitter's Address:2300 NW Corporate Blvd. Suite 141
Boca Raton, FL 33431
Submitter's Telephone:561-995-8883
Contact Name:Keith Kravitz
Date Summary wasPrepared:August 10, 2004
Trade or Proprietary Name:BalanceBack VNG System
Common or Usual Name:VNG
Classification Name:Nystagmograph (21 CFR 882.1460), product code GWN,Class II
Predicate Devices:Device Name510(k) Number
Eye Dynamics, Inc. HouseInfrared/Video Electronystagmographic(ENG) SystemK925111
Sensomotoric Instruments, Inc. 2d Vog- Video-Oculography SystemK972243

Indications for Use:

The BalanceBack VNG device is intended for recording, viewing, and analyzing eye movements in support of identifying balance disorders in human subjects. The VNG is intended for use only by trained physicians or clinicians in an appropriate doctor's office or health care facility. This device provides no diagnoses nor does it provide diagnostic recommendations.

Description of the Device and Summary of the Technological Characteristics:

Eve movements have long been known to provide important clues to the identification and location of balance disorders. Physicians and other clinicians employ the VNG system to conduct a set of established, well-known and well-documented eye movement tests. The results of the VNG eye movement tests are combined with other clinical information so that the attending clinicians may render findings on the health of the patient's primary balance sensors and related balance processing system.

Using the VNG, clinicians guide patients through a series of eye movement tests. A testing battery (i.e., an examination) typically consists of all eight tests supported by the VNG: caloric, positional, Hallpike, gaze, saccades, smooth pursuit, optokinetic, and

1/2

K042529

{1}------------------------------------------------

high-frequency head shake. The attending physician may select one, several, or all tests, and has full flexibility in determining the sequence of the tests to be performed.

K042529

The patient under examination wears the VNG goggles. The goggles contain one IR light source and one IR camera for each eye. The IR video cameras sense the infrared signal reflected from the eyes and create a continuous monochromatic (grayscale) video motion picture stream of the eyes. The infrared signal is used because the human eye cannot detect the presence of the infrared signal and therefore this illumination does not create perturbations in measurements of eye movement. Measurements may therefore be made both in the presence and in the absence of visible light, thus providing clinicians with maximum flexibility for making measurements in a wide variety of situations.

The video stream from the IR cameras is digitized and transferred into the memory of the digital computer where signal processing software tracks, records, analyzes and displays eye movements for the attending clinician. The VNG system makes appropriate calculations and presents reports of well-known and well-documented parameters associated with each eye movement test.

Some eye tests require the patient to follow a visual stimulus. A projector, controlled by the VNG software, displays stimuli on any light-colored surface, such as a wall or projection screen. The goggles and VNG software record the patient's eye movements while he/she follows the stimulus.

Substantial Equivalence:

The BalanceBack VNG device is substantially equivalent to the Eye Dynamics, Inc. House Video/Video Electronystagmographic (ENG) System and the Sensomotoric Instruments, Inc. 2d Vog - Video-Oculography System. All three devices support the recording, viewing, and analyzing of horizontal and vertical eye movements to assist in the identification of balance disorders in human subjects. These devices use similar technology, including workstation, computer monitor, masks mounted with video cameras, and software, for performing caloric, positional, Hallpike, gaze, saccades, smooth pursuit, and optokinetic examinations. All three devices use the same sampling rate and similar illumination wavelengths.

Testing:

Various tests of the hardware and software are being performed to verify system specifications. Verification procedures with pass/fail criteria were developed to ensure that the product met all the specified requirements. As part of this verification, a certified body shall conduct tests to determine the conformance of the device to the recognized standards shown below. The device will not be marketed or sold until it has been certified to pass these tests.

• ANSI Z-136.1 (2000)• IEC 60601-1-1 (2000)• UL 60601-1 (2003)
IEC 60601-1 (2004)• IEC 60601-1-2 (2001)
  • · IEC 60601-1 (2004) · IEC 60601-1-2 (2001)

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 6 - 2004

Fall Prevention Technologies, LLC c/o Mr. Daniel W. Lehtonen Staff Engineer - Medical Devices Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, Massachusetts 01719

Re: K042529

Trade/Device Name: BalanceBack VNG System Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: II Product Code: GWN Dated: September 16, 2004 Received: September 17, 2004

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your be devermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreatives of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may a controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or bain the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reade o a reade a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Daniel W. Lehtonen

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ocgin maneting of substantial equivalence of your device to a legally premarket notification. The PDF Intellig of castion for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire spocific acrive 2017 904 276-0115. Also, please note the regulation entitled, Coliner the Office of Comparket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

D Indications for Use Statement

Ver/ 3 - 4/24/96

Applicant: Fall Prevention Technologies, LLC

510(k) Number (if known): _____

Device Name: BalanceBack VNG System

Indications For Use:

The BalanceBack VNG device is intended for recording, viewing, and analyzing eye movements in support of identifying balance disorders in human subjects. The VNG is intended for use only n support of residentlying on an appropriate doctor's office or health care facility. This device provides no diagnoses nor does it provide diagnostic recommendations.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

er the Counter Use

Miriam C. Provost

(Division Division of General, Restorative, and Neurological Devices

510(k) Number_

Prescription Use
per 21 CFR 801.109

§ 882.1460 Nystagmograph.

(a)
Identification. A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.(b)
Classification. Class II (performance standards).