K Number
K033538
Manufacturer
Date Cleared
2004-02-02

(84 days)

Product Code
Regulation Number
880.6880
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Belimed Steam Sterilizer TOP 5000, series 4-8, is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities. The Belimed Steam Sterilizer TOP 5000 Series 4 - 8 is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1).

Device Description

The Belimed Steam Sterilizer TOP 5000 Series 4 – 8 is intended for use in hospital and health care facilities and is intended to be used in an identical manner as the Belimed Steam Sterilizer TOP 5000 Series 9 - 18. The smaller Chamber size incorporates additional flexibility, in comparison to Series 9 -- 18, and allows to operate the sterilizer in an economical way. Motorized vertical door movement leads to an optimal compact design, and allows the operation of the sterilizer, where space is limited.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the Belimed Steam Sterilizer TOP 5000 Series 4-8.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Sterility Assurance Level (SAL)At least 10^-6 (reduction factor of 10^6)Achieved: All PREVAC, FLASH, and LIQUID cycles demonstrated an SAL of at least 10^-6 through achievement of a time-at-temperature sufficient to produce an F0 of at least 12 by ½ cycle.
Moisture Retention (PREVAC - Fabric Test Pack)Less than 3% increase in pre-sterilization test pack weight and no wet spots.Met: PREVAC cycles using the fabric test pack (Section 5.5.2 AAMI/ANSI-ST8:2001) demonstrated moisture retention of less than 3% increase in pre-sterilization test pack weight and exhibited no wet spots.
Moisture Retention (PREVAC - Full Load Instrument Trays)Less than 20% increase in pre-sterilization weight of the towel and no wet spots on the outer wrapper.Met: PREVAC cycles using full load instrument trays (Section 5.5.4 of ANSI/AAMI-ST8:2001) demonstrated moisture retention of less than 20% increase in pre-sterilization weight of the towel and exhibited no wet spots on the outer wrapper.
Wet Spots (FLASH cycles)No wet spots.Met: FLASH cycles using the unwrapped instrument tray (Section 5.5.5.1 AAMI/ANSI-ST8:2001 and ANSI/AAMI ST37:1996 section 7.7.3) exhibited no wet spots.
LIQUID cycles performanceAchieve and maintain 121°C in the center of the liquid for at least 12 minutes, water loss not exceeding 50 ml, and automatic sealing of the flask closure.Met: LIQUID cycles using three 1'000 ml flasks (Section 5.5.3 of AAMI/ANSI-ST8:2001) demonstrated achieving and maintaining 121°C in the center of the liquid for at least 12 minutes, water loss not exceeding 50 ml, and automatic sealing of the flask closure.
Bowie-Dick (BD) Test PackUniform color change throughout the test sheet.Met: The BD cycle (Section 5.6 of AAMI/ANSI-ST8) demonstrated a uniform color change throughout the test sheet.
Software ValidationPer FDA's moderate level of concern recommendations in "Guidance for the Content for Pre-market Submissions for Software Contained in Medical Devices (5/29/98)".Met: Software validation for cycle operation was performed according to the specified FDA guidance.
Electrical SafetyCompliance with UL 61010A-1:2002 and UL 61010A-2-041:2002.Complies: The device complies with Underwriter Laboratory (UL) Code UL 61010A-1:2002 and UL 61010A-2-041:2002.
Electrical Safety (Canadian)Compliance with CSA C22.2 No. 1010-1 (IEC61010-1:2001).Complies: The device complies with Canadian Standards Association (CSA) Standard C22.2 No. 1010-1 (IEC61010-1:2001).
Pressure Vessel SafetyCompliance with ASME, Section VIII, Division 1 for unfired pressure vessels :2001.Complies: The device complies with American Society of Mechanical Engineers (ASME), Section VIII, Division 1 for unfired pressure vessels :2001.

2. Sample size used for the test set and the data provenance

The document indicates that the sterilizer's cycles were "validated to meet the requirements of AAMI/ANSI-5T8-2001" and that "verification studies" were performed.

