K Number
K040076
Device Name
ANSPACH ADVANCED MICRO SAW ATTACHMENTS
Date Cleared
2004-03-15

(61 days)

Product Code
Regulation Number
882.4310
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Anspach micro saw attachments are indicated for use with eMax electrical and Black Max, MicroMax pneumatic drill systems for surgical cutting, shaping and removal of bone, including bones of the spine and cranium.
Device Description
Keyless driver connects to surgical motor (or eMax Console - electric driver) and holds the micro saw head.. It reduces motor speed and drives the attached saw blade. There designs: Black Max, MicroMaxleMax and "stand alone" electric driver certified to: - UL 2901-1: Second Addition, Rev. 10/24/97 . - CSA-C22.2 No. 601.1- M90, including General Instruction No. 1 . - EN60601-1. Issue 2. (including A1. A2. A11-13) . Drivers are non-invasive and non-patient contact; contain no materials or substances which could be transferred to patient. Drivers cannot internally transfer materials or substances from motor. Drivers are reusable and can be cleaned with soap/water or isopropyl alcohol and can withstand temperatures of up to 278°(F) for up to 30 min. (static). Micro saw head attaches to keyless driver without tools. Head converts motor rotational drive to sagittal, oscillating or reciprocating motions. Heads may be surgically invasive. Unintentional contact may occur due to proximity of blade and tissues of surgical site. Heads are constructed of biocompatible materials and contain no substances which could be transferred to patient. Like drivers, heads cannot internally transfer substances from motor. Heads are reusable and can be cleaned with soap-water or isopropy! alcohol and can withstand temperatures up to 278°(F) for up to 30 min. (static). Saw blades are pre-sterilized, single-use, disposable bone cutting blades. Saw blades are sterilized to an appropriate SAL as specified by recognized standards for sterilization.
More Information

Not Found

Not Found

No
The device description focuses on mechanical components and their function (converting rotational drive to cutting motions). There is no mention of software, algorithms, data processing, or any terms related to AI/ML.

No.
A therapeutic device is one that treats a disease or condition. This device is a surgical tool used for cutting, shaping, and removing bone, which is part of a surgical procedure, not a therapy itself.

No
The device is described as a surgical tool used for cutting, shaping, and removing bone, which is a therapeutic function, not a diagnostic one.

No

The device description clearly details physical hardware components including drivers, saw heads, and saw blades, which are mechanical and not software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "surgical cutting, shaping and removal of bone, including bones of the spine and cranium." This is a surgical procedure performed directly on the patient's body.
  • Device Description: The device is a surgical tool (micro saw attachments, driver, and saw blades) used for physical manipulation of bone during surgery.
  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.

This device is clearly designed for surgical intervention, not for testing biological samples.

N/A

Intended Use / Indications for Use

Anspach micro saw attachments are indicated for use with eMax electrical and Black Max, MicroMax pneumatic drill systems for surgical cutting, shaping and removal of bone, including bones of the spine and cranium.

Product codes (comma separated list FDA assigned to the subject device)

HBE, HEC

Device Description

Keyless driver connects to surgical motor (or eMax Console - electric driver) and holds the micro saw head.. It reduces motor speed and drives the attached saw blade. There designs: Black Max, MicroMaxleMax and "stand alone" electric driver certified to: - UL 2901-1: Second Addition, Rev. 10/24/97 . - CSA-C22.2 No. 601.1- M90, including General Instruction No. 1 . - EN60601-1. Issue 2. (including A1. A2. A11-13) . Drivers are non-invasive and non-patient contact; contain no materials or substances which could be transferred to patient. Drivers cannot internally transfer materials or substances from motor. Drivers are reusable and can be cleaned with soap/water or isopropyl alcohol and can withstand temperatures of up to 278°(F) for up to 30 min. (static). Micro saw head attaches to keyless driver without tools. Head converts motor rotational drive to sagittal, oscillating or reciprocating motions. Heads may be surgically invasive. Unintentional contact may occur due to proximity of blade and tissues of surgical site. Heads are constructed of biocompatible materials and contain no substances which could be transferred to patient. Like drivers, heads cannot internally transfer substances from motor. Heads are reusable and can be cleaned with soap-water or isopropy! alcohol and can withstand temperatures up to 278°(F) for up to 30 min. (static). Saw blades are pre-sterilized, single-use, disposable bone cutting blades. Saw blades are sterilized to an appropriate SAL as specified by recognized standards for sterilization.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

bones of the spine and cranium

Indicated Patient Age Range

Not Found

Intended User / Care Setting

qualified and trained surgeons. with Anspach surgical drill motor systems

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Risk assessment has been accomplished for the micro saws. All identified risks and hazards associated with use of the micro saws were capable of being reduced to acceptable levels through design, process and/or labeling activities.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Anspach Micro Saws.

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.

(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).

0

K040076, 1 of 2

MAR 1 5 2004

Image /page/0/Picture/2 description: The image is a dark, blurry, and indistinct photograph. It is difficult to discern any specific objects or details due to the poor image quality. The overall impression is one of darkness and lack of clarity.

