(256 days)
The Hamilton Medical VENTILAIR II is intended for use in hospitals and other institutions as an alternative to central air or cylinder air to supply medical air at 30 psi nominal to the GALILEO and the RAPHAEL models of Hamilton Medical ventilators which are intended to be operated with 30 psi nominal medical air supply. It can also be used as a backup medical air supply for these ventilators in case of failure of their normal air supply.
The VENTILAIR II is an electromechanical and preumatic device only and has no software-based components. The device is rated at 115 V ± 10 %, 60 Hertz. It requires 466 Watts to power. The enclosure is designed to be mounted either on the GALILEO model or RAPHAEL model ventilator carts. Only one control is accessed during normal device operation, a power switch. The Power switch is switched ON both when the VENTILAIR II is configured and operated as a permanent air supply and when it is used as an emergency backup air supply. An indicator is provided to inform the user of the device performance status. This "Performance Gauge" is a pressure gauge calibrated for three zones of operation: Green (Normal Dry Operation), Yellow (Poor Drying Effect), and Red (Wet with Low Output Pressure). The VENTILAIR II has two air movements. The first is the cool, clean, dry compressed air movement that provides condensation-free compressed air to the ventilator. The second is the ambient air movement - an S-pattern that cools the entire drying system and compressor within the enclosure. The VENTILAIR II has been designed for easy maintenance. The compressor air intake filter element must be checked weekly. The 5 micron filter in the water trap must be replaced annually. Two internal pressure requlators and two other internal components must be checked annually and adjusted if necessary. Compressor overhaul is required periodically using the compressor manufacturer's overhaul kit. If the VENTILAIR II is used as an independent or continuous air source, an annual overhaul is recommended. If the compressor is used only as a backup in case of failure of the main air supply, overhaul is recommended every five years. All maintenance and overhaul procedures are described in the VENTILAIR II Operator's and Service Manuals. The VENTILAIR II is in full compliance with all required safety aspects of the current edition of IEC 60601-1. Several specific safety mechanisms have been incorporated into the VENTILAIR 11.
Here's an analysis of the provided text to extract the acceptance criteria and study information for the Hamilton Medical VENTILAIR II Medical Air Compressor:
The provided text is a 510(k) summary for the Hamilton Medical VENTILAIR II Medical Air Compressor. The purpose of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to an already legally marketed device (predicate device). This typically involves comparing characteristics and performance data. The document does not describe acceptance criteria in the sense of predefined thresholds for a study to pass, but rather presents performance characteristics of the VENTILAIR II and compares them to predicate devices to establish substantial equivalence.
It's important to note that a 510(k) summary often summarizes testing, and a detailed study protocol with explicit acceptance criteria might be in the full submission, but not fully reproduced here.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a substantial equivalence submission, the "acceptance criteria" are implicitly the performance characteristics of the predicate devices, and the "reported device performance" is that of the VENTILAIR II. The goal is to show the VENTILAIR II performs as well as or better than the predicates for the relevant characteristics.
| Characteristic | Predicate Device Performance Examples (Range across predicates) | VENTILAIR II Performance | Implicit Acceptance Criteria (Substantial Equivalence) |
|---|---|---|---|
| Intended Use | Source of compressed air for various medical devices, including ventilators | Independent/backup 30 psi compressed air source for GALILEO and RAPHAEL ventilators | Comparable intended use (providing compressed air for medical ventilators). |
| Filtering | Unspecified to $\le$ 1 micron (Pre-compression HEPA, water trap filter) | $\le$ 5 microns at output; 0.5 micron filter in ventilator input | Effective filtration to ensure clean air supply, comparable to or better than predicates. The 5 micron output, coupled with the ventilator's 0.5 micron filter, likely deemed equivalent or superior. |
| Water Trap | Yes, with automatic vaporization | Yes, with automatic vaporization | Presence and function of water trap for removing moisture, comparable to predicates. |
| Pressure Relief Valve | Set to 59 PSIG - 85 PSIG | Set to 32 PSIG | Presence and functionality of a pressure relief valve for safety. The set point differs, but is appropriate for its intended output pressure. |
| Auto Backup Pressure (On) | Not available, 40.5 PSIG, or optional at 45 PSIG | On at 30 PSIG | Functionality for automatic backup. Set point appropriately aligned with intended operating pressure (30 psi nominal). |
