K Number
K042099
Device Name
TONOPACH ULTRASONIC TONOMETER/PACHYMETER MODEL P-201 (THE TONOPACH)
Date Cleared
2004-12-10

(128 days)

Product Code
Regulation Number
886.1930
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraocular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.
Device Description
The TonoPach is a portable, battery operated, handheld ophthalmic instrument which measures both intraocular pressure (IOP) and corneal thickness (CT) simultaneously and at the same locus on the cornea. The device uses the principles of applanation pneumatometery to measure IOP and the principles of reflection pulsed ultrasound pachymetry to measure CT. The system is comprised of the TonoPach electronics, transducer handpiece, protective membranes and optionally, a foot pedal.
More Information

K 852774, K993674, K020876

Not Found

No
The description focuses on the physical principles of measurement (applanation pneumatometry and reflection pulsed ultrasound pachymetry) and does not mention any AI/ML components or algorithms. The performance studies describe standard clinical and model-based testing, not AI/ML model training or validation.

No
The device is described as a diagnostic instrument used for measurement and screening, not for treatment.

Yes

The "Intended Use / Indications for Use" section explicitly states that the device "is a diagnostic instrument" and is intended for use "as a screening, monitoring or diagnostic aid".

No

The device description explicitly states it is a "portable, battery operated, handheld ophthalmic instrument" comprised of "electronics, transducer handpiece, protective membranes and optionally, a foot pedal," indicating it is a hardware device with integrated software, not a software-only device.

Based on the provided information, the TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is not an In Vitro Diagnostic (IVD).

Here's why:

  • Intended Use: The intended use clearly states that the device is for measuring intraocular pressure and corneal thickness in patients. This involves direct interaction with the human body (the cornea).
  • Device Description: The device is a handheld ophthalmic instrument that uses a transducer handpiece to make measurements on the cornea. This is an in vivo measurement.
  • Anatomical Site: The anatomical site is the cornea, which is part of the living eye.
  • Performance Studies: The performance studies involve testing on live rabbit eyes, freshly enucleated porcine eyes, and human volunteers. This further confirms its use on biological tissues, not in vitro samples.

In Vitro Diagnostics (IVDs) are devices intended for use in the examination of specimens derived from the human body (such as blood, urine, tissue samples) to provide information for diagnostic purposes. The TonoPach does not examine such specimens; it directly measures physical properties of the eye.

N/A

Intended Use / Indications for Use

The TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraocular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.

Product codes

HKX, 90-IYO

Device Description

The TonoPach is a portable, battery operated, handheld ophthalmic instrument which measures both intraocular pressure (IOP) and corneal thickness (CT) simultaneously and at the same locus on the cornea. The device uses the principles of applanation pneumatometery to measure IOP and the principles of reflection pulsed ultrasound pachymetry to measure CT. The system is comprised of the TonoPach electronics, transducer handpiece, protective membranes and optionally, a foot pedal.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Cornea, Eye

Indicated Patient Age Range

Not Found

Intended User / Care Setting

office, bedside or in remote locations, eye clinic volunteers

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Inanimate:

  • Tonometry Model: water-filled spherical membrane attached to a manometer. The TonoPach handpiece tip touches the sphere to mimic intraocular pressure.
  • Pachymetry Model: polycarbonate wafers of varying thicknesses. The TonoPach handpiece tip touches a wet wafer to mimic the cornea for thickness measurement.

Animal:

  • Rabbit Eyes: Live rabbit eyes were used to test the tonometer and pachymeter portions of the TonoPach, evaluate performance on a live eye, and assist in evaluating numerical accuracy and reproducibility.
  • Porcine Eyes: Freshly enucleated porcine eyes were used for several studies, predominately to verify and validate tonometer accuracy and reproducibility. One study demonstrated repeatability with three porcine eyes studied at each 10 mmHg pressure over a range of 5 to 80 mmHg with 10 independent measurements per pressure.

Clinical Trial:

  • Participants: Eye clinic volunteers who required tonometry. All patients who agreed to volunteer were admitted to the study.
  • Sample Size: 347 eyes were tested for tonometry.
  • Protocol: Tonometry testing was the predominant data collected, compared to Goldmann tonometry.

Summary of Performance Studies

Study Type: Inanimate, Animal (Rabbit and Pig), and Clinical Trial (Human) evaluations.

Inanimate Testing:

  • Tonometry: Used a water-filled spherical membrane attached to a manometer to mimic intraocular pressure.
  • Pachymetry: Used polycarbonate wafers of varying thicknesses touched by the TonoPach handpiece tip, mimicking the cornea for thickness measurement.

Animal Testing:

  • Rabbit Eyes: Tested tonometer and pachymeter portions on live eyes to evaluate performance, numerical accuracy, and reproducibility.
  • Porcine Eyes: Primarily used for tonometer accuracy and reproducibility verification and validation. A repeatability study involved 3 porcine eyes, tested at each 10 mmHg pressure from 5 to 80 mmHg with 10 independent measurements per pressure.

