(63 days)
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent ticarcillin/clavulanate at concentrations of 1/2 - 128/2 µg/mL to gram-negative ID/AST or AST only Phoenix panels. Ticarcillin/clavulanate has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
- . BD Phoenix AST Broth used for performing AST tests only.
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
The acceptance criteria and study details for the BD Phoenix™ Automated Microbiology System's performance with ticarcillin/clavulanate are described below, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Performance Metric | Acceptance Criteria (Target) | Reported Device Performance (Ticarcillin/clavulanate) |
|---|---|---|
| Intra-site Reproducibility | > 90% | Greater than 90% |
| Inter-site Reproducibility | > 95% | Greater than 95% |
| Essential Agreement (EA) | Not explicitly stated, but assumed to be high based on "substantially equivalent performance" to reference method. | Not numerically specified in the provided text. The text states "assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested." and "Table 1 summarizes the performance for the isolates tested in this study" but the table content is missing. |
| Category Agreement (CA) | Not explicitly stated, but assumed to be high based on "substantially equivalent performance" to reference method. | Not numerically specified in the provided text. The text states "assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested." and "Table 1 summarizes the performance for the isolates tested in this study" but the table content is missing. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated as an overall number. The study utilized "Clinical, stock and challenge isolates."
- Data Provenance: The isolates were tested "across multiple geographically diverse sites across the United States." The study included both "Clinical isolates" and "Challenge set isolates." This indicates a mix of retrospective (clinical isolates from past patient samples) and potentially prospective (challenge isolates specifically designed for testing) data, originating from the United States.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified. The ground truth was established by comparison to the "NCCLS reference broth microdilution method" and "expected results" for challenge isolates. This suggests expert consensus on the interpretation of reference method results, but the experts themselves are not detailed.
4. Adjudication method for the test set:
- Adjudication Method: Not explicitly described in terms of a specific "X+Y" methodology. The Phoenix System results were "compared to the expected results" (for challenge isolates) and "compared to the results obtained from the NCCLS reference broth microdilution method" (for clinical isolates). This implies the NCCLS reference method and pre-defined expected results served as the definitive ground truth against which the device's performance was measured.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, this does not appear to be an MRMC comparative effectiveness study involving human readers and AI assistance. The device is an automated system for antimicrobial susceptibility testing; it does not involve human readers interpreting images or data with or without AI assistance in the context of a typical MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Yes, this entire study is focused on the standalone performance of the BD Phoenix™ Automated Microbiology System (the "algorithm only") without an explicit "human-in-the-loop" component for interpretation. The system automates the preparation, incubation, reading, and interpretation of results.
7. The type of ground truth used:
- Type of Ground Truth: The ground truth for clinical isolates was the NCCLS reference broth microdilution method. For challenge isolates, the ground truth was expected results.
8. The sample size for the training set:
- Sample Size: The document does not provide details on a separate "training set" or its size. The described studies focus on testing the performance of the system against a reference method. For antimicrobial susceptibility devices, the "training" (or development) of the device's interpretive algorithms would typically occur prior to these validation studies, and the specifics of that development dataset are not included in this 510(k) summary.
9. How the ground truth for the training set was established:
- Ground Truth for Training Set: Not specified, as details about a distinct training set are not provided in this document. The device's internal algorithms would have been developed using some form of reference data, likely the NCCLS reference method for AST, but the specifics are not elaborated here.
{0}------------------------------------------------
| JAN 1 4 2004 | 510(K) SUMMARY |
|---|---|
| SUBMITTED BY: | Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 410-316-4778Fax: (410)-316-4499 |
| CONTACT NAME: | Michelle Bytheway BandyRegulatory Affairs Specialist |
| DATE PREPARED: | November 7, 2003 |
| DEVICE TRADE NAME: | BD Phoenix TM Automated Microbiology System -Ticarcillin/clavulanate 1/2 - 128/2 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK ® System (PMA No. N50510) and BD Phoenix TMAutomated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix TM Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin. |
DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
- . BD Phoenix AST Broth used for performing AST tests only.
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
DEVICE COMPARISON:
The BD Phoenix"™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.
{2}------------------------------------------------
Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
Table 1: Performance of BD Phoenix System for Gram-Negative Organisms by Drug
| Antimicrobial | Concentration | ||||
|---|---|---|---|---|---|
| 1carcillin/ | 17017 | t | 00 | ( | 00 - |
| clavulanate | ug/ml | ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ |
Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Michelle Bytheway Bandy Regulatory Affairs Specialist Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
JAN 1 4 2004
K033557 Re: Trade/Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent ticarcillin/clavulanate(1/2 - 128/2 µg/mL.) -Gram-Negative ID.AST or AST only Phoenix panels Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system. Regulatory Class: Class II Product Code: LON Dated: November 7, 2003 Received: November 12, 2003
Dear Ms. Bandy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Salamat
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: 1033557
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent ticarcillin/clavulanate (1/2 - 128/2 µg/mL) - Gram-Negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent ticarcillin/clavulanate at concentrations of 1/2 - 128/2 µg/mL to gram-negative ID/AST or AST only Phoenix panels. Ticarcillin/clavulanate has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against
| (both β-lactamase producing and non β-lactamase producing): | ||
|---|---|---|
| Citrobacter species | Escherichia coli | Serratia marcescens |
| Enterobacter species | Klebsiella species | |
| Active In Vitro Against: | ||
| (both β-lactamase producing and non β-lactamase producing): | ||
| Acinetobacter baumannii | Morganella morganii | Proteus vulgaris |
| Acinetobacter calcoaceticus | Proteus mirabilis | Providencia rettgeri |
| Acinetobacter haemolyticus | Proteus penneri | Providencia stuartii |
| Acinetobacter lwoffi | ||
| Prescription Use | AND/OR | Over-the-Counter Use |
(Part 21 CFR 801 Subpart D)
AND/OR
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Ludmila Roko
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K033557
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”