| Auto Backup Pressure (Off) | 50 PSIG or 55 PSIG | Off at 38 PSIG | Functionality for automatic backup. Set point appropriately aligned with intended operating pressure (30 psi nominal). |
| Electrical Rating (Voltage) | 102 V - 132 V | 115 V $\pm$ 10 % | Compatibility with standard electrical outlets for medical equipment. |
| Electrical Rating (Hertz) | 60 Hertz | 60 Hertz | Compatibility with standard electrical frequencies. |
| Electrical Rating (Power) | Unspecified to 500 W | 466 W | Reasonable power consumption for the device's function. |
| Overcurrent Protection | Unspecified, Fuse, or Circuit breaker | Fuse | Presence of safety measures for electrical overcurrents. |
| Output | 30 l/min at 44 PSIG; 45 l/min at 50 PSIG; 55 l/min at 50 PSIG | 40 l/min at 30 PSIG | Sufficient air output capacity for its intended use (driving Hamilton ventilators). The pressure output (30 PSIG) is specific to the stated operating pressure of GALILEO and RAPHAEL ventilators. |
| Dew Point Depression | Unspecified, 5 °C below room temperature at 30 l/min, > 3 °F below room temperature at 55 l/min | 3.3 °C below room temperature at 40 l/min | Effective drying capability to prevent condensation and ensure quality medical air. |
| Input/Output Interface | Various DISS fittings | ¾" - 16 male DISS fitting with internal check valve (1 input, 1 output) | Standardized and safe connection for air supply. |
| Ambient Environment (Temp) | 20 to 104 °F | 50 to 104 °F | Operability within typical clinical environmental conditions. |
| Ambient Environment (RH) | 30 to 99 % non-condensing RH | < 85 % RH | Operability within typical clinical environmental conditions regarding humidity. |
| Ambient Environment (Altitude) | Unspecified to 13,000 ft max. altitude | 7,200 ft maximum altitude | Operability across a reasonable range of altitudes. |
| Noise Level | 46 to < 55 dB(A) | < 50 dB(A) | Noise levels acceptable for a clinical environment. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a "sample size" in the context of patients or individual devices tested in a clinical trial. The testing described is "laboratory performance verification and validation test data," which implies engineering and functional testing on units of the device.
- Test Set Sample Size: Not specified as a number of devices or subjects. It is implied that at least one Ventilair II unit was subjected to the listed performance tests.
- Data Provenance: The testing appears to be retrospective in the sense that it evaluates the performance of a manufactured device according to established engineering standards and comparisons. The location of the testing is not explicitly stated but is implicitly associated with Hamilton Medical AG (Switzerland) or its US subsidiary (Reno, NV). It is laboratory-based testing, not clinical data from patients in a specific country.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This section is not applicable as the described testing involves objective engineering and performance verification against known specifications and predicate device characteristics, not expert interpretation of medical images or patient data to establish "ground truth." There is no mention of human experts defining outcomes for test data in this context.
4. Adjudication Method for the Test Set
This section is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies where multiple human readers interpret data, and discrepancies need to be resolved. The testing described here is objective measurement of physical and electrical properties.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This type of study involves multiple human readers evaluating cases, often with and without AI assistance, to measure diagnostic performance. The VENTILAIR II is a medical air compressor, not a diagnostic imaging device or an AI-powered analytical tool, so such a study would not be relevant.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
No, a standalone algorithm performance study was not done. The VENTILAIR II is an electromechanical and pneumatic device with no software-based components, let alone an AI algorithm. Its performance is evaluated through direct measurement of its physical output and adherence to safety standards.
7. The Type of Ground Truth Used
The "ground truth" for the performance evaluation of the VENTILAIR II is based on:
- Engineering specifications and standards: Compliance with IEC 60601-1 and other relevant safety and performance standards.
- Physical measurements: Direct measurements of output pressure, flow rate, dew point, noise level, electrical ratings, etc.
- Comparison to predicate device characteristics: The performance data of the legally marketed predicate devices (Draeger Medical Air Compressor, Infrasonics Air Star Portable Compressor, Bennett MC-2 Mobile Air Compresor) serve as the benchmark for demonstrating substantial equivalence.
8. The Sample Size for the Training Set
This section is not applicable. The VENTILAIR II is an electromechanical device with no AI or machine learning components requiring a "training set" of data.
9. How the Ground Truth for the Training Set was Established
This section is not applicable for the same reason as point 8.
