K Number
K042190

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-09-07

(26 days)

Product Code
Regulation Number
870.3680
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Model 6416 Bipolar Transvenous Temporary Pacing System features an active fixation, bipolar lead and a soft-tipped lubricated guide catheter. The system is designed for temporary intracardiac pacing and/or EGM recording.

The system is disposable, for temporary single patient use with a contemplated implant duration of 7 days or less. The lead and accessories are supplied sterile.

The lead is introduced transvenously using the guide catheter. Once within the appropriate chamber, the helical tip electrode of the lead is actively fixed into the endocardium. After lead placement, the guide catheter is removed by sliding it over the lead's bifurcated connector.

Device Description

The Model 6416 Bipolar Transvenous Temporary Pacing Lead System comprises a temporary lead and a soft-tipped Medtronic guiding catheter. The pacing lead is designed to provide pacing characteristics similar to presently marketed bipolar temporary leads. The lead is introduced percutaneously through the guiding catheter. The lead system is provided in varying lengths to allow for jugular, subclavian, or femoral approach. A removable silicone rubber torque tool is provided on the proximal portion of the lead, to facilitate rotation of the lead during implant and explant. The disposable introducer currently packaged with the system is the Medtronic Vector X coronary guiding catheter, size 6 French. This catheter guides or steers the temporary lead during placement, and shrouds the active fixation helix from blood vessel walls and other inappropriate tissue until actual implantation in the atrium or ventricle. A hemostasis valve tightens and loosens around the lead and controls the passage of the lead through the guide catheter.

AI/ML Overview

This document, K042190, is a Special 510(k) Premarket Notification for the Medtronic Model 6416 Bipolar Transvenous Temporary Pacing System. This submission focuses on a modification to an already marketed device, not the introduction of a completely new device or a typical AI/ML device. Therefore, the standard metrics often associated with AI/ML device performance and acceptance criteria using large datasets, ground truth, and expert adjudication are not applicable here.

The document primarily demonstrates that the modified device is substantially equivalent to its predicate device and that the modifications do not alter its fundamental scientific technology or intended use. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and compatibility tests, not AI model performance evaluations.

Here's an interpretation based on the provided document:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Required for Substantial Equivalence)Reported Device Performance (Summary of Studies)
Maintain existing safety and performance characteristicsMet specified requirements for system compatibility.
Maintain existing technological characteristicsNo change with the modified device.
No change to material type or formulationBiocompatibility testing not repeated; materials unchanged.
Maintain existing sterilization process and effectivenessUses the same 100% Ethylene Oxide (EtO) sterilization process as the predicate device.
Demonstrate compatibility with the guiding catheter and lead"Lead/Catheter Compatibility testing" performed.
Preclinical confirmation of safety and performance"Preclinical confirmation of safety and performance" performed.
Visual verification to ensure no adverse changes"Visual Verification" performed.
Environmental conditioning to ensure device integrity"Environmental Conditioning" performed.

2. Sample sizes used for the test set and the data provenance
The document does not specify exact sample sizes for the "system compatibility testing." The tests listed (Environmental Conditioning, Visual Verification, Lead/Catheter Compatibility testing, Preclinical confirmation) are typically done on a smaller, representative number of devices to demonstrate functionality and safety, not large patient datasets. The data provenance is internal Medtronic testing for device modification.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as the studies are engineering verification and validation tests, not clinical studies requiring expert ground truth for interpretation of medical images or patient data.

4. Adjudication method
Not applicable. The tests performed are objective engineering and performance evaluations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
No, a MRMC study was not done. This type of study is relevant for evaluating human performance with and without AI assistance in diagnostic tasks, which is not the purpose of this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device, not an AI algorithm. The performance evaluation is for the physical device's function, not an algorithm.

7. The type of ground truth used
The "ground truth" for the tests mentioned would be the established engineering specifications, performance standards, and safety requirements for the device. For example:

  • Environmental Conditioning: Ground truth would be the device's ability to withstand specified environmental stressors without degradation in performance.
  • Visual Verification: Ground truth would be adherence to design specifications and absence of manufacturing defects.
  • Lead/Catheter Compatibility testing: Ground truth would be successful and safe interaction between components as per design.
  • Preclinical confirmation of safety and performance: Ground truth would likely be established in animal models or benchtop tests demonstrating the intended function (pacing, EGM recording, fixation) without causing harm.

8. The sample size for the training set
Not applicable. There is no AI model or "training set" in the context of this device and its modification.

9. How the ground truth for the training set was established
Not applicable, as there is no training set mentioned.

{0}------------------------------------------------

Special 510(k) Premarket Notification Attachments

Model 6416 Bipolar Transvenous Temporary Pacing System

K042190

pl /3

SEP = 7 2004

510(K) SUMMARY OF SUBSTANTIAL EQUIVALENCE

Submitter

Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432

Contact: Paula Cordero, Regulatory Affairs Specialist Telephone: (763) 514-8233 Fax: (763) 514-6424 E-Mail: paula.cordero@medtronic.com Date Prepared: August 10, 2004

Name of Device

Trade Name: Model 6416 Bipolar Transvenous Temporary Pacing System Common Name: Temporary Pacing Lead System Classification: Class II

Predicate Devices

The predicate device for the Modified Model 6416 Bipolar Transvenous Temporary Pacing System is the current market released Model 6416 Bipolar Transvenous Temporary Pacing System.