  • Test Set Sample Size:
    • PREVAC cycles (fabric test pack): "verified using the fabric test pack, as described in Section 5,5.2 AAMI/ANSI-ST8:2001". The specific number of fabric test packs used is not explicitly stated.
    • PREVAC cycles (full load instrument trays): "verified using full load instruments trays [...] according to section 5.5.4 of ANSI/AAMI-ST8: 2001". The specific number of instrument trays used is not explicitly stated.
    • FLASH cycles: "verified using the unwrapped instrument tray [...] according to section 5.5.5.1 AAMI/ANSI-ST8:2001 and ANSI/AAMI ST37:1996 section 7.7.3". The specific number of unwrapped instrument trays is not explicitly stated.
    • LIQUID cycles: "verified using three 1'000 ml flasks, as described in section 5.5.3 of AAMI/ANSIe ST8:2001". This indicates a minimum of three 1'000 ml flasks were used per test.
    • BD cycle: "verified using the Bowie-Dick Test Pack were qualified according to section 5.6 . of AAMI/ANSI-ST8". The number of Bowie-Dick Test Packs is not explicitly stated but implies standard usage per the referenced section.
  • Data Provenance: The studies were conducted by SAUTER AG. The data provenance is internal testing and validation performed by the manufacturer in accordance with recognized industry standards (AAMI/ANSI-ST8:2001, ANSI/AAMI ST37:1996, EN285:1996). The context suggests this was prospective testing conducted to validate the device's performance. The country of origin of the data would be associated with the manufacturer, Sauter AG, based in Switzerland.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the device is a steam sterilizer, not an AI or diagnostic imaging device that requires human expert review for ground truth in the typical sense. The ground truth is established through physical measurements and biological indicator kill rates as defined by established standards (e.g., AAMI/ANSI-ST8:2001).

4. Adjudication method for the test set

This section is not applicable. As stated above, the device's performance is assessed against physical and biological criteria defined by standards, not by human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is a steam sterilizer, not an AI-assisted diagnostic or interpretation system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The performance described is standalone. The sterilizer operates autonomously based on its programmed cycles, and its effectiveness is validated through its ability to achieve specific physical and biological endpoints (e.g., SAL, temperature, moisture retention). There is no "human-in-the-loop" performance validation in the sense of an algorithm, other than operators configuring and starting the cycles.

7. The type of ground truth used

The ground truth used for the sterilizer's performance validation is based on objective physical and biological parameters defined by recognized standards:

  • Sterility Assurance Level (SAL) of 10^-6: Demonstrated via complete kill of biological indicators and calculation of F0 values.
  • Time-at-temperature achievement: Verification that specified temperatures were reached and maintained for the required duration.
  • Moisture retention: Quantitative measurement of weight increase (for fabric/towel) and visual inspection for wet spots.
  • Water loss: Quantitative measurement for liquid cycles.
  • Automatic sealing of flask closure: Observed performance for liquid cycles.
  • Uniform color change: For the Bowie-Dick test.
  • Compliance with safety standards (UL, CSA, ASME).

These are empirical measurements and outcomes directly observable or measurable against defined benchmarks, not subjective expert consensus.

8. The sample size for the training set

This section is not applicable. The Belimed Steam Sterilizer TOP 5000 Series 4-8 is a hardware device with programmed cycles, not an AI/ML algorithm that undergoes "training" on a data set. The cycles are "factory-programmed" and validated through engineering design and physical testing.

9. How the ground truth for the training set was established

This section is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K03 3538 21 1.2 510(k) Summary

Date:Nov., 30°.2003FEB - 2 2003
Submitter's Name / Address:Sauter AGZelgstrasse 8CH- 8583 Sulgen / Switzerland
Contact Person:Hans Stadler, Head of Product Development,Email: hst@sauterag.com
Trade Name:Belimed Steam Sterilizer TOP 5000Series 4 - 8
Classification:Steam Sterilizer - Class II, as listed per 21C.F.R. 880.6880
Predicate Device:Belimed Steam Sterilizer TOP 5000Series 9 - 18

DEVICE DESCRIPTION:

The Belimed Steam Sterilizer TOP 5000 Series 4 – 8 is intended for use in hospital and health care facilities and is intended to be used in an identical manner as the Belimed Steam Sterilizer TOP 5000 Series 9 - 18.

The smaller Chamber size incorporates additional flexibility, in comparison to Series 9 -- 18, and allows to operate the sterilizer in an economical way.

Motorized vertical door movement leads to an optimal compact design, and allows the operation of the sterilizer, where space is limited.