{0}------------------------------------------------
| Comments of the contribution of the comments of the minimal concerner of the contraction of the consistence of the consistence of the consistence of the consistence in the co12.00 15.00 | からいつけられるということです。 そして、そこではないというというというというというということです。.AND STATE CALLERY | ||
|---|---|---|---|
| Doc .- Title | ------------------------------------------Ventilair II Section | Joc .- Version ."White the different to the research futured in the first to be the first toCalled College Complete Comments of Children Company Controlled Collection Controller Company of Concello Come of Concerner of Concerner of Concerner of Concerner of Concerne |
510(K) SUMMARY 9
-
- Summary Preparation Date: January 18, 2005
-
- Applicant Information: Name: Hamilton Medical AG Via Nova CH-7403 Rhäzüns Switzerland FDA Establishmen Registration Number: 3001421318 Contact Person:J.. David Thompson, General Manager Hamilton Medical Inc. P.O.Box 30008 Reno, NV 89502 Phone 775-858-3200 Fax 775-856-5621 Email: thompson@hammed1.com
-
- Device Proprietary Name: VENTILAIR II Medical Air Compressor
Common/Usual Name: Air Compressor Classification Name: Compressor. Air. Portable Classification Panel: Anesthesiology Classification Code: BTI
-
- BTI Device Identification Code (per 21 SFR Part 868.6250): A portable air compressor is a device to provide comopressed air for medical purposes, e.g. to drive ventilators and other respiratory devices.
-
- Regulatory Status: Portable air compressors and their accessories (FDA product code BTI) have been classified by the FDA as class II. There are currently no mandatory performance standards or special control requirements for these devices.
-
- Intended Use: The Hamilton Medical VENTILAIR II is intended for use in hospitals and other institutions as an alternative to central air or cylinder air to supply medical air at 30 psi nominal to the GALILEO and the RAPHAEL models of Hamilton Medical ventilators which are intended to be operated with 30 psi nominal medical air supply. It can also be used as a backup medical air supply for these ventilators in case of failure of their normal air supply.
-
- General Device Description: The VENTILAIR II is an electromechanical and preumatic device only and has no software-based components. The device is rated at 115 V ± 10 %, 60 Hertz. It requires 466 Watts to power. The enclosure is designed to be mounted either on the GALILEO model or RAPHAEL model ventilator carts.
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A AN AND A LIES & LINE & LEASE OFAND AND COLLEGE & Access of | Carling of Children College of the | |
|---|---|---|
| 44 5 611.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 800 7 | 2004-03-11 |
| attribution Cadder PartyCALL Co., Co., Canada Co., LEANHOUSE A. F. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. | 10.00THE COLLECTION CALL CALL CALL CALL4 7 4 8 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 | |
| .a new den looked for logically in religion m infrastrial program information of the interest of the interest of the interest of the states of the story of the complete of theclient of BaNds, ac n I (" in a "if("(" ) " = "("("\["+")")")")")")")")")")")")")")")")")")")")")")")")")")")")")")")")")""(""""""""""""""""""""""""""""""""""""""""""""""""----------------------------------------------------------------------------------------------A F E MALL PR . 19 AS, PAR 15 PAR 15 Comment1Acres Courts A. B | Carles Carder Problem Company Comparison. In In Intelligence and a viames Children I stampen below. | |
| I | September 19, 2017 11:40 PM |
{1}------------------------------------------------
| Pinters Manne - Mentilation - HAMLTON MEDCALAG - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -AMILTON MEDICAEAVentilair IF 510 Kr Sur-1130.000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | a consult of the last to | ||
|---|---|---|---|
| Doc .- TitleCallery of Concession | --- " xaction 9: 510(k) SummarvVontilair II . | Onc.-Varsion |
Only one control is accessed during normal device operation, a power switch. The Power switch is switched ON both when the VENTILAIR II is configured and operated as a permanent air supply and when it is used as an emergency backup air supply. An indicator is provided to inform the user of the device performance status. This "Performance Gauge" is a pressure gauge calibrated for three zones of operation: Green (Normal Dry Operation), Yellow (Poor Drying Effect), and Red (Wet with Low Output Pressure).
The VENTILAIR II has two air movements. The first is the cool, clean, dry compressed air movement that provides condensation-free compressed air to the ventilator. The second is the ambient air movement - an S-pattern that cools the entire drying system and compressor within the enclosure.
The VENTILAIR II has been designed for easy maintenance. The compressor air intake filter element must be checked weekly. The 5 micron filter in the water trap must be replaced annually. Two internal pressure requlators and two other internal components must be checked annually and adjusted if necessary. Compressor overhaul is required periodically using the compressor manufacturer's overhaul kit. If the VENTILAIR II is used as an independent or continuous air source, an annual overhaul is recommended. If the compressor is used only as a backup in case of failure of the main air supply, overhaul is recommended every five years. All maintenance and overhaul procedures are described in the VENTILAIR II Operator's and Service Manuals.
