(156 days)
The ZAP™ hCG Test is an immunoassay used for the qualitative detection of human chorionic gonadotropin in human serum, plasma or whole blood and is indicated as an aid for health care professionals in the diagnosis of early pregnancy.
The ZAP™ hCG Test is a one-step immunochromatographic assay for the rapid qualitative detection of hCG in human serum, plasma or whole blood.
Here's an analysis of the provided text regarding the acceptance criteria and study for the ZAP™ hCG Test:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" with numerical thresholds against which the device was measured, as might be found in a highly structured clinical trial. Instead, it describes performance characteristics that were evaluated and compared to a predicate device. The primary success metric appears to be agreement with the predicate device and satisfactory performance in key analytical functional effectiveness studies.
Performance Characteristic/Goal | Stated Performance (ZAP™ hCG Test) |
---|---|
Agreement with Predicate Device (ICON® II HCG ImmunoConcentration Assay) | "A total of 116 serum samples were tested and compared to the results obtained with the predicate device. The results of the two assays were in agreement except for four samples that were positive for hCG by the predicate device only. The discrepant samples were quantitated with the Beckman Access® Total β-hCG assay and the concentrations were in the range 2-6 mIU hCG/mL. Therefore, they were correctly reported as negative based on the 10 mIU hCG/mL cut-off value of the ZAP™ hCG Test." |
Interference | "No interference from potentially interfering substances (albumin at 14 g/dL, bilirubin at 30 ng/dL, haemoglobin at 250 mg/dL, triglycerides at 2000 mg/dL)" |
Cross-reactivity | "No cross-reactivity to LH at 500 mIU/mL, FSH at 1000 mIU/mL, or TSH at 1000 µIU/mL" |
Correlation (Whole Blood and Plasma) | ">90% correlation of results from whole blood and plasma near the cut-off value" |
Blind hCG Study (Spiked Whole Blood) | "100% agreement in a blind hCG study using spiked whole blood and involving three health care sites." |
Conclusion of Equivalence | "The ZAP™ hCG Test performs as well as the ICON® II HCG ImmunoConcentration Assay legally marketed by Beckman Coulter Inc." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Agreement with Predicate Device: 116 serum samples.
- Blind hCG Study: Not explicitly stated, but involved "spiked whole blood."
- For other analytical studies (interference, cross-reactivity, whole blood/plasma correlation), specific sample sizes are not provided, but they are implied through the mention of "a number of evaluation studies."
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective in nature, particularly the comparison with the predicate device and the analytical evaluations using prepared samples (spiked whole blood, interfering substances). There is no mention of prospective clinical recruitment.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable in the conventional sense. For this in vitro diagnostic device, the "ground truth" for the test set was established using:
- The results of a legally marketed predicate device (ICON® II HCG ImmunoConcentration Assay).
- A quantitative reference assay (Beckman Access® Total β-hCG assay) for discrepant samples.
- Laboratory measurements of known concentrations of interfering substances, cross-reactants, and spiked hCG levels.
There were no human experts establishing ground truth through interpretation for the performance studies described.
4. Adjudication Method for the Test Set
Not applicable for this type of in vitro diagnostic device study. Adjudication methods (like 2+1, 3+1) are typically used for studies where human interpretation of medical images or clinical cases forms the basis of ground truth, and discrepancies between readers need to be resolved. Here, the ground truth was based on instrument readings and reference assays.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?
No, an MRMC comparative effectiveness study was not done. This device is an immunoassay for rapid qualitative detection of hCG, not an AI-assisted diagnostic tool that would be used by human readers in the interpretation of complex medical data. Therefore, the concept of human readers improving with AI assistance does not apply.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies described are essentially standalone performance evaluations of the ZAP™ hCG Test as an algorithm/device only. The device is intended to provide a direct qualitative result (positive/negative) without requiring additional human interpretation of the result strip, beyond reading the control and test bands. The "blind hCG study using spiked whole blood and involving three health care sites" further supports this, as it evaluates the device's output accuracy in a practical setting without human interpretive variability being the primary focus.
7. The Type of Ground Truth Used
The ground truth used was a combination of:
- Predicate Device Results: The established performance of a legally marketed, substantially equivalent device (ICON® II HCG ImmunoConcentration Assay).
- Quantitative Reference Assay: The Beckman Access® Total β-hCG assay was used to resolve discrepancies with the predicate device, providing a quantitative "true" value for those samples.
- Known Concentrations: For interference and cross-reactivity studies, the ground truth was known concentrations of substances deliberately added or tested against.
- Spiked Samples: For the blind study, samples were "spiked" with known concentrations of hCG to establish a clear ground truth.
8. The Sample Size for the Training Set
Not applicable. The ZAP™ hCG Test is an immunoassay kit, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The device's performance is determined by its biochemical reagents and physical design, which are developed and validated through laboratory testing and manufacturing controls.
9. How the Ground Truth for the Training Set Was Established
As explained above, there is no "training set" for this type of device. The "ground truth" used during the development and validation of the assay (analogous to training/testing in ML) would involve careful preparation of samples with known hCG concentrations, interfering substances, and cross-reactants, followed by analytical testing to optimize the assay's sensitivity, specificity, and robustness.
§ 862.1155 Human chorionic gonadotropin (HCG) test system.
(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.