COMPARISON TO THE PREDICATE DEVICE:

The Belimed Steam Sterilizer TOP 5000 Series 4 -- 8, is very similar to the predicate device. Modifications made from the predicate device include:

  • ಾ Smaller chamber size
  • t : Vacuum-pump according to smaller chamber size
  • Vertical door movement ନ୍ତ
  • Software updates t

INDICATIONS FOR USE:

The Belimed Steam Sterilizer TOP 5000 Series 4 -- 8 is designed for Sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities.

The Belimed Steam Sterilizer TOP 5000 Series 4 -- 8 is equipped with the following factoryprogrammed Sterilization cycles and cycle values, which are identical with Series 9 - 18 (Table 1).

{1}------------------------------------------------

CYCLESSTERILIZE TEMPSTERILIZE TIME(min)DRY TIME (min)RECOMMENDED LOAD
PREVAC270° F (132°C)270° F(132°C)420Double-wrapped Instrument Trays, max.weight of 17 lbs (7.7 kg) each. Fabric packs.Refer to Table 2 for recommended quantities.
PREVAC270°F (132°C)270° F(132°C)45Fabric Packs.
LIQUID250°F (121°C)250° F(121°C)450Liquids not intended for direct patientcontact!Refer to Table 3 for Guidelines
EXPRESS270°F (132°C)270° F(132°C)43Single Wrapped Instrument Tray with nonporous single instrument
FLASH270° F (132°C)270° F(132°C)31Unwrapped Instrument Tray with a singleInstrument
FLASH 270° F(132°C)270° F(132°C)101Unwrapped Instrument Tray with non porousmultiple instruments (max. weight of 17 lbs)

Table 1: Factory programmed Sterilization Cycles

The following table (Table 2) shows SAUTER AG's recommended load by sterilizer size, Series 4 - 8:

Modelsingle doordouble doorSterilizer Chamber SizeWrappedInstrument Trays20"x10"max. 17lb eachFabric Packs11"x11"x9"max. 6.6lb eachFabric Packs23"x11"x11"max. 17lb each
6-0-6 VS16-0-6 VS226" x 26" x 27.5"(660 x 660 x 700) mm484
6-0-9 VS16-0-9 VS226" x 26 "x 39.5"(660 x 660 x 1000) mm6126
6-0-12 VS16-0-12 VS226" x 26" x 51.5"(660 x 660 x 1300) mm8168

Iable 2: Recommended Loads

The following table (Table 3) is SAUTER AG's guidelines for liquid cycle processing for Series 4 - 8:

Modelsingle doordouble doorSterilizer Chamber SizeVolume of Liquid in OneContainerNumber of containers
6-0-6 VS16-0-6 VS226" x 26" x 27,5"(660 x 660 x 700) mm1000 ml44
6-0-9 VS16-0-9 VS226" x 26" x 39,5"(660 x 660 x 1000) mm1000 ml66
6-0-12 VS16-0-12 VS226" x 26" x 51,5"(660 x 660 x 1300) mm1000 ml88

Table 3: Guidelines for liquid 250°F cycle processing

:

:

{2}------------------------------------------------

The Belimed Steam Sterilizer TOP 5000 Series 4 -- 8 is offered in the following size configurations:

26" x 26" x 27.5" (660 mm x 660 mm x 700 mm)Single Door, Prevacuum
26" x 26" x 27.5" (660 mm x 660 mm x 700 mm)Double Door, Prevacuum
26" x 26" x 39.5" (660 mm x 660 mm x 1000 mm)Single Door, Prevacuum
26" x 26" x 39.5" (660 mm x 660 mm x 1000 mm)Double Door, Prevacuum
26" x 26" x 51.5" (660 mm x 660 mm x 1300 mm)Single Door, Prevacuum
26" x 26" x 51.5" (660 mm x 660 mm x 1300 mm)Double Door, Prevacuum

The Belimed Steam Sterilizer TOP 5000 Series 4 – 8 is designed to be used for the terminal Sterilization of porous and non porcus, heat and moisture stabile materials in the healthcare facilities.

Depending of the chosen cycle materials as different as textiles, glassware, unwrapped or wrapped instrument trays with single or multiple instruments.

The Belimed Steam Sterilizer TOP 5000 Series 4 – 8 are factory equipped with cycles which has been tested in accordance with AAMI/ANSI ST-8:2001 as well as EN285:1996 Standards under defined load conditions.