The VENTILAIR II is in full compliance with all required safety aspects of the current edition of IEC 60601-1. Several specific safety mechanisms have been incorporated into the VENTILAIR 11.
-
- Device Materials: All materials within the VENTILAIR II flow path have been carefully selected to avoid any contamination of the medical air flow path. There are no plastic mold release compounds and no outgassed toxic material present to contaminate the gas flow. A 5 micron filter has been built into the device to capture any particulate impurities. In addition all Hamilton Medical GALILEO and RAPHAEL ventilators have a 0.5 micron filter built into their air inputs to trap any particulate impurities.
Among the information presented in the 510(k) submission to support the equivalency of the VENTILAIR II to these predicate devices is: (a.) device description, (b.) comparison to the legally marketed predicate devices, and (c.) laboratory performance verification and validation test data.
-
- Comparison Table: The table that follows describes some Important characteristics of the VENTILAIR II Medical Air Compressor and its predicate devices: Draeger Medical Air Compressor (K982789), Infrasonics Air Star Portable Compressor (K920954), and Bennett MC-2 Mobile Air Compressor (Pre-Amendment).
| Unio | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------SUFFFFFFULLS |
|---|---|---|
| 1000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 0000 0000 0000 0000 00000000000000000000000000000000000000000000000000000000000004.60STANKS AL I SELECT | Comments of the country of the results of the production of the comments of the production of the----------------------------------------------------------------------.Machel ad 1 1 1 momen a 10 to as 6 a 1 1 2 . 1 .a contract and a season that man and an a magazine rates ab-Contraction of the many would be the love the 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10,. | -------------------------------------------------------------------------------------------------------------------------------- |
{2}------------------------------------------------
| www.rangement.com/sport/supply/supply.com/sport/ | Project Name | ||
|---|---|---|---|
| MALERIAN & AN AND AND AND AND A FEﻟ | Minin It Cantler Or Eાં ની રહ્યપાય | A MARK MANAGER A BRANCH A BREASE A BREASE AND A BALL BALL BE BE BERESS AND | |
| LOC .- PatriB B B B B B | .D:\FDA KORRESPONDENZIVENTILAIR 510(K) RESPONSELE37209------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Called of Children and Children and Children and Children and Children of |
omparison to Legally Marketed Predicate Devices
| # | Characteristic | Bennett MC-2 Air Compressor (Pre-Amendment) | PREDICATE DEVICESDraeger Medical Air Compressor (K982789) | Infrasonics Air Star Compressor (K920954) | VENTILAIR IISimilar or Different(From All Other Predicates Combined) |
|---|---|---|---|---|---|
| 1 | Intended use and applications | Source of compressed air for Bennet IPPB therapy units, respiration units and other devices requiring clean filtered air | Independent or backup compressed air source to power medical ICU ventilators | Independent or backup compressed air source to power Infrasonics Infant Star and other ventilators | Independent/backup 30 psi compressed air source to power GALILEO and RAPHAEL ventilators in hospitals and other institutions at 30 psi nominal operating pressure |
| 2 | User controls/visual indicators | Power switch / pressure gauge | Power switch / power LED, high air temperature LED | Power switch / pressure gauge, power LED, low pressure LED, compressor on LED and wall air on LED | Power switch / pressure performance gauge, power failure and low pressure indicators built into ventilator |
| 3 | Alarms | Audible and visual power failure and low pressure built into ventilator (respiration unit) | Audible and visual high output air temperature | Audible and visual power failure and low pressure | Audible and visual power failure and low pressure indicators built into ventilator |
| 4 | Filtering | Pre-compression HEPA filter; water trap filter (size unspecified in available labeling) | $≤$ 1 micron | Degree of filtering unspecified in available labeling | $≤$ 5 microns at output of device; 0.5 micron filter built into ventilator input |
| 5 | Water trap | Yes, with automatic vaporization | Yes, with automatic vaporization | Yes, with automatic vaporization | Yes, with automatic vaporization |
| 6 | Pressure relief valve | Yes, set to 60 PSIG | Yes, set to 59 PSIG | Yes, set to 85 PSIG | Yes, set to 32 PSIG |
| 7 | Auto backup pressure | Feature not available | On at 40.5 PSIG, off at 50 PSIG | Optional featue, on at 45 PSIG, off at 55 PSIG | On at 30 PSIG, off at 38 PSIG |