510{k} SUMMARY: -- Advanced Micro Saw Attachments

CLASSIFICATION : US Class II

PERFORMANCE STANDARDS:

No known performance standards.

DESIGN, MATERIALS AND OPERATION CHARACTERISTICS;

Keyless driver connects to surgical motor (or eMax Console - electric driver) and holds the micro saw head.. It reduces motor speed and drives the attached saw blade.

There designs: Black Max, MicroMaxleMax and "stand alone" electric driver certified to:

  • UL 2901-1: Second Addition, Rev. 10/24/97 .
  • CSA-C22.2 No. 601.1- M90, including General Instruction No. 1 .
  • EN60601-1. Issue 2. (including A1. A2. A11-13) .

Drivers are non-invasive and non-patient contact; contain no materials or substances which could be transferred to patient. Drivers cannot internally transfer materials or substances from motor. Drivers are reusable and can be cleaned with soap/water or isopropyl alcohol and can withstand temperatures of up to 278°(F) for up to 30 min. (static).

Micro saw head attaches to keyless driver without tools. Head converts motor rotational drive to sagittal, oscillating or reciprocating motions. Heads may be surgically invasive. Unintentional contact may occur due to proximity of blade and tissues of surgical site. Heads are constructed of biocompatible materials and contain no substances which could be transferred to patient. Like drivers, heads cannot internally transfer substances from motor. Heads are reusable and can be cleaned with soap-water or isopropy! alcohol and can withstand temperatures up to 278°(F) for up to 30 min. (static).

Saw blades are pre-sterilized, single-use, disposable bone cutting blades. Saw blades are sterilized to an appropriate SAL as specified by recognized standards for sterilization.

RISK ASSESSMENT:

Risk assessment has been accomplished for the micro saws. All identified risks and hazards associated with use of the micro saws were capable of being reduced to acceptable levels through design, process and/or labeling activities.

INDICATIONS / CONTRAINDICATIONS:

There are no specific indications or contraindications for use of micro saw attachments. Generally, micro saw attachments are indicated for use only by qualified and trained surgeons. with Anspach surgical drill motor systems, for surgical cutting, shaping and removal of bone, including bones of the spine and cranium.

1

K040076 2

CLEANING/STERILIZATION/MAINTENANCE

Cleaning, sterilization and maintenance instructions are provided in product use manuals available to all purchasers and Instructions (directions) For Use (IFU/DFU) that accompany each product. Unauthorized repairs and maintenance can be grounds for termination of warranty benefits.

WARNINGS and CAUTIONS:

Generic Warnings for use of Anspach products are specified in manuals and on product inserts. For safe and effective use of any Anspach product, specialized training is required as surgical techniques are highly specialized procedures. Improper surgical technique or improper use of product can cause injury or death to a user or patient and damage to product.

SUBSTANTIAL EQUIVALENCE:

Advanced micro saw attachments are substantially equivalent to Anspach Micro Saws.

OLD Micro SawNEW Micro Saw
SagittalReciprocatingOscillatingSagittalReciprocatingOscillating
Use w/Electric motorYYYYYY
Use w/Pneumatic motorNNNYYY
Stand alone keyless driverNNNYYY
Tool-less attachmentYYYYYY
90° degree headNNYNNY
180° degree headYNNYNH
20° angled head
Shaft diameterNNNYNN
Speed reduction*
Maximum speed
Degree of ark
Stroke length[PROPRIETARY INFORMATION]
Operating temp.(Max)
Stainless steel constructionYYYYYY
Corrosion resistant bearingsNNNYYY
Immersion cleaning**NNNYYY
4 Min flash sterilization **NNNYYY
Pre-sterilized blades (SAL)YYYYYY

★ OLD: Speed reduction achieved only by power reduction. NEW: Speed reduction achieved with internal gear reduction mechanism.

  • ** Excluding saw blades which are single use and disposable.
    __ End Summary _______________________________________________________________________________________________________________________________________________________________

2

Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

MAR 1 5 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. William G. Conety Director, Regulatory Affairs and Ouality Assurance Anspach Companies 4500 Riverside Drive Palm Beach Gardens, Florida 33410

Re: K040076

Trade/Device Name: Advanced Micro Saw Attachments -Sagittal. Oscillating and Reciprocating Regulation Number: 21 CFR 882.4310, 21 CFR 882.4360 Regulation Name: Powered simple cranial drills, burrs, trephines and their accessories; Electric cranial drill motor Regulatory Class: II Product Code: HBE, HEC Dated: January 12, 2004 Received: January 16, 2004

Dear Mr. Conety:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may. publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

3

Page 2 - Mr. William G. Conety

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost
C.C.

Celia M. Witton, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

510(k) Number: ΚΟΥΟΌ 7 6

Page 1 of 1

Device Name: Advanced Micro Saw Attachments - Sagittal, Oscillating and Reciprocating

INDICATIONS FOR USE:

Anspach micro saw attachments are indicated for use with eMax electrical and Black Max, MicroMax pneumatic drill systems for surgical cutting, shaping and removal of bone, including bones of the spine and cranium.

DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K040076

Prescription Use: V OR Over-the-Counter Use: