K Number
K033712
Device Name
MODIFICATION TO SECTRA IDS5 RADIOLOGY WORKSTATION - VERSION 10.1
Manufacturer
Date Cleared
2003-12-23

(27 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Sectra IDS device is intended for the manipulation and displaying of x-ray images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing. Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
Device Description
The IDS5 Radiology Workstation with IDS5/mx.net is mainly a software product. It is used for visualization and processing of digital radiology images. The system runs under the Window 2000 and Windows XP operating system. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.
More Information

K0033088

Not Found

No
The summary describes a software product for image visualization and processing, but there is no mention of AI, ML, or related terms, nor are there details about training or test sets typically associated with AI/ML development.

No
The device is described as a software product for manipulating and displaying x-ray images. Its function is to visualize and process medical images, which falls under diagnostic imaging rather than direct therapeutic intervention.

No

The device is described as a workstation for manipulating and displaying x-ray images and facilitating their viewing and processing by trained professionals. It does not perform an analysis or provide a diagnosis itself.

Yes

The device description explicitly states it is "mainly a software product" and the hardware requirements are described as "quite ordinary for a system used for displaying images," implying the core medical device functionality resides in the software.

Based on the provided information, the Sectra IDS device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Sectra IDS Function: The description clearly states that the Sectra IDS device is intended for the manipulation and displaying of x-ray images. This involves processing and visualizing medical images acquired from outside the body, not analyzing biological samples.
  • Input Modality: The input is described as "x-ray images, including mammograms; different modalities," which are imaging techniques, not biological samples.

Therefore, the Sectra IDS device falls under the category of medical imaging software/workstations, not in vitro diagnostics.

N/A

Intended Use / Indications for Use

The Sectra IDS device is intended for the manipulation and displaying of x-ray images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.

Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing.

Product codes

90 LLZ

Device Description

The IDS5 Radiology Workstation with IDS5/mx.net is mainly a software product. It is used for visualization and processing of digital radiology images. The system runs under the Window 2000 and Windows XP operating system. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

x-ray images, including mammograms. It can show images from different modalities

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K0033088

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

K03 3712

DEC 2 3 2003

510(k) Summary of Safety & Effectiveness

(as required by 21 CFR 807.92c)

Date Prepared:

December 16, 2003

Submitter's Information:

Sectra Imtec AB Teknikringen 20 SE-583 30 Linköping Sweden Phone: +1 46 13 23 52 00 Fax: +1 46 13 21 21 85

Trade Name, Common Name, Classification:

Trade Name:Sectra IDS5 Radiology Workstation with IDS5/mx.net
Common Name:Picture Archiving and Communications System
Classification Name:Image Processing System (LLZ) (21 CFR § 892.2050)

Predicate Device:

Applicant:Sectra Imtec AB
510(k) Number:K0033088
Device:Sectra IDS5 Radiology Workstation - Version 10.

Device Description:

The IDS5 Radiology Workstation with IDS5/mx.net is mainly a software product. It is used for visualization and processing of digital radiology images. The system runs under the Window 2000 and Windows XP operating system. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.

Indications for Use:

The Sectra IDS device is intended for the manipulation and displaying of x-ray images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.

Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing.

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

1

Technological Characteristics:

The IDS5 Radiology Workstation with IDS5/mx.net system will run on Windows 2000 and Windows XP operating system for PCs, (as a minimum and depending upon system configuration).

Performance Data:

The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.

Conclusion:

Similar to the predicate device, the IDS5 Radiology Workstation with IDS5/mx.net does not contact the patient, nor does it control any life sustaining devices. Images and information being reviewed, processed, relayed, and or transmitted are interpreted by a physician or trained medical personnel, providing ample opportunity for competent human intervention. The device and the predicate device share the same certification or conformance to performance standards and both function as Image Processing System (LLZ). Device failures, which might result in partial or failed transmissions, images, or data, may be recovered from storage or re-transmission after correcting the problem(s). Passwords are required for operation and to protect against unauthorized use.

Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.

2

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the department's name encircling an emblem. The emblem consists of a stylized caduceus-like symbol, representing health and medicine, with three intertwined strands.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Sectra Imtec AB % Mr. Carl Alletto OTech, Inc. 1600 Manchester Way CORINTH TX 76210

DEC 2 3 2003

Re: K033712

Trade/Device Name: Sectra IDS5 Radiology Workstation Version 10.1 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 21, 2003 Received: December 1, 2003

Dear Mr. Alletto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

3

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

4

510(k) Number: Ko 3 3712

Device Name: IDS (Image Display System) by Sectra Imtec AB

Indications For Use:

The Sectra IDS device is intended for the manipulation and displaying of x-ray images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.

Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing.

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

David A. Symm

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over -The-Counter Use

(Optional Format 1-2-96)