H. Pohl
© Copylgitty Hamliton MEDICALAGAGA CARACTER CARACHER CONFECTION - CONFERTRAULICHICONFIDENTAL
{3}------------------------------------------------
| Bennett MC-2 AirCompressor (Pre-Amendment) | PREDICATE DEVICESDraeger Medical AirCompressor (K982789) | Infrasonics Air StarCompressor (K920954) | VENTILAIR IISimilar or Different(From All OtherPredicates Combined) | |
|---|---|---|---|---|
| Characteristic | ||||
| Electrical ratings | 115 V; 60 Hertz; powerconsumption unspecifiedin available labeling | 110 V; 60 Hertz; 500 W | 102 to 132 V; 60 Hertz;power consumptionunspecified in availablelabeling | 115 ± 10 %; 60 Hertz;466 W |
| Overcurrentprotection | Unspecified in availablelabeling | Fuse | Circuit breaker | Fuse |
| Output | 45 l/min at 50 PSIG | 30 l/min at 44 PSIG | 55 l/min at 50 PSIG | 40 l/min at 30 PSIG |
| Dew pointdepression | Unspecified in availablelabeling | 5 °C below roomtemperature at 30 l/min | > 3 °F below roomtemperature at 55 l/min | 3.3 °C below roomtemperture at 40 l/min |
| Input/outputinterface | Two DISS 9/16" - 18threaded outlets (onecapped when not in use) | Quick connect DISScoupling with internalcheck valve (1 input, 1output) | ¾" - 16 male DISSfitting with internal checkvalve (optional input, 1output) | ¾" - 16 male DISS fittingwith internal check valve(1 input, 1 output) |
| Ambientenvironment | Unspecified in availablelabeling | 50 to 104 °F; 30 to 95 %RH, 13,000 ft max.altitude | 20 to 95 °F; ≤ 99 %noncondensing RH;altitude unspecified inavailable labeling | 50 to 104 °F; < 85 % RH;7,200 ft maximum altitude |
| Noise level | Unspecified in availablelabeling | 46 to 49 dB(A) | < 55 dB(A) | < 50 dB(A) |
| Call Concess and Carder of Children | . A fi fi fine for a finance and a minister a a comments of a province and comments of the more and |
|---|---|
| A Property A LEA CONSULT CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSUL | |
| - American American American Comments of Children Children Children Children Children Children Children Children Children Children Children Children Children Children Childre | |
| Commender of Acres of Acres of Canadian Company | |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
| A Children Art | |
| © Copyright by Hamliton MEDICAL AG A CHARLE CONSULTION CONSULTION VERTRAULICH CONFIDENTAL | |
l
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
...... .. ...
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, with three figures representing health, services, and human aspects. The seal is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 4 2005
Hamilton Medical AG C/O Mr. J. David Thompson General Manager Hamilton Medical, Incorporated P.O. Box 30008 Reno, Nevada 89502
Re: K041781
Trade/Device Name: Hamilton Medical VENTILAIR II Medical Air Compressor Regulation Number: 868.6250 Regulation Name: Portable Air Compressor Regulatory Class: II Product Code: BTI Dated: January 27, 2005 Received: January 28, 2005
Dear Mr. Thompson:
We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Fedcral Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Ecderal Register.
{5}------------------------------------------------
Page 2 - Mr. Thompson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| Project Name | Ventilair II 510(k) Submission |
|---|---|
| HAMILTON MEDICAL AG | |
| Doc.-Title: | Ventilair II Section 4: Indications for Use |
| Doc.-Version: | 1.1 |
| Doc.-No: E372047 |
INDICATION FOR USE STATEMENT 1
510(k) Number (if known) _____________________________________________________________________________________________________________________________________________________
Hamilton Medical VENTILAIR II Medical Air Compressor Device Name:
Indications For Use:
The Hamilton Medical VENTILAIR II is intended for use in hospitals and other institutions as an alternative to central air or cylinder air to supply medical air at 30 psi nominal to the GALILEO and the RAPHAEL models of Hamilton Medical ventilators which are intended to be operated with 30 psi nominal medical air supply. It can also be used as a backup medical air supply for these ventilators in case of failure of their normal air supply.
Prescription Use: × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aury Sukum
Sign-Off
of Anesthesiology. General Hospi
tiony. General Hospital, on Control, Dental Devices
Number: K041781
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
| H. Pohl |
|---|
| Page 2 of 2 |
| 2004-03-02 |
| Copyright by HAMILTON MEDICAL AG |
| CONFIDENTIAL |
§ 868.6250 Portable air compressor.
(a)
Identification. A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.