EFFECTIVENESS:

Efficacy of sterilizer function and exposure time recommendations ate ultimately shown by complete kill of biological indicators and verifying an appropriate safety factor or sterility assurance level (SAL) of at least 10° reduction. SAUTER AG validates its sterilization cycles by recommended practices, standards and quidelines developed by various independent organizations such as the Association for Advancement of Medical Instrumentation (AAMI). Prior to release, the Belimed Steam Sterilizer TOP 5000 were validated to meet the requirements of AAMI/ANSI-5T8-2001.

The results of the Belimed Steam Sterilizer TOP 5000 verification studies demonstrate that the sterilizer performs as intended and are summarized as follows:

  • . All PREVAC cycles verified using the fabric test pack, as described in Section 5,5.2 AAMI/ANSI-ST8:2001 were qualified according to section 5.5.2.5 ANSI/AAMI-ST8. These cycles demonstrated a sterility assurance level of at least 10 through achievement of a time-at-temperature sufficient to produce an FO of at least 12 by ½ cycle, a moisture retention of fess than 3% increase in presterilization test pack weight, and exhibited no wet spots.
  • A!! PREVAC cycles verified using full load instruments trays were qualified according to section ୍ଦ 5.5.4 of ANSI/AAMI-ST8: 2001. These cycles demonstrated a sterility assurance level of at least 10 6 through achievement of a time-at-temperature sufficient to produce an F0 of at least 12 by ½ cycle, a moisture retention of less than 20% increase in pre-sterilization weight of the towel, and exhibited no wet spots on the outer wrapper.
  • All FLASH cycles verified using the unwrapped instrument tray were qualified according to section ti 5.5.5.1 AAMI/ANSI-ST8:2001 and ANSI/AAMI ST37:1996 section 7.7.3 . These cycles demonstrated a sterility assurance level of at least 10* through achievement of a time-attemperature sufficient to produce an F0 of at least 12 by ½ cycle and exhibited no wet spots.
  • All LIQUID cycles verified using three 1'000 ml flasks, as described in section 5.5.3 of AAMI/ANSIe ST8:2001, were qualified according to section 5.5.3.5. These cycles demonstrated a sterility assurance level of at least 10€ through achievement of a time-at-temperature sufficient to produce an FO of at least 12 by ½ cycle, a water loss not exceeding 50 ml, and automatic sealing of the

{3}------------------------------------------------

flask closure. A temperature of 121°C was achieved and maintained in the center of the liquid for at least 12 minutes.

  • The BD cycle was verified using the Bowie-Dick Test Pack were qualified according to section 5.6 . of AAMI/ANSI-ST8, and demonstrated a uniform color change throughout the test sheet.
  • The software validation for the cycle operation was performed according to FDA's moderate level . of concern recommendations provided in the document "Guidance for the Content for Pre-market Submissions for Software Contained in Medical Devices (5/29/98)".

SAFETY:

SAUTER AG sterilizers including the Belimed Steam Sterilizer TOP 5000 have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The Belimed Steam Sterilizer TOP 5000 complies with the following requirements:

  • Underwriter Laboratory (UL) Code UL 61010A-1:2002 and Ul. 61010A-2-041:2002 ﺎ .
  • Canadian Standards Association (CSA) Standard C22.2 No. 1010-1 (IEC61010-1:2001) 2.

American Society of Mechanical Engineers (ASME), Section Vill, Division 1 for unfired 3. pressure vessels :2001.

HAZARDS-FAILURE OF PERFORMANCES

Failure of the sterilization process can lead to incidence of cross contamination, the transmission of potentially infectious organisms from one infected person to another who was not otherwise infected prior to the incident.

To avoid failure, the user must ensure the materials, instruments and devices to be sterilized are thoroughly cleaned, that the manufacturer's instructions for use are followed,, that the cycle to be used for each type of sterilizer load has been validated, that the sterilizer has been maintained in accordance with the sterilizer's manufacturer's maintenance schedule and is operating properly, and that each sterilizer load is monitored with available and validated biological and chemical sterilization process indicators.

Today, there are many steam sterilizers in daily use in hospitals throughout the United States. The incident of sterilizer malfunction or sterilization process failure is relatively rare considering the wide spread use of steam sterilizers. Further, there are no known reports in the literature of patient infection that have resulted from steam sterilizer failure. The technology designed in Belimed Steam Sterilizer TOP 5000 provides microprocessor controller safeguards that aborts the cycle and aive appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.