Clinical Trial:

  • Sample Size: 347 eyes.
  • Study Type: Comparison of TonoPach tonometry to Goldmann tonometry in eye clinic volunteers.
  • Key Results: Scatter and Bland & Altman plots, mean, standard deviation, and variance around the mean reveal that the TonoPach meets or exceeds specifications.

Key Metrics

Not Found

Predicate Device(s)

K 852774, K993674, K020876

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 886.1930 Tonometer and accessories.

(a)
Identification. A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.(b)
Classification. Class II.

0

K042099

510(k) SUMMARY

1Submitter (Contact) Information
ACompany Name/Address:RetinaPharma Technologies, Inc.
944 Morgan Road
Jenkintown, PA 19046
BCompany Phone:(215) 885-4558
CCompany Fax:(215) 885-2788
DContact Person:Terry A. Fuller, Ph.D.
President
EManufacturing Location:The Company has not established a manufacturing facility.
It will notify the Agency upon its establishment.
  • 2 Device Information
    A Device Trade Name: TonoPach™ Ultrasonic Tonometer/Pachymeter Model P-201 (the "TonoPach™")

B Device Classification:

The TonoPach is a combination of two devices, a pneumatometer and a pachymeter, which have already been determined to be substantially equivalent to legally marketed predicate devices with regard to safety, effectiveness and intended use. The device classifications are presented in the following table.

PneumatometerPachymeter
Common Name:Tonometer for Measuring
Intraocular PressureUltrasonic Pachymeter
for Measurement of
Corneal Thickness
Classification Name:Tonometer, AC PoweredUltrasonic Pulsed Echo
Imaging System
Product Code:HKX90-IYO
Regulatory Class & Tier:IIII
Regulatory Number:21 CFR 866.193021 CFR 892.1560
Classification & Review
Advisory CommitteesOphthalmicRadiology

3 Substantial Equivalence

The RetinaPharma TonoPach Ultrasonic Tonometer/Pachymeter is judged to be substantially equivalent in safety, effectiveness and intended use to the following legally marketed devices:

1

Predicate DeviceManufacturer510(k)Date Cleared
TonoPen ophthalmic
tonometerOculab, Inc.K 852774August 20, 1985
Pocket Ultrasonic
PachymeterQuantel Medical, Inc.K993674July 13, 2000
Echoscan Model US-1800Nidek, Inc.K020876May 3, 2002

I ist of predicate devices

Device Description. ব

The TonoPach is a portable, battery operated, handheld ophthalmic instrument which measures both intraocular pressure (IOP) and corneal thickness (CT) simultaneously and at the same locus on the cornea. The device uses the principles of applanation pneumatometery to measure IOP and the principles of reflection pulsed ultrasound pachymetry to measure CT. The system is comprised of the TonoPach electronics, transducer handpiece, protective membranes and optionally, a foot pedal.

ട Intended Use

The TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraocular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.

6 Technological Equivalence

Table 6.1 provides a list of predicate devices and the "k" number. In Tables 6.2, 6.3 and 6.4 the TonoPach Pachymeter portion and the Ultrasonic Tonometer are compared, respectively, to predicate devices to show the equivalence. Description statements were relied on to ascertain the intended use and technological features of legally marketed devices, and the substantial equivalence to the TonoPach to such legally marketed devices. The comparison of the intended use and technological features of this device to other legally marketed devices indicates that this device is substantially equivalent to legally marketed predicate devices with regard to safety, effectiveness and intended use. Note: areas of difference between the two devices are indicated in the tables. RetinaPharma believes that these differences are minor and should not raise any concerns regarding the overall safety and efficacy of the TonoPach.

Table 6.1: List of predicate devices
Predicate DeviceManufacturer510(k)Date Cleared
Tonometer
TonoPen ophthalmic tonometerOculab, Inc.K 852774August 20, 1985
Pachymeter
Pocket Ultrasonic Pachymeter
Echoscan Model US-1800Quantel Medical, Inc.
Nidek, Inc.K993674
K020876July 13, 2000
May 3, 2002

2

Table 6.2: Comparison of Tonometer Characteristics
TonoPen vs. TonoPach P 201
Predicate Device
Mentor TonoPen 3RetinaPharma
TonoPach P 201
MeasurementIOPIOP
Pressure measurement deviceMicro strain gaugeMicro strain gauge
Measurement techniqueApplanationApplanation
Pressure range5-80 mm Hg5-80 mm Hg
Pressure contact area*1.5 mm1.8 mm
Measurement frequency500 reading per second500 readings per second
Display/measureLCD / 2 digit readingLCD / 2 digit reading
Statistical storageYesYes
Pressure traceabilityManometer & GoldmannManometer & Goldmann
UsageHandheldHandheld or Slitlamp mounted
Power supplyBatteryBattery
System configuration*Single unit; transducer and
system electronics combinedTethered unit; transducer is
tethered from the system
electronics
VersatilityCan be used with the patient in
any position, making it suitable
for use in the office, bedside or in
remote locations. Suitable for use
in cases of their regular or Hg
corneal astigmatism.Can be used with the patient in
any position, making it suitable
for use in the office, bedside or in
remote locations. Suitable for us
in cases of their regular or Hg
corneal astigmatism.
*Indicates differences between the predicate device and TonoPach

