K Number
K041781
Date Cleared
2005-03-14

(256 days)

Product Code
Regulation Number
868.6250
Panel
AN
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Hamilton Medical VENTILAIR II is intended for use in hospitals and other institutions as an alternative to central air or cylinder air to supply medical air at 30 psi nominal to the GALILEO and the RAPHAEL models of Hamilton Medical ventilators which are intended to be operated with 30 psi nominal medical air supply. It can also be used as a backup medical air supply for these ventilators in case of failure of their normal air supply.

Device Description

The VENTILAIR II is an electromechanical and preumatic device only and has no software-based components. The device is rated at 115 V ± 10 %, 60 Hertz. It requires 466 Watts to power. The enclosure is designed to be mounted either on the GALILEO model or RAPHAEL model ventilator carts. Only one control is accessed during normal device operation, a power switch. The Power switch is switched ON both when the VENTILAIR II is configured and operated as a permanent air supply and when it is used as an emergency backup air supply. An indicator is provided to inform the user of the device performance status. This "Performance Gauge" is a pressure gauge calibrated for three zones of operation: Green (Normal Dry Operation), Yellow (Poor Drying Effect), and Red (Wet with Low Output Pressure). The VENTILAIR II has two air movements. The first is the cool, clean, dry compressed air movement that provides condensation-free compressed air to the ventilator. The second is the ambient air movement - an S-pattern that cools the entire drying system and compressor within the enclosure. The VENTILAIR II has been designed for easy maintenance. The compressor air intake filter element must be checked weekly. The 5 micron filter in the water trap must be replaced annually. Two internal pressure requlators and two other internal components must be checked annually and adjusted if necessary. Compressor overhaul is required periodically using the compressor manufacturer's overhaul kit. If the VENTILAIR II is used as an independent or continuous air source, an annual overhaul is recommended. If the compressor is used only as a backup in case of failure of the main air supply, overhaul is recommended every five years. All maintenance and overhaul procedures are described in the VENTILAIR II Operator's and Service Manuals. The VENTILAIR II is in full compliance with all required safety aspects of the current edition of IEC 60601-1. Several specific safety mechanisms have been incorporated into the VENTILAIR 11.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study information for the Hamilton Medical VENTILAIR II Medical Air Compressor:

The provided text is a 510(k) summary for the Hamilton Medical VENTILAIR II Medical Air Compressor. The purpose of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to an already legally marketed device (predicate device). This typically involves comparing characteristics and performance data. The document does not describe acceptance criteria in the sense of predefined thresholds for a study to pass, but rather presents performance characteristics of the VENTILAIR II and compares them to predicate devices to establish substantial equivalence.

It's important to note that a 510(k) summary often summarizes testing, and a detailed study protocol with explicit acceptance criteria might be in the full submission, but not fully reproduced here.

1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a substantial equivalence submission, the "acceptance criteria" are implicitly the performance characteristics of the predicate devices, and the "reported device performance" is that of the VENTILAIR II. The goal is to show the VENTILAIR II performs as well as or better than the predicates for the relevant characteristics.

CharacteristicPredicate Device Performance Examples (Range across predicates)VENTILAIR II PerformanceImplicit Acceptance Criteria (Substantial Equivalence)
Intended UseSource of compressed air for various medical devices, including ventilatorsIndependent/backup 30 psi compressed air source for GALILEO and RAPHAEL ventilatorsComparable intended use (providing compressed air for medical ventilators).
FilteringUnspecified to $\le$ 1 micron (Pre-compression HEPA, water trap filter)$\le$ 5 microns at output; 0.5 micron filter in ventilator inputEffective filtration to ensure clean air supply, comparable to or better than predicates. The 5 micron output, coupled with the ventilator's 0.5 micron filter, likely deemed equivalent or superior.
Water TrapYes, with automatic vaporizationYes, with automatic vaporizationPresence and function of water trap for removing moisture, comparable to predicates.
Pressure Relief ValveSet to 59 PSIG - 85 PSIGSet to 32 PSIGPresence and functionality of a pressure relief valve for safety. The set point differs, but is appropriate for its intended output pressure.
Auto Backup Pressure (On)Not available, 40.5 PSIG, or optional at 45 PSIGOn at 30 PSIGFunctionality for automatic backup. Set point appropriately aligned with intended operating pressure (30 psi nominal).
Auto Backup Pressure (Off)50 PSIG or 55 PSIGOff at 38 PSIGFunctionality for automatic backup. Set point appropriately aligned with intended operating pressure (30 psi nominal).
Electrical Rating (Voltage)102 V - 132 V115 V ± 10 %Compatibility with standard electrical outlets for medical equipment.
Electrical Rating (Hertz)60 Hertz60 HertzCompatibility with standard electrical frequencies.
Electrical Rating (Power)Unspecified to 500 W466 WReasonable power consumption for the device's function.
Overcurrent ProtectionUnspecified, Fuse, or Circuit breakerFusePresence of safety measures for electrical overcurrents.
Output30 l/min at 44 PSIG; 45 l/min at 50 PSIG; 55 l/min at 50 PSIG40 l/min at 30 PSIGSufficient air output capacity for its intended use (driving Hamilton ventilators). The pressure output (30 PSIG) is specific to the stated operating pressure of GALILEO and RAPHAEL ventilators.
Dew Point DepressionUnspecified, 5 °C below room temperature at 30 l/min, > 3 °F below room temperature at 55 l/min3.3 °C below room temperature at 40 l/minEffective drying capability to prevent condensation and ensure quality medical air.
Input/Output InterfaceVarious DISS fittings¾" - 16 male DISS fitting with internal check valve (1 input, 1 output)Standardized and safe connection for air supply.
Ambient Environment (Temp)20 to 104 °F50 to 104 °FOperability within typical clinical environmental conditions.
Ambient Environment (RH)30 to 99 % non-condensing RH

§ 868.6250 Portable air compressor.

(a)
Identification. A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.