Device Description

The Model 6416 Bipolar Transvenous Temporary Pacing Lead System comprises a temporary lead and a soft-tipped Medtronic guiding catheter. The pacing lead is designed to provide pacing characteristics similar to presently marketed bipolar temporary leads. The lead is introduced percutaneously through the guiding catheter. The lead system is provided in varying lengths to allow for jugular, subclavian, or femoral approach. A removable silicone rubber torque tool is provided on the proximal portion of the lead, to facilitate rotation of the lead during implant and explant. The disposable introducer currently packaged with the system

Image /page/0/Picture/13 description: The image shows the Medtronic logo. The logo consists of a circular symbol on the left, followed by the word "Medtronic" in bold, black letters. To the right of "Medtronic" is the word "Confidential" in a smaller, non-bold font.

{1}------------------------------------------------

Special 510(k) Premarket Notification Attachments

Model 6416 Bipolar Transvenous Temporary Pacing System

KO42190

p.2/3

is the Medtronic Vector X coronary guiding catheter, size 6 French. This catheter guides or steers the temporary lead during placement, and shrouds the active fixation helix from blood vessel walls and other inappropriate tissue until actual implantation in the atrium or ventricle. A hemostasis valve tightens and loosens around the lead and controls the passage of the lead through the guide catheter.

Packaging

The sterile packaging for the Model 6416 Bipolar Transvenous Temporary Pacing System consists of a single pouch configuration. The pouch materials are transparent Tyvek- polyester/polyethylene laminate. The pouches are heat-sealed.

Intended Use

The Model 6416 Bipolar Transvenous Temporary Pacing System features an active fixation, bipolar lead and a soft-tipped lubricated guide catheter. The system is designed for temporary intracardiac pacing and/or EGM recording. The system is disposable, for temporary single patient use with a contemplated implant duration of 7 days or less. The lead and accessories are supplied sterile. The lead is introduced transvenously using the guide catheter. Once within the appropriate chamber, the helical tip electrode of the lead is actively fixed into the endocardium. After lead placement, the guide catheter is removed by sliding it over the lead's bifurcated connector.

Technological Characteristics

The technology used with the Model 6416 Bipolar Transvenous Temporary Pacing System has not changed with the modified Model 6416 Bipolar Transvenous Temporary Pacing System.

Summary of Studies

Medtronic, Inc. performed system compatibility testing to support that the modified Model 6416 Bipolar Transvenous Temporary Pacing System is equivalent to the predicate device. Device testing included:

  • . Environmental Conditioning
  • . Visual Verification
  • Lead/ Catheter Compatibility testing �
  • Preclinical confirmation of safety and performance �

Image /page/1/Picture/15 description: The image shows the Medtronic logo and the word "Confidential" to the right of the logo. The Medtronic logo consists of a stylized image of a person inside of a circle. The word "Medtronic" is in bold, sans-serif font.

{2}------------------------------------------------

Special 510(k) Premarket Notification Attachments

Model 6416 Bipolar Transvenous Temporary Pacing System

K042190

p 3 /3

All system tests performed have demonstrated that the modified Model 6416 Bipolar Transvenous Temporary Pacing System meets the specified requirements.

Sterilization

The Modified Model 6416 Bipolar Transvenous Temporary Pacing System is sterilized using the same100% Ethylene Oxide (EtO) sterilization process as the predicate device. Sterilization certification of the Modified Model 6416 Bipolar Transvenous Temporary Pacing System is based on the manufacturer's determination of substantial equivalence to the predicate device.

Biocompatibility

Biocompatibility testing was not repeated for the modified Model 6416 Bipolar Transvenous Temporary Pacing System. No change was made to material type or material formulation from previously cleared predicate devices.

Conclusion

Through data and information presented, numerous similarities support a determination of substantial equivalence and show the device modification does not affect the intended use of the device or alter the fundamental scientific technology of the device. Market clearance of the Modified Model 6416 Bipolar Transvenous Temporary Pacing System is supported through this Special 510(k) PreMarket Notification.

Image /page/2/Picture/9 description: The image shows the Medtronic logo with the word "Confidential" to the right of it. The Medtronic logo consists of a stylized human figure inside of a circle. The word "Medtronic" is in bold, black letters. The word "Confidential" is in a smaller, non-bold font.

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP - 7 2004

Medtronic, Inc. c/o Ms. Paula Cordero Regulatory Affairs Specialist 7000 Central Avenue NE Minneapolis, MN 55432

Re: K042190

Trade Name: Medtronic® Model 6146 Bipolar Transvenous Temporary Pacing System Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular Permanent or Temporary Pacemaker Electrode Regulatory Class: II (two) Product Code: LDF Dated: August 10, 2004 Received: August 12, 2004

Dear Ms. Cordero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Paula Cordero

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. production will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaince noticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other golferal miormational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Nut RP Ogden
Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Special 510(k) Premarket Notification Indications for Use

Model 6416 Bipolar Transvenous Temporary Pacing System K

INDICATIONS FOR USE

510(k) Number (if known):

Device Name:

Indications For Use:

NA K042190

Medtronic Model 6416 Bipolar Transvenous Temporary Pacing System

The Model 6416 Bipolar Transvenous Temporary Pacing System features an active fixation, bipolar lead and a soft-tipped lubricated guide catheter. The system is designed for temporary intracardiac pacing and/or EGM recording.

The system is disposable, for temporary single patient use with a contemplated implant duration of 7 days or less. The lead and accessories are supplied sterile.

The lead is introduced transvenously using the guide catheter. Once within the appropriate chamber, the helical tip electrode of the lead is actively fixed into the endocardium. After lead placement, the guide catheter is removed by sliding it over the lead's bifurcated connector.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use ________
-------------------------------------------------------------------------------------------------------------------
(Division Sign-Off)Neil Arogen
Division of Cardiovascular Devices
510(k) NumberKO 42190

Image /page/5/Picture/12 description: The image shows the Medtronic logo with the word "Confidential" next to it. The Medtronic logo is a stylized image of a person inside of a circle. The word "Medtronic" is in a bold, sans-serif font.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.