USER INFORMATION

SAUTER AG provides information to the user that is intended to insure safe and effective use of steam sterilization in its detailed Operator's Manual and other labeling. SAUTER AG also recommends the use and periodic review of the AAMI steam sterilization standards to ensure further assurance of the safe and effective use of steam sterilization equipment in health care facilities.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 2 2004

Sauter AG C/O Mr. Stefan Preiss Responsible third Party Official TUV America, Incorporated 1775 Old Highway 8 New Brighton, Minnesota 55112-1891

Re: K033538

Trade/Device Name: Bellmed Steam Sterilizer Top 5000. Series 4- 8 Regulation Number: 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: January 17, 2004 Received: January 22, 2004

Dear Mr. Preiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -Mr. Preiss

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K033538

Device Name: Bellmed Steam Sterlilzer TOP 5000

Indications For Use:

The Belimed Steam Sterilizer TOP 5000, series 4-8, is designed for sterilization of non porous and porous heat and moisture-stable materials used in healthcare facilities.

The Belimed Steam Sterilizer TOP 5000 Series 4 - 8 is equipped with the following factory-programmed Sterilization cycles and cycle values (Table 1),

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use x OTC (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 3

{7}------------------------------------------------

K0355K

CYCLESSTERILIZETEMPSTERILIZETIME(min)DRY TIME(min)RECOMMENDED LOAD
PREVAC270° F (132°C)270° F(132°C)420Double-wrapped Instrument Trays, max.weight of 17 lbs (7.7 kg) each, Fabric packs.Refer to Table 2 for recommended quantities.
PREVAC270° F (132°C)270° F(132°C)45Fabric Packs.
LIQUID250° F (121°C)250° F(121°C)450Liquids not intended for direct patient contact!Refer to Table 3 for Guidelines
EXPRESS270° F (132°C)270°F(132°C)43Single Wrapped Instrument Tray with nonporous single instrument
FLASH270° F (132°C)270° F(132°C)31Unwrapped Instrument Tray with a singleInstrument
FLASH 270° F(132°C)270° F(132°C)101Unwrapped Instrument Tray with non porousmultiple instruments (max. weight of 17 lbs)

Table 1: Factory programmed Sterilization Cycles

The following table (Table 2) shows SAUTER AG's recommended load by sterilizer size, Series 4 - 8:

Modelsingle doordouble doorSterilizer Chamber SizeWrapped InstrumentTrays 20"x10"max. 17lb eachFabric Packs11"x11"x9"max. 6.6lb eachFabric Packs23"x11"x11"max. 17lb each
6-0-6 VS16-0-6 VS226"x 26" x 27.5"(660 x 660 x 700) mm484
6-0-9 VS16-0-9 VS226"x 26"x 39.5"(660 x 660 x 1000) mm6126
6-0-12 VS16-0-12 VS226"x 26" x 51.5"(660 x 660 x 1300) mm8168

Table 2: Recommended Loads

.. .

Page 2 of 3

{8}------------------------------------------------

K033538

The following table (Table 3) is SAUTER AG's guidelines for liquid cycle processing for Series 4 - 8:

Modelsingle doordouble doorSterilizer Chamber SizeVolume of Liquid in OneContainerNumber of containers
6-0-6 VS16-0-6 VS226"x 26" x 27,5"(660 x 660 x 700) mm1000 ml44
6-0-9 VS16-0-9 VS226"x 26" x 39,5"(660 x 660 x 1000) mm1000 ml66
6-0-12 VS16-0-12 VS226"x 26" x 51,5"(660 x 660 x 1300) mm1000 ml88

Table 3: Guidelines for liquid 250°F cycle processing

The Belimed Steam Sterilizer TOP 5000 Series 4 - 8 is offered in the following size configurations:

26"x 26" x 27.5" (660 mm x 660 mm x 700 mm)Single Door, Prevacuum
26"x 26" x 27.5" (660 mm x 660 mm x 700 mm)Double Door, Prevacuum
26"x 26 "x 39.5" (660 mm x 660 mm x 1000 mm)Single Door, Prevacuum
26"x 26 "x 39.5" (660 mm x 660 mm x 1000 mm)Double Door, Prevacuum
26"x 26" x 51.5" (660 mm x 660 mm x 1300 mm)Single Door, Prevacuum
26"x 26" x 51.5" (660 mm x 660 mm x 1300 mm)Double Door, Prevacuum

Pat Ray Anderson Benard 1/30/0k

(Division ... In-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental D

510(k) Number

Page 3 of 3

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).