| Table 6.3: Comparison of Pachymeter Characteristics

"Pocket" Ultrasonic Pachymeter and Echoscan US-1800 vs. TonoPach P 201
Predicate Device
"Pocket"Predicate Device
Echoscan US-1800RetinaPharma
TonoPach P 201
MeasurementCorneal thicknessCorneal thicknessCorneal thickness
Measurement systemUltrasonic pulse
reflectionUltrasonic pulse
reflectionUltrasonic pulse
reflection
Probe typeSolid,
AngledSolid,
Straight, AngledSolid
Straight, Angled
Measurement range*100 to 1300 um200 to 1300 um125 to 1300 um
Clinical accuracy+/- 5 um+/- 5 um+/- 5 um
Least significant display
digit1 um1 um1 um
Default velocity*1620 m/sec1640 m/sec,
changeable1640 m/sec
Exam Mode(s)Single point, 1 mapSingle point, 3 mapsSingle point, 1 map
Data OutputLCD, PrintLCD, PrintLCD, Print
Power requirementsBattery120 VAC, 60 HzBattery
Weight0.46 kg6 kg0.75 kg
*Pachymeter mode. Indicates difference between the predicate device and the TonoPach

3

7 TonoPach Testing

Substantial testing has been performed using the TonoPach. It includes inanimate, animal and human evaluation.

  • A Inanimate: Models for both tonometry and pachymetry have been used.
    • l ) The tonometry model consists of a water-filled spherical membrane attached to a manometer. When the TonoPach handpiece tip touches the sphere, it mimics intraocular pressure.
      1. The pachymetry model consists of polycarbonate wafers of varying thicknesses. When the TonoPach handpiece tip touches a wet wafer, it is suitable to mimic the cornea for thickness measurement purposes.
  • B Animal: Live rabbit eyes and freshly enucleated porcine eyes were used for several studies.
    • l) Rabbit eyes were used to test the tonometer and pachymeter portions of the TonoPach. They were used to evaluate the TonoPach's performance on a live eye. Also, they were used to assist in evaluation of the System's numerical accuracy and reproducibility.
      1. Pig eyes were used predominately to verify and validate tonometer accuracy and reproducibility. One study was used to demonstrate repeatability. Three porcine eyes were studied at each 10 mmHg pressure over a pressure range of 5 to 80 mmHg with 10 independent measurements per pressure. The data exceeds product specifications.
  • C Clinical Crial: The TonoPach was evaluated on eye clinic volunteers who required tonometery. All patients who agreed to volunteer were admitted to the study.
      1. Tonometry testing was the predominated the data collected. 347 eyes were tested for comparison to Goldmann tonometery. Scatter and Bland & Altman plots as well as mean, standard deviation and variance around the mean reveal that the TonoPach meets or exceeds specifications.

8 Statement of Confidentiality

Information contained in this letter and its attachments is considered confidential, proprietary information and a trade secret until RetinaPharma Technologies, Inc. distributes such information publicly. The release of information contained herein is to be governed by applicable provisions of the Food, Drug and Cosmetic Act, the Freedom of Information Act and pertinent FDA regulations in 21 CFR Part 20.

4

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 0 2004

RetinaPharma Technologies, Inc. % Terry Fuller, Ph.D. 944 Morgan Road Jenkintown, PA 19046

Re: K042099

Trade/Device Name: TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and accessories Regulatory Class: Class II Product Code: HKX Dated: July 30, 2004 Received: August 4, 2004

Dear Dr. Fuller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

5

Page 2 - Terry Fuller

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A halyl Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

3) Diagnostic Ultrasound Indications for Use Form

Page 1 of 1

510(k) Number (if known):

Device name: RetinaPharma Model P 201, TonoPachTM System

Intended Use: The TonoPach Ultrasonic TonometerPachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraccular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.

Clinical ApplicationMode of Operation
ABMPWDCWDColor
DopplerAmplitude
DopplerColor
Velocity
ImagingCombined
(specify)Other
(specify)
OphthalmicN
Fetal
Abdominal
Intraoperative (specify)
Intraoperative
Neurological
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral Vascular
Laparoscopic
Musculo-skeletal
Conventional
Musculo-skeletal
Superficial
Other (specify)*

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Clay R. Butterman
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises

510(k) TonoPach Indications for Use modified 041210.doc
From 510(k) 040729_Cover D.doc

510(k) NumberK042099
------------------------